J,E,K Type THERMOCOUPLES FOR GENERAL USE DATA SHEET FTA,B,D These thermocouples are convenient for measuring a wide range of temperature in general plants, and are availble in J, E and K type. They are insulated with a porcelain tube, and then housed in a metal protecting tube for heat-, pressure-and corrosion-proofing. SPECIFICATIONS Applicable standard: JIS C 1602—1995 Thermocouple element: Material; Symbol J Diameter of element; φ1, φ3.2 Normal limit and allowable overheat limit; Material Positive lead Negative lead Steel Alloy mainly made from copper and nickel E Alloy mainly made from nickel and chrome Alloy mainly made from copper and nickel K Alloy mainly made from nickel and chrome Alloy mainly made from nickel Note: Positive lead means the one connected to positive terminal of the instrument measuring thermoelectromotive force, and negative lead means the one connected to the opposite terminal. Class; Class Symbol of material 1 J, E, K 2 J, E, K 3 E, K Note: Class3 is applicable only to low temperatures below 40°C. Temperature accuracy; Symbol Class 1 K 2 3 1 E 2 3 1 J 2 Measured temperature (˚C) -40˚Cormore,lessthan+375˚C Tolerance ±2.5˚C 333˚Cormore,lessthan1200˚C ±0.0075•|t| -167˚Cormore,lessthan+40˚C Normal limit [°C] Allowable overheat limit [°C] J φ1 450 550 φ3.2 600 750 E φ1 500 550 φ3.2 700 800 K φ1 φ3.2 750 950 1000 1200 Note: The normal limit means the temperature limit up to which the thermocouple can be used continuously in air. Allowable overheat limit means the temperature limit up to which the thermocouple can be used for a short time, when necessary. Material of protecting tube: S ; 304 stainless steel R ; 310S stainless steel W; 316 stainless steel U ; 27Cr *T; 316L stainless steel J ; 20Cr-80Ni Mounting and protecting tube form: Protecting tube form ±1.5˚C 375˚Cormore,lessthan1000˚C ±0.004•|t| -40˚Cormore,lessthan+333˚C Diameter of Symbol element [mm] Mounting form Insertion type ±2.5˚C A D F — J — M Q -200˚Cormore,lessthan-167˚C ±0.015•|t| -40˚Cormore,lessthan+375˚C ±1.5˚C 375˚Cormore,lessthan800˚C ±0.004•|t| -40˚Cormore,lessthan+333˚C ±2.5˚C 333˚Cormore,lessthan900˚C ±0.0075•|t| -167˚Cormore,lessthan+40˚C ±2.5˚C Screw-in type Screw-in type -200˚Cormore,lessthan-167˚C ±0.015•|t| -40˚Cormore,lessthan+375˚C ±1.5˚C 375˚Cormore,lessthan750˚C ±0.004•|t| -40˚Cormore,lessthan+333˚C ±2.5˚C 333˚Cormore,lessthan750˚C ±0.0075•|t| Note:Tolerancemeansthemaximumallowablelimitofthe valuecalculatedbysubtractingthetemperatureof thetemperaturemeasuringjunctionfromthe temperatureobtainedbyconvertingthethermal electromotiveforcebythereferencethermal electromotiveforce. |t|isthevalueindicatedbytemperature(˚C)thatis Flange type Note: The thermocouple of diameter φ27 in protecting tube form is applicable to the case where material of protecting tube is J or U. In the case of U, and with mounting form D and Q, the double element is applicable. EDS2-1g Date Jan. 31, 2007 FTA,B,D CODE SYMBOLS Mounting method: Insertion type; Either movable flange (FTR), movable airtight flange (FTT), or movable airtight screw (FTS) is employed for mounting. Screw-in type; G3/4 (PF3/4) or R3/4 (PT3/4) external thread Flange type; JIS 10K-25A • JIS 20K-25A • JIS 30K-25A or JIS 63K-25A Composition: Mounting form Diameter of element Insertion type A φ1 Insertion type D φ3.2 Protecting tube material (Symbol) S • W • T* S • W • T* R • U • J Movable flange or movable airtight flange JIS 5K-25 • 10K-25 Movable airtight screw G1( PF1), G11/4 (PF11/4) S • W • T* Welded screw G3/4 (PF3/4), R3/4 (PT3/4) Flange type M φ1 S • W • T* Welded flange JIS 10K-25 JIS 20K-25 JIS 30K-25 JIS 63K-25 φ3.2 S • W • T* R • U • J 3 - F T B 3 - F TD 3 - 9 10 11 12 Description Kind of thermocouple Jthermocouple Kthermocouple Ethermocouple Element Singleelement Doubleelement Singleelement Doubleelement Singleelement Doubleelement 1 2 3 6 7 8 Movable flange or movable airtight flange JIS 5K-25 • 10K-25 Movable airtight screw G3/4(PF3/4) φ1 F T A F T A F T B F TD Mounting method Screw-in type F•J Flange type Q 1 2 3 4 5 6 7 8 A D F J M Q Mounting method *(1) Insertiontype Insertiontype Screw-intype(Forlowtemperature) Screw-intype(Forhightemperature) Flangetype Flangetype S W * T R U J Protecting tube material *(2) 304stainlesssteel 316stainlesssteel 316Lstainlesssteel 310Sstainlesssteel 27Cr 20Cr-80Ni Mounting dimension FlangeJIS10K-25A FlangeJIS20K-25A FlangeJIS30K-25(OnlyMtype) FlangeJIS63K-25(OnlyMtype) MountingscrewG3/4(PF3/4) MountingscrewR3/4(PT3/4) Insertiontype 1 2 3 4 7 8 0 Welded flange JIS 10K-25 JIS 20K-25 Note: (1) On the J, E thermocouple of mounting method D, Q, protecting tube material is S, W, T. (2) On the thermocouple of mounting method D and protecting tube material J, movable airtight screw is PF1/4. Insertion length (Accordingtotableofinsertionlength) Fillininsertionlengthincmunit 500mm050 Ex.: 1000mm100 015 Insertion length: See other table (page 5) Working conditions: See other table (pages 3 and 4) Terminal box: Material; Diecast aluminum alloy Conduit connection; G1/2 (PF1/2) Coating; Baked melamine resin Finish color; Silver Standard thermoelectromotive force table: See other table (page 5) Scope of delivery: Thermocouple 395 Structure/Explosionproofing Standardtype Flameprooftype(d2G4) Intrinsicallysafeexplosionproofing (i3nG5)forconnectionwithzener barrier Asterisked(*)items:Non-standard A C F (1) Combination of mounting form and protecting tube form (2) Combination of protecting tube material and mounting method Protecting tube form Method Material Mounting form Insertion type A D F — Screw-in type Screw-in type J — M Q Flange type (0.75class) (0.75class) (1.5class) FTD,FTB (1.5class) only (0.4class) (0.4class) Note:The thermocouple of diameter φ27 in protecting tube form is applicable to the case where material of protecting tube is J or U. In the case of U, and with mounting form D and Q, the double element is applicable. A D F J M Q S W T R U J Notes: (1) mark; Manufacturing possible (2) mark; FTA, FTD are applicalbe only to this material (3) mark; Applicalbe to fexplo-sionprooftype Working conditions Working temperature working pressure Application Mass{weight} Outline diagrams 1 S 2 FTA A W 7 T 8 Working temp,; 0 to 450 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm2} Used for corrosive gas or liquid, and when the insertion length is desired to be adjustable 0.5 to 1.4 kg Fig. 1 1 S 2 FTA D W 7 T 8 Working temp,; 0 to 600 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm2} Used for oxidizing atmosphere (gas), and when the insertion length is desired to be adjustable 1 to 6 kg Fig. 2 Used for corrosive gas or liquid, J type is used for comparatively high temp. 0.7 to 1.6 kg Fig. 3 Fig. 4 Model ( 1 S 2 F W 7 J T 8 ) ) 