MB60S - SL Power Electronics

MB60S
60W Single Output
Medical Series
Features
Ultra Small size of 2” x 3” x 1”
For 1U Applications
60W convection cooled
Universal Input 90-264Vac
Approved to IEC60601-1, 3rd Edition with 2 MOPP
Level V Efficiency Compliant Models
Less than 0.5W no-load Power Consumption
3 Year Warranty
Optional LED indicator for power-on
RoHS Compliant
Description
The MB60S Series models provide a reliable power source in high power density in 2” x 3” x 1” package.
Fully compliant to the applicable safety and EMC standards, these models will allow easy integration into many
Medical applications. All 6 models are CE marked to low voltage directive and approved to Medical standards
of IEC60601-1 3rd edition with 2 MOPP.
Model Selection
Model
Number
Volts
Output Current
Convection Cooled
Output Power
Convection Cooled
Ripple &
Noise*
Total
Regulation
OVP
Threshold
MB60S12K
12V
4.58A
55W
120mV pk-pk
±2%
14.4-18Vdc
MB60S15K
15V
4.00A
60W
150mV pk-pk
±2%
18-22.5Vdc
MB60S18K
18V
3.33A
60W
180mV pk-pk
2%
21-25.5Vdc
MB60S24K
24V
2.50A
60W
240mV pk-pk
±2%
28.8-36Vdc
MB60S36K**
36V
1.67A
60W
360mV pk-pk
±2%
42-47Vdc
MB60S48K
48V
1.25A
60W
480mV pk-pk
±2%
57.6-72Vdc
Notes:
* At -20C, the noise and ripple is 2% of the output.
** For product availability, please contact the factory
Input Specifications
PARAMETER
SPECIFICATION
AC Input Voltage:
90-264Vac, single phase
AC Input Frequency:
AC Input Current:
Turn-on Input Voltage:
Turn-off Input Voltage:
Inrush Current:
47-63Hz
120Vac: 1.4A, 240Vac: 0.75A
75V
65V
40A maximum @ 0C
www.slpower.com
MB60S 6-July-15
NOTES
Ramping Up
Ramping Down
1
MB60S
60W Single Output
Medical Series
Leakage Current
(Input–Earth):
<275µA@264Vac, 60 Hz input, NC
Leakage Current
(Output–Earth):
N/A
Leakage Current
(Input-Output):
<90µA@264Vac, 60 Hz input, NC
Input Fuses:
F1, F2: 4A, 250VAC
Fuses provided on all models
Typical
Measured at 120Vac and full load
Efficiency
MB60S12K
83%
MB60S15K
85%
MB60S18K
85%
MB60S24K
MB60S36K
MB60S48K
88%
88%
88%
No Load Input Power:
Turn-on Time:
<0.5W
<2 Seconds at 120Vac.
16mS minimum from loss of ac input at
120 Vac, full load.
Hold-up Time:
rd
IEC 60601-1 3 Ed – 8.7.3.c
24V, 36V, and 48V Models meet Level V
requirement
Meet Level V, standby Power Consumption
55 Watts for 12V output
DC Output Specifications
PARAMETER
SPECIFICATION
Output Power:
60W continuous for operation from -10ºC
to 50ºC 55 Watts for 12V output.
Cooling:
Convection
Total Regulation:
±2% for all models
Overload Protection:
120% - 180% of rated output current
value, Hiccup Mode
Short Circuit
Protection:
Overtemperature
Protection:
Short across the output terminals will not
cause damage to the unit. Hiccup Mode
OVP firing reduces output voltage to
<50% of nominal in <50mS. See chart for
trip range
Automatic Power Shutdown at Tc =
155ºC,
Minimum Load:
No minimum load is required
Overvoltage
Protection:
Ripple and Noise:
Transient Response:
Overshoot:
www.slpower.com
0.5% RMS, 1% pk-pk for all models.
