PSMP20 SERIES
85~264VAC (125~373VDC) Input Voltage Range
Single Outputs, Isolation Class II
20 Watts, Open Frame
Medical AC/DC Switching Power Supplies
MEDICAL
FEATURES
 Isolation Class II
 Withstand 2G Vibration Test
 3.5” x 2.0” x 0.76” Open Frame Package
 Universal Input Voltage: 85~264VAC or 125~373VDC
 Single Outputs
 20 Watts Output Power
 100% Full Load Burn-in Tested
 All Using 105°C Long Life Electrolytic Capacitors
 Cooling by Free Air Convection
 -20°C ~ +70°C Wide Operating Temperature Range
 ±10% Voltage Adjustment
 Short Circuit, Over Load, Over Voltage, and Brown-out
 RoHS Compliant
(Low AC Input Voltage) Protection
 UL60601-1, TUV EN60601-1, and IEC60601-1 Medical
 Energy Star Compliant
 Green Design, No-Load Power Consumption < 0.5W
Approvals
DESCRIPTION
The PSMP20 series of Medical AC/DC switching mode power supplies provides 20 Watts of continuous output power in a
3.5” x 2.0” x 0.76” open frame package. This series consists of 5V, 12V, 15V, and 24VDC output models with a universal
input voltage range of 85~264VAC or 125~373VDC. These power supplies are protected against short circuit, over load,
over voltage, and brown-out (low AC input voltage) conditions and have an MTBF of 188,300 hours using MIL-HDBK217F. This series also has UL60601-1, TUV EN60601-1, and IEC60601-1 medical approvals. All models have been 100%
full load burn-in tested and are RoHS and Energy Star compliant.
12/02/2011
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 1 of 4
PSMP20 Series
20 Watts, Open Frame
Single Outputs, Isolation Class II
Medical AC/DC Switching Power Supplies
Rev. A
SPECIFICATIONS: PSMP20 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT SPECIFICATIONS
Input Voltage Range
TEST CONDITIONS
AC Input Voltage Range
DC Input Voltage Range
Low Line
High Line
Low Line
Inrush Current
High Line
No Load Power Consumption
OUTPUT SPECIFICATIONS
Output Voltage
Voltage Tolerance
Full Load, Vin = 115VAC
Full Load, Vin = 230VAC
Cold Start, Vin = 115VAC
Cold Start, Vin = 230VAC
Line Regulation
Output Power
Output Current
Ripple & Noise (See Note 1)
Low Line
Hold-Up Time
High Line
Setup Time (See Note 3)
Low Line
Rise Time
High Line
Temperature Coefficient
PROTECTION
Over Voltage Protection
Over Load Protection
Short Circuit
Brown-out (Low AC Input Voltage)
GENERAL SPECIFICATIONS
Efficiency
Withstand Voltage
Isolation Resistance
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Vibration
Cooling
MTBF
PHYSICAL SPECIFICATIONS
Weight
12/02/2011
Unit
264
373
63
VAC
VDC
Hz
A
A
0.5
5VDC output model
12V, 15V, & 24VDC output models
5VDC output model
12V, 15V, & 24VDC output models
Low Line to High Line at rated load
0% to 100% rated load
-2
-1
-10
-1
-0.5
-0.5
0
Full Load, Vin = 115VAC
Full Load, Vin = 230VAC
Full Load, Vin = 115/230VAC
Full Load, Vin = 115VAC
Full Load, Vin = 230VAC
0~50°C
-0.03
Latch-off mode
Hiccup mode, recovers automatically after fault condition is removed
115
110
W
See Table
+2
+1
+10
+1
+0.5
+0.5
20
See Table
See Table
18
60
1000
25
35
+0.03
%
%
%
%
W
ms
ms
ms
%/°C
150
%
%
yes
yes
Vin = 230VAC
Input to output
500VDC
With derating (see derating curve)
Non-condensing
MIL-HDBK-217F
See Table
4000
100
VAC
MΩ
-20
+70
°C
-40
+85
°C
20
90
% RH
10
95
% RH
10~500Hz, 2G 10min/1cycle, period for 60 min each along X, Y, and Z axes
Free air convection
188,300 hours
Approximately 2.68oz (76g)
3.5 x 2.0 x 0.76 inches
(88.9 x 50.8 x 19.3 mm)
3 years
Dimensions (L x W x H)
Warranty
SAFETY & EMC
Safety Approvals
EMI Conduction & Radiation
Harmonic Current
EMS Immunity
Max
0.6
0.4
30
65
Voltage Adjustment Range
Load Regulation
Nom
85
125
47
Input Frequency
Input Current
Min
UL60601-1, TUV EN60601-1, and IEC60601-1
EN55011: 2007+A2: 2007 Class B
EN61000-3-2: 2006 Class A, EN61000-3-3: 1995+A1: 2001+A2: 2005
EN60601-1-2: 2001+A1: 2006, IEC61000-4-2,3,4,5,6,8,11 light industry level, criteria A
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 2 of 4
PSMP20 Series
20 Watts, Open Frame
Single Outputs, Isolation Class II
Medical AC/DC Switching Power Supplies
Rev. A
MODEL SELECTION TABLE
Model Number
Input Voltage Range
Output Voltage
Output Current
Ripple & Noise (1)
Efficiency
Output Power
5 VDC
4A
80mVp-p
77%
20W
12 VDC
1.66A
150mVp-p
82%
20W
15 VDC
1.33A
150mVp-p
84%
20W
24 VDC
0.84A
240mVp-p
85%
20W
PSMP-20-05
85 ~ 264 VAC
or
125 ~ 373 VDC
PSMP-20-12
PSMP-20-15
PSMP-20-24
NOTES
1. Ripple & noise is measured at 20MHz bandwidth by using 12" twisted pair-wire terminated with 0.1µF and 47µF
capacitors in parallel.
2. Tolerance includes set up tolerance, line regulation, and load regulation.
3. The length of the setup time is measured a first cold start. Turning the power supply ON and OFF very quickly may lead
to an increase in the setup time.
4. The power supply is considered a component which will be installed into final equipment. The final equipment must be
re-confirmed that it still meets EMC directives.
BLOCK DIAGRAM
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
RECTIFIERS
&
FILTER
POWER
SWITCHING
PWM
CONTROL
+V
-V
DETECTION
CIRCUIT
DERATING CURVE
Load vs Input Voltage
Load vs Temperature
Ta=25°C
rated output voltage, 100% load, 115-230VAC
100
100
90
Load (%)
Load (%)
80
60
80
70
60
40
50
20
40
-20
0
10
20
30
40
50
60
70
85
100
Ambient Temperature (°C)
12/02/2011
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
110
120
264
Input Voltage (VAC)
Page 3 of 4
PSMP20 Series
20 Watts, Open Frame
Single Outputs, Isolation Class II
Medical AC/DC Switching Power Supplies
Rev. A
MECHANICAL DRAWING
3.5 (88.9)
0.15 (3.75)
0.15 (3.75)
0.76 (19.3)
Unit: inches (mm)
1.7 (43.3)
2.0 (50.8)
3-0.13” (3.3mm) Dia.
Mounting Holes
C2
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers
will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just
one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with
its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing
support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
12/02/2011
(603)778-2300
(888)587-9255
(603)778-9797
[email protected]
www.wallindustries.com
5 Watson Brook Rd.
Exeter, NH 03833
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 4 of 4