FUJI EDS5-87

FFI SYSTEM
OPTICAL PRESSURE TRANSMITTER
DATA SHEET
FBL···3
The Model FBL 3 Optical Pressure Transmitter measures
pressures of various fluids accurately, converts them into
optical digital signals and outputs them. This is an intelligent transmitter providing excellent performance and functions due to incorporation of electrostatic capacitance type
silicon sensor and microprocessor.
A fiber optical cable used for the signal transmission line
forms an optical field instrumentation system together with
an optical star coupler and a master station.
FEATURES
SPECIFICATIONS
Functional specifications
Fluids measured:
Liquid, gas or steam
Measuring range, operating pressure and maximum
allowable over-pressure:
Type
Span [kPa]
Operating
pressure
MiniMaxi[MPa]
mum
mum
value
value
FBL
01 -0.1 to 0.13
3.25
130
FBL
02 -0.1 to 0.5
12.5
500
FBL
03 -0.1 to 3
75
3000
FBL
04 -0.1 to 10
250
10000
FBL
05 -0.1 to 50
1250
50000
Range limits
[kPa]
Lower
range
limit
Determined by allowable operating
pressure limit of particular fill fluid
(See table below.)
1. Resistive to noise and lightning
Optical signal ensures a reliable signal transmission, because it is not affected by external noise and inductive
lightning. Use of a nonmetallic optical (fiber) cable prevents propagation of inductive lightning through the cable,
so a signal transmission immune to lightning can be realized.
2. Reliability due to redundant configuration
Host system can be duplicated by using two optical cable
trunk lines (between an optical star coupler and host
system. This enhances reliability of users’ systems.
3. Intrinsic safety type explosion-proof
Each equipment with a built-in battery can be constructed
so as to be an intrinsic safety type individually (intrinsic
safety type barrier unnecessary).
Max.
allowable
overUpper pressure
range [MPa]
limit
130
1
500
1.5
3000
9
10000
15
50000
75
EDS5-87
Date
May 20, 2004
FBL···3
Process temperature, Allowable pressure limit:
For details, refer to Fig. 1.
Fill-fluid
Code
Process
temperature
Allowable
pressure limit
Silicon oil
Y, G, N
-40 to +100˚C
2.7kPa abs
Fluorolube oil
W, A, D
-20 to +80˚C
Atmospheric
pressure
Silicon oil
R
-15 to +100˚C
2.7kPa abs
Self-diagnosis:
Displayed on indication unit (option) and
transmitted to master station.
Item
Host system Indication unit
Measuring range abnormal
Detecting unit failure
Amplifier abnormal
Battery voltage
Battery voltage low alarm
Remote control function:
See Table 1.
Output signal: Optical digital signal
Power supply: Built-in lithium battery (expected life about
4 years)
Optical cable:
Code set type, silica fiber ... core/clad diameter 100/140 µm
Optical connector:
FC connector
Transmission distance:
1.5 km max. (when transmission loss of
optical cable is 4 dB/km)
Damping:
Variable from 0.2 to 32 sec (time constant)
Zero elevation and suppression:
Possible within a range from -0.1 MPa
to maximum span.
Explosion-proof: Intrinsic safety type, JIS ib IIC T3
Ambient temperature:
–30 to +70°C
–10 to +60°C for intrinsic safety explosion-proof type
–20 to +70°C when provided with indicator
–10 to +60°C when filled with fluorolube
oil
Storage temperature:
–40 to +80°C
2
Performance specifications
±0.1% when measuring span is 1/10 or more
of maximum span
±(0.05 + 0.005 x maximum span/measuring
span)% when measuring span is less than
1/10 of maximum span
Note: Percent value with respect to measuring span (including
linearity, hysteresis and repeatability in standard 23°C status)
Accuracy rating
(Note)
Ambient temperature effect:
Zero shift: ± (0.1 + 0.05 × URL/x )%/ 28°C
Overall shift: ± (0.15 + 0.05 × URL/x )%/
28°C
Where; URL: Maximum span
x:
Measuring span
3 times as large as above when the 7th
digit (material) is other than V.
Maximum allowable pressure effect:
Zero shift at maximum span
± 0.2% to maximum allowable pressure
Inclination effect:
0.1 kPa/10° or, when 13th digit code (processing, fill fluid) is W, D or A, double the
above, at maximum span
Measurement period:
0.2 sec
Response time: Time constant 0.2 sec (at 23°C)
Dead time; About 0.2 sec
Physical specifications
Material:
Material
code
Process cover
V
J
H
M
T
B
L
U
SCS14
SCS14
SCS14
SCS14
SCS14
Hastelloy-C lining
Monel lining
Tantalum lining
Detecting unit
Seal diaphragm Other wetted parts
SUS316L
SUS316L · Gold-plated
Hastelloy-C
Monel
Tantalum
Hastelloy-C
Monel
Tantalum
SUS316
SUS316
Hastelloy-C
Monel
Except for
Tantalum
Hastelloy-C FBL 05
Monel
Tantalum
Environmental protection:
Meets JIS C0920, immersion-proof
(equivalent to IEC IP67 or NEMA 6/6P)
Process connection:
Rc1/4 or 1/4-18NPT (whichever selected
by code symbol)
Oval flange threads 7/16-20UNF
Optical cable connection:
G1/2 or 1/2 -14NPT (whichever selected
by code symbol)
Mounting method:
Mounted on the 50A (2B) pipe with Ubolt
Finish:
Epoxy-polyurethane double coat,
Color: silver (blue for amplifier case cover)
External dimensions:
See OUTLINE DIAGRAM.
Mass:
4.5 to 5 kg
Orientation of transmission unit:
Indicator unit turnable 90° upward/downward relative to detection unit.
Optional specifications
Indication unit: 5-digit LCD indication, % or real scale indication (as specified by code symbol)
Operating temperature range: –20 to
+70°C
Oxygen oil-proof processing:

Fluorolube filled.

