FUJI EDS10-77E

PAPERLESS RECORDER
DATA SHEET
PHF
This is a paperless recorder that displays measured data
on the LCD in real time and stores data in CompactFlash.
The type of input such as thermocouple, resistance bulb,
D.C. voltage (current), etc. can be arbitrarily set to 6 channels at the maximum.
The data stored in CompactFlash can be regenerated on
the screen, and the use of supplied support software allows
the data to be regenerated on a PC screen.
The data recorded in ASCII format can be directly read in
a spreadsheet such as Excel, which facilitates the processing on a PC. (The data recorded in binary format cannot be
read in.)
FEATURES
1. Large capacity storage by CompactFlash
Measured data is periodically stored in CompactFlash.
Large storage capacity of up to 512MB allows display files
for approximately 4 years to be recorded continuously at
the display refresh cycle of 30 seconds (in the case of
ASCII data format, 6 channels).
2.Quick search and display of past data
Data stored in CompactFlash can be displayed in succession by scrolling the screen.
3.Various display capability
Depending on the object of measurement, the most
suitable display format can be selected from a variety of
formats including bar graph display, trend display, digital
display, etc.
4.PC support software supplied as standard
Loader software that enables easy display and change of
set data and data viewer software that regenerates the
data stored in CompactFlash are supplied as standard.
5.Compact size
160 (W) × 144 (H) × 185 (D) mm (Panel mounting), 1.5kg
compact size
6.6-point recording (Option)
12 types of thermocouples, 5 types of resistance bulbs
and DC voltage/current input can be recorded up to 6
points.
7. Screen saver function
If the non-operation exceeds the setting value of parameter, “LCD lights-out time”, recorder turns off the
backlight.
Setting range of this parameter is 0 to 60 minutes. If the
setting value is "0", this function doesn’t work, so the
backlight remains on during power on.
Screen saver function makes the life of backlight expand
and power consumption reduce.
8.Ethernet function (Option)
FTP, Web server, e-mail and MODBUS-TCP are available
using 10Base-T.
SPECIFICATIONS
Input system
Number of input points:
3 points or 6 points (Can be selected at
the time of purchase)
Input circuit:
Input mutual isolation
Resistance bulb measured current:
about 1 mA
Measuring cycles:
3 or 6 points….100ms cycles
Recording cycle:1 sec to 12 hours
Input types:
Thermocouple, resistance bulb, dc voltage, and dc current (Shunt resistors are
fitted in input terminals).
Note) Order a shunt resistor (t ype:
PHZP0101) separately.
Measuring range
Input types
Reference range
400.0 to 1760.0°C
Thermocouple B
0.0
to 1760.0°C
R
0.0
to 1760.0°C
S
–200.0 to 1370.0°C
K
–200.0 to 800.0°C
E
–200.0 to 1100.0°C
J
–200.0 to 400.0°C
T
0.0
to 1300.0°C
N
0.0
to 1760.0°C
W
–200.0 to 900.0°C
L
–200.0 to 400.0°C
U
0.0
to 1300.0°C
PN
–200.0 to 600.0°C
Resistance bulb JPt100
–200.0 to 600.0°C
Pt100
–60.0 to 180.0°C
Ni100
–200.0 to 600.0°C
Pt50
–50.0 to 200.0°C
Cu50
0.00
to 50.00mV
DC voltage
50mV
0.0
to 500.0mV
500mV
1.000 to 5.000V
1-5V
0.000 to 5.000V
0-5V
Note) B, R, S, K, E, J, T : JIS C 1602, DIN IEC 584-1
N : NICOSIL-NISIL (IEC 584)
W : 5%Re-26%Re · W (Hoskins Mfg. Co. USA)
L : Fe-Cu · Ni (DIN 43710)
U : Cu-Cu · Ni (DIN 43710)
PN: Platinum
JPt100 : JIS C 1604-1989 (Old JIS Pt 100)
Pt100, Pt50 : JIS C 1604, DIN IEC 751
EDS10-77e
Date
Dec. 5, 2011
PHF
Selection of input types:
By key operation on the front panel. Note
that the same input type (thermocouple,
resistance bulb, voltage) should be selected for channel 4 and 5. Refer to ”Setting method of input types” for details.
