DA22 Power Entry Modules with Line Filter www.schurter.com/PG06 IEC Appliance Inlet C14 with Filter, Fuseholder 1- or 2-pole, for PCB Mounting Standard- or Medical-Filter and -Fuse drawer C14 70° C Description -Panel Mount: Screw-on mounting on PCB, from rear-side -3 Functions: Appliance Inlet, Fuseholder for fuse-links 5 x 20 mm, Line filter in standard and medical version -V-Lock notch standard -PCB terminals Approvals -VDE Certificate Number: 40001522 -UL File Number: E72928 Characteristics -Ultra-compact design -All single elements are already wired -For added safety "Extra-Safe" fuse drawers are available -Suitable for use in equipment according to IEC/UL 60950 Suitable for use in medical equipment according to IEC/UL 60601-1 Other versions on request -Fixing on PCB with self-tapping screws 3 mm or pre-inserted nuts M3 -Fuse drawer 1-pole, plus spare fuse case -Fuse drawer 2-pole, with shorting bar -Medical version M80, 220 nF X capacitors -Class X1- and Y1-capacitors for enhanced withstand voltage -Protection class II, 70°C References Alternative: version for panel mounting KFA Weblinks pdf-datasheet, html-datasheet, General Product Information, Approvals, CE declaration of conformity, RoHS, CHINA-RoHS, REACH, Mating Connectors, e-Shop, SCHURTER-Stock-Check, Distributor-StockCheck, CAD-Drawings, Accessories, Detailed request for product Technical Data Ratings IEC Ratings UL/CSA Leakage Current Dielectric Strength Allowable Operation Temp. Climatic Category Degree of Protection Protection Class Terminal Panel Thickness s Material: Housing 1) 1 - 10 A @ Ta 40 °C / 250 VAC; 50 Hz 1 - 10 A @ Ta 40 °C / 250 VAC; 50/60 Hz ¹⁾ standard < 0.5 mA (250 V / 60 Hz) medical < 5 µA (250 V / 60 Hz) > 1.7 kVDC between L-N > 2.7 kVDC between L/N-PE Test voltage (1 min/50 Hz) -25 °C to 85 °C 25/085/21 acc. to IEC 60068-1 from front side IP 40 acc. to IEC 60529 Suitable for appliances with protection class I acc. to IEC 61140 For PCB mounting ; Additional for ground terminal: Quick connect terminals 6.3 x 0.8 mm angled to pin axis Screw-on mounting: max 3 mm Mounting screw torque max 0.5 Nm Thermoplastic, black, UL 94V-0 Appliance-Inlet/-Outlet Fuseholder Rated Power Acceptance @ Ta 23 °C Power Acceptance @ Ta > 23°C Line Filter MTBF C14 acc. to IEC 60320 UL 498 CSA C22.2 no. 42 (for cold conditions) pin-temperature 70 °C, 10 A, Protection Class I 1 or 2 pole, Shocksafe category PC2 acc. to IEC 60127-6, for fuse-links 5 x 20 mm 5 x 20: 2 W (1 pole)/ 1.4 W (2-pole) per pole Admissible power acceptance at higher ambient temperature see derating curves Standard and Medical Version, IEC 60939, UL 1283, CSA C22.2 no. 8 Technical Details > 2'000'000 h acc. to MIL-HB-217 F UL appovals: 1-pole versions up to 10 A, 2-pole versions up to 8 A Connectors 1 DA22 Power Entry Modules with Line Filter www.schurter.com/PG06 Dimension 46.8 57.6 39 47.6 4.2 2.2 32.5 36.1 2.8 2.5 4.5 4 10.5 19.5 6 34 37.6 32.7 8.1 1.25 30.4±0.10 0.85 39 ±0.15 47.05 16.2 6.3x0.8 6.05 ±0.10 1.6 16.2±0.15 3. 38 ±0.20 4 4. 1.6 41 29 32.7±0.20 R 4 6.6 38 +0 . 0 10 1.6 +0 . 0 10 +0 . 0 10 3.6 Technical Data of Filter-Components Rated Current [A] Filter-Type Inductances L [mH] Capacitance CX [nF] Capacitance CY [nF] R [MΩ] 1 Standard Version with Bleed Resistor 2 x 10 100 2.2 1 2 Standard Version with Bleed Resistor 2x4 100 2.2 1 4 Standard Version with Bleed Resistor 2 x 1.5 100 2.2 1 6 Standard Version with Bleed Resistor 2 x 0.8 100 2.2 1 8 Standard Version with Bleed Resistor 2 x 0.6 100 2.2 1 10 Standard Version with Bleed Resistor 2 x 0.3 100 2.2 1 1 Medical Version (M5) 2 x 10 100 - 1 2 Medical Version (M5) 2x4 100 - 1 4 Medical Version (M5) 2 x 1.5 100 - 1 6 Medical Version (M5) 2 x 0.8 100 - 1 8 Medical Version (M5) 2 x 0.6 100 - 1 10 Medical Version (M5) 2 x 0.3 100 - 1 2 Connectors 3. 6 3. 4 34.4 ±0.10 R 4. 2 ±0 .0 5 4 30 R R 5. 