Wetted parts degreased  Varies with

and cleaned.
 material.
Chlorine service: Fluorolube oil filled.
 Refer to
NACE specification:
 CODE
H2S-proof treatment in  SYMBOLS.

accordance with NACE 
specification

Table 1 Remote Control Function
(Items readable and setting from hand-held
communicator)
Item
Reading Setting
Maximum range
Description
Maximum measuring range
of equipment
Actual measuring range
Variable within 0.2 to 32 sec.
Indication in industrial value
Measuring range
Damping
Real scale
indication
Battery voltage
Error indication
Battery voltage of equipment
Errors of detection unit and
amplifier
Measured data
Zero and span adjustment
Measured value
Adjustment
Note: For operation of the “3” type transmitter (“3” at the 8th digit
of product code), a hand-held communicator is required to
have a version 1.6 or higher, but a communicator before
version 1.6 can be operated with memory data updated.
(Refer to the instruction manual of transmitter.)
Fill-fluid: Silicon oil (Code: Y,G,N)
Fill-fluid: Silicon oil (Code: R)
[kPa abs]
Operating pressure
101
Operating area
20
4
2.7
Non-operating area
–40 –15
60 85 100 [°C]
Process temperature
Fig. 1 Relation between process temperature and
operating pressure
3
FBL···3
SYSTEM BLOCK DIAGRAM
1:1 system
Host system
Master station
Control room
Field
1 to 5V DC
4 to 20mA DC
Optical fiber type
electric converter
(Type; PSN)
Star coupler
Transmitter
Transmitter
Field equipment
(max. 8 units)
Transmission distance: 1.5 km max.
(between master station and field equipment)
4
Transmission distance: 4 km max.
CODE SYMBOLS
1 2 3 4 5 6 7 8
FBL 0
3
9 10111213 1415
F
S
T
Description
Process cover (4th digit)
Process connection Cable lead-in port
Rc1/4
G1/2
1/4-18NPT
1/2-14NPT
Measuring span (6th digit)
3.25....... 130kPa
12.5...... 500kPa
75........ 3000kPa
250.....10000kPa
1250.. 50000kPa
1
2
3
4
5
Material (7th digit)
Process cover
SCS14
SCS14
SCS14
SCS14
SCS14
Hastelloy-C lining
Monel lining
Tantalum lining
V
J
H
M
T
B
L
U
Seal diaphragm
SUS316L
SUS316L · Gold-plated
Hastelloy-C
Monel
Tantalum
Hastelloy-C
Monel
Tantalum
Other wetted parts
SUS316
SUS316
Hastelloy-C
Monel
Not available
Tantalum
when 6th
Hastelloy-C
digit is “5”.
Monel
Tantalum
Indicator (9th digit)
Not provided
Digital, % indication
Digital, real scale indication
A
L
P
Explosion-proof (10th digit)
Non-explosion proof
Intrinsic safety, JIS
A
G
Side vent/drain and fixture (11th digit)
Side vent/drain
None
None
Yes
Yes
A
C
D
F
Fixture
None
Yes (stainless steel)
None
Yes (stainless steel)
Cannot be specified
when 7th digit is “B”,
“L” or “U”.
Treatment and Fill-fluid (13th digit)
D
Treatment
None
None
Degreasing
Oxygen oil-proof
processing
Chlorine service
N
NACE specification
R
None
Y
W
G
A
Fill-fluid
Silicon oil
Fluorolube oil
Silicon oil
Fluorolube oil (when 7th digit is “V”)
Fluorolube oil (unavailable when 7th digit is “V”,
“M” or “L”)
Silicon oil (unavailable when 6th digit is “5”, or 7th
digit is “T” or “U”)
Silicon oil (for vacuum)
O-ring material (for process cover)(14th digit)
Viton
Teflon
A
B
C
D
E
F
Bolt/nut (15th digit)
NACE bolt/nut (ASTM A193 B7M/A194 2HM)
Cannot be specified
NACE bolt/nut (ASTM A320 L7M/A194 2HM)
when 6th digit is “5”.
SUS304/SUS304.... (For general use)
SUS630/SUS304.... Specify when 6th digit is “5”.
5
FBL···3
OUTLINE DIAGRAM
(Unit:mm)
P
241
80
FBL S 0
T
108
70
V
J
H 3–
M
T
Mounting pipe
JIS 50A (2B)
A
C
U-bolt
(M8)
70
φ102
225
8
315
54
15
B
Indicator (option)
Gas vent, drain port
Optical cable lead-in port
A
FBL S 0
T
41.3
Pressure connection
B
L 3–
U
A
C
4–7/16-20UNF
Thread depth 13
(for oval flange)
318
41.3
φ102
27
37 27
47
View from arrow P
φ7
15 3
8
φ15
54
E
Details of “B”
2.5
A
D
φ18.5
Details of “A”
Pressure connection
40
2–7/16–20UNF
Thread depth 10
(for oval flange)
27
47
φ16
+0.2
Type
FBLS
FBLT
D
Rc1/4
1/4–18NPT
E
G1/2
1/2–14NPT
0
2 - 0.1
1
φ12 +0.1
φ4
Details of “A”
6
SCOPE OF DELIVERY
Instrument body and pipe fixture (as specified)
ITEM TO BE PREPARED SEPARATELY
Oval flange:
To be used as a flange of connecting pipe
port.
For details, refer to the DATA SHEET of
oval flange (EDS6-10).
ORDERING INFORMATION
1.
2.
3.
4.
Model type
Measuring range
Indication scale for real scale specification
Others
7
FBL···3
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan