EN

Bridge Amplifier for 2 x 1/4 Strain Gage Bridge
DABI AD2T-2Q
Features
• Industrial bridge amplifier
for 2 x 1/4 S/G bridge
• For cyclical and static applications
with reset function
• Current output
• Protection class IP 65
Electrical Data
Order Code
Current draw
< 90 mA
< 70 mA @ 24 VDC
Bridge excitation
approx. 9 VDC
Gain
0250 0250 µε = 4 - 20 mA
0500 0500 µε = 4 - 20 mA
1000 1000 µε = 4 - 20 mA
/C Tension leads to a positive output signal
Dimensions (mm)
S/G bridge resistance350 Ω
< 400 Ω
Tare accuracy
0250 < 30 µA
0500 < 20 µA
1000 < 16 µA
Reset input active 5 - 33 VDC < 2 mA
inactive< 1 VDC
Tare range
±6 mV/V
Reset puls
> 1 ms
Reset settle time
< 5 ms
4
40
40
M12 x1
Burden
79
79
Delivery Contents
Frequency range (-3 dB) 1'000 Hz
• Mounting screw 2 pcs. M4 x 30
Noise
Accessories (not included in delivery)
(0 ... 5 kHz)
0250 < 30 µApp
0500 < 20 µApp
1000 < 16 µApp
Mechanical Data
Control connection5 pin male (Series 713)
Series 713
Connector female, control side, 5-pin, Part No. 10135462
max. cable length 20 m
Sensor connection4 pin female (Series 712)
M12 x 1
Enclosurealuminum anodised
Environmental Conditions
Connector female with cable, control side, 5-pin
Operating temp. range-25...+85 °C
ESG 34CH0200G 5-pin (shielded) 2 m, PUR,
(Part No. 11046264)
Specified temp. range0...+70 °C
Storage temperature-40...+100 °C
Protection classIP 65
EMCEN 61000-6-2 Immunity
EN 61000-6-3 Emission
6.6
4
26
14
14 - 33 VDC
/C
14
Supply voltage range
DABI AD2T-2Q
29
0250 < 0,5%
0500 < 0,25%
1000 < 0,2%
29
4 - 20 mA calibrated
Characteristic curve deviation
M12 x1
Output signal
ESG 34CH0500G 5-pin (shielded) 5 m, PUR,
(Part No. 11046266)
ESG 34CH1000G 5-pin (shielded) 10 m, PUR,
(Part No. 10155587)
www.baumer.com
26
DABI AD2T-2Q
Electrical Connection
Sensor side Series 712
S/G Bridge
Control side Series 713
EXC.+
Pin
2
3
4 5
3
1
4
1
2
Signal
Pin number
R
DMS 1
EXC.+
1+Vs
2
DMS 2
SIG.-
2n.c.
3
DMS 3
SIG.+
3GND
4
DMS 4
EXC.-
4+IOUT
5
SIG.+
Signal
1
SIG.-
R
DMS 2
C
EXC.-
Reset
Control
DMS 1
C
Reset Function
Transducer
DABI
max. 10m
EXC+ 1
ε
Control Unit
1 +Vs
SIG-
2
3 GND
SIG+
3
2
EXC-
4
4 +IOUT
I
6
max. 20m
5 Reset
14 - 33VDC
RL
4 - 20mA
5 - 33 VDC
Potential equalisation
www.baumer.com
IOUT
Output signal
mV/V
Input signal
Reset
Reset input (active high)
t1
Reset delay (< 0,3 ms)
t2
Reset time (< 5 ms)
t3
Reset impuls (> 1 ms)
6.7