ME40 40W Single Output External Power Medical Series Features Meets UL/EN/IEC60601-1-2, 4th edition for EMC* Approved to EN/IEC/UL60601-1, 3rd edition 2 MOPP Input to Output Isolation Meets DoE Efficiency Level VI Requirements No load input power Average Efficiency Up to 40W of AC-DC Power Universal Input 90-264Vac Input Range Desktop and Wall-Plug versions Meets EN55011/CISPR11, FCC Part 15.109 Class B Conducted & Radiated Emissions, with 6db margin IP22 Rated Enclosure E-cap life of >8 years >1,000,000 hours MTBF 3 Year Warranty Description A high performance AC to DC external power supply family designed for medical applications. The ME40A Medical Series low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels * which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC . The ME40A Series models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up to +40°C and applicable output power derating at 50°C. These models are available in desktop and wall-plug versions, include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at a variety of output connectors. *Consult Factory for Table 9 compliance information. Model Selection Model Number ME40A1203F01 Volts 12.0V Output Current 3.40A Output Power 40W Ripple & Noise¹ 120mV pk-pk Line Regulation ±1% Load Regulation ±5% Overvoltage Trip Range 120%-150% ME40A1803F01 18.0V 2.22A 40W 180mV pk-pk ±1% ±5% 120%-150% ME40A2403F01 24.0V 1.70A 40W 240mV pk-pk ±1% ±5% 120%-140% ME40A1203N01 12.0V 3.40A 40W 120mV pk-pk ±1% ±5% 120%-150% ME40A1803N01 18.0V 2.22A 40W 180mV pk-pk ±1% ±5% 120%-150% ME40A2403N01 24.0V 1.70A 40W 240mV pk-pk ±1% ±5% 120%-140% ME40A1203Q01 12.0V 3.40A 40W 120mV pk-pk ±1% ±5% 120%-150% ME40A1803Q01 18.0V 2.22A 40W 180mV pk-pk ±1% ±5% 120%-150% ME40A2403Q01 24.0V 1.70A 40W 240mV pk-pk ±1% ±5% 120%-140% ME40A1203B01 12.0V 3.40A 40W 120mV pk-pk ±1% ±5% 120%-150% ME40A1803B01 18.0V 2.22A 40W 180mV pk-pk ±1% ±5% 120%-150% ME40A2403B01 24.0V 1.70A 40W 240mV pk-pk ±1% ±5% 120%-140% ME40A1203C01 12.0V 3.40A 40W 120mV pk-pk ±1% ±5% Output Connector Input Configuration 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class I Desktop, IEC60320 C14 Receptacle 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class II Desktop, IEC60320 C8 Receptacle 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class II Desktop, IEC60320 C18 Receptacle 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class II Wall-Plug, Interchangeable Blades² 120%-150% 2.5 x 5.5 x 9.5mm Class II Wall-Plug, Straight Barrel Type, Fixed North ME40A1803C01 18.0V 2.22A 40W 180mV pk-pk 120%-150% ±1% ±5% center positive American Blades³ ME40A2403C01 24.0V 1.70A 40W 240mV pk-pk 120%-140% ±1% ±5% Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. 2. Standard models are fitted with North American blades. Order blade kit KT-1027K for other blades (EU. UK, Australia) 3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”. 4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted. 5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE40B1203F01). www.slpower.com ME40 Rev 2.2 2-MAY-16 1 ME40 40W Single Output External Power Medical Series General Specifications AC Input Input Current Inrush Current Input Fuses Earth Leakage Current Efficiency 100-240Vac, ±10%, 47-63Hz, 1∅ Turn On Time Less than 700mS @115Vac, full load 115Vac: 1.2A, 230Vac: 01.6A Hold-up Time 20mS min., at full Load, 100Vac input 264Vac, cold start: will not exceed 40A F1, F2: 2.0A, 250Vac fuses (line & neutral lines) provided on all models Overtemperature Protection Overload Protection Input-GND: <500μA@264Vac, 60Hz, NC Output-GND: <4mA@264Vac, 60Hz, NC Short Circuit Protection Overvoltage Protection >87%, typical Output Power 40W continuous – See models chart for specific voltage model ratings. No Load Input Power <0.1W per DoE Efficiency Level VI Requirements Isolation Safety Standards Will shutdown upon an overtemperature condition, auto-recovery. 