3630 DATA LOGGER Series

3630 DATA LOGGER Series
Data Logger
For Recording Temperature/Humidity, Instrumentation Readings, Load Current, Leak Current, Voltage,
Pulse Counts, Illumination
Data Loggers for All Types of Measurements
http://www.hioki.co.jp/
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
1
The 3630 DATA LOGGER Series are compact data loggers
that are not much larger than a business card and weigh a mere
70 to 130 grams. Easy to operate, these handy instruments can
record up to 16,000 or 32,000 data elements, and store data in
nonvolatile memory that retains information even if the batteries are dead. The DATA LOGGER Series includes models
that can read temperature/humidity, instrumentation readings,
voltage, current and other special types of data, and can be
used in a broad range of measurements, including HACCP and
ESCO applications. Data recorded in a DATA LOGGER can
transferred to a personal computer through the 3911-20 COMMUNICATION BASE. Once the data is loaded into a personal computer, it can be processed, graphically displayed and
managed on the PC.
For HACCP-related Temperature and Humidity Recording
●HACCP and ESCO applications ●Temperature and humidity management in transport ●Temperature
and humidity management in food processing facilities ●Temperature and humidity management in warehouses or cold storage ●Energy conservation ●Temperature management in cooling processes
●Temperature and humidity management and recording for air conditioning systems
TEMPERATURE/HUMIDITY LOGGER
For Recording Load Current and Monitoring Leak Current
●Recording current of instrumentation signals
●Monitoring abnormal load current ●Management
of plant operation status ●Monitoring leak current
TEMPERATURE LOGGERs
INSTRUMENTATION
LOGGER
CLAMP LOGGER
3641-20
3631-20
3632-20
3633-20
3634-20
3636-20
Can alternately record
temperature and humidity
on two channels for temperature and humidity
measurement
-40 ˚C to 85.0 ˚C
0.0%rh to 100.0%rh
Can alternately record
temperature and humidity
on two channels for temperature and humidity
measurement
0 ˚C to 50.0 ˚C
20.0%rh to 95.0%rh
Waterproof with builtin sensor
for temperature measurement
-20.0˚C to 70.0˚C
External sensor for
temperature measurement
-40.0˚C to 180.0˚C
For measuring typical
instrumentation signals
Range: 20.00mA DC
For measuring alternating current on two channels
Range: 50.00/500.0A AC
(Clamp sensors sold
separately)
For Recording Voltage
For Recording Precipitation or Illumination
●Recording analog output from a variety of sensors ●Monitoring fluctuations in the power supply at a plant or office
●Count recording for precipitation gauges,
water level gauges, etc. ●Measuring illumination in a plant or office
VOLTAGE LOGGERs
VOLTAGE LOGGER
AC VOLTAGE LOGGER
PULSE LOGGER
Illumination Logger
3638-20
3635-24 to 26
3645-20
3637-20
3639-20
3640-20
For measuring alternating
current on two channels
Range: 100.0mA/1000mA
AC
(Clamp sensors sold separately)
For measuring DC voltage
-24: ±500.0mV DC
-25: ±5.000V DC
-26: ±50.00V DC
With preheat function
For measuring DC
voltage
Range: ±50.00mV to
±50.00V DC
For measuring AC
voltage
Range: 600.0V AC
For cumulative pulse
measurement for precipitation gauges, flow
gauges, etc.
For illumination measurement
Range: 2,000 lux to
200,000 lux
LEAK LOGGER
Space-saving, Ultra-compact Temperature Logger
Multichannel Recording of Temperature, Humidity and Voltage
●Temperature recording during transport ●Temperature management in
refrigerators and freezers ●Recording temperatures in food processing
plants ●Recording core temperatures when processing food
●Multipoint simultaneous temperature recording ●Correlated recording of various signals and temperature data ●Management of realtime data collection over a LAN
BUTTON-TYPE TEMPERATURE LOGGER and DATA READER
MULTICHANNEL DATA LOGGER/MEMORY HiCORDER
3650
3920-01 DATA READER (with dedicated analysis software)
Battery, temperature
sensor
Ultra-compact buttontype temperature logger
with built-in memory
-40˚C to 85˚C
8420-01 / 8421-01
Multichannel data loggers for measuring
temperature, humidity
and voltage on up to 8
or 16 channels, respectively
Example of temperature graph produced by
using the dedicated analysis software
Main Screen
Data on a maximum of 256 channels can be collected in real-time through a LAN when using the
optional 9334 LOGGER COMMUNICATOR.
