ASA30M series

ASA30M
30W Me d i c a l Gr ade A dapt er w it h I nt erc ha n g e a b le AC Plu g s
series
Universal 90-264VAC Input
Full Safety Approvals
RoHS Compliant
Interchangeable AC Input Plugs
High Efficiency
2.44”W x 4.28”L x1.44”H
Model Number
Output Voltage
SINGLE OUTPUT
ASA30M-03015
ASA30M-03049
ASA30M-030212
ASA30M-030315
ASA30M-030618
ASA30M-030524
WWW.ASTRODYNETDI.COM
Output Amps
Regulation
Efficiency
Output Connector
4.0
3.0
2.5
2.0
1.65
1.25
±6%
±3%
±2%
±2%
±2%
±2%
70%
73%
76%
78%
78%
80%
Coaxial
Coaxial
Coaxial
Coaxial
Coaxial
Coaxial
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
30W Me d i c a l Gr ade A dapt er w it h I nt erc ha n g e a b le AC Plu g s
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Input Current
Inrush Current
Leakage Current
90-264 VAC
47-63 Hz
0.4~0.8 Amps max.
50 Amps max. @ 264VAC
100uA max.
Size
Construction
Output Connector
Weight
4.28” x2.44” x 1.44”
Wallmount Style
2.1x5.5x12mm Coax
10.58oz (300g)
ENVIRONMENTAL SPECIFICATIONS
See Selection Charts
See Selection Charts ± 1% ± 2%
± 1%
Yes
Continuous; Auto Recovery
10mS typ. @ 115VAC
GENERAL SPECIFICATIONS
Isolation
Efficiency
Safety
series
PHYSICAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Voltage and Current
Load Regulation (Note 4) Line Regulation (Note 3)
Voltage Accuracy (Note 1)
Ripple/Noise (Note 2)
Over Voltage Protection
Short Circuit Protection
Hold Up Time
ASA30M
I/P-O/P: 4000VAC
75% typ. UL60601-1, EN60601-1
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNETDI.COM
Oper. Temperature
Storage Temperature
Conducted EMI
0 to +40°C FL
-20 to +85°C *
CISPR/FCC Class B
Notes:
1. Voltage Setpoint at 60% Load
2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for Ripple & Noise measuring @ 20MHz BW.
3. Line Regulation measured from 100VAC to 240VAC, full load.
4. Load Regulation measured from 60% to full load and from 60% to
20% load (60% ±40%)
5. Minimum load is not required for proper operation.
All specifications are typical at nominal input, 75% load, and 25°C unless
otherwise noted
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
30W Me d i c a l Gr ade A dapt er w it h I nt erc ha n g e a b le AC Plu g s
ASA30M
series
MECHANICAL DIMENSIONS
WWW.ASTRODYNETDI.COM
ASTRODYNE USA: 1-800-823-8082