SITRANS P210 catalog specifications

© Siemens AG 2015
Pressure Measurement
Transmitters for basic requirements
SITRANS P210 for gauge pressure
■ Overview
■ Design
Device structure without explosion protection
The pressure transmitter consists of a piezoresistive measuring
cell with a diaphragm installed in a stainless steel enclosure. It
can be used with a connector per EN 175301-803-A (IP65), a
round plug M12 (IP67), a cable (IP67) or a Quickon cable quick
screw connection (IP67) connected electrically. The output signal is between 4 and 20 mA or 0 and 10 V.
Device structure with explosion protection
The pressure transmitter consists of a piezoresistive measuring
cell with a diaphragm installed in a stainless steel enclosure. It
can be used with a connector per EN 175301-803-A (IP65) or a
round plug M12 (IP67) connected electrically. The output signal
is between 4 and 20 mA.
■ Function
The pressure transmitter SITRANS P210 measures the gauge
pressure of liquids, gases and vapors.
• Stainless steal measuring cell
• Measuring ranges 100 to 600 mbar (1.45 to 8.7 psi) relative
• For low-pressure applications
■ Benefits
•
•
•
•
•
•
High measuring accuracy
Rugged stainless steel enclosure
High overload withstand capability
For aggressive and non-aggressive media
For measuring the pressure of liquids, gases and vapors
Compact design
■ Application
The pressure transmitter SITRANS P210 for gauge pressure is
used in the following industrial areas:
• Mechanical engineering
• Shipbuilding
• Power engineering
• Chemical industry
• Water supply
The pressure transmitter measures the gauge pressure of liquids
and gases as well as the level of liquids.
Mode of operation
U
const.
U
p
I
I0, UB
SITRANS P210 pressure transmitters (7MF1566-...), functional diagram
The stainless steel measuring cell has a thin-film resistance
bridge to which the operating pressure p is transmitted through
a stainless steel diaphragm.
The voltage output from the measuring cell is converted by an
amplifier into an output current of 4 to 20 mA or an output voltage
of 0 to 10 V DC.
The output current and voltage are linearly proportional to the input pressure.
Siemens FI 01 · 2015 US Edition
1/11
1
© Siemens AG 2015
Pressure Measurement
Transmitters for basic requirements
1
SITRANS P210 for gauge pressure
■ Technical specifications
Application
Design
Gauge measurement
Liquids, gases and vapors
Mode of operation
Measuring principle
Piezoresistive measuring cell
(stainless steel diaphragm)
Measured variable
Gauge pressure
Weight
Approx. 0.090 kg (0.198 lb)
Process connections
See dimension drawings
Electrical connections
• Connector per
EN 175301-803-A Form A with
cable inlet M16x1.5 or ½-14 NPT
or Pg 11
• M12 connector
• 2 or 3-wire (0.5 mm2) cable
(  5.4 mm)
• Quickon cable quick screw connection
Inputs
Measuring range
• Gauge pressure
100 … 600 mbar
(1.5 … 8.7 psi)
Output
Current signal
4 ... 20 mA
• Load
(UB - 10 V)/0.02 A
• Auxiliary power UB
DC 7 ... 33 V (10 ... 30 V for Ex)
Voltage signal
0 ... 10 V DC
• Load
 10 k
• Auxiliary power UB
12 ... 33 V DC
• Power consumption
< 7 mA at 10 k
Characteristic curve
Linear rising
Measuring accuracy
Error in measurement at limit setting
incl. hysteresis and reproducibility
Step response time T99
• Typical: 0.25 % of full-scale
value
• Maximum: 0.5 % of full-scale
value
< 5 ms
Long-term stability
• Lower range value and measuring
span
0.25 % of full-scale value/year
Influence of ambient temperature
• Lower range value and measuring
span
• 0.25 %/10 K of full-scale value
• 0.5 %/10K of full-scale value
for a measuring range
100 … 400 mbar
Wetted parts materials
• Measuring cell
Stainless steel, mat.-No. 1.4435
• Process connection
Stainless steel, mat. No. 1.4404
(SST 316 L)
• Gasket
• FPM (Standard)
• Neoprene
• Perbunan
• EPDM
Non-wetted parts materials
• Enclosure
Stainless steel, mat. No. 1.4404
(SST 316 L)
• Rack
Plastic
• cables
PVC
Certificates and approvals
Classification according to pressure
equipment directive
(PED 97/23/EC)
For gases of fluid group 1 and liquids of fluid group 1;