2
Store Up to 32,000 Data Elements in a Small Unit
through Simple Operations
Data is retained even when the batteries are dead
Because nonvolatile memory is used in the DATA LOGGER
Series, data is retained even when the batteries are dead or are
being replaced, ensuring that no valuable data is lost.
Battery power indicator
The battery status is displayed through a four-level indicator.
This can be used as a guide for identifying when the battery
needs to be replaced.
Waterproof construction
Large data storage capacity
A maximum of either 16,000 or 32,000 data elements can be
stored in the DATA LOGGER Series. The 3631 can store 8000
elements of temperature data and 8000 elements of humidity
data.
The 3632 is completely waterproof*, and is ideal for applications
such as temperature management in a refrigerator. The 3631 to
3635 and 3641 are water resistant.
*Not suitable for continual use under water.
Power saving function
Identify data by inputting comments
The dedicated software that is provided with the 3911-20 allows
you to set the current time, recording interval, start of recording,
and the recording method, and also to input comments. By
loading simple comments into the logger, the data can then be
easily identifed after they are transferred to the PC.
The power saving function can be enabled or disabled through
the dedicated software that is provided with the 3911-20. When
enabled, the power saving function turns off the display while
the unit records data. Pressing a key causes the current
measurement to be temporarily displayed on the screen. When
the power saving function is disabled, the current measurement
is always visible on the display.
Use a PC to Analyze and Process Large Volumes of Recorded Data
3911-20 COMMUNICATION BASE
Infrared communication: Approximately 500 data
elements/second (The DATA LOGGER and the
3911-20 perform communications simultaneously.
The photo is shown for illustrative purposes only.)
DATA LOGGER ➔ 3911-20
The 3911-20 COMMUNICATION BASE is used to transfer data from a Data Logger to a
PC. The 3911-20 can collect data on up to 16 channels. Data from multiple DATA
LOGGERs that are installed in fixed positions can be collected by the 3911-20 and then
transferred to a PC for analysis and processing.
●Recording capacity: 32,000 data elements × 8 channels (3631-20 to 3635-20: 16,000 data elements × 16 channels
maximum) ●Communication method: DATA LOGGER ↔ 3911-20: Infrared communications/3911-20 ↔ PC: RS232C ●Power supply: LR03(AAA) alkaline dry cell battery × 4 ●Dimensions and mass: 69 (W) × 92 (H) × 36 (D)
mm; 150g ●Accessories: PC communication software, Windows 95/98/NT4.0 (for DOS/V, PC98); Functions:
Graph display, data list, printing (data and graphs), data processing, file saving (proprietary format or text format)
9637 RS-232C CABLE (9-pin to 9-pin crossed cable/1.8m)
9638 RS-232C CABLE (9-pin to 25-pin crossed cable/1.8m)
* Select cable appropriate for your PC.
3911-20 ➔ PC
RS-232C CABLE
Approximately 1000 data
elements/second
Graph display
Number of data elements
Average value
Recording date and time
Maximum value/minimum value
Measurement data list display
Scaling
Data collected by the 3911-20 can be
processed in a variety of ways on a PC.
Measurement data list
3
For Recording
Temperature and Humidity
Wide measurment
range with long life
( about 5 years)
humidity sensor 9680
3631-20 HUMIDITY LOGGER
Features
Measured items
Temperature and humidity (2 channels) Temperature and humidity (2 channels)
Temperature:
-20.0˚C to 70˚C ( using the built-in temperature sensor)
-40.0˚C to 180.0˚C ( using an external temperature sensor)
-40.0˚C to 85.0˚C (using the 9680 Temp. and Humidity Sensor)
Humidity:
0.0% to 100.0%rh (using the 9680 Temp. and Humidity Sensor)
Temperature:
-20.0˚C to 70˚C ( using the built-in temperature sensor)
-40.0˚C to 180.0˚C ( using an external temperature sensor)
0.0˚C to 50.0˚C (using the 9630 Temp. and Humidity Sensor)
Humidity:
20.0% to 95.0%rh (using the 9630 Temp. and Humidity Sensor)
Temperature:
±0.5˚C (0.0 to 35.0˚C)
±1.0˚C (-40.0 to -0.1˚C / 35.1 to 70.0˚C)
±2.0˚C (70.1 to 120.0˚C)
±5.0˚C (120.1 to 180.0˚C)
Humidity:
See the Accuracy table 1 at bottom right .
