GLD140 Safety Certificates

File E116994
Project 00SC14101
September 27, 2000
REPORT
ON
COMPONENT - POWER SUPPLIES,
MEDICAL AND DENTAL EQUIPMENT
Condor DC Power Supplies, Inc.
Oxnard, California
Copyright © 2000 Underwriters Laboratories Inc.
Underwriters Laboratories Inc. authorizes the above named company to
reproduce this Report provided it is reproduced in its entirety.
Underwriters Laboratories Inc. authorizes the above named company to
reproduce that portion of this Report consisting of this Cover Page through
Page 4.
File E116994
Vol. 1
Sec. 70
and Report
Page 1
Issued:
Revised:
2000-09-27
2005-12-14
DESCRIPTION
PRODUCT COVERED:
USR, CNR - COMPONENT, Switching Power Supply, Medical and Dental,
Models GLD140A-102, GLD140H-101, GLD140X-C or -CF where X equals any alpha
character A through Z which represents various output configurations, -C is
a slotted cover option and –CF is a solid cover with fan option. . Models
may be followed by suffix –XXX and/or G, where XXX may be any number from
001 thru 999 and suffix G indicates compliance to RoHS (RoHS compliance has
not been evaluated by UL).
ELECTRICAL RATINGS:
Input:
Output:
100-240 V ac, 50/60 Hz, 3.0 A.
Output
Output
Output
Output
#1
#2
#3
#4
Volts
5
3.3 to 24
5 to 15
-15 to -12
Convection
Amperes
16 max
10 max
4 max
0.75 max
Airflow
Amperes
20 max
12 max
5 max
1.2 max
Power = 180 Watts maximum with cover/fan option or slotted
Cover with a minimum of 150 LFM airflow.
140 Watts maximum with no airflow and no cover.
Fan output is 12 V dc, 0.25 A
GLD140H, GLD140H-101*: +5 V/6 A, +12 V/2.5 A, +15 V/2 A, -15
V/2A ª Maximum output power = 120W with Convection Cooling
ENGINEERING CONSIDERATIONS (NOT FOR FIELD REPRESENTATIVE'S USE):
USR – indicates investigation to the Standard for Medical Electrical
Equipment, UL 60601-1, First Edition
CNR – indicates investigation to Canadian Standard CSA C22.2 No. 601.1
For use in product where the acceptability of the combination is
determined by Underwriters Laboratories Inc.
File E116994
Vol. 1
Sec. 70
and Report
Page 1A
Issued:
New:
2000-09-27
2005-05-31
ENGINEERING REFERENCES:
Following Illustrations are provided for engineering references:
Ill.
Ill.
Ill.
Ill.
1
2
3
4
–
–
–
–
Isolation Diagram
Installation Instruction Sheet
PWB Trace Layout
Transformer Construction Detail (T3 & T4)
File E116994
Vol. 1
Sec. 70
and Report
Page 2
Issued:
Revised:
2000-09-27
2005-02-02
Condition of Acceptability - When installed in the end-use equipment,
the following are among the considerations to be made:
1.
This component has been judged on the basis of the required spacings
in the First Edition of the Standard for Medical Electrical
Equipment, Part 1: General Requirements for Safety, UL 60601-1,
which covers the end use product for which the component is designed.
2.
The component shall be installed in compliance with the enclosure,
mounting, spacing, casualty markings and segregation requirements of
the end-use application.
3.
Consideration should be given to measuring the temperature on power
electronic components and transformer windings when the power supply
is installed in the end-use equipment.
4.
The input/output connectors are not acceptable for field connections,
they are only intended for connection to mating connectors of
internal wiring inside the end-use machine.
5.
The output circuits have not been evaluated for direct patient
connection (Type B, BF or CF).
6.
The component should be properly bonded to ground in the end-use
equipment.
7.
The Temperature Test was performed in a raised ambient of 50°C.
8.
The isolation transformers, T3 and T4, complies with Class 155
limits.
9.
Leakage current testing should be repeated in the end product
application.
10.
The power supply was evaluated as reinforced insulation between
primary and secondary; basic insulation between primary to ground.
11.
This power supply has been evaluated as Class I, continuous
operation, ordinary equipment and has not been evaluated for use in
the presence of a flammable anesthetic mixture with air, oxygen, or
nitrous oxide.
File E116994
Vol. 1
Sec. 70
and Report
Page 3
Issued:
09-27-00
12.
These power supplies have been evaluated for patient care equipment,
but not patient connected.
13.
Under normal and single fault conditions, the outputs do not exceed
25 V ac or 60 V dc.
14.
The grounding trace on the pwb has not been evaluated as the
Protective Earthing path for any metal parts accessible in the end
product. However, the grounding path was subjected to the Earthing
Test per Clause 18 (30 A for 2 minutes) with acceptable results.
15.
This unit has not been evaluated for use with a slotted cover without
airflow.
File E116994
Vol. 1
Sec. 70
and Report
Page 4
Issued:
Revised:
2000-09-27
2005-12-14
CONSTRUCTION DETAILS:
See Section General, Construction Details.
Trace Layout - Field Representative to insure that the trace layout
has not changed. See Ill. 3.
Model Differences - All models are the same, except for output
ratings and differences in secondary, low voltage circuitry. -C indicates
slotted cover option, -CF indicates cover and fan option.
GLD140H uses a 5 pin regulator for output #4, which requires a
different PWB. New trace layout is added to Ill. 3.
GLD140H-101 is identical to the model GLD140H-101 except it is
provide with a slotted cover (item 23)
GLD140A-102 is identical to model GLD140A except Line-to-Ground
capacitors C3 & C4 value is changed from 1000 pF to 680 pF to reduce earth
leakage current.
Models followed by the suffix –XXX represents value added
configurations that have no impact on safety.
Models followed by the suffix G indicates compliance to RoHS
(RoHS compliance has not been evaluated by UL).