Burn-out function:
Equipped in thermocouple and resistance
bulb inputs as standard, and overswings
the recording to 100% side.
Thermocouple burn-out current:
approx. 0.2 µA
Input filter function:
Settable for each channel (primary delay
filter) Time constants are settable in the
range from 0 to 900 sec.
Scaling function:Possible by dc voltage (current) input
Scaling range:-32767 to 32767
Decimal position: settable at any point
Unit symbol: settable up to 7 digits and
125+12 types
Subtraction function:
Subtraction between each channel is allowed.
Square rooter function:
Square rooter can be performed against
the input value per each channel.
Indication system
5.7” STN color LCD (320 x 240 dots)
with backlight
Note)The LCD may have some pixels that do not stay
on or off.
Due to the characteristics of liquid crystal, the
brightness may not be uniform, which is not a
failure.
Color of indication:
14 colors
Applicable language:
English, French, German, Italian (switchable)
Life of backlight:50,000 hours (20°C)
(the complete indicator unit should be
replaced when replacing backlight).
Trend display: Direction: vertical and horizontal
Number of channels: 6 channels or 4 channels or 3 channels for the screen (Input:
6 points at the maximum).
Display refreshment cycles:
select from 1 second to 12 hours
No numerical value display. Scale display/
no-display can be selected.
Bar graph display:
Direction: vertical
Number of channels: 6 channels or 4 channels or 3 channels for the screen (Input:
6 points at the maximum).
Display refreshment cycles: 1 second.
Digital display: Number of channels: 6 channels or 4 channels or for the screen (Input: 6 points at
the maximum).
Display refreshment cycles: 1 second.
Event summary display:
Alarm summary and message summary
can be displayed.
Indicator:
2
Ethernet log display:
E-mail sending, FTP server log in/off and
MODBUS TCP/IP communication start/
stop can be displayed.
Parameter display/set:
Already-set Data Display and Set Change
Display screen
TAG indication: Number. of characters to be displayed:
Up to 8 characters at 6 channels on one
screen or up to 16 (= 8 × 2) characters
at 4 channels on one screen.
Characters to be displayed:
Alphanumerical characters
Tag, unit and channel number display:
It depends on the screen. See below
table.
Screen
Trend
Bar graph
Number of channel
on one screen
4 or less
more than 5
4 or less
more than 5
Digital
x: only 1 item can be displayed.
+: only 2 items can be displayed.
TAG1
+
x
+
x
Items
TAG2
unit ch number
+
+
+
x
x
+
+
+
x
x
all items are displaed
Historical trend display:
The past data can be displayed from the
Compact Flash or internal memory. The
past data file can be read and displayed
with scroll display function or jump the
cursor to the position which you entered
date and time. Scale display/no-display
can be selected.
Number of screen groups:
1 group (Up to 6 channels per 1 group
can be registered.)
Keyboard
No. of Keys:
Function:
8
Use to select various screens and set
various parameters.
Recording function
External memory media:
Compact Flash card
(Format as FAT32, FAT16 or FAT, or recorder can’t read and write.)
Recording capacity:
2GB maximum (compact flash). Limiting
the recording file to 64MB is recommended (for 112 hours if display refresh
cycle is 1 second. See Table 1 (p. 6).) (When the size of the recording file comes
to be 256MB or more, a new file is created
automatically and recording is maintained.)
* Please change the compact flash every
six month to prevent the data losing.
Recording method:
Turning ON the REC key allows measured
data to be written at fixed cycles.
Recorded as a new file whenever the
recording starts
Data save cycles:
Linked to the display refreshment cycles
on the “Trend display” screen. However,
they are automatically set to about 1
minute if the refreshment cycles are set
to less than 1 minute.
Trend data:
Average, instance or min. and max. measured values out of measured data that
are sampled at the measuring cycles are
saved.
Event data:
Saves alarm data and power ON data
when the power turns off and on during
recording.
Storage capacity:
Approximately 4 years when the display
refresh cycle is 30 seconds (in the case of
6-channel recording in ASCII data format,
and 512MB Compact Flash is used). Refer
to Table 1.