9680 TEMPERATURE AND HUMIDITY SENSOR × 1
Temperature:
±0.5˚C (0.0 to 35.0˚C)
±0.5˚C (0.0 to 35.0˚C)
±1.0˚C (-40.0 to -0.1˚C / 35.1 to
±1.0˚C (-40.0 to -0.1˚C / 35.1 to 70.0˚C)
70.0˚C)
±2.0˚C (70.1 to 120.0˚C)
±5.0˚C (120.1 to 180.0˚C)
Humidity:
See the Accuracy table 2 at bottom right .
9630 TEMPERATURE AND HUMIDITY SENSOR × 1
−
■ Common specifications
●Response time: Built-in sensor: Approximately 25 minutes; External temperature sensor: Depends on response time of sensor (Refer to chart below.)
●Storage capacity: 16,000 data elements (3631-20, 3641-20: 8,000 data elements × 2 channels) ●Recording start: Manual start or timer start ●Recording
interval: 2/5/10/15/20/30 seconds,
1/2/5/10/15/20/30/60 minutes ●Interface: Infrared communications
(Communication between unit and the 3911-20, or between the 3911-20 and a
PC are handled through an RS-232C connection cable.) ●Settings that can be
9631-01/9631-11/9631-21
TEMPERATURE SENSOR
9631-02
TEMPERATURE SENSOR
9631-03
TEMPERATURE SENSOR
Temperature range: 0.0˚C to 50.0˚C
Humidity range: 20.0 to 95.0%rh
Response time:
Temperature: Approximately 100 seconds
Humidity: Approximately 10 minutes
Sensor dimensions:
60 (W) x 25 (H) x 12 (D) mm
Cord length:
9630: Approximately 1 m (provided)
9630-01: Approx. 5 m
9630-02: Approx. 10 m
(Molded plastic type)
Temperature range: -40.0˚C to 180.0˚C
Response time:
Approximately 100 seconds
Sensor dimensions:
5 mm diameter x 28 mm
Cord length:
9631-01: Approx. 1 m
9631-11: Approx. 5 m
9631-21: Approx. 10 m
(Needle type)
Temperature range: -40.0˚C to 120.0˚C
Response time:
Approximately 20 seconds
Dimensions of metallic portion:
1.3 mm diameter x 25 mm
Cord length: Approx. 1 m
(Sheathed type)
Temperature range: -40.0˚C to 120.0˚C
Response time:
Approximately 90 seconds
Dimensions of metallic portion:
4 mm diameter x 180 mm
Cord length: Approx. 1 m
Outer diameter: 7 mm
Inner diameter: 3.2mm
(Lug type)
Temperature range: -30.0˚C to 180.0˚C
Response time: Approximately 45 seconds
Dimensions of metallic portion: 16.5 mm
Cord length:
9631-04: Approx. 1 m
9631-14: Approx. 5 m
9631-24: Approx. 10 m
(Molded plastic type)
Temperature range: -40.0˚C to 180.0˚C
Response time:
Approximately 100 seconds
Sensor dimensions:
5 mm diameter x 28 mm
Cord length: Approx. 30 mm
±0.5˚C (0.0 to 35.0˚C)
±1.0˚C (-40.0 to -0.1˚C /
35.1 to 70.0˚C)
±2.0˚C (70.1 to 120.0˚C)
±5.0˚C (120.1 to 180.0˚C)
−
● 9680 HUMIDITY SENSOR ACCURACY TABLE 1
9630/9630-01/9630-02
HUMIDITY SENSOR
9631-05
TEMPERATURE SENSOR
-20.0˚C to 70.0˚C
(when using the built-in temperature sensor)
-40.0˚C to 180.0˚C
(when using an external temperature sensor)
made through the main unit: Recording interval, recording start/stop ●Settings that can
be made through the 3911-20: Current time, recording interval, recording start time,
recording method, comment ●Power supply: LR03(AAA) alkaline dry cell battery ✕
2; Maximum rated power: 0.1VA; Battery life: Approximately 2 years (3631-20:
Approximately 1 year) when recording interval is set at one minute ●Dimensions and
mass: Approximately 57 (W) ✕ 74 (H) ✕ 19.5 (D) mm; 70g ●Operating environment:
Indoors, at an altitude of 2,000m or less, -20.0˚C to 70.0˚C, 80%rh or less (with no condensation)
Temperature and Humidity Sensors
9631-04/9631-14/9631-24
TEMPERATURE SENSOR
Temperature (1 channel)
-20.0˚C to 70.0˚C (when using the
built-in temperature sensor)
Only the built-in sensor may be hused.
