SITRANS P280 catalog specifications

© Siemens AG 2015
Pressure Measurement
Transmitters with WirelessHART
SITRANS P280 for gauge and absolute pressure
■ Overview
■ Application
The SITRANS P280 is a WirelessHART field device for measuring
absolute and gauge pressure.
The measuring ranges for absolute and gauge pressure measurements are 0 to 1.6, 10, 50, 200 and 320 bar (0 to 23, 145,
725, 2900 and 4641 psi).
The sensor is integrated into the transmitter housing.
On the wireless communication side, the transmitter supports
the WirelessHART standard. A HART modem can be connected
to the transmitter particularly for initial comissioning, alternatively
the device can be commissioned comfortably by means of the
local pushbuttons w/o any additional handset devices.
It can be used in all industries and applications in non-explosive
areas.
■ Design
The SITRANS P280 has a robust aluminum enclosure and is suitable for outside use. It conforms with the IP65 safety class.
The operating temperature range is -40 to +80 °C (-40 to
+176 °F). Power supply is provided through an integrated battery, which is available as an accessory. The device is only approved for operation with this battery.
SITRANS P280 for flexible and cost-effective applications in
pressure monitoring
• Supports the WirelessHART standard (HART V 7.1)
• Very high security level for wireless data transmission
• Built-in local user interface (LUI) with 3-button operation
• Optimum display and readability using graphical display
(104 x 80 pixels) with integrated backlight
• Stand-by (deep sleep phase) can be activated and deactivated device with push of a button
• Battery power supply
• Battery service live up to 5 years
• Extend battery service life with HART modem interface which
can be shut off
• Optimized power consumption through new design, and increase in battery service life.
• Simple configuration thanks to SIMATIC PDM
• Device meets IP65 degree of protection
• Can be used for absolute and gauge pressure measurements
■ Benefits
The SITRANS P280 is a pressure transmitter that features Wireless HART as the standard communication interface.
Also available is a wired interface to connect a HART modem:
• Flexible pressure measurements
• Save costs on writing for difficult installation conditions. Wireless technology offers cost advantages in cases where extensive wiring cost would normally apply.
• It enables additional hitherto unfeasible measuring points,
particularly for monitoring purposes.
• Easy installation on moveable equipment
• Enables cost-effective temporary measurements, for example
for process optimizations.
• Optimum solution in addition to wired communication and new
possibilities for system solutions in process automation
The aerial features a rotatable joint which can be used for directional alignment. Wireless signals can thus be optimally received
and transmitted.
A special highlight is the option for direct operation on the device. The operating strategy used in this case seamlessly integrates into the strategy of all new Siemens field devices.
Using the device’s control buttons, it is easy to turn the HART
modem interface of the device on and off. The device can be put
to passive status and reactivated at any time. This helps to extend the service life of the battery.
The SITRANS P280 transmitter features a ceramic measuring
cell for gauge and absolute pressure measurements.
■ Function
The SITRANS P280 can join to a WirelessHART network. It can
be parameterized and operated through this network. Measured
process values are transported via the network to the
SIEMENS IE/WSN-PA link.
Field device data received by the IE/WSN-PA LINK is transmitted
to the connected systems, for example the process control system SIMATIC PCS 7. For an introduction of WirelessHART,
please see the FI 01 catalogue, section 8 or
http://www.siemens.com/wirelesshart.
Detailed information on IE/WSN-PA can be found in the FI 01 catalogue, section 7 or http://www.siemens.com/wirelesshart.
Siemens FI 01 · 2015 US Edition
1/45
1
© Siemens AG 2015
Pressure Measurement
Transmitters with WirelessHART
SITRANS P280 for gauge and absolute pressure
■ Integration
■ Technical specifications
Connecting to SIMATIC PCS 7
SITRANS P280 WirelessHART pressure transmitter
The integration of field devices in SIMATIC PCS 7 and other process control systems can now be done seamlessly and cost-effectively with wireless technology, especially in situations where
high wiring costs may be expected. Of particular interest are
measuring points which are to be added and for which no MSR
wiring is available.
Mode of operation
Where larger distances between the IW/WSN-PA LINK and control systems need to be overcome, this connection can also be
implemented on a wireless and cost-effective basis using the
products of the SCALANCE W family.
