MENB1020 Series UL Certificate

Issue Date:
2012-11-30
Page 1 of 15
Report Reference #
E302267-A57-UL
UL TEST REPORT AND PROCEDURE
Standard:
Certification Type:
CCN:
Product:
Model:
Rating:
Applicant Name and Address:
ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10)(Medical Electrical
Equipment - Part 1: General Requirements for Basic Safety and
Essential Performance)
CAN/CSA-C22.2 No. 60601-1 (2008) (Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential
Performance)
Component Recognition
QQHM2, QQHM8 (Power Supplies, Medical and Dental)
Medical Power Supply
BP(a)020(b)(c)(e)(f) and (a)ENB1020(b)(c)(d)(e)(f)
(a) can be A to Z for family related designs.
(b) can be S for single output in model BP(a)020 series and A to Z for
design revision changes in model (a)ENB1020 series.
(c) can be 03 for 3.3Vdc, 05 for 5.0Vdc, 06 for 6.0Vdc, 07 for 7.5Vdc,
09 for 9.0Vdc, 12 for 12Vdc, 14 for 14Vdc, 15 for 15Vdc, 18 for 18Vdc,
24 for 24Vdc or 48 for 48Vdc output voltage.
(d) can be 00 thru 99 for standards output cord options ("(d)" is not
provided in model BP(a)020series).
(e) can be F or N or Q or B or H or G or M or C for input plug type.
See Enclosure-Photographs for each plug-type configuration
F-Class I appliance inlet type: IEC60320-C14
Q-Class II appliance inlet type: IEC60320-C18
N-Class II appliance inlet type: IEC60320-C8
B or C-Class I & Class II direct-plug-in for North America, China,
Japan and Argentina (Changeable Direct-plug-in type is only used for
Class II)
H-Class I & Class II direct-plug-in for Australia (AS/NZS 3112)
G-Class I & Class II direct-plug-in for British (BS 1364)
M-Class I & Class II direct-plug-in for European (CEE /16)] & Korea.
(f) can be 00 thru 99 for customer options, not related safety concerns.
Rated Input; 100-240 Vac, 50-60 Hz, 0.5 A(0.5 A-0.3 A)
Rated Output;
3.3Vdc, 3.0A or 5Vdc, 3.0A or 5Vdc, 2.4A or 6Vdc, 2.5A or 7.5Vdc,
2.0A or 9Vdc, 2.0A or 9Vdc, 1.5A or 12Vdc, 1.5A or 14Vdc, 1.2A or
15Vdc, 1.2A or 18Vdc, 1.0A or 24Vdc, 0.75 or 48Vdc, 0.4A or 48Vdc,
0.35A.
(Rated output voltage is designated in the model name designation
system).
BRIDGEPOWER CORP
964 GOSAEK-DONG
GWONSEON-GU
SUWON-SI GYEONGGI-DO 441-813 KOREA
Copyright © 2012
Issue Date:
2012-11-30
Page 2 of 15
Report Reference #
E302267-A57-UL
This is to certify that representative samples of the products covered by this Test Report have been investigated in accordance with the
above referenced Standards. The products have been found to comply with the requirements covering the category and the products are
judged to be eligible for Follow-Up Service under the indicated Test Procedure. The manufacturer is authorized to use the UL Mark on
such products which comply with this Test Report and any other applicable requirements of UL LLC ('UL') in accordance with the FollowUp Service Agreement. Only those products which properly bear the UL Mark are considered as being covered by UL's Follow-Up
Service under the indicated Test Procedure.
The applicant is authorized to reproduce the referenced Test Report provided it is reproduced in its entirety.
UL authorizes the applicant to reproduce the latest pages of the referenced Test Report consisting of the first page of the Specific
Technical Criteria through to the end of the Conditions of Acceptability.
Any information and documentation involving UL Mark services are provided on behalf of UL LLC (UL) or any authorized licensee of UL.
Prepared by:
DongGug Cho
Reviewed by: DongYoul Kim
Copyright © 2012
Issue Date:
2012-11-30
Page 3 of 15
Report Reference #
E302267-A57-UL
Supporting Documentation
The following documents located at the beginning of this Procedure supplement the requirements of this Test
Report:
A. Authorization - The Authorization page may include additional Factory Identification Code markings.
B. Generic Inspection Instructions i. Part AC details important information which may be applicable to products covered by this Procedure.
Products described in this Test Report must comply with any applicable items listed unless otherwise
stated in the body of this Test Report.
ii. Part AE details any requirements which may be applicable to all products covered by this Procedure.
Products described in this Test Report must comply with any applicable items listed unless otherwise
stated in the body of each Test Report.
iii. Part AF details the requirements for the UL Certification Mark which is not controlled by the technical
standard used to investigate these products. Products are permitted to bear only the Certification
Mark(s) corresponding to the countries for which it is certified, as indicated in each Test Report.
Issue Date:
2012-11-30
Page 4 of 15
Report Reference #
E302267-A57-UL
Product Description
Products are component power supplies intended to be used as part of Medical Electrical Equipment. This
AC Input Power Supply provides 2MOPP isolation from Primary to Secondary/Enclosure(for Class II
construction) and/or 1MOPP isolation from Primary to Earth (for Class I construction). It contains the mains
transformer with UL Recognized Insulation System.
This product is the AC-DC Adaptor of the switching type power supply, which electronic components are
mounted on PWB and housed in plastic enclosure and provided with appliance inlet.
Model Differences
The BP-series is the base model. Model ENB1020-series is identical to the base model BP-series except for
the model type designations.
The below information is nomenclature detail for BP(a)020(b)(c)(e)(f) and (a)ENB1020(b)(c)(d)(e)(f):
(a) can be A to Z for family related designs.
(b) can be S for single output in model BP(a)020 series and A to Z for design revision changes in model
(a)ENB1020 series.
(c) can be 03, 05, 06, 07, 09, 12, 14, 15, 18, 24 or 48 for output voltage.
(d) can be 00 thru 99 for standards output cord options ("(d)" is not provided in model BP(a)020series).
(e) can be F or N or Q or B or H or G or M or C for input plug type.
See Enclosure-Photographs for each plug-type configuration
F-Class I appliance inlet type: IEC60320-C14
Q-Class II appliance inlet type: IEC60320-C18
N-Class II appliance inlet type: IEC60320-C8
B or C-Class I & Class II direct-plug-in for North America, China, Japan and Argentina (Changeable Directplug-in type is only used for Class II)
H-Class I & Class II direct-plug-in for Australia (AS/NZS 3112)
G-Class I & Class II direct-plug-in for British (BS 1364)
M-Class I & Class II direct-plug-in for European (CEE /16)] & Korea.
(f) can be 00 thru 99 for customer options, not related safety concerns.
Technical Considerations