Waterproof type Conforms with
IP67 (Models other than the 363220 conform with IP54.)
9680/9680-01/9680-02
HUMIDITY SENSOR
Temperature range: -40.0˚C to 85.0˚C
Humidity range: 0.0 to 100.0%rh
Response time:
Temperature: Approximately 100 seconds
Humidity: Approximately 300 seconds
Sensor dimensions:
30 (W) x 13(H) x 8 (D) mm
Cord length:
9680: Approximately 1 m (provided)
9680-01: Approx. 5 m
9680-02: Approx. 10 m
100
95
80
60
40
20
Out of guaranteed area
Accessories
Temperature (1 channel)
Relative Humidity (% rh )
Accuracy
Temperature logger that supports a builtin temperature sensor or an external
temperature sensor (sold separately)
Waterproof temperature logger
that supports only a built-in temperature sensor.
±10%rh
±8%rh
±10%rh
±8%rh
±6%rh
±8%rh
±6%rh
±5%rh
±6%rh
0
20
10
-40 0
30
40
50
85
Temperature (ºC)
● 9630 HUMIDITY SENSOR ACCURACY TABLE 2
Relative Humidity (% rh )
Measurement range
(resolution: 0.1˚C,
0.1%rh)
3632-20 TEMPERATURE LOGGER 3633-20 TEMPERATURE LOGGER
±10%rh
NEW 3641-20 HUMIDITY LOGGER
Temperature and humidity logger that can record temper- Temperature and humidity logger that can record temperature and humidity on two channels using the 9680 TEM- ature and humidity on two channels using the 9630 TEMPERATURE AND HUMIDITY SENSOR provided.
PERATURE AND HUMIDITY SENSOR provided.
(The value for referance )
Model
100
95
±6%rh
80
60
±8%rh
±10%rh
±5%rh
Out of guaranteed area
±8%rh
±10%rh
40
20
0
±8%rh
Out of guaranteed area
10
20
40
30
Temperature (ºC)
50
NOTE : The response times indicated are reference values for the
time until 90% of the value is indicated for a given change in temperature or humidity.
The temperature sensors are all thermisters, and the humidity sensor 9630 has a polymer structure (resistor type), 9680 has a polymer structure (capacity type).
The 9630 can not use with 3641, the 9680 can not use with 3631.
4
For Recording Instrumentation Readings,
Load Current, and Leak Current
3636-20 CLAMP LOGGER
3638-20 LEAK LOGGER
Can record load current through two channels
using clamp sensors (sold separately)
Can record leak current through two channels
using clamp sensors (sold separately)
Load current (2 channels)
Leak current (2 channels)
3634-20 INSTRUMENTATION LOGGER
Model
Can measure signals up to 20 mA DC, ideal
for measuring instrumentation signals.
Features
For instrumentation/0 to 20 mA DC (1 channel)
Measured items
−
Compatible clamps
Measuring range
DC 0.00 to 20.00 mA
Accuracy
Range in which accuracy
is guaranteed: 23 ± 5ºC
±0.8% rdg. ±5 dgt.
Temperature coefficient: 0.08%/˚C
Rectification method
−
Storage capacity
16,000 data elements
9650 / 9651
9657 / 9658 / 9659
0.00 to 500.0 Arms AC (sine wave)
(Two ranges: 50.00 A/500.0 A)
0.0 to 1000 mArms AC (sine wave)
(Two ranges: 100.0 mA/1000 mA)
±1% rdg. ±5 dgt. (main unit only)
±1% rdg. ±5 dgt. (main unit only)
±2% rdg. ±10 dgt. (main unit + sensor)*1
±2.5% rdg. ±8 dgt. (main unit + sensor)*
±2% rdg. ±6 dgt. (main unit + sensor)*2
(*When range is 50 A/500 A if using the 9650; *1
( When range is 100 mA; *2 When range is
when range is 500 A if using the 9651)
1000 mA)
True RMS
With 50/60Hz filter/True RMS
32,000 data elements (1 channel), 16,000 data elements (2 channels)
Instantaneous value recording/Average value recording Maximum value recording/Average value recording
(average value during the recording interval)
(average value during the recording interval)
Approximately one year with a recording interval of one minute
Approximately one month with a recording interval of one second
Continuous operating time Approximately two years with a recording inter(when using instantaneous value recording)
val of one minute
Approximately ten days with a recording interval of 0.2 seconds
(when power saving function is enabled)
Approximately one month (when using average value recording)
Approx.
57.5
(W)
×
86.5
(H)
× 30.0 (D) mm; 130g
Approx.
57
(W)
×
74
(H)
×
19.5
(D)
mm;
70g
Dimensions and mass
Recording mode
−
■ Common specifications
●Alarm output (3636-20 and 3638-20 only): ON when measured value is outside range set by specified upper and lower limit values (open collector output) ●Recording start: Manual start or timer
start ●Recording interval: 1*/2/5/10/15/20/30 seconds, 1/2/5/10/15/20/30/60 minutes (*1 second setting is supported only by the 3636-20 and the 3638-20) ●Interface: Infrared communications
(Communication between unit and the 3911-20, or between the 3911-20 and a PC are handled
through an RS-232C connection cable.) ●Settings that can be made through the main unit:
Recording interval, recording start/stop ●Settings that can be made through the 3911-20: Current
time, recording interval, recording start time, recording method, comment ●Power supply:
LR03(AAA) alkaline dry cell battery × 2 (× 4 for the 3636-20 and the 3638-20); Maximum rated
power: 0.1VA ●Operating environment: Indoors, at an altitude of 2000m or less, -20.0ºC to 70.0ºC
(0ºC to 50ºC for the 3636-20 and the 3638-20), 80%rh or less (with no condensation) ●Accessories:
9632 CONNECTION CABLE × 1
■ CLAMP ON SENSOR Specifications
Accuracy
Lag current
External magnetic field effect
9650
9651
AC 100 A/AC 100 mA
AC 500 A/AC 500 mA
±1.5% rdg. ±0.03% f.s. (f.s. is rated primary current value)
−
−
9632 CONNECTION CABLE
Cord length: Approx. 1m
(Included with the 3634-20,
3635-24 to -26, 3636-20,
3638-20 and 3639-20; used
for alarm signal output by
the 3636-20, 3638-20, and
3639-20)
(Cord length: Approximately 3m)
For load current measurement (for 3636-20)
Model
Rated primary current/output
Used for input or for
alarm output
−
−
For leak current measurement (for 3638-20)
9659
9658
9657
AC 1 A/AC 25 mV
±1.0%rdg.±12µV
±3.5%rdg.±12µV
±1.0%rdg.±12µV
5 mA (when 100 A AC is input) 1 mA (when 10 A AC is input) 30 mA (when 500 A AC is input)
Equivalent to 5 mA at 400A AC/m; 7.5 mA max.
Frequency characteristics 40 Hz to 1 kHz (within ±8%) 40 Hz to 1 kHz (within ±3%)
45 to 66 Hz
Maximum allowable input 130 A continuous (45 to 66 Hz) 600 A continuous (45 to 66 Hz) 60 A continuous (45 to 66 Hz)
10 A continuous (45 to 66 Hz) 100 A continuous (45 to 66 Hz)
Maximum circuit voltage 300 Vrms AC (insulated conductor) 600 Vrms AC (insulated conductor) 300 Vrms AC (insulated conductor) 150 Vrms AC (insulated conductor) 460 Vrms AC (insulated conductor)
Measurable conductor diameter
Up to φ15 mm
Dimensions and
mass
Approx. 46(W) × 135(H) × 21(D) mm
200g
Up to φ46 mm
Up to φ40 mm
Approx. 77 (W) × 151 (H) × 42 (D) mm Approx. 74 (W) × 145 (H) × 42 (D) mm
340 g
340 g
Up to 12 × 30 mm
Up to 30 × 150 mm
Approx. 65 (W) × 52 (H) × 18 (D) mm Approx. 358 (W) × 108 (H) × 48 (D) mm
2.5 kg
50 g
5
For Recording DC and
AC Voltage
Model
Features
3635-24 to -26 VOLTAGE LOGGER 3645-20 VOLTAGE LOGGER 3637-20 AC VOLTAGE LOGGER
Can measure current DC voltage, ideal for
measuring instrumentation signals or for
measuring analog signals from sensors or
other devices
Voltage logger that permits control of preheating time and is compatible with various
sensors that have different response times
AC voltage measurement logger that can
measure up to 600V AC
Measured items
DC voltage
DC voltage + preheat function
AC voltage
Measuring range
DC ±500.0 mV (-24)/±5.000 V (-25)/
±50.00 V(-26)
DC ±50.00 mV/±500.0 mV/
±5.000V/±50.00V
0.0 to 600 Vrms AC (sine wave)
Accuracy
Range in which accuracy
is guaranteed: 23 ± 5ºC
±0.8% rdg. ±5 dgt.
