P300 for gauge and absolute pressure catalog specifications

© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
■ Overview
The SITRANS P300 is a digital pressure transmitter for relative
and absolute pressure. The conventional thread versions are
available as process connections, as are flush-mounted versions. A large number of the flush-mounted versions are suitable
for food and pharmaceutical applications, and satisfy the
EHEDG and 3A hygiene requirements.
The output signal is a load-independent direct current from 4 to
20 mA or a PROFIBUS PA or FOUNDATION signal, which is linearly proportional to the input pressure. Communication is via
HART protocol or PROFIBUS PA interface. Convenient buttons
for easy local operation of the basic settings of the pressure
transmitter.
The SITRANS P300 has a single-chamber stainless steel casing.
The pressure transmitter is approved with "intrinsically safe" type
of protection. It can be used in zone 1 or zone 0.
■ Benefits
With appropriate parameter settings, it can also be used for the
following additional measurement types:
• Level
• Volume
• Mass
The "intrinsically-safe" Ex version of the transmitter can be installed in hazardous areas (zone 1). The transmitters are provided with an EC type examination certificate and comply with
the respective harmonized European standards of ATEX.
Gauge pressure
This variant measures aggressive, non-aggressive and hazardous gases, vapors and liquids.
• High quality and service life
• High reliability even under extreme chemical and mechanical
loads
• Extensive diagnosis and simulation functions
• Minimum conformity error
• Small long-term drift
• Wetted parts made of high-grade materials (such as stainless
steel, Hastelloy)
• Measuring range 0.008 bar to 400 bar (0.1 psi to 5802 psi)
• High measuring accuracy
• Parameterization over control keys and HART or
PROFIBUS PA or FOUNDATION Fieldbus
■ Application
The pressure transmitter is available in versions for gauge pressure and for absolute pressure. The output signal is always a
load-independent direct current from 4 to 20 mA or a
PROFIBUS PA or FOUNDATION Fieldbussignal, which is linearly
proportional to the input pressure. The pressure transmitter measures aggressive, non-aggressive and hazardous gases, as well
as vapors and liquids.
1/50
It can be used for the following measurement types:
• Gauge pressure
• Absolute pressure
Siemens FI 01 · 2015 US Edition
The smallest span is 0.01 bar (0.15 psi), the largest is 400 bar
(5802 psi).
Level
With appropriate parameter settings, the gauge pressure variant
measures the level of aggressive, non-aggressive and hazardous liquids.
For measuring the level in an open container you require one device; for measuring the level in a closed container, you require
two devices and a process control system.
Absolute pressure
This variant measures the absolute pressure of aggressive, nonaggressive and hazardous gases, vapors and liquids.
The smallest span is 0.008 bar a (0.12 psia), the largest is
30 bar a (435 psia).
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
■ Design
■ Function
The device comprises:
• Electronics
• Housing
• Measuring cell
Operation of electronics with HART communication
9
8
0.0.0.0.0
00
M
10
1
2
3
4
5
IA, UH
μC
5
7
HART interface
4
EEPROM
6
Electronics
6
Sensor
3
1
2
3
4
5
2
Digital display
Process connection
Cable gland
Rating plate
Screw-on lid
The housing has a screw-on lid (5) and, depending on the version, is with or without an inspection window. The electrical terminal housing, the buttons for operation of the device are located under this lid and, depending on the version, the display.
The connections for the auxiliary power UH and the shield are in
the terminal housing. The cable gland is mounted on the side of
the housing. The measuring cell with the process connection (2)
is located on the bottom of the housing. The measuring cell with
the process connection may differ from the one shown in the diagram, depending on the device version.
Example of attached measuring points sign
Y99 = max. 10 char.
Y16 = max. 27 char.
pe Measuring cell
1
2
3
4
5
6
Perspective view of SITRANS P300
Y01 or Y02
= max. 27 char.
Y15 = max. 16 char.
EEPROM
1
.... to .... mbar
Measuring point number (TAG No.)
1234
Measuring point text
Measuring cell sensor
Instrument amplifier
Analog-to-digital converter
Microcontroller
Digital-to-analog converter
One non-volatile memory each in the measuring cell and
electronics
7 HART interface
8 Three input keys (local operation)
9 Digital display
10 Diode circuit and connection for external ammeter
IA Output current
UH Power supply
Pe Input variable
Function diagram of electronics
The input pressure is converted into an electrical signal by the
sensor (1). This signal is amplified by the measuring amplifier (2)
and digitalized in an analog-to-digital converter (3). The digital
signal is analyzed in a microcontroller (4) and corrected according to linearity and thermal characteristics. In a digital-to-analog
converter (5) it is then converted into the output current of 4 to
20 mA. A diode circuit provides reverse polarity protection. You
can make an uninterrupted current measurement with a low-ohm
ammeter at the connection (10). The data specific to the measuring cell, the electronic data and parameter settings are stored in
two non-volatile memories (6). The first memory is linked to the
measuring cell, the second to the electronics.
The buttons (8) can be used to call up individual functions, socalled modes. If you have a device with a display (9), you can
use this to track mode settings and other messages. The basic
mode settings can be changed with a computer via the HART
modem (7).
Siemens FI 01 · 2015 US Edition
1/51
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Operation of electronics with PROFIBUS PA
communication
Operation of electronics with FOUNDATION Fieldbus
communication
9
9
5
μC
EEPROM
2
7
6
pe
1
2
3
4
5
6
7
Power
supply
unit
Coupler
7
FF
interface
12
EEPROM
6
Power supply
10
Power
supply
unit
Coupler
Electronics
BusMaster
1
6
Sensor
pe
Measuring cell
Measuring cell
Measuring cell sensor
Instrument amplifier
Analog-to-digital converter
Microcontroller
Electrical isolation
One non-volatile memory
each in the measuring cell
and electronics
PROFIBUS-PA interface
5
Power supply
10
11
6
Sensor
4
μC
Electronics
1
3
PA
interface
Foundation Fieldbus
4
PROFIBUS DP
3
0.0.0.0.0
00
M
PROFIBUS-PA
M
2
8
0.0.0.0.0
00
EEPROM
8
EEPROM
1
8
9
10
11
12
Three input keys
(local operation)
Digital display
Power supply
DP/PA coupler or link
Bus master
pe
Input variable
1
2
3
4
5
6
7
Measuring cell sensor
Instrument amplifier
Analog-to-digital converter
Microcontroller
Electrical isolation
One non-volatile memory
each in the measuring cell
and electronics
FF interface
8
9
10
Three input keys
(local operation)
Digital display
Power supply
pe
Input variable
Function diagram of electronics
Function diagram of electronics
The input pressure is converted into an electrical signal by the
sensor (1). This signal is amplified by the measuring amplifier (2)
and digitalized in an analog-to-digital converter (3). The digital
signal is analyzed in a microcontroller (4) and corrected according to linearity and thermal characteristics. It is then made available at the PROFIBUS PA over an electrically isolated
PROFIBUS PA interface (7). The data specific to the measuring
cell, the electronic data and parameter settings are stored in two
non-volatile memories (6). The first memory is linked to the measuring cell, the second to the electronics.
The bridge output voltage created by the sensor (1, Figure
"Function diagram of electronics") amplified by the measuring
amplifier (2) and digitized in the analog-to-digital converter (3).
The digital information is evaluated in the microcontroller, its linearity and temperature response corrected, and provided on the
FOUNDATION Fieldbus through an electrically isolated
FOUNDATION Fieldbus interface (7).
The buttons (8) can be used to call up individual functions, socalled modes. If you have a device with a display (9), you can
use this to track mode settings and other messages. The basic
mode settings (12) can be changed with a computer over the
bus master.
The data specific to the measuring cell, the electronics data, and
the parameter data are stored in the two non-volatile memories
(6). The one memory is coupled to the measuring cell, the other