Classification of installation and use : Hand-held or Portable

Device type (component/sub-assembly/ equipment/ system) : Component power supply

Intended use (Including type of patient, application location) : To supply regulated power.

Mode of operation : Continuous

Supply connection : Appliance inlet or direct Plug-in type

Accessories and detachable parts included : None

Other options include : None

The product was investigated to the following additional standards:: ANSI/AAMI ES606011:2005/C1:2009 (includes National Differences for USA); CAN/CSA-C22.2 No. 60601-1:08 (includes
National Differences for Canada), EN 60601-1:2006
Issue Date:
2012-11-30
Page 5 of 15
Report Reference #
E302267-A57-UL

The product was not investigated to the following standards or clauses:: Electromagnetic
Compatibility (IEC 60601-1-2) Clause 14, Programmable Electronic Systems Biocompatibility (ISO
10993-1) Patient applied parts clauses: 4.6, 7.2.10, 8.3, 8.5.2, 8.5.5, 8.7.4.7-8.7.4.9, 8.9.1.15 Battery
related clauses: 7.3.3, 15.4.3 Hand Control related clauses: 8.10.4 Oxygen related clauses: 11.2.2
Fluids related clauses: 11.6.2 - 11.6.4 Sterilization clause: 11.6.7 Motor related clauses: 13.2.13.3,
13.4 Heating Elements related clause: 13.2 Flammable Anesthetic Mixtures Protection: Annex G

The degree of protection against harmful ingress of water is:: Ordinary

The following accessories were investigated for use with the product:: No

The mode of operation is:: Continuous

The product is suitable for use in the presence of a flammable anesthetics mixture with air or oxygen
or with nitrous oxide:: No