Temperature coefficient: 0.08%/ºC
±0.5% rdg. ±5 dgt.
Temperature coefficient: (0.02%rdg.±1.5dgt.)/ºC
±1%rdg.±5dgt.
Rectification method
−
True RMS
True RMS
Storage capacity
16,000 data elements
32,000 data elements
Recording mode
−
Instantaneous value recording/Average value recording (average value during the recording interval)
Preheat function
−
Continuous operating time Approximately two years with a recording inter(when power saving funcval of one minute
tion is enabled)
Dimensions and mass Approx. 57 (W) × 74 (H) × 19.5 (D) mm; 70g
Accessories
9632 CONNECTION CABLE × 1
Open drain output (30 V, 20 mA max.)
Time: 0.5/1/2/5/10/30/60 seconds
−
Approximately one year with a recording interval of one minute (when using instantaneous
value recording)
Approximately one month (when using average value recording)
Approx. 57.5 (W) × 86.5 (H) × 30.0 (D) mm; 70g
Approx. 57.5 (W) × 86.5 (H) × 30.0 (D) mm; 130g
9632/9639 CONNECTION CABLE × 1
9639 CONNECTION CABLE × 1
■ Common specifications
●Recording start: Manual start or timer start ●Recording interval: 1*/2/5/10/15/20/30 seconds, 1/2/5/10/15/20/30/60 minutes (* 1 second setting is not supported by the 3635-24 to 26) ●Interface: Infrared communications (Communication between unit and the 3911-20,
or between the 3911-20 and a PC are handled through an RS-232C connection cable.)
●Settings that can be made through the main unit: Recording interval, recording start/stop
●Settings that can be made through the 3911-20: Current time, recording interval, recording start time, recording method, comment ●Power supply: LR03(AAA) alkaline dry cell
battery × 2 (× 4 for the 3645-20 and the 3637-20); Maximum rated power: 0.1VA
●Operating environment: Indoors, at an altitude of 2000m or less, -20.0ºC to 70.0ºC (0ºC
to 50ºC for the 3645-20 and the 3637-20), 80%rh or less (with no condensation)
■ Voltage logger applications
● For voltage input
9639 CONNECTION CORD
Cord length: Approx. 3 m
(provided with 3637-20 and 3645-20)
● For analog signal input (for the 3635-24 to -26)
φ3.5mm
φ2.5mm
● Recording output from sensors that require preheating (3645-20)
The 3645-20 can be used to record output from various sensors that require preheating,
such as water level sensors and soil sensors. When the 3645-20 is used in combination
with the 3631-20 TEMPERATURE AND HUMIDITY LOGGER, the 3639-20 PULSE
LOGGER or the 3640-20 ILLUMINATION LOGGER, the DATA LOGGER Series
simplifies the construction of databases that can be useful for agricultural civil engineering.
● One unit can record a variety of output
These units can record a broad range of output voltages, from several millivolts to 50V,
even for high sensitivity, low output sensors, such as light sensors.
● For recording analog output from a variety of test equipment
These units can record output from test equipment that outputs analog signals, such as
clamp ammeters, revolution counters, thermometers, and light meters.
9632 CONNECTION
9633 CONNECTION
CABLE
CABLE
Cord length: Approx. 1 m Cord length: Approx. 1 m
9634 CONNECTION
CABLE
Cord length: Approx. 1 m
6
For Recording Precipitation and
Illumination
For recording pulse counts from precipitation gauges, flow meters, etc.
3639-20 PULSE LOGGER
For measuring precipitation
For collecting power data
(Compatible with power meters that
have a pulse output connector)
Pulse logger for counting pulses output from precipitation
gauges, flow meters, etc.
● Compact and light at only 130g, the 3639-20 is easy to install, and yet
can record as much as 32,000 data elements.
● Prevents count errors through a noise filter
● Data that is collected can be put into a time bar graph or a cumulative
bar graph using the dedicated software
■ Basic Specifications
Input/accuracy
: Pulse input: 1 channel (Maximum display: 9999)/ within ±1 dgt.
