RPT-160 series - Total Power International

Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
R P T- 1 6 0
160W Triple Output Medical Type
series
Features :
Universal AC input / Full range
Built-in active PFC function
Protections: Short circuit / Overload / Over voltage / Over temperature
5"x3" compact size
Free air convection for 100W and 145W with 20.5 CFM forced air
With power good and fail signal output
No load power consumption under 0.75W by PS-ON control (G model)
Standby [email protected] with fan, @0.6A without fan (G model)
3 years warranty
RPT G
G: With 5Vsb & no load power
consumption < 0.75 W
Blank: Basic function (without 5Vsb)
160A
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
OUTPUT
CH3
CH1
CH2
CH3
CH1
CH2
CH3
CH1
CH2
CH3
DC VOLTAGE
5V
12V
-5V
5V
12V
-12V
5V
15V
-15V
5V
12V
24V
RATED CURRENT (20.5CFM) 14A
5.5A
1A
14A
5A
1A
14A
3.6A
1A
11A
5A
1.2A
CURRENT RANGE (convection) 0.6 ~ 9A
0.2 ~ 3.8A 0.1 ~ 0.6A 0.6 ~ 9A
0.1 ~ 2.6A 0.1 ~ 0.8A 0.3 ~ 8A
0.6 ~ 14A 0.2 ~ 5A
RATED POWER (convection) Note.7 98.6W
98.4W
99W
98.2W
RATED POWER (20.5CFM) Note.8 145W
146W
143W
147.8W
0.1 ~ 1A
0.2 ~ 2.6A 0.15 ~ 1A
0.6 ~ 14A 0.1 ~ 3.6A 0.1 ~ 1A 0.3 ~ 11A 0.2 ~ 5A
0.15 ~ 1.2A
RIPPLE & NOISE (max.) Note.2 100mVp-p 120mVp-p 120mVp-p 100mVp-p 120mVp-p 120mVp-p 100mVp-p 150mVp-p 150mVp-p 100mVp-p 120mVp-p 200mVp-p
CH1:5 ~ 5.5V
2.0%
5.0%
-5,+7%
2.0%
5.0%
-4,+5%
2.0%
4.0%
8.0%
2.0%
5.0%
LINE REGULATION
0.5%
1.0%
1.0%
0.5%
1.0%
1.0%
0.5%
1.0%
1.0%
0.5%
1.0%
1.0%
LOAD REGULATION
1.5%
3.0%
-5,+6%
1.5%
3.0%
-4,+5%
2.0%
3.0%
8.0%
1.5%
3.0%
-3,+4%
SETUP, RISE TIME
1800ms, 30ms/230VAC
HOLD UP TIME (Typ.)
16ms/230VAC/115VAC at full load
Note.6 90 ~ 264VAC
127 ~ 370VDC
47 ~ 63Hz
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
PF>0.93/230VAC
84%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.8A/115VAC
0.9A/230VAC
COLD START 35A/115VAC
70A/230VAC
PF>0.98/115VAC at full load
84%
83%
LEAKAGE CURRENT
Earth leakage current <300uA / 264VAC, Patient leakage current <100uA/264VAC
OVERLOAD
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE
OVER TEMPERATURE
+7,-5%
3500ms, 30ms/115VAC at full load
FREQUENCY RANGE
PROTECTION
83%
CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
105 (TSW1)detect on heatsink of power transistor
90
(TSW2 )detect on heatsink of power transistor
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover
5VSB : [email protected] without fan, 0.8A with fan 20.5CFM ; tolerance 2%, ripple : 50mVp-p(max.)
5V STANDBY (G model)
PS-ON INPUT SIGNAL (G model) Power on: PS-ON = "Hi" or " > 2 ~ 5V" ; Power off: PS-ON = "Low" or " < 0 ~ 0.5V"
PF>1ms
POWER GOOD / POWER FAIL 500ms>PG>10ms
-20 ~ +70 (Refer to "Derating Curve")
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC
(Note 4)
EMC EMISSION
EMC IMMUNITY
MTBF
NOTE
0.2 ~ 3.4A 0.1 ~ 0.8A 0.6 ~ 9A
CURRENT RANGE (20.5CFM) 0.6 ~ 14A 0.2 ~ 5.5A 0.1 ~ 1A
VOLTAGE RANGE
OTHERS
RPT -160D
CH2
VOLTAGE TOLERANCE Note.3
FUNCTION
RPT -160C
CH1
VOLTAGE ADJ. RANGE
INPUT
RPT -160B
RPT -160A
OUTPUT NUMBER
DIMENSION
PACKING
20 ~ 90% RH non-condensing
-40 ~ +85
, 10 ~ 95% RH
0.03%/ (0 ~ 50 )
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
UL60601-1, TUV EN60601-1 approved
I/P-O/P:4KVAC
I/P-FG:1.5KVAC
O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
191.4Khrs min. MIL-HDBK-217F (25
127*76.2*34.6mm (L*W*H)
0.33Kg; 36pcs/12.9Kg/0.79CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. HS1,HS2 & HS3 can not be shorted.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. The rated power includes 5Vsb @ 0.6A.
8. The rated power includes 5Vsb @ 0.8A.
Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
R P T- 1 6 0
160W Triple Output Medical Type
series
Mechanical Specification
Unit:mm
FAN
AC Input Connector (CN1) : JST B3P-VH or equivalent
5cm
Pin No.
1
2
3
5
20.5CFM min.
SVR1
66.2
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
DC Output Connector (CN2) : JST B8P-VH or equivalent
1
2
76.2
Mating Housing
CN3
HS1
1
2
3
HS3
FS2
Pin No.
1,2,3,4
5,6
7
8
1
2
3
4
5
6
7
8
HS2
FS1
CN1
Assignment
AC/L
No Pin
AC/N
CN2
45
117
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Pin No.
Status
Mating Housing
Terminal
1
2
PG
GND
JST XHP
or equivalent
JST SXH-001T-P0.6
or equivalent
1234
3.5
Mating Housing
Power Good Connector(CN3):JST B2B-XH or equivalent
CN901
LED
CN901
Assignment
COM
CH1
CH2
CH3
5VSB Connector(CN901) : JST B-XH or equivalent
127
Pin No.
1
2,4
3
PS/ON
34.6
5mA(max.)
GND
Assignment
PS/ON
GND
5VSB
Mating Housing
Terminal
JST XHP
or equivalent
JST SXH-001T
or equivalent
3 max.
HS1,HS2,HS3 can not be shorted
Block Diagram
POWER SWITCHING
PWM CONTROL
RECTIFIERS
&
PFC
EMI
FILTER
I/P
FG
POWER
SWITCHING
O.L.P.
fosc :100KHz
+5VSB
DETECTION
CIRCUIT
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
PG
DETECTION
CIRCUIT
CONTROL
+5VSB
V3
+V2
+V1
COM
PG
O.V.P.
PS/ON
CONTROL
Derating Curve
PS/ON
Output Derating VS Input Voltage
Ta=25
100
100
With FAN
90
80
80
70
Without FAN
LOAD (%)
LOAD (%)
60
40
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60
70
(HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPT-160-SPEC 2011-07-05