to the electronics. As the result of this modular design, the electronics and the measuring cell can be replaced separately from
each other.
Using the three input buttons (8) you can parameterize the pressure transmitter directly at the measuring point. The input buttons can also be used to control the view of the results, the error
messages and the operating modes on the display (9).
The results with status values and diagnostic values are transferred by cyclic data transmission on the FOUNDATION Fieldbus. Parameterization data and error messages are transferred
by acyclic data transmission. Special software such as National
Instruments Configurator is required for this.
Mode of operation of the measuring cells
The process connections available include the following:
• G½
• ½-14 NPT
• Flush-mounted diaphragm:
- Flanges to EN
- Flanges to ASME
- NuG and pharmaceutical connections
1/52
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Measuring cell for gauge pressure
Measuring cell for gauge pressure, front-flush diaphragm
1
1
6
5
5
2
2
4
4
3
pe
1
2
3
4
5
6
pe
Reference pressure
Measuring cell
Process connection
Seal diaphragm
Filling liquid
Relative pressure sensor
Pressure as input
variable
1
2
3
4
5
pe
Reference pressure
Measuring cell
Seal diaphragm
Filling liquid
Relative pressure sensor
Pressure as input variable
Measuring cell for gauge pressure, function diagram
3
pe
The input pressure (pe) is transferred to the gauge pressure sensor (6) via the seal diaphragm (4) and the filling liquid (5), displacing its measuring diaphragm. The displacement changes
the resistance value of the four piezo resistors in the measuring
diaphragm in a bridge circuit. The change in the resistance
causes a bridge output voltage proportional to the input pressure.
Transmitters with spans  63 bar (  926.1 psi) measure the input
pressure compared to atmospheric, transmitters with spans of
 160 bar ( 2352 psi) compared to a vacuum.
Measuring cell for absolute pressure
Measuring cell for gauge pressure, front-flush diaphragm, function
diagram
The input pressure (pe) is transferred to the gauge pressure sensor (6) via the seal diaphragm (4) and the filling liquid (5), displacing its measuring diaphragm. The displacement changes
the resistance value of the four piezo resistors in the measuring
diaphragm in a bridge circuit. The change in the resistance
causes a bridge output voltage proportional to the input pressure.
Transmitters with spans  63 bar ( 926.1 psi) measure the input
pressure compared to atmospheric, transmitters with spans of
 160 bar ( 2352 psi) compared to a vacuum.
Measuring cell for absolute pressure, front-flush diaphragm
1
5
4
3
4
1
2
pe
1
2
3
4
5
pe
Measuring cell
Process connection
Seal diaphragm
Filling liquid
Absolute pressure sensor
Pressure as input
variable
Measuring cell for absolute pressure, function diagram
The input pressure (pe) is transferred to the absolute pressure
sensor (5) via the seal diaphragm (3) and the filling liquid (4),
displacing its measuring diaphragm. The displacement
changes the resistance value of the four piezo resistors in the
measuring diaphragm in a bridge circuit. The change in the resistance causes a bridge output voltage proportional to the input
pressure.
3
1
2
3
4
pe
Measuring cell
Seal diaphragm
Filling liquid
Absolute pressure sensor
Pressure as input variable
2
pe
Measuring cell for absolute pressure, front-flush diaphragm, function
diagram
Siemens FI 01 · 2015 US Edition
1/53
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
The input pressure (pe) is transferred to the absolute pressure
sensor (5) via the seal diaphragm (3) and the filling liquid (4),
displacing its measuring diaphragm. The displacement
changes the resistance value of the four piezo resistors in the
measuring diaphragm in a bridge circuit. The change in the resistance causes a bridge output voltage proportional to the input
pressure.
Adjustable parameters on SITRANS P300 with HART
communication
Parameters
Input keys
HART
communication
Start of scale
x
x
Full-scale value
x
x
Parameterization
Electrical damping
x
x
Depending on the version, there are a range of options for parameterizing the pressure transmitter and for setting or scanning
the parameters.
Start-of-scale value without applica- x
tion of a pressure ("Blind setting")
x
Full-scale value without application
of a pressure ("Blind setting")
x
x
Zero adjustment
x
x
current transmitter
x
x
Fault current
x
x
Parameterization using the input buttons (local operation)
With the input buttons you can easily set the most important parameters without any additional equipment.
Parameterization using HART communication
Parameterization using HART communication is performed with
a HART communicator or a PC.
+
230 ... 1100 Ω
SITRANS P
transmitter
Power supply
HART
communicator
Communication between a HART communicator and a pressure
transmitter
When parameterizing with the HART communicator, the connection is made directly to the 2-wire cable.
+
230 ... 500 Ω
SITRANS P
transmitter
HART
modem
Power supply
PC or
laptop
USB/RS 232
HART communication between a PC communicator and a pressure
transmitter
When parameterizing with a PC, the connection is made through
a HART modem.
The signals needed for communication in conformity with the
HART 5.x or 6.x protocols are superimposed on the output current using the Frequency Shift Keying (FSK) method.
1/54
Siemens FI 01 · 2015 US Edition
Disabling of buttons, write protection x
x1)
Type of dimension and actual
dimension
x
x
Input of characteristic
x
Freely-programmable LCD
x
Diagnostic functions
x
1)
Cancel apart from write protection
Diagnostic functions for SITRANS P300 with HART
communication
• Zero correction display
• Event counter
• Limit transmitter
• Saturation alarm
• Slave pointer
• Simulation functions
• Maintenance timer
Available physical units of display for SITRANS P300 with HART
communication
Table style: Technical specifications 2
Physical variable
Physical dimensions
Pressure (setting can also be
made in the factory)
Pa, MPa, kPa, bar, mbar, torr, atm, psi,
g/cm2, kg/cm2, inH2O, inH2O (4 °C),
mmH2O, ftH2O (20 °C), inHg, mmHg
Level (height data)
m, cm, mm, ft, in
Volume
m3, dm3, hl, yd3, ft3, in3, US gallon,
lmp. gallon, bushel, barrel, barrel
liquid
Mass
g, kg, t, lb, Ston, Lton, oz
Temperature
K, °C, °F, °R
Miscellaneous
%, mA
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Parameterization through PROFIBUS PA interface
Physical dimensions available for the display
Fully digital communication through PROFIBUS PA, profile 3.0, is
particularly user-friendly. The PROFIBUS connects the
SITRANS P300 PA to a process control system, e.g.
SIMATIC PSC 7. Communication is possible even in a potentially
explosive environment.
Physical variable
Physical dimensions
Pressure (setting can also
be made in the factory)
Mpa, kPa, Pa, bar, mbar, torr, atm, psi,
g/cm2, kg/cm2, mmH2O, mmH2O (4 °C),
inH2O, inH20 (4 °C), ftH2O (20 °C), mmHg,
inHg
For parameterization through PROFIBUS you need suitable software, e.g. SIMATIC PDM (Process Device Manager).
Level (height data)
m, cm, mm, ft, in, yd
Mass
g, kg, t, lb, Ston, Lton, oz
Parameterization through FOUNDATION Fieldbus interface
Volume
m3, dm3, hl, yd3, ft3, in3, US gallon, lmp.
gallon, bushel, barrel, barrel liquid
volume flow
m3/s, m3/min, m3/h, m3/d, l/s, l/min, l/h, l/ d,
Ml/d, ft3/s, ft3/min, ft3/h, ft3/d, US gallon/s,
US gallon/min, US gallon/h, US gallon/d,
bbl/s, bbl/min, bbl/h, bbl/d
Mass flow
g/s, g/min, g/h, g/d, kg/s, kg/min, kg/h,
kg/d, t/s, t/min, t/h, /t/d, lb/s, lb/min, lb/h,
lb/d, STon/s, STon/min, STon/h, STon/d,
LTon/s, LTon/min, LTon/h, LTon/d
Total mass flow
t, kg, g, lb, oz, LTon, STon
Temperature
K, °C, °F, °R
Miscellaneous
%
Fully digital communication through FOUNDATION Fieldbus is
particularly user-friendly. Through the FOUNDATION Fieldbus
the P300 is connected to a process control system. Communication is possible even in a potentially explosive environment.
For parameterization through the FOUNDATION Fieldbus you
need suitable software, e.g. National Instruments Configurator.
Adjustable parameters for SITRANS P300 with PROFIBUS PA
and FOUNDATION Fieldbus
Adjustable parameters
Input
keys
PROFIBUS PA and
FOUNDATION
Fieldbus interface
Electrical damping
x
x
Zero adjustment (correction of position)
x
x
Buttons and/or function disabling
x
x
Source of measured-value display