NOTES
Total regulation is the maximum deviation from
nominal voltage for all loading conditions
20 MHz Bandwidth, differential mode.
Measured with noise probe directly across
output terminals, and load terminated with
0.1µF ceramic and 10µF low ESR capacitors
500µs typ. response time for return to
within 0.5% of final value for a 50% load
change, ∆i/∆t< 0.2A/µs. Max. voltage
deviation is 3.5%.
5% overshoot at turn-on, 5% overshoot at
turn-off, under all conditions.
MB60S 6-July-15
2
MB60S
60W Single Output
Medical Series
Safety Standard Compliance
Agency
CONDITIONS
rd
UL
ANSI/AAMI ES60101:2005, 3 Edition
CSA
CAN/CSA-C22.2 No. 60601-1 (2008)
Demko
EN 60601-1:2006
rd
CB Report
IEC 60601-1 (3 Edition)
Isolation Type:
B rated
Isolation Specifications
PARAMETER
Insulation Safety
Rating:
Electric Strength Test
Voltage:
CONDITIONS
Rating
Input to Ground
1 MOPP
Input to Output
2 MOPP
Output to Ground
Functional
Input to Ground
1800Vac
Input to Output
4000Vac
Output to Ground
500Vac
NOTES
.
Environmental Specifications
PARAMETER
Operating
Temperature:
Temperature Derating:
Cooling:
Storage Temperature:
Altitude:
Relative Humidity:
Shock:
Vibration:
SPECIFICATION
NOTES
-10°C to +80°C
-40°C Startup guaranteed
For Output Voltage ≥24V, derate output power to
50W @ 60C, 40 Watt @ 70C, and 20 Watts for 80C
Convection
-40°C to +85°C
Operating: -500 to 3,000 meter
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Non-Operating: Half-sine, 40 gpk, 10mS, 3 axes, 6
shocks total
Random vibration per MIL-STD-810E, Method
514.4, Cat. 1, Figure 514.4-1, 1 hr in each of three
axes
<24V will derate to 40W at 60C,
30W at 70C, and 20W at 80C
Reliability Specifications
PARAMETER
SPECIFICATION
NOTES
MTBF:
700,000 hours, 25°C ambient, full load
Calculation is done based on
Telcordia. Reports for each
model is available
Warranty:
3 Years
Limited
HALT Data:
Per SL Power Halt procedure
Report is available
www.slpower.com
MB60S 6-July-15
3
MB60S
60W Single Output
Medical Series
EMI/EMC Compliance
PARAMETER
SPECIFICATION
Conducted Emissions:
EN55011/22 Class B; FCC Part 15
Radiated Emissions:
EN55011/22 Class A; FCC Part 15
Harmonic Current
Emissions
EN61000-3-2, Class A
Voltage Fluctuations &
Flicker
EN61000-3-3
Static Discharge
Immunity:
EN61000-4-2 6kV contact, 8kV air, Criteria A
RF Field Susceptability
EN61000-4-3 (3V/m), Criteria A
Fast Transients/Bursts
Surge Susceptability
Conducted RF
Susceptability
Power Frequency
Magnetic Field Test
Voltage Sags & Surges
NOTES
EN61000-4-4 (PS: 2kV-40A, other lines 1kV20A), Criteria B
EN61000-4-5, Installation Class 3 (1kV diff.
mode, 2kV common mode), Criteria A
EN61000-4-6 (3Vrms), Criteria A
Performance criteria are defined as
following:
A – Normal performance
during and after the test
B – Temporary degradation,
self-recoverable
C – Temporary degradation,
operator intervention required
to recover the operation
EN61000-4-8 (3A/m), Criteria A
EN61000-4-11,
95% dip/0.5 cycle (Criteria A),
60%/5cycles (Criteria B ), 30%/25 cycles
(Criteria A).