Scope of Power Supply evaluation defers the following clauses to the be determined as part of the
end product: Clause 7.5 (Safety Signs), Clause 7.9 (Accompanying Documents), Clause 9
(Mechanical Hazard), Clause 10 (Radiation), Clause 14 (PEMS), Clause 16 (ME Systems)

These power supplies have been previously evaluated by UL to UL 60601-1: 1st ed., 2006-04-26
(includes National Differences for USA), CAN/CSA-C22.2 No. 601.1-M90 (R2005) (includes National
Differences for Canada) under UL Test Report No. E302267-A29, and also by UL to UL 60950-1:
2nd ed., 2007 under UL Test Report No. E300305-A54. All tests conducted per 2nd ed. of UL 606011 and UL 60950-1 were considered representative of the corresponding tests required by 3rd ed. of
AAMI ES 60601-1 and CAN/CSA C22.2 NO.60601-1 as stated under Summary of Testing above.

Software is relied upon for meeting safety requirements related to mechanical, fire and shock: No

The product is suitable for use in the presence of a flammable anaesthetics mixture with air or
oxygen or with nitrous oxide: No

The product has been considered for Pollution Degree 2 and Overvoltage Category II.
Engineering Conditions of Acceptability
For use only in or with complete equipment where the acceptability of the combination is determined by UL
LLC. When installed in an end-product, consideration must be given to the following:

Considerations to the applied parts requirement, to be conducted as end-product

Consideration should be given to measuring the temperature on power electronic components and
transformer windings when the power supply is installed in the end-use equipment. The end-use
product shall ensure that the power supply is used within its ratings.
Issue Date:
2012-11-30
Page 6 of 15
Report Reference #
E302267-A57-UL

The output circuits have not been evaluated for direct patient connection (Type B, BF or CF).

The component shall be installed in compliance with the enclosure, mounting, marking, spacing, and
separation requirements of the end use application.

Power supply provides the following MOPP (means of patient protection): 2 MOPP based upon a
rated voltage 198 Vrms and a working voltage 600 Vpk between Primary and Secondary/Enclosure
and 1 MOPP based on a rated voltage 240 Vrms between Primary and Earth.

Temperature, Leakage Current, Protective Earthing, Dielectric Voltage Withstand, and Marking
Legibility tests should be considered as part of the end product evaluation.

The product was submitted and tested for use at the manufacturer's recommended ambient
temperature (Tmra) of 40 °C at Full Load.

Magnetic devices (T1) employ a Class B (130°C) insulation system

The PWB is rated 105°C minimum.

The products were tested on a 15 A branch circuit. If used on a branch circuit greater than this,
additional testing may be necessary.

The end-product evaluation shall ensure that the requirements related to Accompanying Documents,
Clause 7.9 are met.

End product Risk Management Process to include consideration of requirements specific to the
Power Supply.

End product Risk Management Process to consider the need for different orientations of installation
during testing.

Power Supply tested for 48 hours Humidity Preconditioning. End product Risk Management Process
to determine risk acceptability criteria.

End product to determine the acceptability of risk in conjunction to insulation to resistance to heat,
moisture, and dielectric strength.

Temperature Test was conducted without Test Corner due to no heating elements incorporated in
this power supply. End product to determine the acceptability of risk in conjunction to temperature
testing without test corner as part of the power supply.