x
x
Physical dimension of display
x
x
Position of decimal point
x
x
Bus address
x
x
Adjustment of characteristic
x
x
Input of characteristic
x
Freely-programmable LCD
x
Diagnostic functions
x
Hygiene version
In the case of the SITRANS P300 with 7MF812.-... front-flush diaphragm, selected connections comply with the requirements of
the EHEDG or 3A. You will find further details in the order form.
Please note in particular that the seal materials used must comply with the requirements of 3A. Similarly, the filling liquids used
must be FDA-compliant.
Diagnostic functions for SITRANS P300 with PROFIBUS PA and
FOUNDATION Fieldbus
• Event counter
• Slave pointer
• Maintenance timer
• Simulation functions
• Display of zero correction
• Limit transmitter
• Saturation alarm
Siemens FI 01 · 2015 US Edition
1/55
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
■ Technical specifications
SITRANS P300 for gauge and absolute pressure
HART
PROFIBUS PA and FOUNDATION Fieldbus
Gauge pressure input
Measured variable
Gauge pressure
Spans (infinitely adjustable) or nominal mea- Span (min. ... max.)
suring range and
max. permissible test pressure
Max. perm. test
pressure
Nominal measuring
range
Max. perm. test
pressure
0.01 ... 1 bar
(0.15 ... 14.5 psi)
6 bar
(87 psi)
1 bar
(14.5 psi)
6 bar
(87 psi)
0.04 ... 4 bar
(0.58 ... 58 psi)
10 bar
(145 psi)
4 bar
(58 psi)
10 bar
(145 psi)
0.16 ... 16 bar
(2.3 ... 232 psi)
32 bar
(464 psi)
16 bar
(232 psi)
32 bar
(464 psi)
0.6 ... 63 bar
(9.1 ... 914 psi)
100 bar
(1450 psi)
63 bar
(914 psi)
100 bar
(1450 psi)
1.6 ... 160 bar
(23.2 ... 2321 psi)
250 bar
(3626 psi)
160 bar
(2321 psi)
250 bar
(3626 psi)
4.0 ... 400 bar
(58 ... 5802 psi)
600 bar
(8700 psi)
400 bar
(5802 psi)
600 bar
(8700 psi)
Depending on the process connection, the span
may differ from these values
Depending on the process connection, the nominal measuring range may differ from these values
Lower measuring limit
• Measuring cell with silicone oil
30 mbar a (0.44 psia)
Upper measuring limit
• Measuring cell with silicone oil
100% of max. span
100 % of the max. nominal measuring range
Absolute pressure input
Measured variable
Absolute pressure
Spans (infinitely adjustable) or nominal mea- Span (min. ... max.)
suring range and
max. permissible test pressure
Max. perm. test
pressure
Nominal measuring
range
Max. perm. test
pressure
8 ... 250 mbar a
(0.12...3.63 psia)
6 bar a
(87 psia)
250 mbar a
(3.63 psia)
6 bar a
(87 psia)
43 ... 1300 mbar a
(0.62...18.9 psia)
10 bar a
(145 psia)
1.30 bar a
(19 psia)
10 bar a
(145 psia)
0.16 ... 5 bar a
(2.3 ... 73 psia)
30 bar a
(435 psia)
5 bar a
(73 psia)
30 bar a
(435 psia)
1 ... 30 bar a
(14.5 ... 435 psia)
100 bar a
(1450 psia)
30 bar a
(435 psia)
100 bar a
(1450 psia)
Lower measuring limit
• Measuring cell with silicone oil
0 mbar a (0 psia)
Upper measuring limit
• Measuring cell with silicone oil
100 % of max. span
100 % of the max. nominal measuring range
Input of gauge pressure, with front-flush
diaphragm
Measured variable
Gauge pressure, front-flush
Spans (infinitely adjustable) or nominal mea- Span (min. ... max.)
suring range and max. permissible test pressure
Max. perm. test
pressure
Nominal measuring
range
Max. perm. test
pressure
0.01 ... 1 bar
(0.15 ... 14.5 psi)
6 bar
(87 psi)
1 bar
(14.5 psi)
6 bar
(87 psi)
0.04 ... 4 bar
(0.58 ... 58 psi)
10 bar
(145 psi)
4 bar
(58 psi)
10 bar
(145 psi)
0.16 ... 16 bar
(2.32 ... 232 psi)
32 bar
(464 psi)
16 bar
(232 psi)
32 bar
(464 psi)
0.6 ... 63 bar
(9.14 ... 914 psi)
100 bar
(1450 psi)
63 bar
(914 psi)
100 bar
(1450 psi)
Lower measuring limit
100 mbar a (1.45 psia)
Upper measuring limit
• Measuring cell with silicone oil
1/56
Siemens FI 01 · 2015 US Edition
100% of max. span
100 % of the max. nominal measuring range
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
SITRANS P300 for gauge and absolute pressure
HART
PROFIBUS PA and FOUNDATION Fieldbus
Input of absolute pressure, with
front-flush diaphragm
Measured variable
Absolute pressure, front-flush
Spans (infinitely adjustable) or nominal mea- Span (min. ... max.)
suring range and
max. permissible test pressure
Max. perm. test
pressure
Nominal measuring
range
Max. perm. test
pressure
43 ... 1300 mbar a
(0.62 ... 18.85 psia)
10 bar a
(145 psia)
1300 mbar a
(18.85 psia)
10 bar a
(145 psia)
0.16 ... 5 bar a
(2.32 ... 72.5 psi a)
30 bar a
(435 psia)
5 bar a
(72.5 psia)
30 bar a
(435 psia)
1 ... 30 bar a
(14.5 ... 435 psia)
100 bar a
(1450 psia)
30 bar a
(435 psia)
100 bar a
(1450 psia)
Depending on the process connection, the span
may differ from these values
Lower measuring limit
Depending on the process connection, the nominal measuring range may differ from these values
0 bar a (0 psia)
Upper measuring limit
• Measuring cell with silicone oil
100% of max. span
100 % of the max. nominal measuring range
Output signal
4 ... 20 mA
Digital PROFIBUS PA signal
Physical bus
-
Output
IEC 61158-2
Protection against polarity reversal
Protected against short-circuit and polarity reversal.
Each connection against the other with max. supply voltage.
Electrical damping T63 (step width 0.1 s)
Set to 2 s (0 ... 100 s)
According to IEC 60770-1
Measuring accuracy
Reference conditions
(All error data refer always refer to the set
span)
Increasing characteristic, start-of-scale value 0 bar, stainless steel seal diaphragm,
measuring cell with silicone oil, room temperature 25 °C (77 °F)
Span ratio r = max. span/set span
Nominal measuring range ratio r = nominal
measuring range/set measuring range)
Error in measurement at limit setting incl.
hysteresis and reproducibility
Gauge pressure
Absolute
pressure
Absolute
pressure,
front-flush
Gauge pressure
Absolute
pressure
Absolute
pressure,
front-flush
Linear characteristic
• r + 10
 (0.0029  r + 0.071) %  0.1 %
 0.2 %
 (0.0029  r + 0.071) %  0.1 %
 0.2 %
• 10 < r  30
 (0.0045  r + 0.071) %  0.2 %
 0.4 %
 (0.0045  r + 0.071) %  0.2 %
 0.4 %
• 30 < r  100
 (0.005  r + 0.05) %
-
 (0.005  r + 0.05) %
-
-
Step response time T63
Long-term stability at 30 °C (54 °F)
-
approx. 0.2 s
 (0.25  r) %/5 years
 (0.1  r) %/year
 (0.25  r)/5 years
 (0.1  r) %/year
Influence of ambient temperature
• at -10 ... +60 °C (14 ... 140 °F)
 (0.08 r + 0.1) %1)
 (0.2  r
+ 0 3) %
 (0.08 r + 0.1) %1)
• at -40 ... -10 °C and +60 ... +85 °C
(-40 ... 14 °F and 140 ... 185 °F)
 (0.1  r + 0.15) %/10 K
 (0.2  r +  (0.1  r + 0.15) %/10 K
0.3) %/10 K
 (0.2  r
+ 0 3) %
 (0.2  r +
0.3) %/10 K
Influence of the medium temperature (only
with front-flush diaphragm)
• Temperature difference between medium
temperature and ambient temperature
3 mbar/10 K (0.04 psi/10 K)
Siemens FI 01 · 2015 US Edition
1/57
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
SITRANS P300 for gauge and absolute pressure
HART
PROFIBUS PA and FOUNDATION Fieldbus
Rated conditions
Installation conditions
Ambient temperature
• Measuring cell with silicone oil
• Measuring cell with Neobee oil (FDA-compliant, with flush-mounted diaphragm)
• Measuring cell with inert liquid (not with frontflush diaphragm)
• Display readable
• Storage temperature
Observe the temperature class in areas subject to explosion hazard.
-40 ... +85 °C (-40 ... +185 °F)
-10 ... +85 °C (14 ... +185 °F)
-20 ... +85 °C (-4 ... +185 °F)
-30 ... +85 °C (-22 ... +185 °F)
-50 ... +85 °C (-58 ... +185 °F)
(for Neobee: -20 ... +85 °C (-4 ... +185 °F))
(for temperature oil: -10 ... + 85 °C (14 ... +165 °F))
Climatic class
Condensation
Degree of protection acc. to EN 60529
Relative humidity 0 ... 100 %
Condensation permissible, suitable for use in the tropics
IP65, IP68, NEMA X, enclosure cleaning, resistant to lyes, steam to 150 °C (302 °F)
Electromagnetic Compatibility
• Emitted interference and interference immunity
Acc. to IEC 61326 and NAMUR NE 21
Medium conditions
Temperature of medium
• Measuring cell with silicone oil
-40 ... +100 °C (-40 ... +212 °F)
• Measuring cell with silicone oil (FDA-compliant,
with flush-mounted diaphragm)
• Measuring cell with Neobee oil "Measuring cell
with Neobee oil (FDA-compliant, with flushmounted diaphragm)
• Measuring cell with silicone oil, with temperature decoupler (only for gauge pressure version with flush-mounted diaphragm)
-40 ... +150 °C (-40 ... +302 °F)
• Measuring cell with Neobee oil, with temperature decoupler (only for gauge pressure
version with flush-mounted diaphragm)
• Measuring cell with inert liquid
-10 ... +200 °C (14 ... +392 °F)
