Series AME40-MAZ

Click on Series name for product info on aimtec.com
Series AME40-MAZ
40 Watt | AC-DC / DC-DC Converter
FEATURES:
•
•
•
•
4000VAC I/O isolation
Medical Switching power modules for PCB mounting
Operating temperature: -40 to +85°C
Universal Input: 90-264VAC, 47-440Hz, or 130-370VDC
•
•
•
•
Energy star compliant
RoHS compliant
Low Ripple and Noise
CE, cULus approvals
CB
Models
Single output
Model
Input Voltage
(VAC/Hz)
AME40-3.3SMAZ
AME40-5SMAZ
AME40-12SMAZ
AME40-15SMAZ
AME40-24SMAZ
Input Voltage
(VDC)
Output
Voltage
(V)
Maximum
Capacitive
load (µF)
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
130-370
130-370
130-370
130-370
130-370
3.3
5
12
15
24
8200
8200
2200
1500
560
Input Voltage
(VAC/Hz)
Input Voltage
(VDC)
Output
Voltage
(V)
Maximum
Capacitive
load (µF)
Output Current
max
(A)
8
8
3.33
2.66
1.66
Efficiency
(%)
115
230
VAC
VAC
77
80
84
84
82
75
80
85
84
82
Models
Dual output
Model
Output Current
max
(A)
Efficiency
(%)
115
230
VAC
VAC
AME40-5DMAZ
90-264/47-440
130-370
±5
3300
±4
80
80
AME40-12DMAZ
90-264/47-440
130-370
±12
2200
±1.66
82
80
AME40-15DMAZ
90-264/47-440
130-370
±15
1000
±1.33
82
81
AME40-24DMAZ
90-260/47-440
130-370
±24
470
±0.835
79
77
NOTE: All specifications in this datasheet are measured at an ambient temperature of 25°C, humidity<75%, nominal input voltage and at
rated output load unless otherwise specified.
Input Specifications
Parameters
Current
Inrush current <2ms
Leakage current
External fuse
Input dissipation
Start-up time
Conditions
Typical
115 VAC
230 VAC
115 VAC
230 VAC
Recommended slow blow type
115 / 230 VAC
3.15
≤0.5
150
Conditions
Typical
(LL-HL)
0-100% load single
0-100% load dual
st
nd
25% load - 1 out, 100% load – 2 out
25% load Step
3.3 / 5V models
12 / 15V models
24V models
115VAC
±2
±2
±3
±3
±5
±2
75
100
150
15
0
Maximum
Units
1000
530
30
50
0.15
mA
mA
A
A
mA
A
W
ms
Maximum
Units
Output Specifications
Parameters
Voltage accuracy
Line regulation
Load regulation
Cross regulation
Transient response deviation
Ripple & Noise*
Hold-up time (min)
Minimum Load Current
*Ripple & Noise measured at 20MHz bandwidth with 0.1μF M/C and 47μF E/C
www.aimtec.com
F 052e R8.L
Tel: +1 514 620 2722
1 of 4
%
%
%
%
% of Vout
mV p-p
ms
% of Max
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AME40-MAZ
40 Watt | AC-DC / DC-DC Converter
Isolation Specifications
Parameters
Conditions
Tested I/O voltage
Isolation Resistance
Typical
60 sec
Rated
Units
4000
VAC
MΩ
>1000
General Specifications
Parameters
Conditions
Switching frequency
Protection class
Over current protection
Over voltage protection
Short circuit protection
Operating temperature
Maximum case temperature
Storage temperature
Temperature coefficient
Cooling
Humidity
Case material
Weight
Dimensions (L X H X W)
Typical
47
Class II
110
Zener diode clamp
Auto recovery
-40 - +85
Foldback
With derating above 50ºC
Maximum
KHz
140
110
%
%
o
C
C
o
C
o
%/ C
o
100
-40 to +95
0.02
Free air convection
Non condensing
MTBF
Units
95
Plastic (flammability to UL 94V-0)
300
3.56 x 2.48 x 1.18 inches
90.5 x 63 x 30 mm
o
> 800,000 hrs (MIL-HDBK -217F, t=+25 C)/Full Load
> 200,000 hrs (MIL-HDBK -217F, t=at highest operating temperature)/Full Load
% RH
g
Environment Approval
Test
Shock
Vibration
Parameters
Wave form
Acceleration amplitude
Bump duration
Converter operation
Number of bumps
Test mode
Displacement
Acceleration
Converter operation
Conditions
Half sine wave
5gn
30 ms
Before and after test, body mounted (on chassis)
18 (3 in each direction for every axis)
Sweep sine, 10-100Hz, speed 0.05Hz/s