meets requirements as per article
3, paragraph 3 (good engineering
practice)
Lloyd‘s Register of Shipping (LR)
12/20010
Germanischer Lloyd (GL)
GL19740 11 HH00
American Bureau of Shipping (ABS) ABS_11_HG 789392_PDA
Bureau Veritas (BV)
BV 271007A0 BV
Det Norske Veritas (DNV)
A 12553
Conditions of use
Drinking water approval (ACS)
ACS 11 ACC NY 055
Process temperature with gasket
made of:
GOST
GOST-R
• Influence of power supply
0.005 %/V
• FPM (Standard)
-15 ... +125 °C (+5 ... +257 °F)
• Neoprene
-35 ... +100 °C (-31 ... +212 °F)
• Perbunan
-20 ... +100 °C (-4 ... +212 °F)
• EPDM
-40 ... +145 °C (-40 ... +293 °F),
usable for drinking water
Ambient temperature
-25 ... +85 °C (-13 ... +185 °F)
Storage temperature
-50 ... +100 °C (-58 ... +212 °F)
Degree of protection (to EN 60529)
• IP 65 with connector per
EN 175301-803-A
• IP 67 with M12 connector
• IP 67 with cable
• IP 67 with cable quick screw
connection
Electromagnetic compatibility
• acc. IEC 61326-1/-2/-3
• acc. NAMUR NE21, only for
ATEX versions and with a max.
measuring deviation 1 %
Mounting position
upright
1/12
Siemens FI 01 · 2015 US Edition
Underwriters Laboratories (UL)
• for USA and Canada
UL 20110217 - E34453
• worldwide
IEC UL DK 21845
Explosion protection
Intrinsic safety "i"
(only with current output)
Ex II 1/2 G Ex ia IIC T4 Ga/Gb
Ex II 1/2 D Ex ia IIIC T125 °C
Da/Db
EC type-examination certificate
SEV 10 ATEX 0146
Connection to certified intrinsicallysafe resistive circuits with maximum values:
Ui  30 V DC; Ii  100 mA;
Pi  0.75 W
Effective internal inductance and
capacity for versions with plugs per
EN 175301-803-A and M12
Li = 0 nH; Ci = 0 nF
© Siemens AG 2015
Pressure Measurement
Transmitters for basic requirements
SITRANS P210 for gauge pressure
■ Selection and ordering data
Article No.
Order code
7M F 1 5 6 6 - 7 7 7 7 7 - 7 7 7 7 7 7 7
SITRANS P 210 pressure transmitters for gauge pressure for low pressure applications
Accuracy typ. 0.25 %
Wetted parts materials: Stainless steel + sealing material
Non-wetted parts materials: stainless steel
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Measuring range
Overload limit
min.
Burst pressure
max.
For gauge pressure
0...100 mbar
0...160 mbar
0...250 mbar
0...400 mbar
0...600 mbar
(1.45 psi)
(2.32 psi)
(3.63 psi)
(5.8 psi)
(8.7 psi)
-400 mbar
-400 mbar
-800 mbar
-800 mbar
-1000 mbar
(-5.8 psi)
(-5.8 psi)
(-11.6 psi)
(-11.6 psi)
(-14.5 psi)
400 mbar
400 mbar
1000 mbar
1000 mbar
2000 mbar
(5.8 psi)
(5.8 psi)
(14.5 psi)
(14.5 psi)
(29.0 psi)
1 bar
1 bar
2 bar
2 bar
3 bar
3 AA
3 AB
3 AC
3 AD
3 AG
(14.5 psi)
(14.5 psi)
(29.0 psi)
(29.0 psi)
(43.5 psi)
9 AA
Other version, add Order code and plain text:
Measuring range: ... up to ... mbar (psi)
H1Y
Output signal
0
10
4 ... 20 mA; two-wire system; power supply 7 ... 33 V DC (10 ... 30 V DC for ATEX versions)
0 ... 10 V; three-wire system; power supply 12 ... 33 V DC
Explosion protection (only 4 ... 20 mA)
0
1
None
With explosion protection Ex ia IIC T4
Electrical connection
Connector per DIN EN 175301-803-A, stuffing box thread M16 (with coupling)
Connection via fixed mounted cable, 2 m (not for type of protection "Intrinsic safety i")
Quickon cable quick screw connection PG9 (not for type of protection "Intrinsic safety i")
Connector per DIN EN 175301-803-A, stuffing box thread 1/2"-14 NPT (with coupling)
Connector per DIN EN 175301-803-A, stuffing box thread PG11 (with coupling)
Fixed mounted cable, length 5 m
Special version
0
0
0
1
3
4
5
6
7
9
N1Y
Process connection
G½" male per EN 837-1 (½" BSP male) (standard for metric pressure ranges mbar, bar)
G½" male thread and G1/8" female thread
G¼“ male per EN 837-1 (¼“ BSP male)
7/16"-20 UNF male
A
B
C
D
¼"-18 NPT male (standard for pressure ranges inH2O and psi)
¼"-18 NPT female
½"-14 NPT male
½"-14 NPT female
7/16"-20 UNF female
M20x1.5 male
E
F
G
H
J
P
Special version
Z
P1Y
Sealing material between sensor and enclosure
A
B
C
D
Z
Viton (FPM, standard)
Neoprene (CR)
Perbunan (NBR)
EPDM
Special version
Version
Standard version
Q1Y
1
Further designs
Supplement the Article No. with "-Z" and add Order code.
