FREESCALE MPXC2011DT1_12

Pressure
Freescale Semiconductor
Data Sheet: Technical Data
MPXC2011DT1
Rev 9, 10/2012
High Volume Sensor for Low
Pressure Applications
MPXC2011DT1
Freescale Semiconductor has developed a low cost, high volume, miniature
pressure sensor package which is ideal as a sub-module component or a
disposable unit. The unique concept of the Chip Pak allows great flexibility in
system design while allowing an economic solution for the designer. This new
chip carrier package uses Freescale Semiconductor's unique sensor die with its
piezoresistive technology, along with the added feature of on-chip, thin-film
temperature compensation and calibration.
NOTE:Freescale Semiconductor is also offering the Chip Pak package in
application-specific configurations, which will have an “SPX” prefix,
followed by a four-digit number, unique to the specific customer
Features
•
•
•
•
•
Low Cost
Integrated Temperature Compensation and Calibration
Ratiometric to Supply Voltage
Polysulfone Case Material (ISO 10993)
Provided in Easy-to-Use Tape and Reel
Pressure Sensors
0 to 75 mmHg
(0 to 10 kPa)
Application Examples
•
•
•
•
Respiratory Diagnostics
Air Movement Control
Controllers
Pressure Switching
ORDERING INFORMATION
Device Name
MPXC2011DTI
Package Options
Tape and Reel
Case
No.
423A
Gauge
Pressure Type
Differential
•
Absolute
Device Marking
Date Code, Lot ID
CHIP PAK PACKAGE
MPXC2011DT1
CASE 423A-03
NOTE: The die and wire bonds are exposed on the front side of the Chip Pak (pressure is applied to the backside of the device).
Front side die and wire protection must be provided in the customer's housing. Use caution when handling the devices
during all processes.
© 2006, 2008-2010, 2012 Freescale Semiconductor, Inc. All rights reserved.
Pressure
Freescale Semiconductor's MPXC2011DT1 Pressure
Sensor has been designed for medical usage by combining
the performance of Freescale Semiconductor's shear stress
pressure sensor design and the use of biomedically
approved materials. Materials with a proven history in
medical situations have been chosen to provide a sensor that
can be used with confidence in applications, such as invasive
blood pressure monitoring. It can be sterilized using ethylene
oxide. The portions of the pressure sensor that are required
to be biomedically approved are the rigid housing and the gel
coating.
The rigid housing is molded from a white, medical grade
polysulfone that has passed extensive biological testing
including: 10993-5:1999, 10993-10:2002, and 1099311:1993.
The MPXC2011DT1 contains a silicone dielectric gel
which covers the silicon piezoresistive sensing element. The
gel is a nontoxic, nonallergenic elastomer system which
meets all USP XX Biological Testing Class V requirements.
The properties of the gel allow it to transmit pressure
uniformly to the diaphragm surface, while isolating the
internal electrical connections from the corrosive effects of
fluids, such as saline solution. The gel provides electrical
isolation sufficient to withstand defibrillation testing, as
specified in the proposed Association for the Advancement of
Medical Instrumentation (AAMI) Standard for blood pressure
transducers. A biomedically approved opaque filler in the gel
prevents bright operating room lights from affecting the
performance of the sensor.
MPXC2011DT1
2
Sensors
Freescale Semiconductor, Inc.
Pressure
MAXIMUM RATINGS
Table 1. Maximum Ratings(1)
Rating
Symbol
Value
Unit
Maximum Pressure (Backside)
Pmax
75
kPa
Storage Temperature
Tstg
-25 to +85
C
Operating Temperature
TA
+15 to +40
C
1. Exposure beyond the specified limits may cause permanent damage or degradation to the device.
OPERATING CHARACTERISTICS
Table 2. Operating Characteristics (VS = 10 Vdc, TA = 25C unless otherwise noted, P1 > P2)
Characteristic
Symbol
Min
Typ
Max
Unit
Pressure Range(1)
POP
0
—
10
kPa
Supply Voltage(2)
VS
—
3
10
Vdc
Supply Current
Io
—
6.0
—
mAdc
VFSS
24
25
26
mV
Voff
-1.0
—
1.0
mV
Sensitivity
V/P
—
2.5
—
mV/kPa
Linearity(5)
—
-1.0
—
1.0
%VFSS
Pressure Hysteresis(5) (0 to 10 kPa)
—
—
0.1
—
%VFSS
Temperature Hysteresis(5) (+15C to +40C)
—
—
0.1
—
%VFSS
TCVFSS
-1.0
—
1.0
%VFSS
TCVoff
-1.0
—
1.0
mV
Input Impedance
Zin
1300
—
2550