• Measuring cell with high-temperature oil (only
for gauge pressure version with flush-mounted
diaphragm)
-10 ... +250 °C (14 ... 482 °F)
-10 ... +150 °C (-14 ... +302 °F)
-40 ... +200 °C (-40 ... +392 °F)
-20 ... +100 °C (-4 ... +212 °F)
Design (standard version)
Weight (without options)
Enclosure material
Material of parts in contact with the medium
• Connection shank
• Oval flange
• Seal diaphragm
• Measuring cell filling
Process connection
1/58
Siemens FI 01 · 2015 US Edition
Approx. 800 g (1.8 lb)
Stainless steel, mat. no. 1.4301/304
Stainless steel, mat. no. 1.4404/316L or Hastelloy C276, mat. no. 2.4819
Stainless steel, mat. no. 1.4404/316L
Stainless steel, mat. no. 1.4404/316L or Hastelloy C276, mat. no. 2.4819
•Silicone oil
•Inert filling liquid
•G½B to EN 837-1
•Female thread ½-14 NPT
•Oval flange PN 160 (MAWP 2320 psi) with fastening thread:
-7/16 -20 UNF to IEC 61518
•M10 as per DIN 19213
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
SITRANS P300 for gauge and absolute pressure
HART
PROFIBUS PA and FOUNDATION Fieldbus
Design (version with front-flush diaphragm)
Weight (without options)
approx. 1 ... 13 kg (2.2 ... 29 lb)
Enclosure material
Stainless steel, mat. no. 1.4301/304
Material of parts in contact with the medium
• Process connection
Stainless steel, mat. no. 1.4404/316L
• Seal diaphragm
Stainless steel, mat. no. 1.4404/316L
•Silicone oil
•Inert filling liquid
•FDA compliant fill fluid (Neobee oil)
• Measuring cell filling
Process connection
•Flanges as per EN and ASME
•F&B and pharmaceutical flanges
Ra-values  0.8 µm (32 µ-inch)/welds Ra)  1.6 µm (64 µ-inch)
(Process connections acc. to 3A; Ra-values  0.8 µm (32 µ-inch)/welds Ra  0.8 µm (32 µ-inch)
Surface quality touched-by-media
Power supply UH
Terminal voltage on transmitter
10.5 ... 42 V DC
for intrinsically safe operation: 10.5 ... 30 V DC
Supplied through bus
Separate power supply
-
Not necessary
• Without Ex
-
9 ... 32 V
• With intrinsically-safe operation
-
9 ... 24 V
Bus voltage
Current consumption
• Max. basic current
-
12.5 mA
• Start-up current  basic current
-
Yes
• Max. fault current in the event of a fault
-
15.5 mA
Fault disconnection electronics (FDE)
-
Available
Certificates and approvals
Classification according to PED 97/23/EC
For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of Article 3,
paragraph 3 (sound engineering practice)
In preparation
Water, waste water
Explosion protection
Intrinsic safety "i"
PTB 05 ATEX 2048
Ex II 1/2 G Ex ia/ib IIB/IIC T4, T5, T6
• Marking
• Permissible ambient temperature
- Temperature class T4
-40 ... +85 °C (-40 ... +185 °F)
- Temperature class T5
-40 ... +70 °C (-40 ... +158 °F)
- Temperature class T6
• Connection
-40 ... +60 °C (-40 ... +140 °F)
To certified intrinsically-safe circuits with peak
values:
Ui = 30 V, Ii = 100 mA,
Pi = 750 mW, Ri = 300 
To certified intrinsically-safe circuits with peak
values:
FISCO supply unit:
Ui = 17.5 V, Ii = 380 mA,
Pi = 5.32 W
Linear barrier:
Ui = 24 V, Ii = 250 mA, Pi = 1.2 W
• Effective inner capacitance:
Ci = 6 nF
Ci = 1.1 nF
• Effective internal inductance:
Li = 0.4 mH
Li  7 H
Explosion protection to FM for USA and Canada
(cFMUS)
• Identification (DIP) or (IS); (NI)
• Identification (DIP) or (IS)
Certificate of Compliance 3025099
CL I, DIV 1, GP ABCD T4 ... T6; CL II, DIV 1, GP EFG; CL III; CL I, ZN 0/1 AEx ia IIC T4 ... T6; CL I,
DIV 2, GP ABCD T4 ... T6; CL II, DIV 2, GP FG; CL III
Certificate of Compliance 3025099C
CL I, DIV 1, GP ABCD T4 ... T6; CL II, DIV 1, GP EFG; CL III; Ex ia IIC 4 ... T6; CL I, DIV 2,
GP ABCD T4 ... T6; CL II, DIV 2, GP FG; CL III
Siemens FI 01 · 2015 US Edition
1/59
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
SITRANS P300 for gauge and absolute pressure
HART
Dust explosion protection for zone 20/21/22
PROFIBUS PA and FOUNDATION Fieldbus
PTB 05 ATEX 2048
Ex II 1D Ex ia D 20 T 120 °C
Ex II 2D Ex ib D 21 T 120 °C
Ex II 3D Ex ib D 21 T 120 °C
• Marking
• Permissible ambient temperature
- Temperature class T4
-40 ... +85 °C (-40 ... +185 °F) (in the case of mineral glass windows only -20 ... +85 °C (-4 ... +185 °F))
- Temperature class T5
-40 ... +70 °C (-40 ... +158 °F) (in the case of mineral glass windows only-20 ... +70 °C (-4 ... +158 °F))
- Temperature class T6
-40 ... +60 °C (-40 ... +140 °F) (in the case of mineral glass windows only -20 ... +60 °C (-4 ... +140 °F))
• Connection
To certified intrinsically-safe circuits with peak
values:
Ui = 30 V, Ii = 100 mA, Pi = 750 mW
To certified intrinsically-safe circuits with peak
values:
Ui = 24 V, Ii = 380 mA, Pi = 5.32 mW
• Effective inner capacitance:
Ci = 6 nF
Ci = 5 nF
• Effective internal inductance:
Li = 0.4 H
Li = 10 H
PTB 05 ATEX 2048
Type of protection Ex nA/nL/ic (Zone 2)
• Marking
II 2/3 G Ex nA T4/T5/T6
II 2/3 G Ex nL IIB/IIC T4/T5/T6
• Permissible ambient temperature
- Temperature class T4
-40 ... +85 °C (-40 ... +185 °F) (in the case of mineral glass windows only -20 ... +85 °C (-4 ... +185 °F))
- Temperature class T5
-40 ... +70 °C (-40 ... +158 °F) (in the case of mineral glass windows only -20 ... +70 °C (-4 ... +158 °F))
- Temperature class T6
-40 ... +60 °C (-40 ... +140 °F) (in the case of mineral glass windows only -20 ... +60 °C (-4 ... +140 °F))
• Ex nA/nL connection
To certified intrinsically-safe circuits with peak
values:
Um = 45 V
To certified intrinsically-safe circuits with peak
values:
Um = 32 V
• Ex ic connection
To certified intrinsically-safe circuits with
peak values:
Ui = 45 V
To certified intrinsically-safe circuits with
peak values:
Ui = 32 V
• Effective inner capacitance:
Ci = 6 nF
Ci = 5 nF
• Effective internal inductance:
Li = 0.4 mH
Li = 20 H
1)
Conversion of temperature error per 28 °C. Valid for temperature range -3 ... +53 °C < (0.064  r + 0.08) %/28 °C (50 °F).
1/60
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
HART Communication
HART communication
230 ... 1100 
Protocol
HART Version 5.x
Software for computer
SIMATIC PDM
PROFIBUS PA communication
Simultaneous communication with
master class 2 (max.)
4
The address can be set using
Configuration tool or
local operation
(standard setting Address 126)
FOUNDATION Fieldbus
communication
Function blocks
3 function blocks analog input,
1 function block PID
• Analog input
- Adaptation to customer-specific process variables
Yes, linearly rising or falling
characteristic
- Electrical damping, adjustable
0 ... 100 s
- Simulation function
Output/input (can be locked
within the device with a bridge)
- Failure mode
parameterizable (last good
value, substitute value, incorrect
value)
- Limit monitoring
Yes, one upper and lower warning limit and one alarm limit
respectively
- Square-rooted characteristic
for flow measurement
Yes
Cyclic data usage
• Output byte
• Input byte
5 (one measured value) or
10 (two measured values)
0.1 or 2 (totalizer mode and
reset function for dosing)
• Internal preprocessing
Device profile
PROFIBUS PA Profile for Process Control Devices Version
3.0, class B
Function blocks
2
• PID
Standard FOUNDATION Fieldbus function block
• Physical block
1 resource block
Transducer blocks
1 transducer block Pressure with
calibration, 1 transducer block
LCD
• Analog input
- Adaptation to customer-specific process variables
Yes, linearly rising or falling
characteristic
- Electrical damping adjustable
0 … 100 s
- Simulation function
Input /Output
- Failure function
parameterizable (last good
value, substitute value, incorrect
value)
- Limit monitoring
Yes, one upper and lower warning limit and one alarm limit
respectively
• Register (totalizer)
Can be reset, preset, optional
direction of counting, simulation
function of register output
- Failure mode
parameterizable (summation
with last good value, continuous
summation, summation with
incorrect value)
- Limit monitoring
One upper and lower warning
limit and one alarm limit respectively
• Physical block
1
Transducer blocks
2
• Pressure transducer block
- Can be calibrated by applying
two pressures
Yes
- Monitoring of sensor limits
Yes
- Simulation function: Measured
pressure value, sensor temperature and electronics temperature
Constant value or over parameterizable ramp function
• Pressure transducer block
- Can be calibrated by applying
two pressures
Yes
- Monitoring of sensor limits
Yes
- Specification of a container
characteristic with
Max. 30 nodes
- Simulation function for measured pressure value and sensor temperature
Constant value or over parameterizable ramp function