130 to 180% of rating, Hiccup Mode Hiccup Mode, auto recovery. Hiccup mode. See models chart for trip ranges. Input-Output: 2 MOPP Input-Ground: 1 MOPP Output-Ground: 1 MOPP EN/IEC/UL60601-1, 3rd edition and Ripple and Noise See models chart on pg 1. Operating Temperature -20°C to +70°C Output Voltage See models chart on pg 1. Temperature Derating See derating chart Transient Response Regulation Drop Test Vibration Dimensions Weight 500µs response time for return to within 0.5% of final value for any 50% load step over the range of 5% to 100% of rated load, ∆i/∆t< 0.2A/µs. Max. voltage deviation is +/-3.5%. Storage Temperature Altitude See models chart on pg 1. 1.4m from table top to wooden platform, 6 faces. Relative Humidity -40°C to +85°C Operating: to 5000m. Non-operating: -500 to 40,000 ft. 5% to 95%, non-condensing Operating: 0.003g/Hz, 1.5grms overall, 3 axes, 10 min/axis, 1-500Hz. Non-Oper.: random waveform, 3 minutes per axis, 3 axes and Sine waveform, Vib. frequency/acceleration: 10-500Hz/1g, sweep rate of 1 octave / minutes, Vibration time of 10 sweeps / axes, 3 axes Shock Operating: Half-sine, 20gpk, 10mS, 3 axes, 6 shocks total Non-Operating: Half-sine waveform, impact acceleration of 100G, Pulse duration of 6 mS, Number of shocks: 3 for each of the three axis See outline drawings MTBF >1,000,000 hours, full load, 110 & 220Vac input, 25°C amb., per Telcordia 332 Issue 6. E-Cap Life 250g >8 year life based on calculations at 115Vac/60Hz & 230Vac/50Hz, ambient 25°C at 24 hrs per day, 365 days/year, 6 power up cycles per day. All specifications are typical at nominal input, full load, at 25°C ambient unless noted. www.slpower.com ME40 Rev 2.2 2-MAY-16 2 ME40 40W Single Output External Power Medical Series EMI/EMC Compliance Conducted Emissions: EN55011/CISPR22 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac Radiated Emissions: EN55022/CISPR22 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac Common Mode Noise: High Frequency (100kHz-20MHz): <40mA pk-pk Electro-Static Discharge (ESD) Immunity on Power ports: EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A IEC60601-1-2, 4th Edition, Table 4 Radiated RF EM Fields Susceptibility EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz IEC60601-1-2, 4th Edition, Table 4 Electrical Fast Transients (EFT) /Bursts: EN55024/IEC61000-4-4, Level 4, +/- 4kV, 100Khz rep rate, 40A, Criteria A IEC60601-1-2, 4th Edition, Table 5 Surges, Line to Line (Diff Mode) and Line to GND (CMN Mode) EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A Surpasses IEC60601-1-2, 4th Edition requirements. Conducted Disturbances induced by RF Fields EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz IEC60601-1-2, 4th Edition, Table 5. Rated Power frequency magnetic fields EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz IEC60601-1-2, 4th Edition, Table 4 Voltage Interruptions, Dips, Sags & Surges EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225, 270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A --100% dip for 5000mS (250/300 cycles), Criteria B --60% dip for 100mS, Criteria B --30% dip for 500mS, Criteria A IEC60601-1-2, 4th Edition, Table 5 Harmonic Current Emissions EN55011/EN61000-3-2, Class A Flicker Test EN61000-3-3 All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special environments. Mechanical Drawing 54 33.6 1450±50 1450±50 96 95.5 55 IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector Notes: www.slpower.com Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector 1. All dimensions in mm. 2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K. ME40 Rev 2.2 2-MAY-16 3 ME40 40W Single Output External Power Medical Series Derating Chart: ME40 Series Power Derating 45 Output Power- Watts 40 35 30 25 12V 20 18V 15 24V 10 5 0 -20 -10 0 10 20 30 40 50 60 70 Ambient Temperature- ͦC Connector Information Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below. The “03” in the standard model number is replaced by the applicable digits below: Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice. www.slpower.com ME40 Rev 2.2 2-MAY-16 4