Memory usage: Indicates the memory which has already
used on the screen. When all the memory
is used up, you can stop recording or
delete the oldest recording file to save
the newest data.
Compact flash card form:PHZP2801
(CF card)
(If a card other than the above is used,
no operation assurance is ensured.
Meanwhile, as for other CF cards for
which operation check will have been
completed, the results will be posted on
our company's homepage sequentially.
Please refer to this website.)
Data format:
Either of ASCII or binary format can be
selected. (Switching cannot be made while
the recording is in progress. In the case
of ASCII format, the data can be directly
read on Excel, etc.)
Note: The data recorded in binary format
cannot be read directly.
Approximately 118 bytes per sample (for
6-channel input in ASCII format) or approximately 28 bytes (for 6-channel input
in binary format)
Alarm function
No. of settings: Up to 4 alarms for each channel are settable.
Type of alarm: High/Low limits
Indication:
Status (alarm types) is displayed on digital
display unit when an alarm occurs.
History display on alarm summary (Alarm
start/cancel time and alarm types)
Hysteresis:
Set within the recording range of 0 to 100%
(it is effective only in case of high/low limit
alarm)
Relay output:
Number of points; 10
Alarm latch function:
Keeps alarm indication and alarm output
turning on after alarm reset.
ON/OFF operation is performed according
to key setting.
Power supply
Rated power voltage:
100 to 240V AC
Range of operating voltage:
90 to 264V AC
Supply frequency:
50/60Hz (both employable)
Power consumption
100V AC
About 32VA
240V AC
About 42VA
Structure
Mounting method:
Panel-mounted (vertical panel)
Thickness of panel:
2 to 26 mm
Materials:
PC-ABS for case and bezel
Color:
Black
External dimensions:
Panel-mounted:160 (W) × 144 (H) × 185
(D) mm
Mass:
Approx. 1.5 kg (no option)
External terminal board:
Screw terminals (M3 thread)
RJ45 : Ethernet terminal (option)
Normal operating condition
Power voltage: 90 to 264V AC
Supply frequency:
50/60 Hz ±2% (both employable)
Ambient temperature: Panel-mounted
0 to 50°C
(without Ethernet option*1)
0 to 40°C
(with Ethernet option*2)
Note)In case of 30°C or more of ambient temperature,
this display might be fogged little bit (This is not
out of order).
Ambient humidity:
20 to 80%RH
Vibration:
10 to 60Hz 0.2m/s2 or less
Shock:
None
Magnetic field: 400 A/m or less
Signal source resistance:
Thermocouple input …. 1kΩ or less
Resistance bulb input… 10Ω/wire or less
(resistance of each wire of 3-wire system
should be balanced).
Voltage input… 0.1% or less of input
resistance
Mounting posture:
Forward tilt 0°, backward tilt within 30°,
horizontal 0°
Warm-up time: One hour or more after power ON
Environmental protection:
IEC IP50 (Front)/20 (Terminal) Installation category:
II
Pollution degree:2
Operating altitude:
2000m max.
*1: In case of the 12th digit of ordering code is ”Y”.
*2: In case of the 12th digit of ordering code is “E”.
3
PHF
Safety and EMC standard
Safety standard:Based on IEC61010-1
EMC standard: Based on EN61326
Transportation/storage conditions
Temperature:
Humidity:
Vibration:
Shock:
–10 to +60°C
5 to 90%RH, no condensation
10 to 60Hz, 2.45m/s2 or lower
294m/s2 or lower (packed state)
Reference standard
Accuracy/resolution:
Measuring conditions (23±2°C, 65±10%
RH, power voltage, frequency fluctuation
within ±1%, no external noise, warm-up
time of 1 hour or more, vertical mounting,
standard values of signal source resistance
and wiring resistance... within 1% )
Input types
Thermocouple B
R
S
K
E
J
T
N
W
L
U
PN
Digital indication
accuracy
Note 1
Digital indication
resolution
± (0.15%+1 digit)
±(0.3%+1 digit)
for the range shown below
Thermocouple B :
400 to 600°C
Thermocouples R and S :
0 to 300C
Thermocouples K, E, J, T,
L and U : -200 to -100°C
0.1°C
Resistance
bulb
JPt100
Pt100 ± (0.15%+1 digit)
Pt50
0.1°C
Ni100
Cu50 ± (0.5%+1 digit)
DC voltage 50mV
10µV
500mV
100µV
±
(0.15%+1
digit)
1-5V
1mV
0-5V
1mV
Note 1) Digital indication accuracy is a percentage (%) of the value in the
measuring range on page 1.
