Appendix DEMKO CERTIFICATE

DEMKO CERTIFICATE
Certificate No.
Page
Date of Issue
Certificate Holder
Manufacturer
Production site
Certified Product
Model
Trademark
Rated Voltage / Frequency
Rated Current
Insulation Class
Degree of protection (IP)
Tested acc. to
Test Report No.
Additional
Expire date
Certification Manager
Jan-Erik Storgaard
Certification Body
146059-02
1/3
2008-11-04
SL POWER ELECTRONICS CORP.
6050 KING DR. BLDG A
VENTURA CA 93003, USA
SL POWER ELECTRONICS CORP.
6050 KING DR. BLDG A
VENTURA CA 93003, USA
SL POWER ELECTRONICS XIANGHE
ANPING ECONOMIC & TECH DEVELOPING ZONE
XIANGHE, HEBEI 065402, CHINA
Power Supply
See appendix
SL (LOGO)
100-240 V~, 50-60 Hz
0.5-0.25 A
EN 60601-1:1990 + A1:1993 + A2:1995
E116994-A47-CB-1, Issue Date 2008-10-06
for building-in
2009-09-12
The Manufacturer complies with the Production Surveillance Requirements. Products included in this certificate are allowed
to carry the registered approval marks of UL International Demko A/S, d or for cables <DEMKO>.The name of
UL International Demko A/S can be used in the marketing of the products. This Statement is only valid for products, which are
identical to the tested product, and manufactured at the above mentioned production site(s). UL International Demko A/S
has to be informed in writing about any changes, in accordance with the " UL International Demko A/S Standard Terms and
Conditions" for UL International Demko A/S services. The validity of this certificate is shortend if the EU legislation require retesting and re-certification, due to new standards or amemdments coming into force, before the expiry date.
UL International Demko A/S, Lyskaer 8, P.O. Box 514, DK-2730
Herlev, Denmark, Tel. +45 44 85 65 65, [email protected]
www.ul-europe.com
Appendix DEMKO CERTIFICATE
Certificate No.
Page
Date of Issue
146059-02
2/3
2008.11.04
The Certificate covers the following:
Model/Type reference
MINT1022AWXYZ, where W represents the output voltage which may be any number
from 05 thru 24; X indicates the type of output
connector which may be any number from 01 thru 99; Y indicates the
input connector options which may be any letter from A thru Z; and Z
indicates the configuration options which may be the number 01 for
standard configuration or 02 thru 99 for modifications.
Output:
MINT1022A05XYZ: 5 V dc, 3.2 A
MINT1022A12XYZ: 12 V dc, 1.83 A
MINT1022A24XYZ: 24 V dc, 0.92 A
This certificate replaces certificate No. 146059-01, dated 2008-09-25.
UL International Demko A/S has issued a new certificate as due to correct output information.
This certificate has been issued on the basis of CB Certificate (CB Test Certificate) No.
US/13072A/UL, issued by Underwriters Laboratories Inc., dated 2008-10-06.
Certification Body
UL International Demko A/S, Lyskaer 8, P.O. Box 514, DK-2730
Herlev, Denmark, Tel. +45 44 85 65 65, [email protected]
www.ul-europe.com
Appendix DEMKO Certificate
Certification Mark
Certificate No.
Page
Date of Issue
D-mark
146059-02
3/3
2008-11-04
Issue Date:
2008-09-18
Amendment 1
2009-02-06
Page 1 of 8
Report Reference #
E116994-A47-UL-1
SPECIFIC TECHNICAL CRITERIA
TEST REPORT
UL 60601-1
Medical Electrical Equipment
Part 1: General requirements for safety
Report Reference No ........................: E116994-A47-UL-1
Compiled by .....................................: Ahmad Daoudi
Reviewed by ....................................: Elizabeth Drew
Date of issue ....................................: 2008-09-18
Standards .........................................: UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment,
Part 1: General Requirements for Safety)
CAN/CSA-C22.2 No. 601.1-M90, 2005 (Medical Electrical Equipment Part 1: General Requirements for Safety)
Test procedure .................................: Component Recognition
Non-standard test method ...............: N/A
Test item description .......................: Power Supply
Trademark ........................................:
Model and/or type reference ............: MINT1022AWXYZ, where W represents the output voltage which may
be any number from 05 thru 24; X indicates the type of output
connector which may be any number from 01 thru 99; Y indicates the
input connector options which may be any letter from A thru Z; and Z
indicates the configuration options which may be the number 01 for
standard configuration or 02 thru 99 for modifications.
Rating(s) ...........................................: Input: 100-240 V~, 50-60 Hz, 0.5-0.25 A
Output:
MINT1022A05XYZ: 5 V dc, 3.2 A
MINT1022A12XYZ: 12 V dc, 1.83 A
MINT1022A24XYZ: 24 V dc, 0.92 A
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2008-09-18
Amendment 1
2009-02-06
Page 2 of 8
Report Reference #
E116994-A47-UL-1
GENERAL INFORMATION
Test item particulars (see also clause 5):
Classification of installation and use ......................... :
For building-in
Supply connection ..................................................... :
Header or Terminal Block or other options
Accessories and detachable parts included in the
evaluation .................................................................. :
None
Options included ....................................................... :
None
Possible test case verdicts:
- test case does not apply to the test object ............. :
N/A
- test object does meet the requirement ................... :
P(Pass)
- test object does not meet the requirement ............. :
F(Fail) (acceptable only if a corresponding, less
stringent national requirement is "Pass")
Abbreviations used in the report:
- normal condition ............................................ :
N.C. - single fault condition ................ :
S.F.C.
