UL TEST REPORT AND PROCEDURE

Issue Date:
2011-03-04
Page 1 of 12
Report Reference #
E302267-A32-UL
UL TEST REPORT AND PROCEDURE
Standard:
Certification Type:
CCN:
Product:
Model:
Rating:
Applicant Name and Address:
UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment,
Part 1: General Requirements for Safety)
CAN/CSA-C22.2 No. 601.1-M90, 2005 (Medical Electrical Equipment Part 1: General Requirements for Safety)
Component Recognition
QQHM2, QQHM8 (Power Supplies, Medical and Dental)
AC-DC Adaptor
BP(a)040(b)(c)(d)(e)
(a) can be A to Z for family related designs.
(b) can be S for single.
(c) can be 05, 09, 12, 15, 18, 24 or 48 for output voltage.
(d) can be F or N or Q for input plug type.
(e) can be 00 thru 99 for customer options, not related safety
concerns.
(a)ENB1040(b)(c)(d)(e)(f)
(a) can be A to Z for family related designs.
(b) can be A to Z for design revision changes.
(c) can be 05, 09, 12, 15, 18, 24 or 48 for output voltage.
(d) can be 00 thru 99 for standard output cord options.
(e) can be F or N or Q for input plug type.
(f) can be 00 thru 99 for customer options, not related safety concerns.
Rated Input; 100-240Vac, 50-60Hz, 1.2 A.
Rated Output; 5Vdc, 5.0A or 9Vdc, 4.0A or 12Vdc, 3.4A or 15Vdc,
2.7A or 18Vdc, 2.2A or 24Vdc, 1.7A or 48Vdc, 0.83A. Rated output
voltage is designated in the model name designation system).
BRIDGEPOWER CORP
964 GOSAEK-DONG
GWONSEON-GU
SUWON-SI GYEONGGI-DO 441-813 KOREA
This is to certify that representative samples of the products covered by this Test Report have been investigated in accordance with the
above referenced Standards. The products have been found to comply with the requirements covering the category and the products are
judged to be eligible for Follow-Up Service under the indicated Test Procedure. The manufacturer is authorized to use the UL Mark on
such products which comply with this Test Report and any other applicable requirements of Underwriters Laboratories Inc. ('UL') in
accordance with the Follow-Up Service Agreement. Only those products which properly bear the UL Mark are considered as being
covered by UL's Follow-Up Service under the indicated Test Procedure.
The applicant is authorized to reproduce the referenced Test Report provided it is reproduced in its entirety.
UL authorizes the applicant to reproduce the latest pages of the referenced Test Report consisting of the first page of the Specific
Technical Criteria through to the end of the Conditions of Acceptability.
Any information and documentation involving UL Mark services are provided on behalf of Underwriters Laboratories Inc. (UL) or any
authorized licensee of UL.
Copyright © 2011
Issue Date:
Prepared by:
Reviewed by:
2011-03-04
Page 2 of 12
KiSoo Ahn
Underwriters Laboratories Inc.
KiSoo Ahn
Underwriters Laboratories Inc.
Copyright © 2011
Report Reference #
E302267-A32-UL
Issue Date:
2011-03-04
Page 3 of 12
Report Reference #
E302267-A32-UL
Supporting Documentation
The following documents located at the beginning of this Procedure supplement the requirements of this Test
Report:
A. Authorization - The Authorization page may include additional Factory Identification Code markings.
B. Generic Inspection Instructions i. Part AC details important information which may be applicable to products covered by this Procedure.
Products described in this Test Report must comply with any applicable items listed unless otherwise
stated in the body of this Test Report.
ii. Part AE details any requirements which may be applicable to all products covered by this Procedure.
Products described in this Test Report must comply with any applicable items listed unless otherwise
stated in the body of each Test Report.
iii. Part AF details the requirements for the UL Certification Mark which is not controlled by the technical
standard used to investigate these products. Products are permitted to bear only the Certification
Mark(s) corresponding to the countries for which it is certified, as indicated in each Test Report.
Issue Date:
2011-03-04
Page 4 of 12
Report Reference #
E302267-A32-UL
Product Description
Switching type power supply, which electronic components are mounted on PWB and housed in plastic
enclosure and provided with appliance inlet. In addition, some series of power supply are the direct plug-in
units. See "model differences" for details.
Model Differences
- The BPM040 series is the basic model. Model (a)ENB1040 series is identical to the basic model series
except for the model type designations.
- The marked models as 05, 09, 12, 15, 18, 24 and 48 on nomenclature are similar, except for Main
transformer (T1) windings, output rating and minor secondary circuit options not affecting safety.
-For each plug-type configuration
F-Class I appliance inlet type: IEC60320-C14,
Q-Class II appliance inlet type: IEC60320-C18,
N-Class II appliance inlet type: IEC60320-C8.
Technical Considerations

