GLD140 Instructions

DC POWER SUPPLIES INC.
2311 Statham Pkwy
Oxnard, CA 93033 + 805-486-4565
www.condorpower.com
GLD140 SERIES
INSTALLATION INSTRUCTIONS
MODEL NUMBERS: GLD140X-C or -CF where X equals any alpha character A through Z, -C is a slotted cover option
and –CF is a solid cover with fan option. Models may be followed by suffix –XXX and/or G, where XXX may be any
number from 001 thru 999. The –XXX suffix are used for value added configurations that have no impact on safety and
suffix G indicates compliance to RoHS.
RATINGS:
Input: 100-240 V ac, 3.0 A, 50/60 Hz
Output:
Max Output Power at ambient of 50°C:
180 Watts maximum with cover/fan option or slotted cover with a minimum of 150 LFM airflow.
140 Watts maximum with no airflow and no cover.
Max Output Current in table below is with Convection / 150 LFM airflow cooling except GLD140H which is convection
only.
Model
GLD140A
GLD140B
GLD140C
GLD140D
GLD140E
GLD140F
GLD140G
GLD140H
Output #1 Note 2
Volts
Current
5
16 / 20 A
5
16 / 20 A
5
16 / 20 A
5
16 / 20 A
5
16 / 20 A
5
16 / 20 A
5
16 / 20 A
5
6A
Output #2
Volts
Current
12
6/8A
12
6/8A
12
6/8A
24
4/6A
24
4/6A
15
5/7A
3.3
10 / 12 A
12
2.5 A
Output #3
Volts Current
12
4/5A
5
4/5A
15
4/5A
12
4/5A
15
4/5A
12
4/5A
12
4/5A
15
2A
Output #4
Volts
Current
-12
0.75 / 1.2 A
-12
0.75 / 1.2 A
-15
0.75 / 1.2 A
-12
0.75 / 1.2 A
-15
0.75 / 1.2 A
-12
0.75 / 1.2 A
-12
0.75 / 1.2 A
-15
2A
Notes:
1. For all models, fan output is 12 V dc, 0.25 A max.
2. Minimum output current of 3 A is required on output #1 for proper regulation.
3. Maximum Operating Relative Humidity 96 %, no condensation.
4. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing conditions before application of
power.
CERTIFICATION: All models are Certified to be in compliance with the applicable requirements of UL 60601-1, CSA
22.2 No. 601.1, EN 60601-1, and IEC 60601-1.
(5.1) Protection against electric shock = Class I
CLASSIFICATION:
(In accordance with sub- (5.2) Degree of protection against electric shock = Not acceptable for applied part without
clause 5 of IEC 60601-1) additional isolation (contact factory for details)
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anaesthetic mixture with
air, oxygen, or nitrous oxide. This evaluation is to be made on the end equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION: Condor DC Power Supplies, Inc. declares under our sole responsibility that all
models listed above are in conformity with the applicable requirements of EN 60950-1 following the provisions
of the Low Voltage Directive 73/23/EEC.
GROUNDING: Protection Class I requires that the ground terminal be bonded to Protective Earth in the end application.
Using this terminal for the primary system earthing terminal is not recommended.
OUTPUTS: It is recommended that the output common or return be connected to Protective Earth in the end application.
The output is not acceptable for patient connection without additional isolation. The DC output is SELV under normal and
single fault conditions. Output #3 is isolated from Ground and may be referenced positive or negative.
41-34338-0001 Rev. G 11/30/05
Page 1 of 2
GLD140 SERIES INSTALLATION INSTRUCTIONS
OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. In some applications where an
overvoltage condition could result in a hazard as defined in applicable safety standards, redundant or additional overvoltage
protection may be required. Consult factory for details.
CAUTION: When performing Dielectric Strength Tests, catastrophic failure of the unit may result if a Dielectric Strength
test voltage greater than 1800 V ac is applied between primary and secondary circuits. The components providing isolation
from primary to secondary cannot be tested while installed in the power supply without overstressing basic (primary to
ground) insulation. All isolating components are individually 100 % tested at 4800 V ac prior to installation.
TEMPERATURES: The maximum operating temperatures of certain safety components, as defined in the applicable safety
standards, must not be exceeded after installation to preserve the intended safety. The output power, ambient air temperature
and the availability, amount, direction and/or restriction of airflow influence the temperatures of these components.
FUSING: Fuses for both Line and Neutral are provided in the power supply. A blown fuse is an indication of catastrophic
failure of circuit component(s). Repair must be performed by Condor authorized personnel. Refer to fuse markings for type
and rating.
WARNING! SHOCK HAZARD! Dangerous voltages are present on some components, printed wiring traces and heatsinks.
EXPLANATION OF SYMBOLS
Alternating Current
Direct Current
Attention, Consult
Accompanying Documents
Attention, Dangerous Voltages
Earth (Ground)
CONNECTIONS
J1
1
3
5
AC Input
Earth
AC Neutral
AC Line
J2
1
2
3
4
5
6
Signal Out
+ Sense
- Sense
Power Fail
Power Good
Common
Inhibit
J3
9, 10, 19, 20
5, 14, 15
4
1
11
2
3, 6-8, 12, 13, 16-18
DC Output
V1
V2
V2 Sense
V3
V3 Return
V4
Common
1
2
J5 FAN
Fan Return
+ Fan
MATING CONNECTORS
J1: Molex P/N 39-01-4051 (Housing)
Molex P/N 39-00-0038 (Pins)
J2 and FAN: Amp MTA-100 Receptacle
J3: Molex P/N 39-01-2200 (Housing)
Molex P/N 39-00-0038 (Pins)
CAUTION: Do not exceed 7 A per pin on connector J3
Condor DC Power Supplies Inc. will not be liable for the safety, reliability or performance of these power supplies if a) any changes, modifications or repairs are carried out by other
than authorized agents of Condor DC Power Supplies Inc., or b) the installation of the supply is not in accordance with these installation instructions and the applicable UL, CSA,
and EN/IEC safety standards.
41-34338-0001 Rev. G 11/30/05