• No voltage contact input (Count is incremented when short between terminals changes to open)
• Voltage inputHigh: +1.5V to +45V; Low: 0.0V to +0.2V(Count is incremented when voltage level goes high)
Recording interval : 1/2/5/10/15/20/30 seconds, 1/2/5/10/15/20/30/60 minutes, 1 unit (±100 ppm)
(precision)
Recording capacity : 32,000 data elements
: Infrared communications; data is collected in the 3911-20 COMMUNICATION
Interface
BASE (sold separately) and is then transferred to a PC through an RS-232C
interface.
: Status is output each recording interval when a set value is exceeded (open drain output).
Alarm output
: LR03(AAA) alkaline dry cell battery × 4
Power supply
: Approximately three months with 10 minute measurement interval (display off, no
Battery life
voltage contact measurement)
Dimensions and mass : Approx. 58 (W) × 87 (H) × 30 (D) mm; 130g (not waterproof)
: 9629 CONNECTION CABLE × 1; 9632 CONNECTION CABLE × 1
Accessories
9629 CONNECTION CABLE
(5 m/for measurement)
* The 3911-20 (sold separately) and a connecting cable are both required in order to analyze
measurement data on a PC. See page 2 for details.
For Recording Fluctuations in Illumination
3640-20 LUX LOGGER
To quickly log
fluctuations in illumination and then
process the data on
a PC
■ Basic Specifications
Optical element
Measuring range
Display
Accuracy
Recording interval
Recording capacity
Interface
:
:
:
:
:
:
:
3911-20 settings
Power supply
Battery life
:
:
:
Dimensions and mass :
:
Accessories
Silicon photo diode
2000/20,000/200,000 lx (manual range)
LCD 2.00/20.00/200.0 (measured value is display value × 1000)
±4% rdg. ±5 dgt.
1/2/5/10/15/20/30 seconds, 1/2/5/10/15/20/30/60 minutes
32,000 data elements
Infrared communications; data is collected in the 3911-20 COMMUNICATION BASE (sold separately) and is then transferred to a PC through an
RS-232C interface.
Current time, recording interval, start time, recording method, comment, range
LR03(AAA) alkaline dry cell battery × 4
Approximately one year with 1 minute recording interval (in power
saving mode)
Approx. 58 (W) × 87 (H) × 30 (D) mm; 130g
9662 LUX SENSOR (cord length: 2m) × 1
* The 3911-20 (sold separately) and a connecting cable are both required in order to analyze
measurement data on a PC. See page 2 for details.
● Broad measurement range from 2000 to 200,000 lux
● Light and compact illumination logger with a large memory
● In addition to measuring illumination on site, can also download collected data to a PC for processing
7
For Recording Temperature During Transport
Compact Logger That Revolutionizes Temperature Control Systems
■ Basic Specifications of Main Unit
3650 TEMPERATURE LOGGER
: Temperature -40.0ºC to 85.0ºC (resolution: 0.5ºC), 1 channel
Measuring range
Recording capacity : 2,048 data elements (approximately 1 year with the maximum recording interval
of 4 hours and 15 minutes)
Recording interval : 1 minute to 255 minutes
Recorded informa- : Temperature history, temperature distribution, temperature warning history, simple notes:
tion
: Data collected in DATA READER is transferred to a PC through an RS-232C interface
Interface
Power supply : Internal battery; can measure temperature at least 500,000 times
: Approximately 4 years of continuous use with a measuring interval of 5 minutes
Battery life
(Varies according to measuring conditions; battery cannot be replaced.)
Dimensions and mass : Approximately 17.35 mm in diameter × 5.89 mm thick; 3.3g
: Special antibacterial holder
Accessories
Actual size
■ Basic Specifications of the 3920-01 DATA READER
17.35 mm
Button-type logger with built-in
temperature sensor
Special mount provided
Compact button-sized temperature logger
● Sensor, memory and power supply are all encased in a unit with a diameter
of a mere 17.35mm
● Can easily control temperatures in HACCP processes over a range from 40ºC to 85.0ºC
● Can record data for up to 2048 temperature readings
● Temperature data can be collected in a special reader (sold separately), and
can then be used to create graphs or for other processing on a PC
● Easy to check for temperature abnormalities in distribution chain for foodstuffs by using a checker (sold separately) at the receiving destination
● Can be easily installed in a special (antibacterial) mount
Communication method
Analysis software
Functions
Data storage format
Accessories
:
:
:
:
:
3920-01
RS-232C (D-sub 9 pin male)
For PC running Windows 95/98
Measurement start setting/analysis, 16-channel graph display, etc.