End product to determine the acceptability of risk in conjunction to the results of Mechanical Testing
conducted.
Issue Date:
2012-11-30
Page 7 of 15
Report Reference #
E302267-A57-UL
Additional Information
N/A
Markings and instructions
Clause Title
Marking or Instruction Details
Company identification
Classified or Recognized company's name, Trade name, Trademark or File
Model
Model number
Supply Connection
Voltage range, ac/dc, phases if more than single phase
Alternating current
Direct current
Supply Frequency
Rated frequency range in hertz
Class II equipment
Power Input
Amps, VA, or Watts
Output
Rated output voltage, power, frequency.
Special Instructions to UL Representative
N/A
Production-Line Testing Requirements
Test Exemptions - The following models are exempt from the indicated test
Model
Grounding Continuity
Dielectric Voltage
Withstand
Patient Circuit Dielectric
Voltage Withstand
All models
Not Except
Not Except
Except
Solid-State Component Test Exemptions - The following solid-state components may be disconnected
from the remainder of the circuitry during either Dielectric Voltage Withstand Test:
Component
N/A
Sample and Test Specifics for Follow-Up Tests at UL
The following tests shall be conducted in accordance with the Generic Inspection Instructions
Plastic Enclosure or Part
N/A
Test
Sample(s)
Test Specifics
Issue Date:
2012-11-30
Page 8 of 15
Report Reference #
E302267-A57-UL
TABLE: List of Critical Components
Object/part or
Description
Manufacturer/
trademark
type/model
technical data
CCN
Marks of
Conformity
Enclosure
Cheil Industries
HN-1064(+)
Overall Sized approx 84.5 by
46.5 by 32mm & 113 by 48 by
33.5mm.Min 2.0 mm
thickness, V-0. A Recognized
Fabricated Part by 'QMMY2'
Manufacturer.
Min.10A, 250V
Min.2.5A, 250V
Rated 250V, 3.15A. Marked as "250V, T3.15AL" on
PWB.
Rated 250V, 3.15A. Marked as "250V, T3.15AL" on
PWB.
Rated 250V, 3.15A. Marked as "250V, T3.15AL" on
PWB.
Rated Min. 470 V, (line-to-line), Overall dimension
14 mm.
Rated Min. 470 V, (line-to-line), Overall dimension
14 mm.
NTC, 5ohm at 25°C.
QMFZ2
UL
AXUT2
AXUT2
JDYX2
UL
UL
UL
JDYX2
UL
JDYX2
UL
VZCA2/8
UL/cUL
VZCA2
UL
Tested in the
equipment
FOWX2
-
FOWX2, FOWX8
UL/cUL
FOWX2, FOWX8
UL/cUL
FOWX2, FOWX8
UL/cUL
Tested in the
equipment
Tested in the
equipment
Tested in the
-
Appliance Inlet-Class I
Appliance Inlet-Class II
Fuse (F1, F2)
Rong Feng Industrial
Rong Feng Industrial
Wickmann-Werke
GMBH
Fuse (F1, F2) - Alternate Save Fusetech Inc.
SS-5
Fuse (F1, F2) - Alternate Hollyland
5ET
Varistor (ZNR4)
(Optional)
Varistor (ZNR4)
(Optional) - Alternate
Thermister (TH1)
Success Electronics
SVR14D471K
AMOTECH CO LTD
INR14D471
Various
Various
Carli
MPX
Pilkor
Sunil
PCX2 335M or
PCX2 337
436D
Tenta
MEX
Various
Various
250V, max.0.1 uF. Marked with X1 or X2. Meets
IEC384-14.
250V, max.0.1 uF. Marked with X1 or X2. Meets
IEC384-14.
250V, max.0.1 uF. Marked with X1 or X2. Meets
IEC384-14.
250V, max.0.1 uF. Marked with X1 or X2. Meets
IEC384-14.
Min.1/8W, 510Kohm.
Various
Various
Min.1/8W, 1.5Mohm.
Bridgepower Corp or
3025517
Core: Ferrite, Coils: Polyurethane wire min130°C.
X-capacitor (CX1) (Line
to Line)
X-capacitor (CX1) (Line
to Line) - Alternate
X-capacitor (CX1) (Line
to Line) - Alternate
X-capacitor (CX1) (Line
to Line) - Alternate
Discharge resistor (PR1,
PR2)
Discharge resistor (PR1,
PR2)-24V & 48V only
Line Filter (LF1)
SS-120
RF-180
392
UL
-
Issue Date:
2012-11-30
Object/part or
Description
Page 9 of 15
Manufacturer/
trademark
Report Reference #
type/model
Wendeng Jeil
Electronics Co Ltd
Y-Capacitors (CY1)
Success Electronics
SE
Y-Capacitors (CY1) Alternate
Y-Capacitors (CY1) Alternate
Y-Capacitors (CY1) Alternate
Y-Capacitors (CY1) Alternate
Y-Capacitors (CY1) Alternate
TDK Corp.
CD
Dongil
DA
Netron tech
AA
Samwha Capacitor
SD
KUNSHAN
TONGFENG
ELECTRONICS
CO LTD
Success Electronics
CT7
SE or SB
TDK Corp.
CD or CS
Dongil
DA or DS
Netron tech
AA or AD