1 5.8 DA22 Power Entry Modules with Line Filter www.schurter.com/PG06 Diagrams Standard version with bleed resistor and Medical version M80 1-/2-pole Medical version M5 1-/2-pole PE’’ PE’’ 1) L PE R F2 N L’ L1 F1 Cx Cy L2 PE’ Cy 2) 1) N’ L PE N 1) Line 2) Load L1 F1 R PE’ Cx F2 L’ L2 2) N’ 1) Line 2) Load Derating Curves 2-pole 2.0 1.5 I = 1.0 In I = 0.7 In 1.0 I = << In 0.5 0.0 0 20 40 60 80 100 Admissible power acceptance in Watt Admissible power acceptance in Watt 1-pole Ambient air temperature Ta °C 1.4 1.2 1.0 I = 1.0 In 0.8 I = 0.7 In 0.6 I = << In 0.4 0.2 0.0 0 20 40 60 80 100 Ambient air temperature Ta °C Attenuation Loss - - - - differential mode _____ common mode Standard version 1A 2A 8A 10 A 4A 6A Connectors 3 DA22 Power Entry Modules with Line Filter www.schurter.com/PG06 Medical version (M5) 1A 2A 8A 10 A 4A 6A All Variants Rated current Filter-Type Fuseholder Comment Order Number 1 Standard Version with Bleed Resistor 1-pole - DA22.1111.11 2 Standard Version with Bleed Resistor 1-pole - DA22.2111.11 4 Standard Version with Bleed Resistor 1-pole - DA22.4111.11 6 Standard Version with Bleed Resistor 1-pole - DA22.6111.11 8 Standard Version with Bleed Resistor 1-pole - DA22.8111.11 10 Standard Version with Bleed Resistor 1-pole 1) DA22.9111.11 1 Standard Version with Bleed Resistor 2-pole - DA22.1121.11 2 Standard Version with Bleed Resistor 2-pole - DA22.2121.11 4 Standard Version with Bleed Resistor 2-pole - DA22.4121.11 6 Standard Version with Bleed Resistor 2-pole - DA22.6121.11 8 Standard Version with Bleed Resistor 2-pole - DA22.8121.11 10 Standard Version with Bleed Resistor 2-pole 1) DA22.9121.11 1 Medical Version (M5) 2-pole - DA22.1321.11 2 Medical Version (M5) 2-pole - DA22.2321.11 4 Medical Version (M5) 2-pole - DA22.4321.11 6 Medical Version (M5) 2-pole - DA22.6321.11 8 Medical Version (M5) 2-pole - DA22.8321.11 10 Medical Version (M5) 2-pole 1) DA22.9321.11 Availability for all products can be searched real-time:http://www.schurter.com/en/Stock-Check/StockCheck-SCHURTER 1) UL approvals: versions with 1-pole fuseholder are approved up to 10 A, 2-pole versions up to 8 A. Packaging unit 20 Pcs Required Accessory Description Fusedrawer 2 Fusedrawer für Fuse Links 5x20 mm, with or without Voltage Selector Insert Fingergrip, 2-pole 4301.1403 Fingergrip, 1-pole 4301.1405 Extra-Safe, 1-pole 4301.1407 ... 4 4301.1401 Extra-Safe, 2-pole Connectors DA22 Power Entry Modules with Line Filter www.schurter.com/PG06 Accessories Cord retaining kits Cord retaining strain relief Flat head, H 4700.0008 Mating Outlets/Connectors Category / Description Appliance Outlet Overview complete IEC Appliance Outlet F, Screw-on Mounting, Front Side, Solder Terminal 4787 IEC Appliance Outlet F, Snap-in Mounting, Front Side, Solder or Quick-connect Terminal 4788 IEC Appliance Outlet F or H, Screw-on Mounting, Front Side, Solder, PCB or Quick-connect Terminal 5091 Appliance Outlet further types to DA22 Connector Overview complete 4782 Mounting: Power Cord, 3 x 1 mm² / 3 x 18 AWG, Cable, Connector: IEC C13 4782 4022 Mounting: Power Supply Cord, 3 x 1.5 mm², Screw clamps, Connector: IEC C13 4022 4785 Mounting: Power Cord, 3 x 1 mm² / 3 x 18 AWG, Cable, Connector: IEC C13 4785 4300-06 Mounting: Power Cord, 3 x 1 mm² / 3 x 18 AWG, Cable, Connector: IEC C13 4300-06 4012 Mounting: Power Supply Cord, 3 x 1.5 mm², Screw clamps, Connector: IEC C13 4012 Connector further types to DA22 ... Mating Outlets/Connectors shuttered Power Cord Overview complete VAC17KS, North America, 2.0 m, Connector C17, cord end: 6051.2801 VAC17KS, Europlug, 2.0 m, Connector IEC C17, cord end: 6051.2803 VAC17KS, Europlug, 2.0 m, Connector IEC C17, cord end: 6051.2813 VAC13KS, Overview, diverse Connector IEC C13, cord end: VAC13KS Power Cord further types to DA22 24.08.2015 The specifications, descriptions and illustrations indicated in this document are based on current information. All content is subject to modifications and amendments. Information furnished is believed to be accurate and reliable. However, users should independently evaluate the suitability and test each product selected for their own applications. Connectors 5