Proprietary format, or CSV format
Dedicated analysis software (one CD-R)
■ Basic Specifications of the 3921 DATA CHECKER
: Place on the 3650; if the stored temperature values are normal, a green
LED lights; if any of the stored temperature values are abnormal, a red
LED lights.
* The threshold levels for abnormal values must be
set in the main unit beforehand.
Power supply : Button-type lithium battery (CR-2032) × 1
Dimensions and mass : Approx. 40 (W) × 58 (H) × 19 (D) mm; 43 g
Functions
* The 3920-01 (sold separately) is required in order to analyze measurement
3921
data on a PC.
For Multichannel Recording of Temperature, Humidity, and Voltage
Multi-channel Logger with PC Network Connectivity, Fully Isolated Channels
8420-01, 8421-01 MEMORY HiLOGGER
* The optional 8992 PRINTER or the 8993 DIGITAL I/O UNIT can be mounted in this unit.
LAN-compatible Data Logger That Displays Recorded Trends
● Two types with 8 channels (8420-01) or 16 channels (8421-01) of insulated analog inputs
● 5.7-inch color STN LCD and LAN functions (10BASE-T connector) included as standard features
● Simultaneous input of voltage, thermocouple, temperature resistor and humidity sensor;
count is maintained of pulse input; fluctuations in RPMs can also be measured simultaneously
● With optional 8993 unit, 16 digital input channels and 16 alarm output channels are also
available
● Through a LAN connection, a maximum of 256 channels of real-time data can be collected in
a PC (using the optional 9334 unit)
■ Basic Specifications
Number of
channels
: 8420; 8 analog channels (scanning method) + 4 pulse channels (inputs
and outputs and each channel are insulated from each other)
8421; 16 analog channels (scanning method) + 4 pulse channels (inputs
and outputs and each channel are insulated from each other)
Input
: [Voltage] 100 mV to 100 V f.s. 5 ranges (maximum resolution 5 µV,
100 mV f.s. range)
[Thermocouple] K, E, J, T, N, R, S, B, W (WRe5-25) (100ºC to 2000ºC
f.s. 3 ranges, maximum resolution 0.01ºC, 100ºC f.s. range)
[Temperature resistor] Pt100, JPt100 (100ºC to 2000ºC f.s. 3 ranges, maximum resolution 0.01ºC, 100ºC f.s. range)
[Humidity] 100%rh f.s. 1 range (0.1% resolution)
[Pulse] Cumulative counter, RPM fluctuation counter; DC to 5 kHz
Recording interval : 100 ms to 1 h (16 ranges)
Memory capacity : 4M word DRAM
External storage : Flash ATA card (up to 528MB, saving in real time possible)
Interface
: for LAN cable connection (10BASE-T connector), RS-232C
Recording section : Uses 8992 PRINTER UNIT (sold separately) to record on thermal paper roll
Power supply : 9447 BATTERY PACK (provides approximately 5 hours of continuous
operation), or 9418-10 AC ADAPTER
Dimensions and mass : Approx. 324 (W) × 170 (H) × 52 (D) mm; 1.4 kg (when 8992 and 8993
are not mounted)
Accessories
: 9418-10 AC ADAPTER (12 V - 2.5 A) × 1
8992 PRINTER UNIT
Removable
Permits printing on site
8993 DIGITAL I/O UNIT
Removable
OPTIONS
9234 RECORDING PAPER (18m, 10rolls/1set)
9334 LOGGER COMMUNICATOR
9447 BATTERY PACK (Ni-MH)
9648 CARRYING CASE
9653 HUMIDITY SENSOR
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@ hiokiusa.com
Europe Representative Office :
Meineckestrasse 48, 40474 Dusseldorf, Germany
TEL / FAX +49-211-4544153
E-mail: hioki-eu@ doitsu.de
Shanghai Representative Office :
1108 Union Building, 100 Yan An Road (East),
Shanghai, 200002, P.R.China
TEL +86-21-6328-9947/4938
FAX +86-21-6328-2064
E-mail: [email protected]
All information correct as of May. 30, 2002. All specifications are subject to change without notice.
3630SE3-25E-03P
Printed in Japan