cover page for test report

Issue Date:
2007-09-21
Page 1 of 2
Report Reference #
E116994-A37-CB-1
COVER PAGE FOR TEST REPORT
Test Item Description:
Model/Type Reference:
Rating(s):
Power Supply
GEM600-VV-WXY-ZZ G and GEM750-VV-WXY-ZZ G. Where VV is the output
Voltage 24-48VDC, may also be followed by "-XYZ", where W is an Alpha
character A or T, which represents the Input Connector, X is an alpha character
A, B or T which represents the Output Connector, Y is an Alpha Character D,
which represents the Optional OR-ing Diode, may also be followed by "-ZZ" is
any numeric character 00-99 which represents Value Added Configurations not
related to safety and "G" indicates compliance with the RoHS directive. (RoHS
compliance has not been evaluated by UL.)
GEM600-VV-WXY-ZZ G
Input: 100-240 V~, 50/60 Hz, 9 A/4.5 A
GEM750-VV-WXY-ZZ G
Input: 100-240 V~, 50/60 Hz, 11 A/5.5 A
Output:
GEM600
V
A
24
25
28
21.5
36
16.7
48
12.5
GEM750
V
24
28
32
36
48
Standards:
Applicant Name and
Address:
Factory Location(s):
A
31.2
26.8
23.4
20.8
15.6
+ Maximum current output with Airflow of 42 CFM with an integral fan at an
ambient of 50°C.
IEC 60601-1:1988 + A1:1991 + A2:1995
SL POWER ELECTRONICS CORP
6050 KING ST
VENTURA CA 93003
INDUSTRIAS S L S A DE C V
COSTA RICA #60
COL CUAHUTEMOC
MEXICALI
BAJA CALIFORNIA N MEXICO
SL POWER ELECTRONICS XIANGHE
ANPING ECONOMIC & TECH DEVELOPING ZONE
XIANGHE,
Underwriters Laboratories Inc.
Issue Date:
2007-09-21
Page 4 of 65
Report Reference #
E116994-A37-UL-1
SPECIFIC TECHNICAL CRITERIA
TEST REPORT
UL 60601-1
Medical Electrical Equipment
Part 1: General requirements for safety
Report Reference No ........................: E116994-A37-UL-1
Compiled by .....................................: Ahmad Daoudi
Reviewed by ....................................: Kent C. Jones
Date of issue ....................................: 2007-09-21
Standards .........................................: UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment,
Part 1: General Requirements for Safety)
CAN/CSA-C22.2 No. 601.1-M90, 2005 (Medical Electrical Equipment Part 1: General Requirements for Safety)
Test procedure .................................: Component Recognition
Non-standard test method ...............: N/A
Test item description .......................: Power Supply
Trademark ........................................:
Model and/or type reference ............: GEM600-VV-WXY-ZZ G and GEM750-VV-WXY-ZZ G. Where VV is
the output Voltage 24-48VDC, may also be followed by "-XYZ", where
W is an Alpha character A or T, which represents the Input Connector,
X is an alpha character A, B or T which represents the Output
Connector, Y is an Alpha Character D, which represents the Optional
OR-ing Diode, may also be followed by "-ZZ" is any numeric character
00-99 which represents Value Added Configurations not related to
safety and "G" indicates compliance with the RoHS directive. (RoHS
compliance has not been evaluated by UL.)
Rating(s) ...........................................: GEM600-VV-WXY-ZZ G
Input: 100-240 V~, 50/60 Hz, 9 A/4.5 A
GEM750-VV-WXY-ZZ G
Input: 100-240 V~, 50/60 Hz, 11 A/5.5 A
Output:
GEM600
V
A
24
25
28
21.5
36
16.7
48
12.5
GEM750
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2007-09-21
Page 5 of 65
V
24
28
32
36
48
Report Reference #
E116994-A37-UL-1
A
31.2
26.8
23.4
20.8
15.6
+ Maximum current output with Airflow of 42 CFM with an integral fan
at an ambient of 50°C.
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2007-09-21
Page 6 of 65
Report Reference #
E116994-A37-UL-1
GENERAL INFORMATION
Test item particulars (see also clause 5):
Classification of installation and use ......................... :
For building-in
Supply connection ..................................................... :
Terminal block
Accessories and detachable parts included in the
evaluation .................................................................. :
None
Options included ....................................................... :
Terminal Block, Bussbars, Phoenix Contact, or
N+1 OR-ing Diode.
Possible test case verdicts:
- test case does not apply to the test object ............. :
N/A
- test object does meet the requirement ................... :
P(Pass)
- test object does not meet the requirement ............. :
F(Fail) (acceptable only if a corresponding, less
stringent national requirement is "Pass")
Abbreviations used in the report:
- normal condition ............................................ :
N.C. - single fault condition ................ :
S.F.C.
- operational insulation ..................................... :
OP
- basic insulation ........................ :
BI
- basic insulation between parts of opposite
polarity:
BOP - supplementary insulation ......... :
SI
- double insulation ............................................ :
DI
RI
- reinforced insulation ................ :
General remarks:
- "(see Enclosure #)" refers to additional information appended to the Test Report
- "(see appended table)" refers to a table appended to the Test Report
- Throughout the Test Report a point is used as the decimal separator
General Product Information:
CA1.0
Report Summary
CA1.1
N/A
CB1.0
Product Description
CB1.1
The equipment (DC power supplies) covered by this report are components which are intended
for use in end-product equipment used in a hospital or related health care facility, evaluated to
the standard for Medical Equipment.