Note 2) No error of reference contact compensation of thermocouple is
included.
Error of reference contact compensation:
K, E, J, T, N, L, U, PN: ­±0.5°C
R, S, B, W: ±1.0°C
(when measured at 0°C or more)
Max. input voltage:
Thermocouple, resistance bulb, dc voltage: ±10V DC (continuous)
Input resistance:Thermocouple, DC voltage:
About 1MΩ
Withstand voltage:
Power terminal – ground: 2000V AC, 1 min
Input terminal – ground: 500V AC, 1 min
Alarm terminal – ground: 2000V AC, 1 min
Alarm terminal – alarm terminal:
750V AC, 1 min
Communication terminal – ground:
500V AC, 1 min
Effect on operation
Effect of power supply fluctuation conditions:
For the fluctuation in the range from 90
to 264V AC (frequency: 50/60Hz)
Reading change (100V AC base): ±(0.2%+1
digit) or lower.
For the fluctuation in the range from 47
to 63Hz (power voltage: 100V AC)
Reading change (50Hz base): ±(0.2%+1
digit) or lower.
Effect of input signal resistance:
Thermocouple input: (0.5µV/Ω)+1 digit
or less
DC voltage: Fluctuation for resistance
value equivalent to 0.1% of the input
resistance: ±(0.2%+1 digit) or lower.
Resistance bulb (for wiring resistance of
10W for 1 line (the same for 3 lines))
Reading change: ±(0.2%+1 digit) or lower.
Effect of ambient temperature:
Reading change: ±(0.3%+1 digit)/10°C
or lower.
Effect of Mounting position:
For the backward 30° slant
Reading change: ±(0.2%+1 digit) or lower.
Effect of vibration:
When sine wave of 10 to 60Hz with the
acceleration of 0.2m/s 2 is applied in each
direction for 2 hours.
Reading change: ±(0.2%+1 digit) or lower.
Effect of external noise:
Normal mode noise (50, 60Hz±0.1Hz)
…20dB or more
Common mode noise (50, 60Hz±0.1Hz)
…120dB or more
(Thermocouple input: minus terminalground)
(Resistance bulb input: b Line-ground)
Additional function (option)
Others
With calendar function (Christian era)
Accuracy: ±100ppm or less (monthly
error: about 4 minutes)
However, time error at ambient temperature 23±2°C and power ON/OFF is
not included.
Memory backup:Parameters are saved to the internal nonvolatile flash memory.
The clock and totalized data are backed
up with built-in lithium battery.
Insulation resistance:
100 MΩ or more (when measured between each terminal and ground by using
a 500V DC megger)
Clock:
4
■Alarm relay output/DI (11th digit of code symbols: “1”)
A card with 10-point relay output and 5-point DI input
can be mounted.
Terminal structure: M3 screw terminal
Alarm relay output:
Contact output (SPST:10 points),
Individual channel or common output (OR
output) allowed.
DO1: Contact capacity; 150V/3A AC,
30V/3A DC (resistive load)
DO2-10: Contact capacity; 240/3A AC,
30V/3A DC (resistive load)
DI input:
No-voltage contact input (5 points)
The following control is allowed by contact input.
(1) Recording start/stop
(2) LCD turns on
(3) E-mail sending
On pulse width: 200msec or longer
Off pulse width: 200msec or longer
Ethernet (Option)
The following can be performed through the Ethernet
function.
■HTTP server (Internet Explorer 6 is compatible) Note 1
Measurement display:
Digitally displays the measurement of
each channel of the recorder and alarm
occurrence status.
Event summary display:
Displays event summary including alarm
ON/OFF.
Main unit information display:
Displays memory use conditions and
information on the main unit such as the
battery end warning.
Integrated value display:
Digitally displays the integrated value of
each channel of the recorder.