Output Impedance
Zout
1400
—
3000

Response Time(6) (10% to 90%)
tR
—
1.0
—
ms
Warm-Up
—
—
20
v
ms
Offset Stability(7)
—
—
0.5
—
%VFSS
Full Scale Span(3)
Offset(4)
Temperature Effect on Full Scale Span(5)
Temperature Effect on Offset(5)
1. 1.0 kPa (kiloPascal) equals 0.145 psi.
2. Device is ratiometric within this specified excitation range. Operating the device above the specified excitation range may induce additional
error due to device self-heating.
3. Full Scale Span (VFSS) is defined as the algebraic difference between the output voltage at full rated pressure and the output voltage at the
minimum rated pressure.
4. Offset (Voff) is defined as the output voltage at the minimum rated pressure.
5. Accuracy (error budget) consists of the following:
Linearity:
Output deviation from a straight line relationship with pressure, using end point method, over the specified
pressure range.
Temperature Hysteresis: Output deviation at any temperature within the operating temperature range, after the temperature is cycled to
and from the minimum or maximum operating temperature points, with zero differential pressure applied.
Pressure Hysteresis:
Output deviation at any pressure within the specified range, when this pressure is cycled to and from the
minimum or maximum rated pressure, at 25C.
TcSpan:
Output deviation at full rated pressure over the temperature range of 0 to 85C, relative to 25C.
TcOffset:
Output deviation with minimum rated pressure applied, over the temperature range of 0 to 85C, relative to 25C.
6. Response Time is defined as the time for the incremental change in the output to go from 10% to 90% of its final value when subjected to a
specified step change in pressure.
7. Offset stability is the product's output deviation when subjected to 1000 hours of Pulsed Pressure, Temperature Cycling with Bias Test.
MPXC2011DT1
Sensors
Freescale Semiconductor, Inc.
3
Pressure
PACKAGE DIMENSIONS
A
M
C
L
F
N
B
1
2 3
4
V
K
DETAIL A
–T–
D1
G
J
H
FRONT VIEW
E
END VIEW
AC
NOTES:
1. DIMENSIONING AND TOLERANCING PER ANSI
3.
Y14.5M, 1982.
2. CONTROLLING DIMENSION: INCH.
4.
F
AA
AB
AD
D2
DETAIL A
BACK VIEW
DIM
A
B
C
D1
D2
E
F
G
H
J
K
L
M
N
V
AA
AB
AC
AD
INCHES
MIN
MAX
0.240
0.260
0.350
0.370
0.140
0.150
0.012
0.020
0.014
0.022
0.088
0.102
0.123
0.128
0.045
0.055
0.037
0.047
0.007
0.011
0.120
0.140
0.095
0.105
0.165
0.175
0.223
0.239
0.105
0.115
0.095
0.107
0.015
0.035
0.120
0.175
0.100
0.115
MILLIMETERS
MIN
MAX
6.10
6.60
8.89
9.40
3.56
3.81
0.30
0.51
0.36
0.56
2.24
2.59
3.12
3.25
1.14
1.40
0.94
1.19
0.18
0.28
3.05
3.56
2.41
2.67
4.19
4.45
5.66
6.07
2.67
2.92
2.41
2.72
0.38
0.89
3.05
4.45
2.54
2.92
STYLE
2: 1:
STYLE
PINPIN
1. 1.
GND
VCC
2. 2.S++OUT
3. 3.Vs–OUT
4. 4.S-GROUND
CASE 423A-03
ISSUE C
CHIP PAK PACKAGE
MPXC2011DT1
4
Sensors
Freescale Semiconductor, Inc.
Pressure
Table 3. Revision History
Revision
number
Revision
date
Description of changes
9
10/2012
• Deleted references to device number MPXC2012DT (a no-gel option) throughout the document
MPXC2011DT1
5
Sensors
Freescale Semiconductor, Inc.
How to Reach Us:
Information in this document is provided solely to enable system and software
implementers to use Freescale products. There are no express or implied copyright
Home Page:
freescale.com
Web Support:
freescale.com/support
licenses granted hereunder to design or fabricate any integrated circuits based on the
information in this document.
Freescale reserves the right to make changes without further notice to any products
herein. Freescale makes no warranty, representation, or guarantee regarding the
suitability of its products for any particular purpose, nor does Freescale assume any
liability arising out of the application or use of any product or circuit, and specifically
disclaims any and all liability, including without limitation consequential or incidental
damages. “Typical” parameters that may be provided in Freescale data sheets and/or
specifications can and do vary in different applications, and actual performance may
vary over time. All operating parameters, including “typicals,” must be validated for each
customer application by customer’s technical experts. Freescale does not convey any
license under its patent rights nor the rights of others. Freescale sells products pursuant
to standard terms and conditions of sale, which can be found at the following address:
freescale.com/salestermsandconditions.
Freescale, the Freescale logo, Energy Efficient Solutions logo, are trademarks of
Freescale Semiconductor, Inc., Reg. U.S. Pat. & Tm. Off. Xtrinsic is a trademark of
Freescale Semiconductor, Inc. All other product or service names are the property of
their respective owners.
© 2012 Freescale Semiconductor, Inc.
MPXC2011DT1
Rev. 9
10/2012