The GEM Series or power supplies are designed for building-in to an end product. They are
designed with 1 cover mounted integral fan rated at 42 CFM. Rated Ambient is 50°C.
CC1.0
Model Differences
CC1.1
All models within the Model Series GEM600 are similar to each other except for Main Power
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2007-09-21
Page 7 of 65
Report Reference #
E116994-A37-UL-1
Transformers and minor component changes.
All models within the Model Series GEM750 are similar to each other except for the Main Power
Transformers and minor component changes.
Model Series GEM600 and Model Series GEM750 are similar, except for the Power FET's, which
are more robust in the GEM750 Series.
Nomenclature for Model Series GEM600-VV-WXY-ZZ G and Model Series GEM750-VV-WXY-ZZ
G, where
VV = the output Voltage 24-48VDC
W = an Alpha character A or T, which represents the Input Connector,
X = an alpha character A, B or T which represents the Output Connector,
Y = an Alpha Character D, which represents the Optional OR-ing Diode,
-ZZ= any numeric character 00-99 which represents Value Added Configurations not related to
safety
G = indicates compliance with the RoHS directive. (RoHS compliance has not been evaluated
by UL.)
OPTIONS:
A - Terminal Block (EBY P/N 5002-03-N-12)
B - Bussbars
T - Phoenix Contact (P/N 173204)
D - N+1 OR-ing Diode. Will allow 2 or more power supplies to be connected in parallel and share
the output load.
CD1.0
Additional Information
CD1.1
The schematics for these models are kept on file at the CB Testing Laboratory mentioned in the
first page of this test report, and can be provided by the applicant upon request by
CBTL's/NCB's.
Only limited testing was considered necessary on Model GEM750-48 due to testing conducted
as part of a similar already evaluated Model Series GPHP600 and GPHP750 covered under Test
Report Reference E116994-A16-CB-1, CB Certificate US/8708/UL .
CE1.0
Technical Considerations
CE1.1
The product was investigated to the following
additional standards:
EN 60601-1: 1990 + A1:1993 + A2:1995 +
A13:1996, CAN/CSA C22.2 No. 601.1-M90
(R1997), CAN/CSA C22.2 No. 601.1S1-94, and
CAN/CSA C22.2 No. 601.1B-98 (National
Differences for Canada), (except EMC
limitations, EN 60601-1-2, Biocompatibility, EN
10993-1, Programmable Electronic Systems,
IEC 60601-1-4)
CE1.2
The product was not investigated to the
following standards or clauses:
Clause 36, Electromagnetic Compatibility (IEC
601-1-2), Clause 48, Biocompatibility (ISO
10993-1), Clause 52.1, Programmable
Electronic Systems (IEC 601-1-4)
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2007-09-21
Page 8 of 65
Report Reference #
E116994-A37-UL-1
CE1.3
The product is Classified only to the following
hazards:
Shock, Fire,
CE1.4
The degree of protection against harmful
ingress of water is:
Ordinary
CE1.5
The following accessories were investigated for None
use with the product:
CE1.6
The mode of operation is:
Continuous
CE1.7
Software is relied upon for meeting safety
requirements related to mechanical, fire and
shock:
No
CE1.8
The product is suitable for use in the presence
of a flammable anesthetics mixture with air or
oxygen or with nitrous oxide:
No
CF1.0
Engineering Conditions of Acceptability
CF1.1
For use only in or with complete equipment where the acceptability of the combination is
determined by Underwriters Laboratories Inc.
When installed in an end-product, consideration must be given to the following:
CF2.0
This power supply has been evaluated as Class -I with no applied parts, continuous operation,
ordinary equipment and has not been evaluated
for use in the presence of a flammable
anesthetic mixture with air, oxygen or nitrous
oxide.
CF2.1
This component has been judged on the basis -of the required spacings in the Standards for
Medical Equipment, Part 1: General
Requirements for Safety, UL60601-1, First
Edition (2003) , IEC60601-1, Second Edition
(1998) and CSA 22.2 No. 601.1, which covers
the end use product for which the component is
designed.
CF2.2
The component shall be installed in compliance -with the enclosure, mounting, spacings,
casualty markings and segregation
requirements of the end-use application.
CF2.3
Consideration should be given to measuring the -temperature on power electronic components
and transformer windings when the power
supply is installed in the end-use equipment.
CF2.4
The input/output are not acceptable for field
connection, they are only intended for
connection to mating connectors of internal
wiring inside the end-use machine. The output
circuits have not been evaluated for direct
TRF No.: IEC60601_1C
--
Underwriters Laboratories Inc.
Issue Date:
2007-09-21
Page 9 of 65
Report Reference #
patient connection (Type B, BF or CF).
CF2.5
The component should be properly bonded to
the ground in the end-use equipment.
CF2.6
The temperature test was performed in a raised -ambient of 50°C.
CF2.7
Leakage Current testing should be repeated in
the end-product application.
--
CF2.8
The main Power Transformer (T6) and Bias
Transformer (T7), comply with Class F (155°C)
limits.
--
CF2.9
Additional fusing should be considered in the
-end product since this power supply was tested
with only one internal fuse, rated T 12A, 250 V.
CF3
Humidity and Blocked Ventilation Cooling
-Abnormal Testing should be considered as part
of the end-use application.
TRF No.: IEC60601_1C
--
Underwriters Laboratories Inc.
E116994-A37-UL-1