■FTP server (Internet Explorer 6 is compatible.) Note 1
File download: Record files stored in compact flash (CF)
can be downloaded from the browser.
File delete:
Record files stored in CF can be deleted
from the browser.
Access authentication:
Authenticates access authority to FTP
server.
■SMTP (e-mail client)
Transmits e-mails to specified address
under the following conditions.
(1) When an alarm turns on or off
(2) When DI is set to ON or OFF
(3) When an error occurs to the main unit
(such as low battery or no memory
space)
(4) At specified intervals
■MODBUS TC/IP
Data read:
Settings can be read through MODBUS
TCP/IP communication.
Data write:
Settings can be written through MODBUS
TCP/IP communication.
Note1: Neither Netscape nor Mozilla Firefox
is available.
Required memory:
64MB or larger
Disk drive:
Windows 2000/XP-capable CD-ROM drive
Hard disk capacity:
Free capacity of 30MB or larger required
Printer: Windows 2000/XP-capable printer and
printer driver
Note) PC loader communication cable (type
PHZP1801) is separately required.
■Data viewer software
Major function: Regenerates the past trend record on the
PC from the data in the Compact Flash.
Provided with historical trend display and
event display functions.
O/S:
Windows 2000/XP
Required memory:
64MB or larger
Disk drive:
Windows 2000/XP-complaint CD-ROM
drive
Hard disk drive: Free capacity of 30MB or larger required
Printer:
Windows 2000/XP-capable printer and
printer driver
Standard functions
Function
Record range
voluntary setting
Input type setting
Skip function
Trend display
TAG name display
Screen name
display
Unit creation
Scaling function
Description
Recording range can be set by channel.
Input can be set by channel.
(Key operation on the front face)
The same input type is selected for channel 4 and 5.
See “SELECTING INPUT TYPE” on the last page.
Skips arbitrary channel display/recording.
Time display: Time is displayed at the top of the
trend display screen.
Alarm display: On occurrence of an alarm and the
restoration, alarm is displayed in the
alarm display field.
The Compact Flash usage is displayed at the top of
the bargraph.
By channel, Maximum of 8 characters.
Displays the screen name (maximum of 16
characters).
Industrial units can be arbitrarily created, Maximum
of 7 digits, 12 types.
Arbitrary scaling is allowed in the case of DC
voltage input. Decimal point position can also be
arbitrarily set in the range from -32767 to 32767.
PV shift
Shift the zero point and slant of the reading.
Input filter
Prevents sudden fluctuation of input for each
channel (primary delay filter).
Time constant: 0 to 900 seconds.
Burnout function
Displays the break of thermocouple/resistance bulb
input by scaling out to 100% side.
Historical
trend display
Regenerates and displays the data stored in the
compact flash by scrolling the screen or jump to
time when you entered.
Support software
The following software is provided as standard.
•PC/AT-compatible machine
•Operation on PC98-series machines by NEC is not guaranteed.
•Operation on self-made or shop-brand PCs is not guaranteed.
■Loader software for PC
Major function: Performs various parameter setting/change
of the main unit
O/S: Windows 2000/XP
5
PHF
Table 1. Recording capacity
Input point: 6
Data format: ASCII
The recording can be made for the period of time listed in
CompactFlash
size below. When the number
64MB
the
tables shown
of input points
isDisplay
3, therefreshment
period is approximately
1.6
those10listed
1 sec
10 sec
30times
sec of
1 min
min
incycle
the table.
capacitythe period is approximately 4 times as long
InRecordable
binary format,
159 hours 66 days 199 days 398 days 10.9 years
(about)
as those listed in the table.
CompactFlash size
Display refreshment cycle
Recordable capacity(about)
256MB
1 sec
10 sec
30 sec
1 min
ORDERING CODE
PHF
4 5 6 7 8 9 10 11 12 13
1 B 1 1 - E 1
V
Digit
Specifications
Note
4 <Number of input points>
5
3
6
6
11 <Alarm (relay) output/DI input>
Without
With
12 <Communication>
Without any communication
With Ethernet communication
0
1
Y
E
26 days 265 days 2.1 years 4.3 years
When Compact Flash is not used, up to 600K bytes of the
recording data and the event data can be stored in the
main unit. (In case of 6-channel in Max./Min. recording, approximately 21,000 data can be stored. For 5 hour at the
display refresh cycle of 1 second. The number of the save
data varies depending on the number of the event data.