1 mm
3g, 3 loops 30min one cycle, 3h total, every axis tested
Before and after test, body mounted (on chassis)
Safety Specifications
Parameters
Agency approvals
Standards
www.aimtec.com
F 052e R8.L
cULus, CE, CB
Medical Electrical Equipment
Information technology Equipment
Harmonic Current Emissions
EMI - Conducted and radiated emission
Voltage fluctuations and flicker
Electrostatic Discharge Immunity
RF, Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
RF, Conducted Disturbance Immunity
Power frequency Magnetic Field Immunity
Voltage dips, Short Interruptions Immunity
IEC\EN\UL 60601-1, 2 x MOOP, CSA-C22.2 No. 601.1-M90
EN 60950-1:2006+A11:2009
IEC/EN 61000-3-2, Class A
EN55011, class B
IEC/EN 61000-3-3, (EN60555-3)
IEC 61000-4-2 Level 3
IEC 61000-4-3 Level 2
IEC 61000-4-4 Level 3
IEC 61000-4-5 Level 2
IEC 61000-4-6 Level 2
IEC 61000-4-8 Level 2
IEC 61000-4-11
Tel: +1 514 620 2722
2 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AME40-MAZ
40 Watt | AC-DC / DC-DC Converter
Pin Out Specifications
Pin
Single
Dual
1
2
3
4
5
AC Input (N)
AC Input (L)
+V Output
-V Output
No pin
AC Input (N)
AC Input (L)
+V Output
Common
-V Output
3.56
(0.14)
3.56
(0.14)
Dimensions
90.50
(3.56)
3
Top View
4
27.94
(1.10)
55.88
(2.20)
63.00
(2.48)
27.94
(1.10)
2
5
3.56
(0.14)
3.56
(0.14)
1
81.30
(3.20)
4.60
(0.18)
5Ø 1.30 (Ø 0.05)
Dimensions mm (inch)
Case Tolerance ±0.50 (±0.02)
Pin Diameter 1.30 ± 0.05 (0.05 ± 0.002)
Pin Pitch Tolerance ±0.35 (±0.014)
Side View
30.00
(1.18)
DIV: 10.00 (0.40)
4.60
(0.18)
10.00
(0.40)
1.30
(0.05)
Derating
Free Air Convection
50º C
Output Power (%)
100
80
Safe Operating Area
60
40
20
0
- 40
- 20
0
20
40
60
80
85
100
Ambient Temperature º C
www.aimtec.com
F 052e R8.L
Tel: +1 514 620 2722
3 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AME40-MAZ
40 Watt | AC-DC / DC-DC Converter
Earth/Ground Connection EMC EN55022 class B compliant Application circuit
L1
35mH
ASE2501V-G
+Vout
L
90-264VAC
CX1
0.6uF
CX2
0.6uF
AMExw40-MAZ
N
-Vout
The Application circuit is EMC compliant for any type of Earth/Ground connection: Input Ground connection, Output Ground connection as
shown or both sides, which is not recommended if the product Isolation is used as a Safety feature.
NOTE: 1. Datasheets are updated as needed and as such, specifications are subject to change without notice. Once printed or downloaded, datasheets are no longer controlled
by Aimtec; refer to www.aimtec.com for the most current product specifications. 2. Product labels shown, including safety agency certifications on labels, may vary based on the
date manufactured. 3. Mechanical drawings and specifications are for reference only. 4. All specifications are measured at an ambient temperature of 25°C, humidity<75%,
nominal input voltage and at rated output load unless otherwise specified. 5. Aimtec may not have conducted destructive testing or chemical analysis on all internal components
and chemicals at the time of publishing this document. CAS numbers and other limited information are considered proprietary and may not be available for release. 6. This
product is not designed for use in critical life support systems, equipment used in hazardous environments, nuclear control systems or other such applications which necessitate
specific safety and regulatory standards other the ones listed in this datasheet. 7. Warranty is in accordance with Aimtec’s standard Terms of Sale available at www.aimtec.com.
www.aimtec.com
F 052e R8.L
Tel: +1 514 620 2722
4 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only