APM-40 SERIES - MINMAX Technology

APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
FEATURES
►Fully encapsulated Plastic Case
►3 Mounting Versions:
- PCB Mounting with Solder Pins
- Chassis Mounting with Screw Terminals
- DIN-Rail Mounting
► Package Dimension 88.9x63.5x30.0 mm (PCB Version)
►Universal Input 85-264VAC, 47-440 Hz
►Protection Class ll
►Extended Operating Temp.Range -40°C to +60°C at full Load
►LED Output Indicator (Chassis Version Models)
►Industrial Safety to UL/cUL/IEC/EN 60950-1 and UL508
►Medical Safety Approval to ANSI/AAMI ES60601-1 2XMOPP
IEC/EN 60601-1 3rd Edition CB Report
►Over Load and Over Temperature Protection
►3 Year Product Warranty
PRODUCT OVERVIEW
The new MINMAX APM-40 series is a range of fully encapsulated AC/DC power modules. These high performance products feature an extended operating
temperature range of -40°C to +80°C. Universal input voltage 85-264VAC and UL/IEC/EN safety approvals including medical safety and UL508 listing
qualify these power supplies modules for applications in products with worldwide markets.EMI-filter meets EN55022, class B and FCC,part15, class B.
The APM-40 series power modules provide an economical solution for many space critical applications in commercial, medical and industrial electronic
equipment.
Model Selection Guide
Model Number PCB Mounting
Output
(For model with Chassis
Voltage
Output Current
Input Current
Max. capacitive
Load
Max.
115VAC, 60Hz
230VAC, 50Hz
@Max. Load
Mounting, add suffix C)
VDC
5
12
15
24
±12
±15
APM-40S05
APM-40S12
APM-40S15
APM-40S24
APM-40D12
APM-40D15
mA
8000
3330
2660
1660
±1660
±1330
mA(typ.)
716
689
680
687
687
680
(typ.)
@Max. Load, 115VAC
μF
8000
3900
3900
680
1500#
1000#
# For each output
429
414
408
413
413
408
Efficiency
%
81
84
85
84
84
85
Input Specifications
Parameter
Model
AC Voltage Input Range
Input Frequency Range
All Models
DC Voltage Input Range
No-Load Power Consumption
Inrush Current (Cold Start at 25℃)
E-mail:[email protected]
2014/08/07 REV:1
Typ.
Max.
Unit
85
---
264
VAC
47
---
440
Hz
120
---
370
VDC
---
---
0.3
W
115VAC
---
---
30
A
230VAC
---
---
60
A
Tel:886-6-2923150
Page 1 of 5
Min.
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
Output Specifications
Parameter
Conditions
Min.
Typ.
Max.
Unit
-----
±2.0
±0.5
-----
--±1.0
--±2.0
No minimum Load Requirement
5VDC Output Models
1.5
--0-20 MHz Bandwidth
Other Output Models
1.0
--Zener diode clamp
120
----±0.02
----85VAC, Hiccup Mode, auto-recovery
105
--(long term overload condition may cause damage)
Hiccup mode, indefinite (automatic recovery)
-----
%
%
%
%
1.8
1.3
----5
%VPP of Vo
%VPP of Vo
% of Vo
%/℃
%
---
% Inom.
Max.
Unit
Output Voltage Accuracy
Line Regulation
Single Output Model
Dual Output Models
Load Regulation
Min.Load
Ripple & Noise
Over Voltage Protection
Temperature Coefficient
Overshoot
Current Limitation
Short Circuit Protection
General Specifications
Parameter
Conditions
I/O Isolation Voltage (reinforced)
Leakage Current
I/O Isolation Resistance
Switching Frequency
Min.
Typ.
4000
----VACrms
--80
--μA
500 VDC
1000
----MΩ
--130
--KHz
115VAC, 60Hz
--25
--ms
230VAC, 50Hz
--80
--ms
MIL-HDBK-217F@25℃, Ground Benign
200,000
----Hours
According IEC/EN 60536
IEC/EN 60950-1 CB Report, cUL/UL 60950-1, CSA-C22.2 No.60950-1, UL508 listed certificate
IEC/EN 60601-1 3rd Edition CB Report, ANSI/AAMI ES60601-1 2XMOPP, CAN/CSA-C22.2 No.60601-1 certificate
Hold-up Time
MTBF (calculated)
Protection Class II
Safety Approvals
EMC Specifications
Parameter
Standards & Level
Performance
EN55011, EN55022, EN61000-6-4, EN61000-6-3, FCC part 15
EN60601-1-2, EN55024, EN61000-6-2, EN61000-6-1
ESD
EN61000-4-2 air ± 8kV , Contact ± 4kV
Radiated immunity
EN61000-4-3 10V/m
Fast transient
EN61000-4-4 ±2kV
Surge
EN61000-4-5 ±1kV
EN61000-4-6 10Vrms
Conducted immunity
EN61000-4-8 30A/M
PFMF
EN61000-4-11 30% 10ms
Dips
EN61000-4-11 >95% 5000ms
Interruptions
EMI
EMS
Class B
A
A
A
A
A
A
A
B
Environmental Specifications
Parameter
Temperature Range (operational)
Power Derating
Storage Temperature Range
Over Temperature Protection
Humidity (non condensing)
Cooling
E-mail:[email protected]
2014/08/07 REV:1
Conditions
Min.
