installation instructions mint1065b single output series

SL Power Electronics, Corp..
6050 King Dr. Bldg. A
VENTURA, CA 93003 + 805-486-4565
MODEL NUMBERS:
INSTALLATION INSTRUCTIONS
MINT1065B SINGLE OUTPUT SERIES
MINT1065BVWXYZ Series.
down as follows:
Sample Model # is MINT1065B1275C01 which breaks
MINT
1
065
A
12
75 C 01
1
2
3
4
5
6
7
8
1 – Medical Internal Model # Prefix
2 – Signifies # of outputs: i.e. 1, 2, 3, 4, etc.
3 – Output Wattage: 065 is 65W
4 – Signifies Generational Differences such as Class I, ClassII or Energy Star level changes. May be any Letter
A-Z. Where “A” represents Class I Power Supplies and “B” represents Class II Power Supplies.
5 – Output Voltage: Numeric indicator from 12 to 48. i.e. 12 = 12Vdc
6 – Output Connector options. 75 = 4 pin Tyco/AMP MTA156 or equivalent
7 – Input Connector options. C = 2 pin Tyco/Amp MTA156 or equivalent.
8 – Configuration: 01=Standard. 02-99 for modifications.
Unit is RoHS Compliant.
RATINGS
Input: 100-240 VAC, 1.3 A, 50-60 Hz (Korea input is 110V~, 1.27A)
Output: 12Vdc to 48Vdc at 5.25A to 1.35A.
:
Model
MINT1065B 12
MINT1065B 15
MINT1065B 18
MINT1065B BASE MODELS AND RATINGS
Watts
Output
Model
Watts
63
12V/5.25A
MINT1065B 24
65
65
15V/4.33A
MINT1065B 48
65
63
18V/3.5A
Output
24V/2.7A
48V/1.35A
Notes:
1. Maximum operating ambient is 40°C.
2. Maximum Relative Humidity 96%, no condensation.
CERTIFICATION:
CLASSIFICATION:
(In accordance with subclause 5 of IEC 60601-1)
All models are certified to be in compliance with the applicable requirements of
UL/CSA/EN/IEC 60601-1.
(5.1) Protection against electric shock = Class II
(5.2) Degree of protection against electric shock = Signal output or intermediate
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anesthetic mixture
with air, oxygen, or nitrous oxide. This evaluation is to be made on the end
equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION: SL Power Electronics Corp (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable requirements of
EN60950 following the provisions of the Low Voltage Directive 2006/95/EC.
GROUNDING: Protection Class II requires that the Input remains ungrounded in the end application. .
Page 1 of 2
41-36127-0002-1
OUTPUTS: The output must not be connected to Earth ground or to metallic enclosure in the end application.
The output has not been evaluated for direct patient connection (consult factory for details).
SELV under normal and single fault conditions.
The DC output is
OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety standards,
redundant or additional overvoltage protection may be required. Consult factory for details.
CAUTION: When performing Dielectric Strength Tests, catastrophic failure of the unit may result if a
Dielectric Strength test voltage greater than 1800 V ac is applied between primary and secondary circuits. The
components providing isolation from primary to secondary cannot be tested while installed in the power supply
without overstressing basic (primary to ground) insulation. All isolating components are individually 100 %
tested at 4800 V ac prior to installation.
ISOLATION: The creepage distance between primary and secondary circuits is 8 mm minimum. The
required creepage and clearance distances from primary circuits to secondary circuits must be maintained after
installation to preserve the intended safety. This may require the use of plastic attaching hardware and standoffs
in the end application. The use of an insulator between the PWB Bottom and a metal chassis should be
considered.
TEMPERATURES: The maximum operating temperatures of certain safety components, as defined in the
applicable safety standards, must not be exceeded after installation to preserve the intended safety. The output
power, ambient air temperature and the availability, amount, direction and/or restriction of airflow influence the
temperatures of these components.
OVERCURRENT PROTECTION: The MINT1065B series has dual internal fusing as required by IEC/EN/UL
60601-1.
WARNING! RISK OF FIRE! A blown internal fuse is an indication of catastrophic failure of circuit
component(s). Refer to fuse marking on the supply for rating. Repair must be performed by SLPE authorized
personnel.
WARNING! SHOCK HAZARD!
and heatsinks.
Dangerous voltages are present on some components, printed wiring traces
CONNECTIONS
EXPLANATION OF SYMBOLS
Alternating Current
Direct Current
Attention, Consult
Accompanying Documents
J1 Pin
1
3
AC Input
Line
Neutral
J2 Pin
1
2
3
4
DC Output
Output #1 (+)
Output #1 (+)
Return
Return
Attention, Dangerous Voltages
Class II
RECOMMENDED MATING CONNECTORS
J1 AMP 640250-3 (center pin not used)
J2 AMP 640250-4
CAUTION
Do not exceed 5 Amps per contact
SLPE will not be liable for the safety, reliability or performance of these power supplies if a) any changes, modifications or repairs are carried out by other
than authorized agents of SLPE, or b) the installation of the supply is not in accordance with these installation instructions and the applicable UL, CSA, and
EN/IEC safety standards.
41-36127-0002-1