FEATURES
•
•
•
•
•
•
•
•
•
Class I
Single Outputs
RoHS Compliant
Input to Output: 2MOPP
Up to 90% High Efficiency
IEC-320-C14 AC Inlet
Active Power Factor Correction
Free Air Convection Cooling
Up to 100 Watts Output Power
• 90-260VAC Input Voltage Range
• Over Current, Short Circuit, and Over Voltage
Protection
• Energy Star 2.0, Efficiency Level V
• IEC60601-1 Edition 3.1, ES60601-1:2005(R2012),
CSAC22.2 NO.60601-1:14, EN60601-1:2006/A1:2013,
UL/cUL, TUV T-Mark, CE Safety Approvals
• Optional Output Connectors Available
• 100% Burn-in Tested
DESCRIPTION
Size:
5.75 x 2.99 x 1.69 inches
146.0 x 76.0 x 43.0 mm
The DTHPU101 series of class I medical AC/DC desktop power supplies provides up to
100 Watts of continuous output power in a 5.75” x 2.99” x 1.69” package. This series
consists of single output models with a 90~260VAC wide input range and an IEC-320-C14
AC inlet. Some features include high efficiency up to 90%, 2MOPP insulation, and over
current and over voltage protection. All units are UL94V-1, RoHS, and Energy Star 2.0
Level V compliant and meet FCC Part-18 class B and CISPR-11 EN55011 class B
emission limits. This series also has IEC60601-1 Edition 3.1, ES60601-1:2005(R2012),
CSAC22.2 NO.60601-1:14, EN60601-1:2006/A1:2013, UL/cUL, TUV T-Mark, CE safety
approvals. All models have been 100% burn-in tested.
MODEL SELECTION TABLE
Output Voltage (1)
Output Current
DTHPU101-105
Input Voltage
Range
90 ~ 260VAC
9.09 ~ 7.69 A
Total
Regulation
±5%
No Load Power
Consumption
0.5W
Typical
Efficiency
81%
11 ~ 13 VDC
DTHPU101-106
90 ~ 260 VAC
13 ~ 16 VDC
7.69 ~ 6.25 A
±4%
0.5W
84%
100W
DTHPU101-107
DTHPU101-108
90 ~ 260 VAC
16 ~ 21 VDC
6.25 ~ 4.76 A
±4%
0.5W
87%
100W
90 ~ 260 VAC
21 ~ 27 VDC
4.76 ~ 3.70 A
±4%
0.5W
87%
100W
DTHPU101-109
90 ~ 260 VAC
27 ~ 33 VDC
3.70 ~ 3.03 A
±3%
0.5W
87%
100W
DTHPU101-110
90 ~ 260 VAC
33 ~ 40 VDC
3.03 ~ 2.50 A
±3%
0.5W
87%
100W
DTHPU101-111
90 ~ 260 VAC
40 ~ 48 VDC
2.50 ~ 2.08 A
±3%
0.5W
87%
100W
Model Number
01/12/2016
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Output Power
100W
Page 1 of 4
Rev B
DTHPU101 SERIES
100 Watts
Medical AC/DC Desktop Power Supplies
Single Outputs
SPECIFICATIONS: DTHPU101 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
INPUT SPECIFICATIONS
Safety Approvals Input Voltage Range
Operating Input Voltage Range
Input Voltage
100
90
47
Input Frequency
Low Line
Input Current
High Line
Low Line
Inrush Current
High Line
No Load Power Consumption
Power Factor Correction
100VAC, full load
240VAC, full load
115VAC, full load, 25°C, cold start
230VAC, full load, 25°C, cold start
230VAC, no load
230VAC, full load
0.95
LL to HL, full load
11~13V
13~27V
27~48V
0.5
-5
-4
-3
90VAC, full load
110VAC, full load
100VAC, full load
100VAC, Full load to half load
16
0.3
240
260
63
1.2
0.5
50
120
See Table
1
VAC
Hz
A
A
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation(3)
Total Regulation
Output Power
Output Current
Ripple & Noise (peak to peak)(4)
Hold-up Time(5)
Start-up Time
Transient Response Time
Temperature Coefficient
See Table
1
+5
+4
+3
See Table
See Table
100
%
%
mVp-p
ms
s
ms
%/°C
2
4
+0.04
-0.04
PROTECTION
Over Voltage Protection
Over Current Protection
Short Circuit Protection
Recovers automatically after fault condition is removed
112
110
132
150
Automatic Recovery
%
%
GENERAL SPECIFICATIONS
Efficiency
Dielectric Withstanding Voltage
Insulation Resistance
Safety Ground Leakage Current
230 VAC, full load
Primary to Secondary, limit current <10mA
Primary to PE, limit current <10mA
Primary to Secondary, 500VDC, 25ºXC/70% RH
240VAC/60Hz
See Table
4000
1500
VDC
MΩ
mA
50
0.25
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Cooling
Electro Static Discharge
Vibration
Surge Voltage
Flammability Rating
MTBF
Derating linearly from 100% Load at 40°C to 50% load at 70°C
10~95% RH
Non-condensing
-10
-40
0
0
All conditions
Air Discharge, IEC61000-4-2
Contact Discharge, IEC61000-4-2
10~500Hz, 10min./1cycle, 60min. each along X,Y,Z axes
Line-Neutral
Line-PE & Neutral-PE
MIL-HDBK-217F, 25°C
+70
+85
95
95
Up to 3000m
Free air convection
15
8
5
1
2
UL94V-1
100,000
°C
°C
%
%
kV
G
kV
hours
PHYSICAL SPECIFICATIONS
Weight
Dimensions (L x W x H)
1.08~1.48 lbs (490~670g)
5.75 x 2.99 x 1.69 inches (146.0 x 76.0 x 43.0 mm)
SAFETY & EMC
Safety Approvals
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
EMI Requirements for EN55011
01/12/2016
220VAC
110VAC
IEC60601-1 Edition 3.1, ES60601-1:2005(R2012), CSAC22.2 NO.60601-1:14,
EN60601-1:2006/A1:2013, UL/cUL, TUV T-Mark, CE
B
Class
B
Class
B
Class
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 4
Rev B
DTHPU101 SERIES
100 Watts
Medical AC/DC Desktop Power Supplies
Single Outputs
NOTES
1. The output voltage is specified as a range (ex: 40~48VDC); the customer must specify what they would like the output voltage set at.
2. This series needs to use AWG#18x3C + AWG#16x3C/4FT output cable in order to meet the total regulation specified. The regulation will
change if a different output cable is used.
3. Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
4. Ripple and noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at
rated load and nominal line.
5. Hold up time is measured from end of the last charging pulse to the time when the main output drops down to low limit of main output at
rated load and nominal line.
6. Optional output connectors available. Please call factory for ordering details.
DERATING CURVES
Input Voltage Derating Curve
Temperature Derating Curve
MECHANICAL DRAWING
Net Weight: (490~670g)
Optional output connectors available
01/12/2016
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 4
Rev B
DTHPU101 SERIES
100 Watts
Medical AC/DC Desktop Power Supplies
Single Outputs
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and
development engineers will provide a solution that exceeds your performance requirements on-time and on
budget. Our ISO9001-2008 certification is just one example of our commitment to producing a high quality, welldocumented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for
working closely with its customers to ensure each solution meets or exceeds form, fit and function requirements.
We will continue to provide ongoing support for your project above and beyond the design and production phases.
Give us a call today to discuss your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
01/12/2016
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 4