5 4 MPa FTA ( 0 0 100 450°C 1 2 3 4 Working temp,; 0 to 450 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 JIS 30K-25 • JIS 63K-25 Material; 304, 316, 316L stainless steel Used for corrosive gas or liquid, and when airtightness is especially required 1.4 to 2.3 kg (JIS 10K) 1.4 to 2.3 kg (JIS 20K) 2 to 2.9 kg (JIS 30K) 2.9 to 3.8 kg (JIS 63K) Fig. Fig. Fig. Fig. 1 S 1 2 FTA Q W 2 7 T 8 Working temp,; 0 to +600 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 Material; 304, 316, 316L stainless steel Used for oxidizing atmosphere (gas) 2 to 7 kg Fig. 9 Fig. 10 1 2 S 3 A FTD W 6 T 7 8 Working temp,; ミ200 to +500 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm2} Used for corrosive gas or liquid, and when the insertion length is desired to be adjustable 0.5 to 1.4 kg Fig. 1 1 2 S 3 D W FTD 6 T 7 8 Working temp,; ミ200 to +700 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm2} Used for oxidizing atmosphere (gas), and when the insertion length is desired to be adjustable 1 to 6 kg Fig. 2 Used for corrosive gas or liquid, J type is used for comparatively high temp. 0.7 to 1.6 kg Fig. 3 Fig. 4 Working temp,; ミ200 to +500 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 JIS 30K-25 • JIS 63K-25 Material; 304, 316, 316L stainless steel Used for corrosive gas or liquid, and when airtightness is especially required 1.4 to 2.3 kg (JIS 10K) 1.4 to 2.3 kg (JIS 20K) 2 to 2.9 kg (JIS 30K) 2.9 to 3.8 kg (JIS 63K) Fig. Fig. Fig. Fig. Working temp,; ミ200 to +700 。C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 Material; 304 stainless steel Used for oxidizing atmosphere (gas), and when airtightness is FTA. especially required 2 to 7 kg Fig. 9 Fig. 10 FTA 1 S 2 M W 7 T 8 ( ( 1 2 S 3 F FTD W 6 J T 7 8 1 2 S 3 FTD M W 6 T 7 8 5 MPa 4 0 -200 1 2 3 4 1 2 S 3 FTD Q W 1 6 2 T 7 8 100 ) ) 5 6 7 8 500°C FJ 5 6 7 8 FTA,B,D Working temperature, working pressure Application Mass{weight} Outline diagrams 1 2 S 3 FTB A W 6 T 7 8 Working temp.; –200 to +750 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm2} Used for corrosive gas or liquid, and when the insertion length is desired to be adjustable 0.5 to 1.4 kg Fig. 1 1 2 S 3 FTB D W 6 T 7 8 Working temp,; –200 to +900 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm2} Used for oxidizing atmosphere (gas), and when the insertion length is desired to be adjustable 1 to 6 kg Fig. 2 1 2 FTB D U 7 8 Working temp,; 0 to 1000 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm 2} 1 2 FTB D R 7 8 Working temp,; 0 to 1000 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm 2} 1 2 D J 7 8 Working temp,; 0 to 1000 °C Working pressure; Atmospheric pressure When both movable airtight screw and flange are used, working pressure can be up to 0.1MPa {1.02kgf/cm 2} J type is used for comparatively high temp. Used for corrosive gas or liquid, 0.7 to 1.6 kg Fig. 3 Fig. 4 FTB 1 2 S 3 F FTB W 6 J T 7 8 ( ( ( ( ( 5 4 MPa Model ) ) ) ) ) Used for high temp. gas including sulfur gas and carbon gas Used for high temp. gas exclud- ing sulfur gas Used when mechanical strength is required and for high temp. gas excluding sulfur gas Used for corrosive gas or liquid, 0.5 0 -200 100 500 750°C 1 2 1 S 3 2 FTB M W 6 3 T 7 4 8 Working temp,; –200 to +750 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 JIS 30K-25 • JIS 63K-25 Material; 304, 316, 316L stainless steel and when