installation instructions mint1065a single output series

SL Power Electronics, Corp..
6050 King Dr. Bldg. A
VENTURA, CA 93003 + 805-486-4565
MODEL NUMBERS:
INSTALLATION INSTRUCTIONS
MINT1065A SINGLE OUTPUT SERIES
MINT1065BVWXYZ Series.
down as follows:
Sample Model # is MINT1065B1275C01 which breaks
MINT
1
065
A
12
75 C 01
1
2
3
4
5
6
7
8
1 – Medical Internal Model # Prefix
2 – Signifies # of outputs: i.e. 1, 2, 3, 4, etc.
3 – Output Wattage: 065 is 65W
4 – Signifies Generational Differences such as Class I, Class II or Energy Star level changes. May be any Letter
A-Z. Where “A” represents Class I Power Supplies and “B” represents Class II Power Supplies.
5 – Output Voltage: Numeric indicator from 12 to 48. i.e. 12 = 12Vdc
6 – Output Connector options. 75 = 4 pin Tyco/AMP MTA156 or equivalent
7 – Input Connector options. C = 2 pin Tyco/Amp MTA156 or equivalent.
8 – Configuration: 01=Standard. 02-99 for modifications.
Unit is RoHS Compliant.
RATINGS
Input: 100-240 VAC, 1.3 A, 50-60 Hz (Korea input is 110V~, 1.27A)
Output: 12Vdc to 48Vdc at 5.25A to 1.35A.
Model
MINT1065A 12
MINT1065A 15
MINT1065A 18
MINT1065A BASE MODELS AND RATINGS
Watts
Output
Model
Watts
63
12V/5.25A
MINT1065A 24
65
65
15V/4.33A
MINT1065A 48
65
63
18V/3.5A
Output
24V/2.7A
48V/1.35A
Notes:
1. Maximum operating ambient is 40°C.
2. Maximum Relative Humidity 96%, no condensation.
CERTIFICATION:
CLASSIFICATION:
(In accordance with subclause 5 of IEC 60601-1)
All models are certified to be in compliance with the applicable requirements of
UL/CSA/EN/IEC 60601-1.
(5.1) Protection against electric shock = Class I
(5.2) Degree of protection against electric shock = Signal output or intermediate
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anesthetic mixture
with air, oxygen, or nitrous oxide. This evaluation is to be made on the end
equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION: SL Power Electronics Corp (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable requirements of
EN60950 following the provisions of the Low Voltage Directive 2006/95/EC.
GROUNDING: The Protective Earth (ground) terminal must be bonded to Protective Earth in the host
equipment. Using the Protective Earth terminal on the supply for grounding the host equipment is not
recommended. A separate dedicated grounding point should be used.
Page 1 of 2
41-36127-0001 Rev. 3
OUTPUTS: Output common should be connected to Protective Earth in the end application. The output is
intended for Protectively Earthed Signal Output and Intermediate Circuits only. The output is not acceptable for
patient connection without additional isolation. The DC output is SELV under normal and single fault
conditions.
OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety standards,
redundant or additional overvoltage protection may be required. Consult factory for details.
CAUTION: When performing Dielectric Strength Tests, catastrophic failure of the unit may result if a
Dielectric Strength test voltage greater than 1800 V ac is applied between primary and secondary circuits. The
components providing isolation from primary to secondary cannot be tested while installed in the power supply
without overstressing basic (primary to ground) insulation. All isolating components are individually 100 %
tested at 4800 V ac prior to installation.
ISOLATION: The creepage distance between primary and ground is 4 mm minimum; between primary and
secondary circuits is 8 mm minimum. Secondary to ground creepage is not defined or controlled. The output
common is bypassed to ground using a 1000 pF 400V Y2 Type capacitor. The required creepage and clearance
distances from primary circuits to ground and secondary circuits must be maintained after installation to preserve
the intended safety.
TEMPERATURES: The maximum operating temperatures of certain safety components, as defined in the
applicable safety standards, must not be exceeded after installation to preserve the intended safety. The output
power, ambient air temperature and the availability, amount, direction and/or restriction of airflow influence the
temperatures of these components.
OVERCURRENT PROTECTION: The MINT1065A series has dual internal fusing as required by IEC/EN/UL
60601-1.
WARNING! RISK OF FIRE! A blown internal fuse is an indication of catastrophic failure of circuit
component(s). Refer to fuse marking on the supply for rating. Repair must be performed by SLPE authorized
personnel.
WARNING! SHOCK HAZARD!
and heatsinks.
Dangerous voltages are present on some components, printed wiring traces
CONNECTIONS
EXPLANATION OF SYMBOLS
Alternating Current
Direct Current
Attention, Consult
Accompanying Documents
J1 Pin
1
3
AC Input
Line
Neutral
J2 Pin
1
2
3
4
DC Output
Output #1 (+)
Output #1 (+)
Return
Return
Attention, Dangerous Voltages
Earth (Ground)
RECOMMENDED MATING CONNECTORS
J1 AMP 640250-3 (center pin not used)
J2 AMP 640250-4
CAUTION
Do not exceed 5 Amps per contact
SLPE will not be liable for the safety, reliability or performance of these power supplies if a) any changes, modifications or repairs are carried out by other
than authorized agents of SLPE, or b) the installation of the supply is not in accordance with these installation instructions and the applicable UL, CSA, and
EN/IEC safety standards.
41-36127-0001 Rev. 3