Classification of installation and use : Portable

Supply connection : Appliance Inlet

Accessories and detachable parts included in the evaluation : None

Options included : None

The product was investigated to the following additional standards:: EN 60601-1: 1990 + A1:1993 +
A2:1995

The product was not investigated to the following standards or clauses:: Clause 36, Electromagnetic
Compatibility (IEC 60601-1-2), Clause 48, Biocompatibility (ISO 10993-1), Clause 52.1,
Programmable Electronic Systems (IEC 60601-1-4)

The product is Classified only to the following hazards:: Casualty, Fire, Shock

The degree of protection against harmful ingress of water is:: Ordinary

The mode of operation is:: Continuous

Software is relied upon for meeting safety requirements related to mechanical, fire and shock:: No

The product is suitable for use in the presence of a flammable anesthetics mixture with air or oxygen
or with nitrous oxide:: No
Engineering Conditions of Acceptability
Issue Date:
2011-03-04
Page 5 of 12
Report Reference #
E302267-A32-UL
For use only in or with complete equipment where the acceptability of the combination is determined by
Underwriters Laboratories Inc. When installed in an end-product, consideration must be given to the
following:

Unit has not been evaluated for use in or likely to be used in the patient vicinity.

Instructions shall be considered in the end product application.

Marking shall be considered in the end product application.

Leakage current measurements shall be considered in the end-use product.

