Bridge Amplifier for Strain Gage Full Bridge DABU AD2T-FB Features • Industrial bridge amplifier for S/G full bridge • For cyclical and static applications with reset function • Voltage output • Protection class IP 65 • Analog signal path Order Code Output signal ±10 V calibrated (max. ±12 V) Characteristic curve < 0,2% deviation Supply voltage range 18 - 33 VDC Current draw < 60 mA < 40 mA @ 24 VDC Bridge excitation approx. 7 VDC S/G bridge resistance350 Ω (RC) Output impedance 22 Ω Tare accuracy 0.25 < 15 mV 0.50 < 10 mV 1.00 < 5 mV 2.00 < 5 mV Reset input active 5 - 33 VDC < 2 mA inactive < 1 VDC Tare range ±6 mV/V Reset puls > 1 ms Reset settle time < 5 ms Frequency range (-3 dB) 1'000 Hz Signal polarity Bipolar Noise (0 ... 5 kHz) 0.25 < 15 mVpp 0.50 < 7,5 mVpp 1.00 < 5 mVpp 2.00 < 5 mVpp DABU AD2T-FB Mechanical Data Control connection5 pin male (Series 713) Gain 0.50 1.00 1.25 2.00 0,50 mV/V = 0 - 10 V 1,00 mV/V = 0 - 10 V 1,25 mV/V = 0 - 10 V 2,00 mV/V = 0 - 10 V Dimensions (mm) 4 40 40 79 Delivery Contents • Mounting screw 2 pcs. M4 x 30 Accessories (not included in delivery) Series 713 Connector female, control side, 5-pin, Part No. 10135462 max. cable length 10 m M12 x 1 Enclosurealuminum anodised Connector female with cable, control side, 5-pin Environmental Conditions ESG 34CH0200G 5-pin (shielded) 2 m, PUR, (Part No. 11046264) Specified temp. range0...+70 °C Storage temperature-40...+100 °C Protection classIP 65 EMCEN 61000-6-2 Immunity EN 61000-6-3 Emission 6.8 14 14 29 29 79 Sensor connection4 pin female (Series 712) Operating temp. range-25...+85 °C 4 26 M12 x1 M12 x1 Electrical Data ESG 34CH0500G 5-pin (shielded) 5 m, PUR, (Part No. 11046266) ESG 34CH1000G 5-pin (shielded) 10 m, PUR, (Part No. 10155587) www.baumer.com 26 DABU AD2T-FB Electrical Connection Sensor side Series 712 Pin Control side Series 713 2 3 4 5 3 1 4 1 2 Pin number Signal EXC.+ Signal 1 Full bridge EXC.+ 1+Vs 2 Full bridge SIG.- 2-VOUT 3 Full bridge SIG.+ 3GND 4 Full bridge EXC.- 4+VOUT 5 S/G Bridge SIG.+ EXC.- Reset Control SIG.- Reset Function Transducer DABU max. 10m Control Unit mV/V EXC.+ 1 1 +VS SIG.- 2 3 GND SIG.+ 3 2 -VOUT EXC.- 4 4 +VOUT U 6 max. 10m 5 Reset 18 - 33VDC RL ±10V 5 - 33 VDC Potential equalisation www.baumer.com VOUT Output signal mV/V Input signal Reset Reset input (active high) t1 Reset delay (< 0,3 ms) t2 Reset time (< 5 ms) t3 Reset impuls (> 1 ms) 6.9