ES
OS
Measuring principle
piezo-resistive
Measured variable
Gauge and absolute pressure
Gauge pressure input
Measuring range
0 ... 1.6 bar (0 ... 23 psi)
0 ... 10 bar (0 ... 145 psi)
0 ... 50 bar (0 ... 725 psi)
0 ... 200 bar (0 ... 2900 psi)
0 ... 320 bar (0 ... 4641 psi)
Overload limit/Bursting pressure
4 bar (58 psi)
20 bar (290 psi)
100 bar (1450 psi)
400 bar (5801 psi)
640 bar (9282 psi)
Units
mbar, bar, m4H2O, i4H2O, atm,
Torr, gcm², kgcm², Pa, kPa, Mpa,
psi, mmHG, mmH2O, ftH2O,
inHG, inH2O
MS
Absolute pressure input
Industrial Ethernet
SIMATIC
system
SITRANS
AW200
SIMATIC ET 200
with HART support
Wireless
HART
IE/WSNPA LINK
DP/PA
LINK
SITRANS SITRANS
P280
TF280
PROFIBUS
1
Measuring range
0 ... 1.6 bar a (0 ... 23 psia)
0 ... 10 bar a (0 ... 145 psia)
0 ... 50 bar a (0 ... 725 psia)
0 ... 200 bar a (0 ... 2900 psia)
0 ... 320 bar a (0 ... 4641 psia)
Overload limit/Bursting pressure
4 bar a (58 psia)
20 bar a (290 psia)
100 bar a (1450 psia)
400 bar a (5801 psia)
640 bar a (9282 psia)
Units
mbar, bar, m4H2O, i4H2O, atm,
Torr, gcm², kgcm², Pa, kPa, MPa,
psi, mmHG, mmH2O, ftH2O,
inHG, inH2O
Output
Output signal
2.4 GHz Wireless signal with
TSMP (Time Synchronized Mesh
Protocol)
Measuring accuracy
as per IEC 60770-1
Error in measurement at limit setting
incl. hysteresis and reproducibility
typ. 0.17 % of sensor’s span
max. 0.25 % of sensor’s span
Long-term stability
max.  0.25 % of sensor/year
span
Influence of ambient temperature
typ. 0.07 %/10K, max. 0.2 %/10 K
of sensor’s span
Rated conditions
Ambient conditions
• Ambient temperature
-40 ... +80 °C (-40 ... +176 °F)
(in ambient temperatures below 20 °C (-4 °F) and above +70 °C
(158 °F), readability of the display
is limited.)
• Storage temperature
-40 ... +85 °C (-40 ... +185 °F)
• Relative humidity
< 95 %
Climatic class
4K4H in accordance with EN
60721-3-4(stationary use at locations not protected against
weather)
Degree of protection
IP65/NEMA 4
Allowable media temperature
-40 ... 85 °C (-40 ... +185 °F)
PROFIBUS PA
Integration of a meshed network in SIMATIC PCS7
■ Configuration
Configuration of the SITRANS P280 may be carried out as follows:
• Initial comissioning for the SITRANS P280 with SIMATIC PDM
is generally carried out via a HART modem or the integrated
local user interface, since the network ID and join key must be
set up on the device before it can be accepted and integrated
into the WirelessHART network.
• Once it is integrated into the network, the device can be conveniently operated with the WirelessHART network, the onsite
HART modem or via the local user interface.
• Siemens WirelessHART devices operate with optimum coexistence to SCALANCE W family products.
1/46
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Pressure Measurement
Transmitters with WirelessHART
SITRANS P280 for gauge and absolute pressure
Design
Enclosure material
low-copper die-cast aluminum,
AC-AlSi12(Fe)
Shock resistance
in accordance with
DIN EN 60068-2-29 / 03.95
Resistance to vibration
in accordance with
DIN EN 60068-2-6/ 12.07
Weight
• without battery
1.5 kg (3.31 lb)
• With battery
1.6 kg (3.53 lb)
Dimensions (W x H x D)
See Dimensional drawing
Process connection
• G½B male thread as per
EN837-1
• ½-14 NPT
Sensor break
Is recognized
Displays and controls
Display (with illumination)
• Size of display
104 x 80 pixels
• Number of digits
adjustable
• Number of spaces after comma
adjustable
Setting options
• on site with 3 buttons
• with SIMATIC PDM or HARTCommunicator
Power supply
Battery
Selection and Ordering data
Article No.