Unit was evaluated for use in a 40° C ambient.
Additional Information
Max. Normal Load Condition: Rated output current
Markings and instructions
Clause Title
Marking or Instruction Details
Company identification
Classified or Recognized company's name, Trade name, Trademark or File
Model
Model number
Supply Connection
Voltage range, ac/dc, phases if more than single phase
Alternating current
Supply Frequency
Rated frequency range in hertz
Power Input
Amps, VA, or Watts
Class II equipment
Output
Rated output voltage, power, frequency.
Attention, consult
accompanying
documents
Special Instructions to UL Representative
Samples specified as below shall be selected from current production or stock for the tests indicated. Product
Name: BPMseries and MENB series, CCL Item: Transformer (T1), Dielectric Voltage Withstand: Primary lead
to secondary lead with 4000 V ac for 1 min.
Issue Date:
2011-03-04
Page 6 of 12
Report Reference #
E302267-A32-UL
Production-Line Testing Requirements
Test Exemptions - The following models are exempt from the indicated test
Model
Grounding Continuity
Dielectric Voltage
Withstand
Patient Circuit Dielectric
Voltage Withstand
All models
Not Exempt
Not Exempt
Exempt
Solid-State Component Test Exemptions - The following solid-state components may be disconnected
from the remainder of the circuitry during either Dielectric Voltage Withstand Test:
N/A
Sample and Test Specifics for Follow-Up Tests at UL
The following tests shall be conducted in accordance with the Generic Inspection Instructions
Model
Samples
Test
Test Details
N/A
N/A
N/A
N/A
Issue Date:
2011-03-04
Page 7 of 12
Report Reference #
E302267-A32-UL
TABLE: List of Critical Components
Object/part or
Description
Manufacturer/
trademark
Appliance Inlet (Class I)
Rong Feng Industrial
Co., Ltd.
Appliance Inlet (Class II) Rong Feng Industrial
Co., Ltd.
Appliance Inlet (Class II) Rong Feng Industrial
- Alternate
Co., Ltd.
Enclosure(Fire/Mech./El CHEIL INDUSTRIES.
ec.)
Fuse (F1)
Fuse (F1)
- Alternate
Fuse (F1)
- Alternate
Fuse (F1)
- Alternate
Fuse (F2)
Fuse (F2)
- Alternate
Fuse (F2)
- Alternate
Fuse (F2)
- Alternate
Varistor (ZNR4)
(Optional)
Varistor (ZNR4)
(Optional) - Alternate
Varistor (ZNR4)
(Optional) - Alternate
type/model
technical data
CCN
Marks of
Conformity
SS-120
Rated 15 A / 250 V.
AXUT2/8
USR/CNR
RF-180
Rated 2.5 A / 250 V.
AXUT2/9
USR/CNR
SS-120A
Rated 15 A / 250 V.
AXUT2/8
USR/CNR
HN-1064W(+)
Overall Sized approx. 102.0 by 60.0 by 32.5mm.
Min 2.0mm thickness, V-1, RTI 80 degree C.
Composed of two pieces, secured together by
ultrasonic welding.
Rated 250V, T3.15AL
Rated 250V, T3.15AL
QMFZ2/8
USR/CNR
JDYX2/8
JDYX2/8
USR/CNR
USR/CNR
Rated 250V, T3.15AL
JDYX2/8
USR/CNR
Save fusetech Inc
SS-5
LITTELFUSE
392
WICKMANN WERKE
Hollyland co., ltd.
5ET
NIPPON SEISEN
CABLE, LTD.
Save fusetech Inc
LITTELFUSE
WICKMANN WERKE
Hollyland co., ltd.
SLT
Rated 250V, T3.15AL
JDYX2/8
USR/CNR
SS-5
392
Rated 250V, T3.15AL
Rated 250V, T3.15AL
JDYX2/8
JDYX2/8
USR/CNR
USR/CNR
5ET
Rated 250V, T3.15AL
JDYX2/8
USR/CNR
NIPPON SEISEN
CABLE, LTD.
Success Electronics
SLT
Rated 250V, T3.15AL
JDYX2/8
USR/CNR
SVR14D471K
VZCA2/8
USR/CNR
AMOTECH CO LTD
INR14D471
Rated 470 V, (line-to-line), Overall dimension 14
mm.
Rated 470 V, (line-to-line), Overall dimension 14
mm.
Rated 470 V, (line-to-line), Overall dimension 14
mm.
VZCA2
USR
VZCA2
USR
HONGZHI
HEL14D471K
ENTERPRISES LTD
Issue Date:
2011-03-04
Page 8 of 12
Report Reference #
E302267-A32-UL
Object/part or
Description
Manufacturer/
trademark
type/model
technical data
CCN
Marks of
Conformity
Thermistor (TH1)
Various
Various
NTC, 5 ohm or 10 ohm at 25 degree C
-
Line Filter (LF1)
Bridgepower or
Wendeng Jeil
3025634
X-capacitor (CX1) (Line
to Line)
X-capacitor (CX1) (Line
to Line) - Alternate
X-capacitor (CX1) (Line
to Line) - Alternate
X-capacitor (CX1) (Line
to Line) - Alternate
X-capacitor (CX1) (Line
to Line) - Alternate
X-capacitor (CX1) (Line
to Line) - Alternate
BridgeDiode(BD1)
Carli
MPX
FOWX2
USR/CNR
Iskra
FOWX2
USR/CNR
FOWX2
USR/CNR
Okaya
KNB 1530 or
1562 or 1563
PCX2 335M or
PCX2 337
LE
FOWX2
USR/CNR
Sunil
436D
FOWX2
USR/CNR
Tenta
MEX
FOWX2
USR/CNR
Various
Various
Core: Ferrite,
Coils: Polyurethane Enamelled Wire
Min.130degree C. Bobbin: (QMFZ2) Bakelite, type
PM9820@, V-0, 150 deg C
Rated 250V, Max.. 0.47 uF. Marked with X1 or X2.
Meets IEC/EN60384-14.
Rated 250V, Max.. 0.47 uF. Marked with X1 or X2.
Meets IEC/EN60384-14.
Rated 250V, Max.. 0.47 uF. Marked with X1 or X2.
Meets IEC/EN60384-14.
Rated 250V, Max.. 0.47 uF. Marked with X1 or X2.
Meets IEC/EN60384-14.
Rated 250V, Max.. 0.47 uF. Marked with X1 or X2.
Meets IEC/EN60384-14.
Rated 250V, Max.. 0.47 uF. Marked with X1 or X2.
Meets IEC/EN60384-14.
Min. 4.0A, Min. 600V.
Tested in the
equipment
Tested in the
equipment
-
Various
Various
510Kohm 1/8W
