ME60 Series Datasheet

ME60
60W Single Output External Power
Medical Series
Features

Meets UL/EN/IEC60601-1-2, 4th edition for
EMC

Approved to EN/IEC/UL60601-1, 3rd edition
and EN/IEC/UL60601-1-11:2010 for Home
Healthcare (non-hospital use) applications with
and isolation levels which satisfy the 2 MOPP
requirements

Meets DoE Efficiency Level VI Requirements
 No load input power
 Average Efficiency

Up to 60W of AC-DC Power

Universal Input 90-264Vac Input Range

Desktop Style Package

Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin

E-cap life of >7 years

3 Year Warranty

RoHS Compliant
Description
A high performance AC to DC external power supply family designed for medical applications. The ME60A Medical Series
low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition and EN/IEC/UL606011-11:2010 for Home Healthcare (non-hospital use) applications with and isolation levels which satisfy the 2 MOPP
requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC. The ME60A Series models will operate at
universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output
power up to +40°C and applicable output power derating at 50°C. These models are available in desktop versions,
include an IP22 rating per IEC60529 for the enclosure, and the output cable can be terminated at a variety of output
connectors.
Model Selection
Model
Number
ME60A1203F01
ME60A2403F01
ME60A4803F01
ME60A1203N01
ME60A2403N01
ME60A4803N01
ME60A1203Q01
ME60A2403Q01
ME60A4803Q01
Notes:
Volts
12.0V
24.0V
48.0V
12.0V
24.0V
48.0V
12.0V
24.0V
48.0V
Output
Current
5.00A
2.70A
1.35A
5.00A
2.70A
1.35A
5.00A
2.70A
1.35A
Output
Power
60W
60W
60W
60W
60W
60W
60W
60W
60W
Ripple &
Noise¹
120mV pk-pk
240mV pk-pk
480mV pk-pk
120mV pk-pk
240mV pk-pk
480mV pk-pk
120mV pk-pk
240mV pk-pk
480mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Output
Connector
Input
Configuration
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
www.slpower.com
ME60 Rev 2.0 4-SEP-15
1
ME60
60W Single Output External Power
Medical Series
General Specifications
AC Input
Input Current
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time
Less than 1 sec @115Vac, full load
115Vac: 1.5A, 230Vac: 0.75A
Hold-up Time
20mS min., at full Load, 100Vac input
264Vac, cold start: will not exceed 40A
F1, F2: 2A, 250Vac fuses (line & neutral lines)
provided on all models
Overvoltage
Protection
No Load Input
Power
<0.210W (meets DoE Efficiency Level VI
Requirements)
Transient
Response
Regulation
Drop Test
Vibration
Dimensions
Weight
www.slpower.com
Short Circuit
Protection
>88%, typical
60W continuous – See models chart for
specific voltage model ratings.
Output Voltage
Overload Protection
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
Output Power
Ripple and Noise
Overtemperature
Protection
Isolation
Safety Standards
See models chart on pg 1.
See models chart on pg 1.
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Hiccup Mode, auto recovery.
130 to 150% of output voltage (max. 60V on
48V model), hiccup mode
Input-Output: 1 MOPP
Input-Ground: 2 MOPP
Output-Ground: 1 MOPP
EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
-20°C to +50°C. Derate above 40°C.
Start Up at -40°C, full load, (warmup period before
Temperature
Derating
Derate output power above 40°C to TBD at
50°C
Altitude
1.4m from table top to wooden platform, 4
faces.
130 to 180% of rating, Hiccup Mode
Operating
Temperature
Storage Temperature
See models chart on pg 1.
Will shutdown upon an overtemperature
condition, auto-recovery.
Relative Humidity
all parameters are within published specifications).
-40°C to +85°C
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
W: 2.67” x L: 4.25” x H: 1.29”
W: 67.9mm x L: 108mm x H: 32.7mm
MTBF
>250,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
E-Cap Life
400g
ME60 Rev 2.0 4-SEP-15
>7 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day. (80% load on 12V model)
2
ME60
60W Single Output External Power
Medical Series
EMI/EMC Compliance
Conducted Emissions:
EN55011/CISPR22 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions:
EN55022/CISPR22 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise:
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4.4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5 & 6
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
Conducted Disturbances induced by RF
Fields
Rated Power frequency magnetic fields
Voltage Interruptions, Dips, Sags &
Surges
EN55024/IEC61000-4-5, Level 4, +/-1kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5, 6, & 8.
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
EN55024/IECEN61000-4-11: --100% dip for 20mS, Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions
EN55011/EN61000-3-2, Class A
Flicker Test
EN61000-3-3
Mechanical Drawing
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Notes: 1) All dimensions in mm.
2) 2.5mm barrel connector shown, other options are
available.
3) The unit should not be covered or enclosed to protect
against excessive case temperature rise.
www.slpower.com
ME60 Rev 2.0 4-SEP-15
3
ME60
60W Single Output External Power
Medical Series
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2015 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
ME60 Rev 2.0 4-SEP-15
4