Manufacturer’s test certificate M per IEC 60770-2 (calibration certificate) supplied
C11
Siemens FI 01 · 2015 US Edition
1/13
1
© Siemens AG 2015
Pressure Measurement
Transmitters for basic requirements
SITRANS P210 for gauge pressure
■ Dimensional drawings
G1/2"
25
(1.0)
25
(1.0)
20
(0.8)
G1/8"
M16x1.5
or
1/2-14 NPT
20
(0.8)
M12x1/Fixcon
51 (2.0)
10.7 (0.42)
36 (1.42)
25.5 (1.0)
M20 x 1.5
G1/2"
1/2-14 NPT
max. 30 Nm
24 (0.94)
SITRANS P210, electrical connections, dimensions in mm (inch)
20
(0.8)
16
(0.6)
Cable gland
with cable
16
(0.6)
30.2 (1.19)
G1/4"
7/16"-20 UNF
1/4-18 NPT
25
(1.0)
max. 20 Nm
7/16"-20 UNF
max. 20 Nm
23
(0.9)
30
(1.2)
1
1/2"-14 NPT
1/2"-18 NPT
max. 20 Nm
SITRANS P210, process connections, dimensions in mm (inch)
1/14
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Pressure Measurement
Transmitters for basic requirements
SITRANS P210 for gauge pressure
■ Schematics
1+
1
2
IO
Connection: 1 (+),
2 (-)
+
2-
UB
RL
IO Output current
RL Load
UB Power supply
Connection with current output and connector per EN 175301
1+
IO
UB
RL
3-
br
IO
gn
UB
RL
1
1+
IO
2-
4 3
1
UB
2
UO
+
UB
RL
4+
3-
Connection: 1 (+UB),
3 (-),
4 (+UO)
UO Output voltage
RL Load
UB Power supply
Connection with voltage output and connector M12x1
br
UO
gn
+
UB
RL
ws
Connection: br (+UB, brown),
gn (+UO, green),
ws (-, white)
UO Output voltage
UB Power supply
RL Load
Connection with voltage output and cable
1
Connection: 1 (+),
2 (-)
+
RL
1+
IO Output current
UB Power supply
RL Load
Connection with current output and cable
Connection: 1 (+UB),
2 (-),
3 (+UO)
UO Output voltage
RL Load
UB Power supply
Connection with voltage output and connector per EN 175301
Connection: br (brown),
gn (green)
+
UB
RL
2-
IO Output current
RL Load
UB Power supply
Connection with current output and connector M12x1
+
UO
3+
Connection: 1 (+),
3 (-)
+
3
1
1+
1
3
2
IO Output current
RL Load
UB Power supply
Connection with current output and Quickon cable quick screw connection
1+
3+
2
3
+
UO
RL
UB
2-
Connection: 1 (+UB),
3 (-),
2 (+UO)
UO Output voltage
RL Load
UB Power supply
Connection with voltage output and Quickon cable quick screw connection
Version with explosion protection: 4 ... 20 mA
The grounding connection is conductively bonded to the transmitter enclosure
1+
1
2
2-
+
IO
RL
UB
Connection: 1 (+),
2 (-)
IO Output current
RL Load
UB Power supply
Connection with current output and connector per EN 175301 (Ex)
1+
4 3
1
34
IO
+
RL
UB
Connection: 1 (+),
3 (-),
4( )
IO Output current
RL Load
UB Power supply
Connection with current output and connector M12x1 (Ex)
Siemens FI 01 · 2015 US Edition
1/15
1