Tested in the
equipment
Tested in the
equipment
Tested in the
equipment
Tested in the
equipment
Tested in the
equipment
Tested in the
equipment
Pilkor
Discharging Resister
(PR1, PR2)
Electrolytic Capacitor
(C1)
Switching IC (U1)
Various
Various
Max. 82 uF, Min. 400 V, min. 105 degree.
Various
Various
Max 30V, Max.4mA
FET (Q1)
Various
Various
Min 650V, Max.7.2A
Main Transformer(T1)
for output 5 V
Bridgepower or
Wendeng Jeil
3025688001
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
-
-
Issue Date:
2011-03-04
Page 9 of 12
Report Reference #
Object/part or
Description
Manufacturer/
trademark
type/model
Main Transformer(T1)
for output 9V
Bridgepower or
Wendeng Jeil
3025688002
Main Transformer(T1)
for output 12 V
Bridgepower or
Wendeng Jeil
Main Transformer(T1)
for output 15V
Bridgepower or
Wendeng Jeil
Main Transformer(T1)
for output 18 V
Bridgepower or
Wendeng Jeil
Main Transformer(T1)
for output 24 V
Bridgepower or
Wendeng Jeil
Main Transformer(T1)
for output 48 V
Bridgepower or
Wendeng Jeil
Optical Isolator (U2)
Vishay
Semiconductor
COSMO
ELECTRONICS
CORP
SHARP CORP
ELECTRONIC
COMPONENTS
Optical isolator (U2) Alternate
Optical isolator (U2) Alternate
E302267-A32-UL
technical data
CCN
Marks of
Conformity
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
3025688003
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
3025688004
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
3025688005
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
3025688006
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
3025688007
JEC(B), Class B. Core: Ferrite, size 28 by 20 mm.
Coils: TIW:Furukawa, TEXE, 130 deg
C.Bobbin:(QMFZ2), V-0, 130 degC. see Enclosure
for details.
TCET1103(G)D or Double protection optical isolator. Providing
TCET1103
isolation voltage 5000 Vac
KP1010X
Double protection optical isolator. Providing
isolation voltage 5000 Vac
Tested in the
equipment
-
Tested in the
equipment
-
Tested in the
equipment
-
Tested in the
equipment
-
Tested in the
equipment
-
Tested in the
equipment
-
FPQU2/8
USR/CNR
FPQU2/8
USR/CNR
PC123
FPQU2
US
Double protection optical isolator. Providing
isolation voltage 5000 Vac
Issue Date:
2011-03-04
Object/part or
Description
Optical isolator (U2) Alternat
Page 10 of 12
Manufacturer/
trademark
GROUP
LITE-ON
TECHNOLOGY
CORP
KODENSHI KOREA
Optical isolator (U2) Alternate
Bridging Capacitor (CY1, SUCCESS
CY2)
ELECTRONICS CO
LTD
PWB
Various
Heatsink(HS1)-Primary Various
E302267-A32-UL
type/model
technical data
CCN
Marks of
Conformity
LTV-817
Double protection optical isolator. Providing
isolation voltage 5000 Vac
FPQU2/8
USR/CNR
PC-17K
Double protection optical isolator. Providing
isolation voltage 5000 Vac.
250Vmin, 1000 pF. Marked with Y1 & Y2.
FPQU2/8
USR/CNR
FOWX2
USR
ZPMV2
Tested in the
equipment
USR
-
Tested in the
equipment
-
AVLV2 or ZJCZ
USR
PGDQ2, PGAA
USR
QMFZ2
USR
AVLV2
USR
AVLV2
USR
SE & CB
Various
Various
Heatsink(HS2)Secondary
Various
Various
Output cable(LPS)
Various
Various
Nameplate Label
BETHEL
BG-1000
Bonding Glue
Various
Various
Wiring, internal (Primary) Various
Various
Bonding conductor
Various
Various
Report Reference #
Min. V-1, min 130°C.
Metal, overall sized approx.
54 by 22 mm,
2 mm thickness.
Wound by polyester film
tape OANZ2), Min. 130 deg C,
Metal, overall sized approx.
35 by 22 mm,
2 mm thickness.
Wound by polyester film
tape OANZ2), Min. 130 deg C,
For use of external interconnection, max 3.05 m
long, max. 300 V, 80 °C, min.20 AWG, Marked with
VW-1 or FT-1.
Suitable for use on surface of Polycarbonate (PC)
with max. 60 °C surface temperature.
Min. V-2, min. 100 °C for additional secureness of
internal conductor.
Max. 300 V, 80 °C, min.20 AWG, Marked with VW1 or FT-1.
Mechanically clamped or secured on PWB from
Appliance Inlet. Min 18 AWG, Green-and-Yellow
Insulation.
Issue Date:
2011-03-04
Page 11 of 12
Report Reference #
E302267-A32-UL
Issue Date:
2011-03-04
Page 12 of 12
Report Reference #
E302267-A32-UL
Enclosures
Type
Supplement Id
Description
Collateral
Particular
Photographs
3-45
External view of F-Class I appliance inlet type: IEC60320-C14.
Photographs
3-46
External view of N-Class II appliance inlet type: IEC60320-C8.
Photographs
3-47
External view of Q-Class II appliance inlet type: IEC60320-C18.
Schematics + PWB
5-06
Circuit Diagram
Schematics + PWB
5-07
PWB Pattern
Miscellaneous
7-02
T1 transformer Spec(For 5Vdc output)
Miscellaneous
7-09
T1 transformer Spec(For 9Vdc output)
Miscellaneous
7-10
T1 transformer Spec(For 12Vdc output)
Miscellaneous
7-11
T1 transformer Spec(For 15Vdc output)
Miscellaneous
7-12
T1 transformer Spec(For 18Vdc output)
Miscellaneous
7-13
T1 transformer Spec(For 24Vdc output)
Miscellaneous
7-14
T1 transformer Spec(For 48Vdc output)
Miscellaneous
7-16
Enclosure drawing
Miscellaneous
7-34
Marking plate
Diagrams
Manuals