STANDARD ACCESSORY
Quantity
Panel
mounting
Item
Recorder (PHF)
1
Panel mounting bracket
2
CD-ROM PC support software
instruction manual
1
Noise filter for the power supply
1
OPTIONAL ITEMS
Item
Code
Specification
Shunt resistor for DC
current input
PC loader communication
cable
CD-ROM with Instruction
manual and 2 support
software
PHZP0101
10Ω ±0.1%
PHZP1801
With USB-A and
USB miniB * 3m
PHZP2101
PC card adapter
PHZP0501
For Compact Flash
Compact Flash
PHZP2801-512
PHZP2801-01G
512MB
1GB
* Shape of this cable is shown below
USB (A) Plug – USB (Mini-B ) Plug
6
OUTLINE DIAGRAMS (Unit : mm)
Panel Mounting
In the case of 3, 6-point input
2
Ethernet terminal (Option)
Digital input & alarm output
(Option)
Source terminal
Input terminal
T 26
T
Mtg.bracket
L
REC
DISP
R-MODULE
Rcj
Fuji Electric
143.6
136.5
161
N
100∼240VAC
Panel
ENT
SEL
160
140
21.6
185
136.5
In the case of 3,6-point input
(Note) When placing the main unit on another instrument
or on the floor, allow a space of 100mm or more between
the unit and instrument or the floor.
Panel cutout
Mounting n unit
(160 × n-22) +20
n : Number of units to be mounted
137+1.5
0
137+1.5
0
Mounting one unit
+1.5
137 0
Do not use the water proof packing in case of mounting n unit
EXTERNAL CONNECTION DIAGRAMS (M3 screw)
In the case of 3, 6-point input
Alarm (relay) output /
digital Input terminal
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
DI1
DI2
DI3
DI4
DI5
DO1
DO2
DO3
DO4
DO5
DO6
DO7
DO8
DO9
DO10
Source terminal
Input terminal
Voltage
Thermocouple
41
42
43
CH4
RCJ
CH5
CH6
(Note1)
(Note2)
Resistance Thermocouple Voltage
bulb
Resistance
bulb
11
12
13
CH1
21
22
23
CH2
51
52
53
31
61
32
62
33
63
AC100 to 240V
50/60Hz
CH3
For current input, connect an optional shunt resistance to a voltage input terminal.
Do not use any input terminal which is not needed.
7
PHF
SELECTING INPUT TYPE
The input types of channel 4 and 5 is the same.
Channel 5 can only be allocated the input type that is the same as channel 4.
The following input types are available.
Input category
Details
Thermocouple, 50mV
Resistance bulb
500mV
5V
K, E, J, T, R, S, B, N, W, L, U, and PN thermocouples, 50mV
Pt100, JPt100, Ni100, Pt50, Cu50
500mV
1 to 5V, 0 to 5V
Note) Arbitrary input type can be selected for any channels other than channel 4 and 5 irrespective of the type
allocated to other channels.
Example of channel input type selection
Input type
Input category
Description
Thermocouple, 50mV
Channel 1
K thermocouple
Channel 2
1-5V
Channel 3
500mV
Channel 4
K thermocouple
Channel 5
50mV
Thermocouple,
50mV
Channel 6
Pt100
Resistance bulb
5V
500mV
The input type of the thermocouple and 50mV is the same.
Note 1) Windows 2000/XP, Excel and Internet Explorer are the registered trademarks of Microsoft Corporation in the U.S.A.
Note 2) Compact Flash is the registered trademark of Sandisk Corporation.
Note 3) Modbus® is the trade mark or registered trade mark of AEG Schneider Automation International.
Note 4) PC98 series are the trade mark or registered trade mark of NEC Corp.
Note 5) Netscape is the trade mark or registered trade mark of Netscape Communication Corp.
Note 6) Mozilla Firefox is the registered trade mark of Mozilla Foundation.
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan