Outline of the 9267 SAFETY TEST DATA MANAGEMENT SOFTWARE

2
System & Application Software
Enhanced Efficiency in System Measurements
Efficient testing can be consistently performed using a controller (custom-built)
to conduct protective grounding, insulation resistance, and voltage withstand
testing all at once. Contact us for system upgrades.
3157-01 AC GROUNDING HiTESTER
● Measurement current: Max. 31 A
● Resistance measurement range: 0 to 1.800 W
3153 AUTOMATIC INSULATION / WITHSTANDING HiTESTER
● Withstanding test : Testing voltage; 0.2 to 5.00 kV AC 500 VA
: Measurement range; Current :0.01 to 100.0 mA
● Insulation test : Testing voltage; 500 V DC / 1000 V DC
: Measurement range; 0.1 to 9999 MΩ
Outline of the 9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
■ Test Types and Corresponding Test Instrument
• Insulation Withstand Test: 3153/3159/3158 (withstand voltage only)
• Protective Conductivity Test: 3157 • Live Line Testing: 3332
■ Test Sequence Operation
The above three tests can be performed in any desired sequence.
However, the same test should not be performed more than once.
Also, at least one of the tests must be performed.
■ Interface for connecting test instruments to the PC
RS-232C only
(Multiple test instrument can be connected through a commonly available multiplexer.)
■ Test Data Recording
Test results are recorded as a specified text file (in CSV format).
■ Connection Switching Function
Connections can be switched prior to each test by sending commands to
the PLC (programmable logic controller).
■ ID Number Management
ID numbers can be managed for test targets.
Automatic ID Incrementing
ID Number Digit Count Setting
ID Header Appending Function
■ 3153 Program Setting
A program setting screen is provided for the 3153.
Program setting can be saved as or loaded from text files.
[Program Setting]
[Load Data]
[Save Program Setting as Text]