- operational insulation ..................................... :
OP
- basic insulation ........................ :
BI
- basic insulation between parts of opposite
polarity:
BOP - supplementary insulation ......... :
SI
- double insulation ............................................ :
DI
RI
- reinforced insulation ................ :
General remarks:
- "(see Enclosure #)" refers to additional information appended to the Test Report
- "(see appended table)" refers to a table appended to the Test Report
- Throughout the Test Report a point is used as the decimal separator
General Product Information:
CA1.0
Report Summary
CA1.1
N/A
CB1.0
Product Description
CB1.1
The equipment (DC power supplies) covered by this report, are components, which are intended
for use in end-product equipment used in a hospital or related health care facility, evaluated to
standard Medical Equipment.
The MINT1022 Series is designed for building-in to an end piece of equipment.
CC1.0
Model Differences
CC1.1
The power supplies in the MINT1022 Series are identical to each other with exception to
Transformer, T2, and minor component changes in the secondary circuit for the different outputs.
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2008-09-18
Amendment 1
2009-02-06
Page 3 of 8
Report Reference #
E116994-A47-UL-1
The MINT1022 Series is available with different types of input and output connector.
Model Number System:
Example:
MINT 1 022 A 05 05 I 01
1 2 3 4 5 6 7 8
1. Medical Internal Model number prefix.
2. Signifies number of outputs: 1
3. Output wattage: 022 = 22 W
4. Signifies generational differences such as energy star level changes, EMC level changes;
may be any letter from A to Z.
5. Output voltage: 05=5V, 12 = 12 V, 24 = 24 V
6. Output connector options: 5 = 4 Pin SPOX type or equiv. May be any Numeric Characters
00-99.
7. Input connector options: I = 4 Pin SPOX or equiv. May be any Alpha Character A-Z.
8. Configuration: 01 = standard, 02-99 for modifications.
CD1.0
Additional Information
CD1.1
The schematics for these models are kept on file at the CB Testing Laboratory mentioned in the
first page of this test report, and can be provided by the manufacturer upon request by CBTLs.
CE1.0
Technical Considerations
CE1.1
The product was investigated to the following
additional standards:
UL 60601-1, 1st Edition, 2006-04-26 (includes
National Differences for USA), EN 60601-1:
1990 + A1:1993 + A2:1995 + A13:1996,
CAN/CSA-C22.2 No. 601.1-M90 (R2005)
(includes National Differences for Canada),
(except EMC limitations, EN 60601-1-2,
Biocompatibility, EN 10993-1, Programmable
Electronic Systems, IEC 60601-1-4)
CE1.2
The product was not investigated to the
following standards or clauses:
Clause 36, Electromagnetic Compatibility (IEC
601-1-2), Clause 48, Biocompatibility (ISO
10993-1), Clause 52.1, Programmable
Electronic Systems (IEC 601-1-4)
CE1.3
The product is Classified only to the following
hazards:
Casualty, Shock, Fire
CE1.4
The degree of protection against harmful
ingress of water is:
Ordinary
CE1.5
The following accessories were investigated for None
use with the product:
CE1.6
The mode of operation is:
Continuous
CE1.7
Software is relied upon for meeting safety
requirements related to mechanical, fire and
shock:
No
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.
Issue Date:
2008-09-18
Amendment 1
2009-02-06
Page 4 of 8
Report Reference #
E116994-A47-UL-1
No
CE1.8
The product is suitable for use in the presence
of a flammable anesthetics mixture with air or
oxygen or with nitrous oxide:
CE2.0
The Accompanying Documents referred to by
-this standard refer to the Installation
Instructions. The installation instructions are not
shipped with this unit. They can be found on the
company website. http//www.slpower.com or
can request for a copy by writing to SL Power
Electronics Corp. at: 6050 King Drive, Bldg. A,
Ventura, CA 93003.
CF1.0
Engineering Conditions of Acceptability
CF1.1
For use only in or with complete equipment where the acceptability of the combination is
determined by Underwriters Laboratories Inc.
When installed in an end-product, consideration must be given to the following:
CF2.0
This power supply has been evaluated as Class -I, continuous operation, ordinary equipment and
has not been evaluated for use in the presence
of a flammable anesthetic mixture with air,
oxygen or nitrous oxide.
CF2.1
This component has been judged on the basis
of the required spacings in the Standard for
Medical Equipment, Part 1: General
Requirements for Safety, UL60601-1 and CSA
22.2 No. 601.1, which covers the end use
product for which the component is designed.
CF2.2
The component shall be installed in compliance -with the enclosure, mounting, spacings,
casualty markings and segregation
requirements of the end-use application.
CF2.3
Consideration should be given to measuring the -temperature on power electronic components
and transformer windings when the power
supply is installed in the end-use equipment.
CF2.4
The input/output connectors are not acceptable -for field connection, they are only intended for
connection to mating connectors of internal
wiring inside the end-use machine. the output
circuits have not been evaluated for direct
patient connection (Type B, BF or CF).
CF2.5
The component should be properly bonded to
the ground in the end-use equipment.
--
CF2.6
The maximum recommended ambient
temperature was considered to be 40 °C.
--
CF2.7
Leakage Current testing should be repeated in
--
TRF No.: IEC60601_1C
--
Underwriters Laboratories Inc.
Issue Date:
2008-09-18
Amendment 1
2009-02-06
Page 5 of 8
Report Reference #
the end-product application.
CF2.8
The Power Transformer (T2) comply with Class -A (105°C) limits.
CF2.9
Only one fuse is provided on the line side of the -input. The need for additional fusing shall be
provided as part of the end-product.
CF3
Consideration in conducting the Earthing Test
on Heat Sink, HS1, should made as part of the
end-use product.
TRF No.: IEC60601_1C
--
Underwriters Laboratories Inc.
E116994-A47-UL-1