Notes:
1. Specifications subject to change without notice.
2. Specifications are for convection rating at factory settings with 115Vac input and 25°C ambient unless
otherwise stated.
www.slpower.com
MB60S 6-July-15
4
MB60S
60W Single Output
Medical Series
Mechanical Drawing
Connector Information
Input Connector
J100
PIN 1) AC LINE
PIN 2) EMPTY
PIN 3) AC NEUTRAL
Mating Connector:
Tyco/AMP 640250-3
Pins = 770461-1
DC Output
Connector
J2
PIN 1) +Vout
PIN 3) -Vout
PIN 2) +Vout
PIN 4) -Vout
Mating Connector:
AMP 640250-4
Pins = 770461-1
Ground
(FG)
19-30258-0187 (Keystone 1285)
(Zierick 895)(.187*0.020)
Mating Connector
Molex 19002-0005
1. Mounting holes should be connected together for EMI purpose
2. FG is safety ground connection
3. This power supply requires mounting on metal standoffs 0.20” (5mm) in height
www.slpower.com
MB60S 6-July-15
5
MB60S
60W Single Output
Medical Series
Characteristic Curves
Output vs. Temperature
-40C start up. At -20C, the supply meet its full spec except ripple & noise might be increased from 1% to 2% of the output
voltage
55W convection cooled, derating output power to 30W at 70°C for outputs 12V and 15V
60W convection cooled, derating output power to 50W at 60°C and 40W at 70°C for Output Voltages ≥ 24V
20W convection cooled at 80C
Output Power(W)
Output Power vs . Temperature
Convection for
12V Model
Convection for
24V and above
70
50
30
10
-10
0
10 20 30 40 50 60 70 80
Temperature (°
°C)
Fig.1
Efficiency vs. Loading
Efficiency vs. Output Power
90.0%
80.0%
70.0%
60
45
30
15
6
Efficiency (%)
100.0%
Output Power (W)
120Vac
240Vac
Fig.2
www.slpower.com
MB60S 6-July-15
6
MB60S
60W Single Output
Medical Series
Ripple & Noise
To verify that the output ripple and noise does not exceed the level specified in the product specification, measured using a scope probe socket with
0.1uF ceramic and a 10uF electrolytic capacitor connected in parallel across it, 20MHz BW.
24V OUT, NO LOAD, 90VAC, 60HZ
24V OUT, FULL LOAD, 90VAC, 60HZ
24V OUT, NO LOAD, 264VAC, 50HZ
www.slpower.com
24V OUT, FULL LOAD, 264VAC, 50HZ
MB60S 6-July-15
7
MB60S
60W Single Output
Medical Series
Output Transient Response
50% load step within the regulation limits of minimum and maximum load, dI/dt< 0.2A/µSec. Recovery time not specified as there is no laps in regulation
with a 50% Load Step. Maximum voltage deviation is 3.5%
24V OUT, 120VAC, 25% TO 75% LOAD STEP
24V OUT, 240VAC, 25% TO 75% LOAD STEP
Output Overload Characteristic
24V OUT, 90VAC
24V OUT, 264VAC
Short Circuit Protection
Supply shall protect itself against Short Circuit conditions. No damage will occur if the output is shorted.
24V OUT, 90VAC
24V OUT, 264VAC
www.slpower.com
MB60S 6-July-15
8
MB60S
60W Single Output
Medical Series
Overvoltage Protection
OVP firing reduces output voltage to <50% of nominal in <50ms. See models chart for trip ranges.
24V OUT, FULL LOAD, 90VAC, 60HZ
24V OUT, FULL LOAD, 264VAC, 50HZ
Turn On Time
24V OUT, FULL LOAD, 90VAC, 60HZ
24V OUT, FULL LOAD, 264VAC, 50HZ
Hold Up Time
24V OUT, FULL LOAD, 120VAC, 60HZ
Data Sheet © 2012 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication.
for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
MB60S 6-July-15
However, SL Power accepts no responsibility