airtightness is especially required Used for oxidizing atmosphere 1.4 to 2.3 kg (JIS 10K) 1.4 to 2.3 kg (JIS 20K) 2 to 2.9 kg (JIS 30K) 2.9 to 3.8 kg (JIS 63K) Fig. Fig. Fig. Fig. 1 2 S 1 3 FTB Q W 2 6 T 7 8 Working temp,; –200 to +900 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 Material; 304 stainless steel (gas) Used for high temp. gas includ- 2 to 7 kg Fig. 9 Fig. 10 1 2 1 FTB Q U 7 2 8 Working temp,; 0 to 1000 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 Material; 304 stainless steel ing sulfur gas and carbon gas Used for high temp. gas excluding sulfur gas 1 2 1 FTB Q R 7 2 8 Working temp,; 0 to 1000 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 Material; 304 stainless steel Used when mechanical strength is required and for high temp. 1 2 1 FTB Q J 7 2 8 Working temp,; 0 to 1000 °C Working pressure; Below flange (JIS) nominal pressure JIS 10K-25 • JIS 20K-25 Material; 304 stainless steel gas excluding sulfur gas FTB7. FJ 5 6 7 8 Insertion length [Dimension "L" of "Outline Diagrams"] Insertion length "L" Mounting method A D F Insertion length 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 Insertion length "L" J M 150 200 250 300 350 400 450 500 550 600 650 700 750 Q 150 200 250 300 350 400 450 500 550 600 650 700 750 800 200 250 300 350 400 450 500 550 600 650 700 750 800 Mounting A method D 950 1050 1150 1250 1350 1450 F 950 1050 1150 1250 1350 1450 1550 1650 1750 1850 1950 2450 2950 3450 3950 Insertion length J 900 1000 1100 1200 1300 1400 M 800 900 1000 1100 1200 1300 Q 850 950 1050 1150 1250 1350 Standard thermoelectromotive force of J thermocouple (Unit: µV) 850 950 1050 1150 1250 1350 1450 1550 1650 1750 1850 2350 2850 3350 3850 JIS C 1602-1995 Temp. –100 [°C] 0 Temp. [°C] 0 100 200 300 0 –4633 404 0 501 0 0 507 5269 10779 545 555 16327 554 21848 27393 552 560 33102 39132 587 623 45494 647 –10 –5037 389 –501 494 10 507 512 5814 546 11334 555 16881 553 22400 27953 552 563 33689 39755 590 627 46141 645 10 –20 –5426 375 −995 487 20 1019 518 6360 549 11889 556 17434 552 22952 28516 552 564 34279 40382 594 630 46786 645 20 –30 –5801 358 −1482 479 30 1537 522 6909 12445 550 555 17986 552 23504 29080 553 567 34873 597 41012 633 47431 643 30 –40 –6159 341 −1961 470 40 2059 526 7459 13000 551 555 18538 552 24057 29647 553 569 35470 41645 601 636 48074 641 40 –50 –6500 321 –2431 462 50 2585 531 8010 13555 552 555 19090 552 24610 30216 554 572 36071 42281 604 638 48715 638 50 –60 –6821 302 –2893 451 60 3116 534 8562 553 14110 555 19642 552 25164 30788 556 574 36675 42919 609 640 49353 636 60 –70 –7123 280 –3344 442 70 3650 537 9115 14665 554 554 20194 551 25720 31362 556 577 37284 43559 612 644 49989 633 70 –80 –7403 256 –3786 429 80 4187 539 9669 555 15219 554 20745 552 26276 31939 558 580 37896 44203 616 645 50622 629 80 –90 –7659 231 –4215 418 90 4726 543 10224 15773 555 554 21297 551 26834 32519 559 583 38512 44848 620 646 51251 626 90 –100 –7890 –4633 100 5269 10779 16327 21848 27393 33102 39132 45494 51877 100 400 500 600 700 800 Temp. [°C] 0 Remarks: Thermoelectromotive force values shown in the above table are when temperature of reference junction is 0 °C Standard thermoelectromotive force of E thermocouple (Unit: µV) Temp. –100 [°C] 0 Temp. [°C] 0 0 −5237 444 0 582 0 0 591 –10 −5681 426 −582 570 10 –20 −6107 −1152 409 557 –30 Temp. [°C] 200 300 400 500 600 700 800 900 