Samwha Capacitor
SD, SC
KUNSHAN
TONGFENG
ELECTRONICS
CO LTD
Various
CT7
Y-Capacitors (CY1,
CY2) - Alternate
Y-Capacitors (CY1,
CY2) - Alternate
Y-Capacitors (CY1,
CY2) - Alternate
Y-Capacitors (CY1,
CY2) - Alternate
Y-Capacitors (CY1,
CY2) - Alternate
Y-Capacitors (CY1,
CY2) - Alternate
Bridge Resistor (R3)
Bridge Resistor (R3, R4) Various
- Alternate
E302267-A57-UL
technical data
CCN
Bobbin: (QMFZ2), Hexian Specialty Chemicals,
type PM9820@, V-0, 150°C. See Enclosure for
details.
250V, max.1000 pF. Marked with Y1. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1. Meets
IEC384-14.
equipment
Marks of
Conformity
FOWX2
UL
FOWX2
UL
FOWX2/8
UL
FOWX2
UL
FOWX2
UL
FOWX2
UL
250V, max.1000 pF. Marked with Y1 or Y2. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1 or Y2. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1 or Y2. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1 or Y2. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1 or Y2. Meets
IEC384-14.
250V, max.1000 pF. Marked with Y1 or Y2. Meets
IEC384-14.
FOWX2
UL
FOWX2
UL
FOWX2
UL
FOWX2
UL
FOWX2
UL
FOWX2
UL
Various
1/4W, 10MOhm
-
Various
1/4W, 4.7MOhm
Tested in the
equipment
Tested in the
equipment
-
Issue Date:
2012-11-30
Page 10 of 15
Report Reference #
E302267-A57-UL
Object/part or
Description
Manufacturer/
trademark
type/model
technical data
CCN
Marks of
Conformity
Bridge diode (BD1)
Various
Various
Min 600V, max. 0.8A.
-
Electrolytic Capacitor
(C1, C2)
Switching IC (U1)
Various
Various
Rated 400 V or 450V, max. 22 uF, min. 105°C.
Various
Various
Max30V, Max4mA.
FET (Q1)
Various
Various
Max 650 V, 4.2A. Secured to HS1 by a screw.
Main Transformer (T1)
for 3.3V output
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685001)
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Tested in the
equipment
Tested in the
equipment
Tested in the
equipment
Tested in the
equipment
-
-
UL
-
UL
UL
Winding: R/C (OBJT2), Triple Insulated Wire, 130
deg C, Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Main Transformer (T1)
for 5 V & 6 Voutput
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685002 or
3025685002S)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Winding: R/C (OBJT2), Triple Insulated Wire, 130
deg C, Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Main Transformer (T1)
for 7.5 V & 9 Voutput
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685003)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Winding: R/C (OBJT2), Triple Insulated Wire, 130
deg C, Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Issue Date:
2012-11-30
Object/part or
Description
Page 11 of 15
Manufacturer/
trademark
Report Reference #
type/model
E302267-A57-UL
technical data
CCN
Marks of
Conformity
-
UL
-
UL
-
UL
Polyurethane wire Min. 130 deg C.
Main Transformer (T1)
for 12 Voutput
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685004)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Winding: R/C (OBJT2), Polyurethanewire Min.
130deg C.Triple Insulated Wire, 130 deg C,
Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Main Transformer (T1)
for 14 V & 15 V output
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685005)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Winding: R/C (OBJT2), Polyurethanewire Min.
130deg C.Triple Insulated Wire, 130 deg C,
Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Main Transformer (T1)
for 18 V output
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685006)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Winding: R/C (OBJT2), Polyurethanewire Min.
130deg C.Triple Insulated Wire, 130 deg C,
Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Issue Date:
2012-11-30
Page 12 of 15
Report Reference #
E302267-A57-UL
Object/part or
Description
Manufacturer/
trademark
type/model
technical data
CCN
Marks of
Conformity
Main Transformer (T1)
for 24 V output
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685007)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