Max.
Ambient
Above +60℃
-40℃
+80℃
1.5W / ℃
+95℃
-40℃
Shutdown at 90℃ (automatic recovery at approx. 67℃)
--Free-Air convection
Tel:886-6-2923150
Page 2 of 5
95% rel. H
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
Notes
1
This product is not designed for use in critical life support systems, equipment used in hazardous environment, nuclear control systems or other
2
Specifications typical at Ta=+25℃, resistive load, 115VAC, 60Hz input voltage, after warm-up time rated output current unless otherwise noted.
3
Ripple & Noise measured with a 0.1μF/50V MLCC and a 1μF/50V Aluminum electrolytic.
such applications which necessitate specific safety and regulatory standards other the ones listed in this datasheet.
4
Safety approvals cover frequency 47-63 Hz.
5
We recommend to protect the converter by a slow blow fuse in the input supply line.
6
Other input and output voltage may be available, please contact factory.
7
To order the module with chassis mount package, please add a suffix C (e.g. APM-40S05C).
8
Part number for DIN-Rail mounting bracket: AC-DIN-02
9
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
10
Specifications are subject to change without notice.
Package Specifications PCB Mounting
Mechanical Dimensions
Pin Connections
5
4
0.7
[0.03]
3
Pin
Single Output
Dual Output
1
AC (N)
AC (N)
2
AC (L)
AC (L)
3
+Vout
+Vout
4
No Pin
No Pin
5
-Vout
Common
6
No Pin
No Pin
7
NC
-Vout
[
7
6
6.0
[0.24]
15.24
[0.60]
12.7
[0.50]
12.7
[0.50]
1.0
0.04]
15.24
[0.60]
30.0
[1.18]
15.9 [0.63]
88.9 [3.50]
81.3 [3.20]
31.6 [1.24]
NC: No Connection
53.3 [2.10]
Mounting M4.0 Thread
Bottom View
2
1
5.0
[0.20]
►All
55.88 [2.20]
63.5 [2.50]
Physical Characteristics
Case Size
:
88.9x63.5x30.0mm (3.50x2.50x1.18 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Pin Material
:
Copper Alloy with Gold Plate Over Nickel Subplate
Weight
:
310g
E-mail:[email protected]
2014/08/07 REV:1
Tel:886-6-2923150
Page 3 of 5
dimensions in mm (inches)
►Tolerance: ±0.5
►Pin
(±0.02)
diameter  1.0 ±0.1 (0.04±0.004)
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
Package Specifications Chassis Mounting (order code suffix C)
Mechanical Dimensions
Connections
Indication LED
50.0 [1.97]
7
6
5
4
3
2
Top view
63.8 [2.51]
4 x 3.5
1
DCON
POWER "GOOD" INDICATOR
Pin
Single Output
Dual Output
1
AC (N)
AC (N)
2
AC (L)
AC (L)
3
+Vout
+Vout
4
NC
NC
5
-Vout
Common
6
NC
NC
7
NC
-Vout
100.0 [3.94]
112.0 [4.41]
NC: No Connection
92.0 [3.62]
16.8 [0.66]
5.0 [0.20]
34.1 [1.34]
10.0 [0.39]
Physical Characteristics
Case Size
:
112.0x63.8x34.1mm (4.41x2.51x1.34 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Weight
:
320g
Package Specifications with DIN Rail Mounting Bracket
Mechanical Dimensions
7
6
5
4
3
Indication LED
Top view
2
1
63.8 [2.51]
3.5
50.0 [1.97]
4x
DCON
POWER "GOOD" INDICATOR
100.0 [3.94]
112.0 [4.41]
34.1 [1.34]
11.0 [0.43]
3.0 [0.12]
5.0
[0.20]
10.0 [0.39]
16.8 [0.66]
92.0 [3.62]
Physical Characteristics
Case Size
:
112.0x63.8x34.1mm (4.41x2.51x1.34 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Weight
:
374g
E-mail:[email protected]
2014/08/07 REV:1
Tel:886-6-2923150
Page 4 of 5
►All
dimensions in mm (inches)
►Tolerance: ±0.5
(±0.02)
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
DIN-Rail Mounting Bracket (Order code for Kit : AC-DIN-02)
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: [email protected]
Minmax Technology Co., Ltd.
2014/08/07 REV:1
Page 5 of 5