SITRANS P280 WirelessHART
pressure transmitter
(Required battery not included with delivery, see
accessories)
7 MP 1 1 2 0 -
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Measuring cell filling
Dry measuring cell
0
Measuring span
Gauge pressure
0 ... 1.6 bar (0 ... 23 psi)
0 ... 10 bar (0 ... 145 psi)
0 ... 50 bar (0 ... 725 psi)
0 ... 200 bar (0 ... 2900 psi)
0 ... 320 bar (0 ... 4641 psi)
D
E
F
G
H
Absolute pressure
0 ... 1.6 bar a (0 ... 3 psia)
0 ... 10 bar a (0 ... 145 psia)
0 ... 50 bar a (0 ... 725 psia)
0 ... 200 bar a (0 ... 2900 psia)
0 ... 320 bar a (0 ... 4641 psia)
M
N
P
Q
R
Wetted parts
K
Ceramic
Display
1
Display, visible
3.6 V DC
Communication
Enclosure
1
Die-cast aluminum
Radio
WirelessHART V7.1 conforming
Process connection
Transmission frequency band
2.4 GHz (ISM-Band)
Transmission range under reference
conditions
Up to 250 m (line of sight) in outside areas
Up to 50 m (greatly dependent on
obstacles) in inside areas
G½ as per EN 837-1
½-14 NPT
Communication interfaces
77777 777 0
• HART communication with
HART modem
• WirelessHART
Certificates and approvals
Wireless communication approvals
R&TTE, FCC
General Product Safety
CSAUS/C, CE, UL
Classification according to pressure
equipment directive
(PED 97/23/EC)
Gases: Fluid group 1
Liquids: Fluid group 1;
meets requirements as per Section 3, Subsection 3 (sound engineering practice)
0
1
Explosion protection
A
Without
Antenna
A
Variable, attached to device
Order code
Further designs
Please add "-Z" to Article No. and specify Order
code(s) and plain text.
Stainless steel tag plate (measuring point
description)
max. 16 digits entered in plain text
Y15: ....................
Y15
Measuring point message
max. 27 characters entered in plain text: Y16:
....................
Y16
Accessories
Article No.
Lithium battery for SITRANS TF280/P280
7MP1990-0AA00
Mounting bracket, steel
7MF4997-1AC
Mounting bracket, stainless steel
7MF4997-1AJ
Cover, die-cast aluminum, without window
7MF4997-1BB
Cover, die-cast aluminum, with window
7MF4997-1BE
IE/WSN-PA LINK
see Sec. 7
HART modem with USB interface
7MF4997-1DB
SIMATIC PDM
see Sec. 8
Siemens FI 01 · 2015 US Edition
1/47
1
© Siemens AG 2015
Pressure Measurement
Transmitters with WirelessHART
SITRANS P280 for gauge and absolute pressure
■ Dimensional drawings
Ø 80 (3.15)
270 (10.6)
90
°
Antenna
130 (5.12)
Ø 80 (3.15)
G½B
SW27
52 (2.05)
1
Pressure balance element
M12 x 1.5
238 (9.4) for absolute pressure
246 (9.7) for gauge pressure
SITRANS P280 WirelessHART pressure transmitter, process connection G½", dimensions in mm (inch)
The dimensional drawing of the mounting bracket see on page 1/171.
1/48
Siemens FI 01 · 2015 US Edition
154 (6.06)
© Siemens AG 2015
Pressure Measurement
Transmitters with WirelessHART
SITRANS P280 for gauge and absolute pressure
Ø 80 (3.15)
264 (10.4
90
°
Antenna
130 (5.12)
½ - 14 NPT
Ø 80 (3.15)
SW27
52 (2.05)
154 (6.06)
Pressure balance element
M12 x 1.5
238 (9.4) for absolute pressure
246 (9.7) for gauge pressure
SITRANS P280 WirelessHART pressure transmitter, process connection ½ - 14 NPT, dimensions in mm (inch)
The dimensional drawing of the mounting bracket see on page 1/171.
Siemens FI 01 · 2015 US Edition
1/49
1