-
UL
-
UL
FPQU2/8
UL/cUL
FPQU2/8
UL/cUL
FPQU2
UL
FPQU2/8
UL/cUL
Tested in the
-
Winding: R/C (OBJT2), Polyurethanewire Min.
130deg C.Triple Insulated Wire, 130 deg C,
Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Main Transformer (T1)
for 48 V output
Bridgepower Corp or
Wendeng Jeil
Electronics Co Ltd
JEC(B) (Part No.:
3025685008)
Bobbin: (QMFZ2), V-0, 130 deg C.
Class B: R/C (OBJY2), Class 130(B) Insulation
System, BridgePower Corp, Designated JEC(B)
Winding: R/C (OBJT2), Polyurethanewire Min.
130deg C.Triple Insulated Wire, 130 deg C,
Furukawa Electric, Cat. No: TEX-E
Core: Ferrite, size 25 by 19mm. Coils:
Polyurethane wire Min. 130 deg C.
Optical Isolator (U2)
Vishay
Semiconductor
TCET1103(G)
Optical Isolator (U2)Alternate
COSMO
ELECTRONICS
CORP
Sharp Corp
KP1010
Optical Isolator (U2)Alternate
Optical Isolator (U2)Alternate
Heatsink(HS1)(For Q1)
PC123
KODENSHI KOREA
PC-17K
PC-17K1C
Various
Various
Bobbin: (QMFZ2), V-0, 130 deg C.
Double protection optical isolator. Providing
isolation voltage 5000 Vac. External dcr min. 8.0
mm.
Double protection optical isolator. Providing
isolation voltage 5000 Vac. External dcr min. 8.0
mm.
Double protection optical isolator. Providing
isolation voltage 5000 Vac.
Double protection optical isolator. Providing
isolation voltage 5000 Vac. External dcr min. 8.0
mm.
Overall Sized Min. ø 18
Issue Date:
2012-11-30
Object/part or
Description
Heatsink(HS2)(For D3)
Page 13 of 15
Manufacturer/
trademark
type/model
Various
Various
Insulation
Various
Sheet(Between T1 & U2)
Various
PWB
Output Cable
Various
Various
Various
Various
Bonding Glue
Various
Various
Nameplate Label
BETHEL
BG-1000
Report Reference #
E302267-A57-UL
technical data
CCN
by 20 mm , Min.
1.5 mm thick
Overall Sized Min. ø 20
by 21 mm , Min.
1.5 mm thick
Overall Sized Min. ø 31
by 20 mm , Min.
0.4 mm thick
Min. V-1, 130°C.
SPT1. 300 V, 80°C, min 20 AWG, VW-1, FT-1.
Max. 3.05 m long
Min. V-2, min. 100 °C for additional secureness of
internal conductor.
Suitable for use on surface of Polycarbonate (PC)
with max.80°C surface temperature.
equipment
Marks of
Conformity
Tested in the
equipment
-
QMFZ2
UL
ZPMV2
AVLV2
UL
UL
QMFZ2
UL
PGDQ2, PGAA
UL
Issue Date:
2012-11-30
Page 14 of 15
Report Reference #
E302267-A57-UL
Enclosures
Type
Supplement Id
Description
Collateral
Particular
Photographs
3-04
External View (Changeable direct plug-in type for Australia)
Photographs
3-05
External View of B Type For China (See the nomenclature of
model designation)
Photographs
3-06
External View of B Type for North America - Class II (See the
nomenclature of model designation)
Photographs
3-07
External View of H Type - Class II (See the nomenclature of model
designation)
Photographs
3-08
External View of G Type - Class I (See the nomenclature of model
designation)
Photographs
3-09
External View of M Type For Europe - Class II(See the
nomenclature of model designation)
Photographs
3-10
External View of N Type 1 - Class II(See the nomenclature of
model designation)
Photographs
3-11
External View of Q Type - Class II (See the nomenclature of model
designation)
Photographs
3-12
External View (Changeable direct plug-in type for UK)
Photographs
3-13
External View of F Type - Class I (See the nomenclature of model
designation)
Photographs
3-14
Internal View of F Type - Class I (See the nomenclature of model
designation)
Photographs
3-15
Internal View of F Type - Class I (See the nomenclature of model
designation)
Photographs
3-16
External View of POF B type
Photographs
3-17
Internal top view of POE A type
Photographs
3-18
Internal bottom view of POE A type
Photographs
3-19
Internal top view of POE B type
Photographs
3-20
Internal top bottom of POE B type
Schematics + PWB
5-03
Circuit Diagram
Schematics + PWB
5-04
Pattern Diagram
Miscellaneous
7-01
Marking Plate_1
Miscellaneous
7-02
Marking Plate_2
Diagrams
Manuals