Siemens FI 01 · 2015 US Edition
1/61
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
Selection and Ordering data
Article No.
Selection and Ordering data
SITRANS P300 pressure transmitters for relative and absolute pressure, single-chamber measuring housing, rating plate inscription in English
Article No.
SITRANS P300 pressure transmitters for relative and absolute pressure, single-chamber measuring housing, rating plate inscription in English
4 ... 20 mA/HART
7MF 8 0 2 3 -
4 ... 20 mA/HART
7MF 8 0 2 3 -
PROFIBUS PA
7MF 8 0 2 4 -
PROFIBUS PA
7MF 8 0 2 4 -
FOUNDATION Fieldbus (FF)
7MF 8 0 2 5 -
FOUNDATION Fieldbus (FF)
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Measuring cell filling
Silicone oil
Inert liquid
Measuring cell cleaning
normal
Cleanliness level 2 to
DIN 25410
77777 - 7777
Display
• Without display, with keys, closed lid
• With display and keys, closed lid11)
1
3
max. span (min. ... max.)
0.01 ... 1 bar
(0.145 ... 14.5 psi)
0.04 ... 4 bar
(0.58 ... 58 psi)
0.16 ...16 bar
(2.32 ... 232 psi)
0.63 ... 63 bar
(9.14 ... 914 psi)
B
C
D
E
1.6 ... 160 bar
4 ... 400 bar
2.5 ... 250 mbar a
13 ... 1300 mbar a
(23.2 ... 2320 psi)
(58 ... 5802 psi)
(0.04 ... 3.63 psia)
(0.19 ... 18.86 psia)
F
G
Q
S
0.05 ... 5 bar a
0.3 ... 30 bar a
(0.7 ... 72.5 psia)
(4.35 ... 435 psia)
T
U
• With display and keys, lid with Makrolon pane
(setting on HART devices: mA,
with PROFIBUS PA and FOUNDATION Fieldbus
equipment: pressure units)11)
• With display and keys (setting acc. to specifications, Order code "Y21" or "Y22" required), lid with
Makrolon pane11)
• With display and keys, lid with glass pane (setting
on HART devices: mA, with
PROFIBUS and FOUNDATION Fieldbus equipment: pressure units)11)
• With display and keys (setting acc. to specifications, Order code "Y21" or "Y22" required), lid with
glass pane11)
Process connection
• Connection shank G½B to EN 837-1
• Female thread ½-14 NPT
• Stainless steel oval flange with process connection (Oval flange has no female thread) 6)
- Mounting thread 7/16-20 UNF to EN 61518
- Mounting thread M10 to DIN 19213
- Mounting thread M12 to DIN 19213
• Male thread M20 x 1.5
• Male thread ½ -14 NPT
Non-wetted parts materials
• Stainless steel, deep-drawn and electrolytically
polished
Version
• Standard versions
Explosion protection
• None
• With ATEX, Type of protection:
- "Intrinsic safety (Ex ia)"
• Zone 20/21/227)
• Ex nA/nL (Zone 2)8)
• with FM "intrinsic safety" (cFMUS)
Electrical connection / cable entry
• Screwed gland M20x1.5 (polyamide)9)
• Screwed gland M20x1.5 (metal)
• Screwed gland M20x1.5 (stainless steel)
• M12 connectors (metal), without cable socket)
• M12 connectors (stainless steel), without cable
• Screwed gland ½-14 NPT metal thread 10)
• Screwed gland ½-14 NPT stainless steel thread
1/62
Siemens FI 01 · 2015 US Edition
1
2
4
5
6
7
Power supply units see Chap. 7 "Supplementary Components".
Wetted parts materials
Seal diaphragm
Measuring cell
Stainless steel
Stainless steel
Hastelloy
Stainless steel
Hastelloy
Hastelloy
Version for diaphragm seal1) 2) 3) 4) 5)
7MF 8 0 2 5 77777 - 7777
Included in delivery of the device:
• Brief instructions (Leporello)
• CD-ROM with detailed documentation
A
B
C
Y
0
1
2
3
4
5
6
4
1
A
B
C
E
M
A
B
C
F
G
H
J
1)
When the manufacture’s certificate (calibration certificate) has to be
ordered for transmitters with diaphragm seals according to IEC 60770-2, it
is recommended only to order this certificate exclusively with the diaphragm seals. The measuring accuracy of the total combination is certified here.
2)
If the acceptance test certificate 3.1 is ordered for the transmitter with
mounted diaphragm seals this certificate must also be ordered with the
respective remote seals.
3)
The diaphragm seal is to be specified with a separate order number and
must be included with the transmitter order number, for example
7MF802.-..Y..-.... and 7MF4900-1...-.B
4)
The standard measuring cell filling for configurations with remote seals (Y)
is silicone oil.
5)
Remote seal for direct mounting only available in combination with process connection ½-14 NPT.
6)
M10 fastening thread: Max. span 160 bar (2320 psi)
7/16-20 UNF and M12 fastening thread: Max. span 400 bar (5802 psi)
7)
Only available together with electrical connection option A
8)
Only available together with electrical connection options B, C or G.
9)
Only together with HART electronics.
10) Without
cable gland.
11) Display
cannot be turned.
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Selection and Ordering data
Article No.
Selection and Ordering data
SITRANS P300 pressure transmitters for relative
and absolute pressure with front-flush membrane, single-chamber measuring housing, rating
plate inscription in English
Article No.
SITRANS P300 pressure transmitters for relative
and absolute pressure with front-flush membrane, single-chamber measuring housing, rating
plate inscription in English
4 ... 20 mA/HART
7MF 8 1 2 3 -
4 ... 20 mA/HART
7MF 8 1 2 3 -
PROFIBUS PA
7MF 8 1 2 4 -
PROFIBUS PA
7MF 8 1 2 4 -
FOUNDATION Fieldbus (FF)
7MF 8 1 2 5 -
FOUNDATION Fieldbus (FF)
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Measuring cell filling
Silicone oil
Inert liquid
FDA compliant fill fluid
• Neobee oil
77777 - 7777
Display
• Without display, with keys, closed lid
Measuring cell cleaning
normal
Cleanliness level 2 to
DIN 254101)
1
3
normal
4
(0.15 ... 14.5 psi)
(0.58 ... 58 psi)
(2.32 ... 232 psi)
(9.14 ... 914 psi)
B
C
D
E
13 ... 1300 mbar a2)
0.05 ... 5 bar a2)
0.03 ... 30 bar a2)
(0.19 ... 18.9 psia)2)
(0.7 ... 72.5 psia)2)
(4.35 ... 435 psia)2)
S
T
U
Wetted parts materials
Seal diaphragm
Measuring cell
Stainless steel
Stainless steel
Process connection
• Flange version with Order code M.., N.., R.. or Q..
(see "Further designs")
Non-wetted parts materials
• Stainless steel, deep-drawn and electrolytically
polished
Version
• Standard versions
Explosion protection
• None
• With ATEX, Type of protection:
- "Intrinsic safety (Ex ia)"
• Zone 20/21/224)
• Ex nA/nL (Zone 2)5)
• with FM "intrinsic safety" (cFMUS)
Electrical connection / cable entry
• Screwed gland M20x1.5 (polyamide)6)
• Screwed gland M20x1.5 (metal)
• Screwed gland M20x1.5 (stainless steel)
• M12 connectors (without cable socket)
• M12 connectors (stainless steel), without cable
socket)
• Screwed gland ½-14 NPT metal thread 7)
• Screwed gland ½-14 NPT stainless steel thread 7)
1
2
• With display and keys, closed lid8)
max. span
0.01 ... 1 bar
0.04 ... 4 bar
0.16 ... 16 bar
0.63 ... 63 bar
Stainless steel
Hastelloy3)
7MF 8 1 2 5 77777 - 7777
• With display and keys, lid with Makrolon pane
(setting on HART devices: mA, with PROFIBUS PA
and FOUNDATION Fieldbus equipment: pressure
units)8)
4
• With display and keys (setting acc. to specifications, Order code "Y21" or "Y22" required), lid with
Makrolon pane8)
5
• With display and keys, lid with glass pane (setting
on HART devices: mA, with PROFIBUS PA and
FOUNDATION Fieldbus equipment: pressure
units)8)
6
• With display and keys (setting acc. to specifications, Order code "Y21" or "Y22" required), lid with
glass pane8)
7
Power supply units see Chap. 7 "Supplementary Components"
Included in delivery of the device:
• Brief instruction (Leporello)
• CD-ROM with detailed documentation
A
B
7
4
1
A
1)
Not suitable for oxygen applications.
2)
Not with temperature decoupler P00 and P10, not for process connections
R01, R02, R04, R10 and R11, and can only be ordered in conjunction with
silicone oil.
3)
Only available for flanges with options M.., N.. and Q..
4)
Only together with electrical connection option A.
5)
Only available together with electrical connection options B, C or G.
6)
Only together with HART electronics.
7)
Without cable gland.
8)
Display cannot be turned.
B
C
E
M
A
B
C
F
G
H
J
Siemens FI 01 · 2015 US Edition
1/63
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 for gauge and absolute pressure
Selection and Ordering data
Order code
Further designs
Add "-Z" to Article No. and
specify Order code.
Selection and Ordering data
HART PA
Pressure transmitter with mounting
A02
bracket (2 shackles, 4 nuts, 4 U-plates,
1 angle) made of:
made completely of stainless steel, for wall or
pipe mounting
Cable socket for M12 plug
• Stainless steel
A51
Rating plate inscription
(instead of English)
• German
• French
• Spanish
• Italian
B10
B12
B13
B14
English rating plate
Pressure units in inH20 and/or psi