6319 679 13421 743 21036 781 28946 801 37005 810 45093 807 53112 796 61017 784 68787 767 591 601 6998 687 14164 748 21817 783 29747 803 37815 809 45900 53908 805 795 61801 782 69554 765 10 20 1192 609 7685 694 14912 752 22600 786 30550 804 38624 810 46705 54703 804 794 62583 781 70319 763 20 −6516 −1709 391 546 30 1801 619 8379 702 15664 756 23386 788 31354 805 39434 809 47509 55497 804 792 63364 780 71082 762 30 –40 −6907 −2255 372 532 40 2420 628 9081 708 16420 761 24174 790 32159 806 40243 810 48313 56289 803 791 64144 778 71844 759 40 –50 −7279 −2787 353 519 50 3048 637 9789 714 17181 764 24964 793 32965 807 41053 809 49116 57080 801 790 64922 776 72603 757 50 –60 −7632 −3306 331 505 60 3685 645 10503 721 17945 768 25757 795 33772 807 41862 809 49917 57870 801 789 65698 775 73360 755 60 –70 −7963 −3811 310 491 70 4330 655 11224 727 18713 771 26552 796 34579 808 42671 808 50718 58659 799 787 66473 773 74115 754 70 –80 −8273 −4302 288 475 80 4985 663 11951 733 19484 775 27348 798 35387 809 43479 806 51517 59446 798 786 67246 771 74869 752 80 –90 −8561 −4777 264 460 90 5648 671 12684 737 20259 777 28146 800 36196 809 44285 808 52315 60232 797 785 68017 770 75621 752 90 –100 −8825 −5237 100 6319 13421 21036 28946 37005 45093 53112 68787 76373 100 100 JIS C 1602-1995 61017 0 Remarks: Thermoelectromotive force values shown in the above table are when temperature of reference junction is 0 °C FTA,B,D Standard thermoelectromotive force of K thermocouple (Unit: µV) JIS C 1602-1995 Temp. [°C] –100 0 Temp. [°C] –90 −3554 298 −3852 286 −4138 273 −4411 258 −4669 244 −4913 228 −5141 213 −5354 196 −5550 180 −5730 161 0 392 −392 386 −778 378 −1156 371 −1527 362 −1889 354 −2243 344 −2587 333 −2920 323 −3243 311 –100 −5891 0 –10 –20 –30 –40 –50 –60 –70 –80 0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 52410 349 52759 347 53106 345 53451 344 53795 343 54138 341 54479 340 54819 90 0 397 397 401 798 405 1203 409 1612 411 2023 413 2436 415 2851 416 3267 415 3682 414 4096 413 4509 411 4920 408 5328 407 5735 403 6138 402 6540 401 6941 399 7340 399 7739 399 8138 401 8539 401 8940 403 9343 404 9747 406 10153 408 10561 410 10971 411 11382 413 11795 414 12209 415 12624 416 13040 417 13457 417 13874 419 14293 420 14713 420 15133 421 15554 421 15975 422 16397 423 16820 423 17243 424 17667 424 18091 425 18516 425 18941 425 19366 426 19792 426 20218 426 20644 427 21071 426 21497 427 21924 426 22350 426 22776 427 23203 426 23629 426 24055 425 24480 425 24905 425 25330 425 25755 424 26179 423 26602 423 27025 422 27447 422 27869 420 28289 421 28710 419 29129 419 29548 417 29965 417 30382 416 30798 415 31213 415 31628 413 32041 412 32453 412 32865 410 33275 410 33685 408 34093 408 34501 407 34908 405 35313 405 35718 403 36121 403 36524 401 36925 401 37326 399 37725 399 38124 398 38522 396 38918 396 39314 394 39708 393 40101 393 40494 391 40885 391 41276 389 41665 388 42053 387 42440 386 42826 385 43211 384 43595 383 43978 381 44359 381 44740 379 45119 378 45497 376 45873 376 46249 374 46623 372 46995 372 47367 370 47737 368 48105 368 48473 365 48838 364 49202 363 49565 361 49926 360 50286 358 50644 356 51000 355 51355 353 51708 352 52060 350 −3554 100 4096 8138 12209 16397 20644 24905 29129 33275 37326 41276 45119 48838 52410 0 10 20 30 40 50 60 70 80 Temp. [°C] 10 20 30 40 50 60 70 80 90 100 Remarks: Thermoelectromotive force values