ME 40 DS

ME40
40W Single Output External Power
Medical Series
Features

Meets UL/EN/IEC60601-1-2, 4th edition for
EMC*

Approved to EN/IEC/UL60601-1, 3rd edition

2 MOPP Input to Output Isolation

Meets DoE Efficiency Level VI Requirements


No load input power
Average Efficiency

Up to 40W of AC-DC Power

Universal Input 90-264Vac Input Range
 Desktop and Wall-Plug versions

Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin

IP22 Rated Enclosure

E-cap life of >8 years

>1,000,000 hours MTBF

3 Year Warranty
Description
A high performance AC to DC external power supply family designed for medical applications. The ME40A Medical Series
low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels
*
which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC . The ME40A Series
models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C,
delivering full rated output power up to +40°C and applicable output power derating at 50°C. These models are available
in desktop and wall-plug versions, include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at
a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
ME40A1203F01
Volts
12.0V
Output
Current
3.40A
Output
Power
40W
Ripple &
Noise¹
120mV pk-pk
Line
Regulation
±1%
Load
Regulation
±5%
Overvoltage
Trip Range
120%-150%
ME40A1803F01
18.0V
2.22A
40W
180mV pk-pk
±1%
±5%
120%-150%
ME40A2403F01
24.0V
1.70A
40W
240mV pk-pk
±1%
±5%
120%-140%
ME40A1203N01
12.0V
3.40A
40W
120mV pk-pk
±1%
±5%
120%-150%
ME40A1803N01
18.0V
2.22A
40W
180mV pk-pk
±1%
±5%
120%-150%
ME40A2403N01
24.0V
1.70A
40W
240mV pk-pk
±1%
±5%
120%-140%
ME40A1203Q01
12.0V
3.40A
40W
120mV pk-pk
±1%
±5%
120%-150%
ME40A1803Q01
18.0V
2.22A
40W
180mV pk-pk
±1%
±5%
120%-150%
ME40A2403Q01
24.0V
1.70A
40W
240mV pk-pk
±1%
±5%
120%-140%
ME40A1203B01
12.0V
3.40A
40W
120mV pk-pk
±1%
±5%
120%-150%
ME40A1803B01
18.0V
2.22A
40W
180mV pk-pk
±1%
±5%
120%-150%
ME40A2403B01
24.0V
1.70A
40W
240mV pk-pk
±1%
±5%
120%-140%
ME40A1203C01
12.0V
3.40A
40W
120mV pk-pk
±1%
±5%
Output
Connector
Input
Configuration
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades²
120%-150%
2.5 x 5.5 x 9.5mm
Class II Wall-Plug,
Straight Barrel Type,
Fixed North
ME40A1803C01
18.0V
2.22A
40W
180mV pk-pk
120%-150%
±1%
±5%
center positive
American Blades³
ME40A2403C01
24.0V
1.70A
40W
240mV pk-pk
120%-140%
±1%
±5%
Notes:
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
2. Standard models are fitted with North American blades. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE40B1203F01).
www.slpower.com
ME40 Rev 2.2
2-MAY-16
1
ME40
40W Single Output External Power
Medical Series
General Specifications
AC Input
Input Current
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time
Less than 700mS @115Vac, full load
115Vac: 1.2A, 230Vac: 01.6A
Hold-up Time
20mS min., at full Load, 100Vac input
264Vac, cold start: will not exceed 40A
F1, F2: 2.0A, 250Vac fuses (line & neutral
lines) provided on all models
Overtemperature
Protection
Overload Protection
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
Short Circuit
Protection
Overvoltage
Protection
>87%, typical
Output Power
40W continuous – See models chart for
specific voltage model ratings.
No Load Input
Power
<0.1W per DoE Efficiency Level VI
Requirements
Isolation
Safety Standards
Will shutdown upon an overtemperature
condition, auto-recovery.
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
Hiccup mode. See models chart for trip
ranges.
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
EN/IEC/UL60601-1, 3rd edition and
Ripple and Noise
See models chart on pg 1.
Operating
Temperature
-20°C to +70°C
Output Voltage
See models chart on pg 1.
Temperature
Derating
See derating chart
Transient
Response
Regulation
Drop Test
Vibration
Dimensions
Weight
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Storage Temperature
Altitude
See models chart on pg 1.
1.4m from table top to wooden platform, 6
faces.
Relative Humidity
-40°C to +85°C
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
See outline drawings
MTBF
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
E-Cap Life
250g
>8 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
www.slpower.com
ME40 Rev 2.2
2-MAY-16
2
ME40
40W Single Output External Power
Medical Series
EMI/EMC Compliance
Conducted Emissions:
EN55011/CISPR22 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions:
EN55022/CISPR22 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise:
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
Conducted Disturbances induced by RF
Fields
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5.
Rated Power frequency magnetic fields
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
Voltage Interruptions, Dips, Sags &
Surges
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions
EN55011/EN61000-3-2, Class A
Flicker Test
EN61000-3-3
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Mechanical Drawing
54
33.6
1450±50
1450±50
96
95.5
55
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Notes:
www.slpower.com
Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
1. All dimensions in mm.
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
ME40 Rev 2.2
2-MAY-16
3
ME40
40W Single Output External Power
Medical Series
Derating Chart:
ME40 Series Power Derating
45
Output Power- Watts
40
35
30
25
12V
20
18V
15
24V
10
5
0
-20
-10
0
10
20
30
40
50
60
70
Ambient Temperature- ͦC
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
ME40 Rev 2.2
2-MAY-16
4