Order code
FF
Further designs
Add "-Z" to Article No. and
specify Order code.
HART PA
FF

Sanitary process connection according
DIN 11851 (Dairy connection with slotted
union nut)
Certified to 3A6)
• DN 50, PN 25
• DN 80, PN 25
N04
N06






Tri-Clamp connection according
DIN 32676/ISO 2852
Certified to 3A6)
• DN 50/2", PN 16
• DN 65/3", PN 10
N14
N15






N28

















B21



Varivent connection
Certified to 3A and EHEDG6)
• Type N = 68 for Varivent housing
DN 40 ... 125 und 1½" ... 6", PN 40


Temperature decoupler up to 200 °C7)
for front-flush diaphragm version



C11

P00
Quality inspection certificate (Five-step
factory calibration) to IEC 60770-21)
Inspection certificate2)
Acc. to EN 10204-3.1
C12



Temperature decoupler up to 250 °C
Measuring cell filling: High-temperature oil
(Silicone oil)
P10



Factory certificate
Acc. to EN 10204-2.2
C14



Degree of protection IP65/IP68
(only for M20x1.5 and ½-14 NPT)
D12



Bio-Control sanitary process connection
Certified to 3A and EHEDG6)
• DN 50, PN 16
• DN 65, PN 16
Q53
Q54






Degree of protection IP6k9k
(only for M20x1.5)
D46



Sanitary process connection to DRD
• DN 50, PN 40
M32



Export approval Korea
E11



Ex Approval IEC Ex (Ex ia)
(only for transmitter 7MF8...-.....-.B..)
E45



Ex Approval Ex ia/ib NEPSI
E55



SMS socket with union nut
• 2“
• 2½“
• 3“
M67
M68
M69









SMS threaded socket
• 2“
• 2½“
• 3“
M73
M74
M75









IDF socket with union nut ISO 2853
• 2“
• 2½“
• 3“
M82
M83
M84









IDF threaded socket ISO 2853
• 2“
• 2½“
• 3“
M92
M93
M94









Sanitary process connection to NEUMO
Bio-Connect screw connection
Certified to 3A and EHEDG6)
• DN 50, PN 16
• DN 65, PN 16
• DN 80, PN 16
• DN 100, PN 16
• DN 2", PN 16
• DN 2½", PN 16
• DN 3", PN 16
• DN 4", PN 16
Q05
Q06
Q07
Q08
Q13
Q14
Q15
Q16
























Sanitary process connection to NEUMO
Bio-Connect flange connection
Certified to 3A and EHEDG6)
• DN 50, PN 16
• DN 65, PN 16
• DN 80, PN 16
• DN 100, PN 16
• DN 2", PN 16
• DN 2½", PN 16
• DN 3", PN 16
• DN 4", PN 16
Q23
Q24
Q25
Q26
Q31
Q32
Q33
Q34
























Only for SITRANS P300 with front-flush
diaphragm (7MF81..-...)
Flange to EN 1092-1, Form B1
• DN 25, PN 403)
• DN 25, PN 1004)
• DN 40, PN 40
• DN 40, PN 100
• DN 50, PN 16
• DN 50, PN 40
• DN 80, PN 16
• DN 80, PN 40
M11
M21
M13
M23
M04
M14
M06
M16
























Flanges to ASME B16.5
• 1", class 1504)
• 1½", class 150
• 2", class 150
• 3", class 150
• 4", class 150
• 1", class 3004)
• 1½", class 300
• 2", class 300
• 3", class 300
• 4", class 300
M40
M41
M42
M43
M44
M45
M46
M47
M48
M49






























Threaded connector to DIN 3852-2, form A,
thread to ISO 228
• G ¾"-A, front-flush4)
R01
• G 1"-A, front-flush4)
R02
R04
• G 2"-A, front-flush4)









Tank connection5)
Sealing is included in delivery
• TG 52/50, PN 40
• TG 52/150, PN 40






1/64
Siemens FI 01 · 2015 US Edition
R10
R11
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Selection and Ordering data
Order code
Further designs
Add "-Z" to Article No. and
specify Order code.
Selection and Ordering data
HART PA
FF
Order code
Additional data
Please add "-Z" to Article No. and specify
Order code(s) and plain text.
HART PA
Measuring range to be set
Specify in plain text (max. 5 characters):
Y01: ... up to ... mbar, bar, kPa, MPa, psi
Y01

8)
Stainless steel tag plate and entry in
device variable (measuring point description)
Max. 16 characters, specify in plain text:
Y15: ...........................................
Y15



Measuring point text (entry in device variable)
Max. 27 characters, specify in plain text:
Y16: ...........................................
Y16



Entry of HART TAG
Y17

Setting of the display in pressure units
Specify in plain text (standard setting: bar):
Y21: mbar, bar, kPa, MPa, psi, ...
Note: The following pressure units can be
selected:
bar, mbar, mm H2O*), inH2O*), ftH2O*),
mmHG, inHG, psi, Pa, kPa, MPa, g/cm2,
kg/cm2, Torr, ATM or %
*) ref. temperature 20 °C
Y21



Setting of the display in nonpressure units9)
Specify in plain text:
Y22: ..... up to ..... l, m3, m, USg, ...
(specification of measuring range in
pressure units "Y01" is essential, unit with
max. 5 characters)
Y22 +
Y01

Preset bus address
(possible between 1 ... 126)
Specify in plain text:
Y25: .....................
Y25


Sanitary process connection to NEUMO
Bio-Connect clamp connection
Certified to 3A and EHEDG6)
• DN 50, PN 16
• DN 65, PN 10
• DN 80, PN10
• DN 100, PN 10
• DN 2½", PN 16
• DN 3", PN 10
• DN 4", PN 10
Q39
Q40
Q41
Q42
Q48
Q49
Q50





















Sanitary process connection to NEUMO
Bio-Connect S flange connection
Certified to 3A and EHEDG
• DN 2", PN 16
Q72



Aseptic threaded socket to DIN 11864-1 Form A
Certified to 3A and EHEDG
• DN 50, PN 25
• DN 65, PN 25
• DN 80, PN 25
• DN 100, PN 25
N33
N34
N35
N36












Aseptic flange with notch to DIN 11864-2
Form A
Certified to 3A and EHEDG
• DN 50, PN 16
• DN 65, PN 16
• DN 80, PN 16
• DN 100, PN 16
N43
N44
N45
N46





















Factory mounting of valve manifolds, see accessories.