shown in the above table are when the temperature of reference junction is 0 °C OUTLINE DIAGRAM [Unit:mm] (Refer to working conditions of page 3) Determine insertion length "L" by collating the 5th digit of code symbols with the table shown on page 5. For example, when the 5th digit of code symbol is A, select "L" from 150mm to 1450mm (twenty two varieties) by the table shown on page 5. Insertion type Fig. 1 1 2 A S A 3 A W 03 - ••• FT B F 6 D T 7 8 Fig. 2 1 S, R 2 A A W, J 3 D 03 - ••• F FT B T 6 D U 7 8 Note: φD is as follows: 0 Element Protecting tube material Single and double 304, 22 316, 316L stainless steel φD Single 27Cr 21.3 Double 27Cr 27 Single and double 20Cr-80Ni 27 Screw-in type Fig. 3 Fig. 4 Flange type 1 2 A S A 3 F W 73 - ••• FT B F 6 J D T 7 8 Fig. 5 1 2 A S A 3 M W 13 - ••• FT B F 6 D T 7 8 Fig. 6 1 2 A S A 3 M W 23 - ••• FT B F 6 D T 7 8 Fig. 7 1 2 A S A 3 M W 43 - ••• FT B F 6 D T 7 8 1 2 A S A 3 F W 83 - ••• FT B F 6 J D T 7 8 Note: Dimension "A" in case of screw-in type Dimension "A" For low temp. 50mm For high temp. 150mm FTA,B,D Terminal box for flameproof type 100 125 G1/2 (PF1/2) Fig. 8 1 2 A S A 3 M W 33 - ••• FT B F 6 D T 7 8 Movable airtight screw (type FTS) 2, 4, 6 in 4th digit of code Type Fig. 9 1 S, R 2 A A W, J 3 Q 13 - ••• F FT B T 6 D U 7 8 Protecting tube diameter of Fig. 9, 10 Protecting tube material Element φD φD1 304, 316, 316L stainless steel Single Double 22 29 27Cr Single Double 21.3 27 29 20Cr-80Ni Single Double 27 34 φd1 φd2 a b Protecting tube Approx. ℓ outer dia. (φD) Mass {weight} [kg] FTS6E 31 R1 1/4 58 (PT1 1/4) 30 100 30 1.2 FTS6D 28 R1 1/4 58 (PT1 1/4) 30 100 25 to 27 1.2 FTS4C 24 R1 (PT1) 50 20 80 21.3 to 22 0.8 FTS2A 13 R3/4 (PT3/4) 20 80 12 0.8 38 Note: Material is SS400 1, 3, 5 in 4th digit of code Type 1 2 A S, U A 3 Fig. 10 FT B 6 Q W,R 23 - ••• F D T J 7 8 Cap φd1 φd2 φa b c Protecting Mass Approx. ℓ tube outer {weight} dia. (φD) [kg] FTS5E 31 G1 1/4 69 (PF1 1/4) 58 30 100 30 1.2 FTS5D 28 G1 1/4 69 (PF1 1/4) 58 30 100 25 to 27 1.2 FTS3C 24 G1 (PF1) 59 50 20 80 21.3 to 22 0.8 FTS1A 13 G3/4 (PF3/4) 38 20 80 12 0.8 48 Note: Material is SS400 Movable airtight flange (type FTT) Movable flange (type FTR) JIS 10K-50 flange φd Protecting tube outer dia. (φD) 31 30 28 26.8 • 27 24 21.3 • 21.7 • 22 Material SS400 JIS 10K-25 flange φd Protecting tube outer dia. (φD) FTT2D 28 26.8 • 27 FTT2C 24 21.3 • 21.7 • 22 Type FTT2A 13 FC200 JIS 5K-25 flange φd Protecting tube outer dia. (φD) FTT1D 28 26.8 • 27 FTT1C 24 21.3 • 21.7 • 22 FTT1A 13 12 φd Protecting tube outer dia. (φD) 50 49 31 30 28 25 to 27 24 21.3 to 22 13 12 Material Material SS400 JIS 10K-25 flange Material 12 Type JIS 10K-50 flange Type φd Protecting tube outer dia. (φD) FTR2D 28 25 to 27 FTR2C 24 21.3 to 22 FTR2A 13 12 Material FC200 JIS 5K-25 flange Type φd Protecting tube outer dia. (φD) FTR1D 28 25 to 27 FTR1C 24 21.3 to 22 FTR1A 13 12 Material FC200 FC200 FTA,B,D CautiononSafety *Beforeusingthisproduct,besuretoreaditsinstructionmanualinadvance. Head Office GateCityOhsaki,EastTower,11-2,Osaki1-chome, Shinagawa-ku,Tokyo141-0032,Japan http://www.fesys.co.jp/eng Instrumentation Div. International Sales Dept. No.1,Fuji-machi,Hino-city,Tokyo,191-8502Japan Phone:81-42-585-6201,6202Fax:81-42-585-6187 http://www.fic-net.jp/eng Information in this catalog is subject to change without notice. PrintedinJapan