Only Y01, Y15, Y16, Y17, Y21, Y22 and Y25 can be factory preset
Aseptic flange with groove to DIN 11864-2
Form A
Certified to 3A and EHEDG
• DN 50, PN 16
N43 +
P11
• DN 65, PN 16
N44 +
P11
N45 +
• DN 80, PN 16
P11
• DN 100, PN 16
N46 +
P11
Aseptic clamp with groove to DIN 11864-3
FormA
Certified to 3A and EHEDG
• DN 50, PN 25
• DN 65, PN 25
• DN 80, PN 16
• DN 100, PN 16
N53
N54
N55
N56
FF
Max. 8 characters, specify in plain text:
Y17: ...........................................
 = available












Ordering example
Item line:
7MF8023-1DB24-1AB7-Z
B line:
A02 + Y01 + Y21
C line:
Y01: 1 ... 10 bar (14.5 ... 145 psi)
C line:
Y21: bar (psi)
1)
When the manufacture’s certificate (calibration certificate) has to be
ordered for transmitters with diaphragm seals according to IEC 60770-2, it
is recommended only to order this certificate exclusively with the diaphragm seals. The measuring accuracy of the total combination is certified
here.
2)
If the acceptance test certificate 3.1 is ordered for the transmitter with
mounted diaphragm seals this certificate must also be ordered with the
respective remote seals.
3)
Special seal in Viton included in the scope of delivery
4)
Cannot be combined with Order codes P00 and P10. Can only be ordered
with silicone oil measuring cell filling.
5)
The weldable socket can be ordered under accessories.
6)
3A certification only if used in conjunction with 3A-compliant sealing rings.
7)
Certified to 3A and EHEDG. The maximum permissible temperatures of the
medium depend on the respective cell fillings (see medium conditions).
8)
Measuring accuracies for PROFIBUS PA transmitters with Option Y01 are
calculated in the same way as for HART devices.
9)
Preset values can only be changed over SIMATIC PDM.
Siemens FI 01 · 2015 US Edition
1/65
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
■ Dimensional drawings
approx. 78 (3.07)
approx. 28 (1.1)
H2 = approx. 80 (3.15)
H1 = approx. 120 (4.72)
ø 91 (3.58)
approx. 66 (2.6)
SITRANS P300, with oval flange, dimensions in mm (inch)
Ø
approx. 69 (2.72)
approx. 125 (4.92)
82
approx. 41 (1.61)
°
90 (3.54)
approx. 118 (4.65)
approx. 66 (2.6)
approx.
28
(1.1)
1
approx. 105 (4.13)
Ø 50...60 (1.97...2.36)
72 (2.83)
approx. 131 (5.16)
90 (3.54)
SITRANS P300, process connection M20 x 1.5, with mounted mounting bracket, dimensions in mm (inch)
1/66
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Ø 91 (3.58)
26,5 (1.0
4)
H1 = approx. 120 (4.72)
69 (2.72)
26 (1.02)
66 (2.6)
50 (1.97)
H2 = approx. 49 (1.93)
Temperature decoupler
°
82
SITRANS P300, front-flush, dimensions in mm (inch)
The diagram shows a SITRANS P300 with an example of a
flange. In this drawing the height is subdivided into H1 and H2.
H1 = Height of the SITRANS P300 up to a defined cross-section
H2 = Height of the flange up to this defined cross-section
Only the height H2 is indicated in the dimensions of the flanges.
Siemens FI 01 · 2015 US Edition
1/67
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Flanges as per EN and ASME
Sanitary process screw connection to NEUMO Bio-Connect
Flange to EN
Order DN
code
H2
Order DN
code
M11
M21
M13
M23
M04
M14
M06
M16
D
25
25
40
40
50
50
80
80
PN
D
H2
40
100
40
100
16
40
16
40
115 mm (4.5“)
140 mm (5.5“)
150 mm (5.9“)
170 mm (6.7“)
165 mm (6.5“)
165 mm (6.5“)
200 mm (7.9“)
200 mm (7.9“)
Approx.
52 mm (2“)
H2
EN 1092-1
D
H2
Order DN
code
M40
M41
M42
M43
M44
M45
M46
M47
M48
M49
1“
1½“
2“
3“
4“
1“
1½“
2“
3“
4“
H2
PN
D
150
150
150
150
150
300
300
300
300
300
110 mm (4.3“) Approx.
130 mm (5.1“) 52 mm (2“)
150 mm (5.9“)
190 mm (7.5“)
230 mm (9.1“)
125 mm (4.9“)
155 mm (6.1“)
165 mm (6.5“)
210 mm (8.1“)
255 mm (10.0“)
H2
D
DIN 11851 (milk pipe union with slotted union nut)
N04
N06
50
80
PN
D
H2
25
25
92 mm (3.6“)
127 mm (5.0“)
Approx.
52 mm (2“)
D
Q23
Q24
Q25
Q26
Q31
Q32
Q33
Q34
50
65
80
100
2“
2½“
3“
4“
Order DN
code
Connections to DIN
Order DN
code
D
H2
16
16
16
16
16
16
16
16
82 mm (3.2“)
105 mm (4.1“)
115 mm (4.5“)
145 mm (5.7“)
82 mm (3.2“)
105 mm (4.1“)
105 mm (4.1“)
145 mm (5.7“)
Approx.
52 mm (2“)
PN
D
H2
16
16
16
16
16
16
16
16
110 mm (4.3“)
140 mm (5.5“)
150 mm (5.9“)
175 mm (6.9“)
100 mm (3.9“)
110 mm (4.3“)
140 mm (5.5“)
175 mm (6.9“)
Approx.
52 mm (2“)
Sanitary process connection to NEUMO Bio-Connect clamp
connection
H2
NuG and pharmaceutical connections
H2
50
65
80
100
2“
2½“
3“
4“
Order DN
code
ASME B16.5
D
Q05
Q06
Q07
Q08
Q13
Q14
Q15
Q16
PN
Sanitary process connection to NEUMO Bio-Connect flange
connection
Flanges to ASME
D
Q39
Q40
Q41
Q42
Q47
Q48
Q49
Q50
50
65
80
100
2“
2½“
3“
4“
PN
D
16
10
10
10
16
16
10
10
77.4 mm (3.0“) Approx.
90.9 mm (3.6“) 52 mm (2“)
106 mm (4.2“)
119 mm (4.7“)
77.4 mm (3.0“)
90.9 mm (3.6“)
106 mm (4.2“)
119 mm (4.7“)
H2
Sanitary process connection to NEUMO Bio-Connect S flange
connection
Order DN
code
Order DN
code
H2
N14
N15
50
65
PN
D
H2
16
10
64 mm (2.5“)
91 mm (3.6“)
Approx.
52 mm (2“)
H2
Tri-Clamp nach DIN 32676
Q72
2“
PN
D
H2
16
125 mm (4.9“)
Approx.
52 mm (2“)
D
Threaded connection G¾", G1" and G2" acc. to DIN 3852
D
D
H2
Varivent connection
R01
¾“
60
37 mm (1.5“)
R02
1“
60
48 mm (1.9“)
R04
2“
60
78 mm (3.1“)
Approx.
45 mm
(1.8“)
Approx.
47 mm
(1.9“)
Approx.
52 mm (2“)
H2
Order DN
code
PN
Other connections
Order DN
code
H2
N28
PN
D
H2
40
84 mm
(3.3“)
Approx.
52 mm (2“)
D
40 ... 125
D
Sanitary process connection to DRD
Order DN
code
H2
1
M32
50
PN
D
H2
40
105 mm (4.1“)
Approx.
52 mm (2“)
D
1/68
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 for gauge and absolute pressure
Tank connection TG 52/50 and TG52/150
PN
D
H2
Order DN
code
PN
D
H2
Approx.
63 mm
(2.5“)
Approx.
170 mm
(6.7“)
N33
N34
N35
N36
50
65
80
100
25
25
25
25
78 x 1/6“
95 x 1/6“
110 x ¼“
130 x ¼“
Approx.
52 mm (2“)
25
40
63 mm (2.5“)
R11
25
40
63 mm (2.5“)
H2
R10
D
H2
Order DN
code
Aseptic threaded socket to DIN 11864-1 Form A
D
SMS socket with union nut
M67
M68
M69
PN
2“
25
2½“ 25
3“
25
D
H2
84 mm (3.3“)
100 mm (3.9“)
114 mm (4.5“)
Approx.
52 mm (2“)
Aseptic flange with notch to DIN 11864-2 Form A
H2
H2
Order DN
code
D
D
SMS threaded socket
H2
M73
M74
M75
PN
2“
25
2½“ 25
3“
25
H2
70 x 1/6 mm
85 x 1/6 mm
98 x 1/6 mm
Approx.
52 mm (2“)
D
D
IDF socket with union nut
H2
Order DN
code
M82
M83
M84
PN
2“
25
2½“ 25
3“
25
D
H2
77 mm (3“)
91 mm (3.6“)
106 mm (4.2“)
Approx.
52 mm (2“)
H2
H2
M92
M93
M94
2“
25
2½“ 25
3“
25
D
H2
50
65
80
100
16
16
16
16
94
113
133
159
Approx.
52 mm (2“)
N43
N44
N45
N46
H2
Order DN
code
PN
D
H2
N43 + 50
P11
N44 + 65
P11
N45 + 80
P11
N46 + 100
P11
16
94
Approx.
52 mm (2“)
16
113
16
133
16
159
Order DN
code
PN
D
H2
50
65
80
100
25
25
16
16
77.5
91
106
130
Approx.
52 mm (2“)
N53
N54
N55
N56
IDF threaded socket
PN
D
Aseptic clamp with groove to DIN 11864-3 Form A
D
Order DN
code
PN
Aseptic flange with groove to DIN 11864-2 Form A
D
H2
Order DN
code
Order DN
code
D
64 mm (2.5“)
Approx.
77.5 mm (3.1“) 52 mm (2“)
91 mm (3.6“)
D
Siemens FI 01 · 2015 US Edition
1/69
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
1
SITRANS P300 Accessories/Spare parts
Selection and Ordering data
Article No.
Mounting bracket and fastening parts kit
made of stainless steel
7MF8997-1AA
Lid without window
gasket not included
7MF8997-1BA
Lid with glass window
gasket not included
7MF8997-1BD
NBR enclosure sealing
7MF8997-1BG
Measuring point label
unlabeled
7MF8997-1CA
Cable gland
• metal
• plastic (blue)
7MF8997-1EA
7MF8997-1EB
Weldable sockets for PMC connection
• PMC Style Standard: Thread 1½"
• PMC Style Minibolt: front-flush 1"
7MF4997-2HA
7MF4997-2HB
Gaskets for PMC connection
(packing unit = 5 units)
• PTFE seal for PMC Style Standard:
Thread 1½"
• Gasket made of Viton for PMC Style Minibolt:
front-flush 1"
Weldable socket for TG 52/50 and
TG 52/150 connection
• TG 52/50 connection
• TG5 2/150 connection
Seals for TG 52/50 and TG 52/150 made of
silicone
Seals for flange connection with front-flush
diaphragm
Material FPM (Viton), 10 units
• DN 25, PN 40 (M11)
• DN 25, PN 100 (M21)
• 1", class 150 (M40)
• 1", class 300 (M45)
Selection and Ordering data
Article No.
Operating Instructions1)
Spare parts / Accessories
7MF4997-2HC
7MF4997-2HD
7MF4997-2HE
7MF4997-2HF
7MF4997-2HG
7MF4997-2HH
7MF4997-2HJ
7MF4997-2HK
7MF4997-2HL
• for SITRANS P300 series with HART
- German
- English
- French
- Spanish
- Italian
- Leporello German/English
A5E00359580
A5E00359579
A5E00359578
A5E00359576
A5E00359577
A5E00359581
• for SITRANS P300 series with PROFIBUS PA
- German
- English
- French
- Spanish
- Italian
- Leporello German/English
A5E00414587
A5E00414588
A5E00414589
A5E00414590
A5E00414591
A5E00414592
Compact operating instructions
• English, german, spanish, french, italian,
dutch
• English, estonian, latvian, lithuanian, polish,
romanian
• English, bulgarian, czech, finnish, slovakian,
slovenian
• English, danish, greek, portuguese, swedish, hungarian
• Korean
The compact operating instructions are available in 21 EU languages on the product CD
supplied with each transmitter. They can also
be downloaded from the SITRANS P web
page.
Brief instructions (Leporello)
• for SITRANS P300 with HART
- German/English
• for SITRANS P300 with PROFIBUS PA
- German/English
• for SITRANS P300 with FOUNDATION Fieldbus
- German/English
A5E03434626
A5E03434631
A5E03434645
A5E03434656
A5E03693760
A5E00359581
A5E00414592
A5E01176733
CD with SITRANS P documentation
• German, English, French, Spanish, Italian
including compact operating instructions in
21 EU languages
A5E00090345
Certificates (order only via SAP)
instead of Internet download
• hard copy (to order)
A5E03252406
• on CD (to order)
A5E03252407
HART modem
with USB interface
7MF4997-1DB
Power supply units see Chap. 7 "Supplementary Components".
1)
1/70
Siemens FI 01 · 2015 US Edition
You can download these operating instructions free-of-charge from our
Internet site at www.siemens.com/sitransp.
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 - Factory-mounting of valve manifolds on transmitters
■ Overview
The SITRANS P300 transmitter for gauge and absolute pressure
can be delivered factory-fitted with the following valve manifolds:
• 7MF9011-4EA and 7MF9011-4FA valve manifolds for gauge
pressure and absolute pressure transmitters
■ Design
The 7MF9011-4EA valve manifolds are sealed with gaskets
made of PTFE between transmitter and the valve manifold as
standard. Soft iron, stainless steel and copper gaskets are also
available for sealing purposes if preferred.
The 7MF9011-4FA valve manifolds are sealed with PTFE sealing
tape between the transmitter and the valve manifold.
Once installed, the complete unit is checked under pressure for
leaks (compressed air 6 bar (87 psi))and is certified leak-proof
with a test report to EN 10204 - 2.2.
All valve manifolds should preferably be secured with the respective mounting brackets. The transmitters are mounted on
the valve manifold and not on the unit itself.
If you order a mounting bracket when choosing the option "Factory mounting of valve manifolds", you will receive a mounting
bracket for the valve manifold instead of a bracket for mounting
the transmitter.
If you order an acceptance test certificate 3.1 to EN 10204 when
choosing the option "Factory mounting of valve manifolds", a
separate certificate is provided for the transmitters and the valve
manifolds respectively.
■ Selection and Ordering data
7MF9011-4FA valve manifold on gauge and absolute
pressure transmitters
Add -Z to the Article No. of the transmitter Order
and add Order codes
code
SITRANS P300
7MF802.-...1.-...
With process connection
female thread ½-14 NPT
in-sealed with PTFE sealing tape
Delivery incl. high-pressure test certified
by test report to EN 10204-2.2
T03
Further designs:
Delivery includes mounting brackets and
mounting clips made of stainless steel
(instead of the mounting bracket supplied
with the transmitter)
A02
Supplied acceptance test certificate to
EN 10204- 3.1 for transmitters and
mounted valve manifold
C12
7MF9011-4EA valve manifold on gauge and absolute
pressure transmitters
Add -Z to the Article No. of the transmitter Order
and add Order codes
code
SITRANS P300
7MF802.-...0.-...
T02
with process connection
collar G½ A to EN 837-1
with gasket made of PTFE between valve
manifold and transmitter
Alternative sealing material:
• Soft iron
A70
• Stainless steel, Mat. No. 14571
A71
• copper
A72
Delivery incl. high-pressure test certified
by test report to EN 10204-2.2
Further designs:
Delivery includes mounting brackets and
mounting clips made of stainless steel
(instead of the mounting bracket supplied
with the transmitter)
A02
Supplied acceptance test certificate to
EN 10204- 3.1 for transmitters and
mounted valve manifold
C12
Siemens FI 01 · 2015 US Edition
1/71
1
© Siemens AG 2015
Pressure Measurement
Transmitters for food, pharmaceuticals and biotechnology
SITRANS P300 - Factory-mounting of valve manifolds on transmitters
■ Dimensional drawings
195 (7.7)
160 (6.3)
7MF9011-4EA valve manifold with mounted gauge pressure and absolute pressure transmitters, dimensions in mm (inch)
160 (6.3)
7MF9011-4FA valve manifold with mounted gauge pressure and absolute
pressure transmitters
1/72
Siemens FI 01 · 2015 US Edition
270 (10.6)
7MF9011-4EA valve manifold with mounted gauge pressure and absolute pressure transmitters
270 (10.6)
270 (10.6)
Valve manifolds mounted on SITRANS P300
270 (10.6)
1
180 (7.1)
7MF9011-4FA valve manifold with mounted gauge pressure and absolute
pressure transmitters, dimensions in mm (inch)