FUJI EDS5-88

FFI SYSTEM
OPTICAL ABSOLUTE PRESSURE TRANSMITTER
DATA SHEET
FBK···3
The Model FBK 3 Optical Absolute Pressure Transmitter
measures absolute pressures of various fluids accurately,
converts them into optical digital signals and outputs them.
This is an intelligent transmitter providing excellent performance and functions due to incorporation of electrostatic
capacitance type silicon sensor and microprocessor.
A fiber optical cable used for the signal transmission line
forms an optical field instrumentation system together with
an optical star coupler and a master station.
FEATURES
1. Resistive to noise and lightning
Optical signal ensures a reliable signal transmission, because it is not affected by external noise and inductive
lightning. Use of a nonmetallic optical (fiber) cable prevents propagation of inductive lightning through the cable,
so a signal transmission immune to lightning can be realized.
2. Reliability due to redundant configuration
Host system can be duplicated by using two optical cable
trunk lines (between an optical star coupler and host
system). This enhances reliability of users’ systems.
3. Intrinsic safety type explosion-proof
Each equipment with a built-in battery can be constructed
so as to be an intrinsic safety type individually (intrinsic
safety type barrier unnecessary).
SPECIFICATIONS
Functional specifications
• Fluids measured: Liquid, gas or steam
• Measuring range, operating pressure and maximum
allowable over-pressure:
Type
Span
[kPa abs]
Operating
pressure
MiniMaxi[kPa abs]
mum
mum
value
value
Range limits
[kPa abs]
FBK 01
0 to 16
1.6
16
0
16
0.5
FBK 02
0 to 130
3.25
130
0
130
0.5
FBK 03
0 to 500
12.5
500
0
500
1.5
FBK 04
0 to 3000
75
3000
0
3000
9
Lower
range
limit
Max.
allowable
overUpper pressure
range [MPa abs]
limit
• Process temperature: –40 to +85°C
EDS5-88
Date
April 15, 2005
FBK···3
• Self-diagnosis: Displayed on indication unit (option) and
transmitted to master station.
Item
Host system Indication unit
Performance specifications
Accuracy rating
Measuring range abnormal
(Note)
±0.2% when measuring span is 1/10 or more
of maximum span
± 0.1+ 0.01
Detecting unit failure
Amplifier abnormal
• Remote control function:
See Table 1.
• Output signal: Optical digital signal
• Power supply: Built-in lithium battery (expected life about
4 years)
• Optical cable: Code set type, silica fiber ... core/clad diameter 100/140 µm
• Optical connector:
FC connector
• Transmission distance:
1.5 km max. (when transmission loss of
optical cable is 4 dB/km)
• Damping:
Variable from 0.2 to 32 sec (time constant)
• Zero elevation and suppression:
Possible within a range from 0 kPa abs
to maximum span.
• Explosion-proof:
Intrinsic safety type, JIS ib IIC T3
• Ambient temperature:
–30 to +70°C
–10 to +60°C for intrinsic safety explosion-proof type
–20 to +70°C when provided with indicator
• Storage temperature:
–40 to +80°C
%
when measuring span is less than
1/10 of maximum span
Battery voltage
Battery voltage low alarm
Maximum span
Measuring span
Note: Percent value with respect to measuring span (including
linearity, hysteresis and repeatability in standard 23°C status)
• Ambient temperature effect:
Zero shift: ± (0.125 + 0.2 × URL/x )%/ 28°C
Overall shift: ± (0.175 + 0.2 × URL/x )%/
28°C
Where; URL: Maximum span
x:
Measuring span
2 times as large as above when the 7th
digit (material) is other than V.
• Maximum allowable pressure effect:
Zero shift at maximum span
± 0.2% to maximum allowable pressure
• Inclination effect:
0.1 kPa/10°
• Measurement period:
0.2 sec
• Response time:Time constant 0.2 sec (at 23°C)
Dead time; About 0.2 sec
Physical specifications
Material:
Material
code
V
H
M
T
Process cover
SCS14
SCS14
SCS14
SCS14
Detecting unit
Seal diaphragm Other wetted parts
SUS316L
Hastelloy-C
Monel
Tantalum
SUS316
Hastelloy-C
Monel
Tantalum
• Environmental protection:
Meets JIS C0920, immersion-proof
(equivalent to IEC IP67 or NEMA 6/6P)
• Process connection:
Rc1/4 or 1/4-18NPT (whichever selected
by code symbol)
Oval flange threads 7/16-20UNF
• Optical cable connection:
G1/2 or 1/2 -14NPT (whichever selected
by code symbol)
• Mounting method:
Mounted on the 50A (2B) pipe with Ubolt
• Finish:
Epoxy-polyurethane double coat,
Color: silver (blue for amplifier case cover)
• External dimensions:
See OUTLINE DIAGRAM.
• Mass:
4.5 kg
• Orientation of transmission unit:
Indicator unit turnable 90° upward/downward relative to detection unit.
2
Optional specifications
• Indication unit: 5-digit LCD indication, % or real scale indication (as specified by code symbol)
Operating temperature range: –20 to
+70°C
• NACE specification:
H2S-proof treatment in
accordance with NACE
specification
Table 1 Remote Control Function
(Items readable and setting from hand-held
communicator)
Item
Maximum range
Measuring range
Damping
Real scale
indication
Battery voltage
Error indication
Measured value
Adjustment
Reading Setting
Description
Maximum measuring range
of equipment
Actual measuring range
Variable within 0.2 to 32 sec.
Indication in industrial value
Battery voltage of equipment
Errors of detection unit and
amplifier
Measured data
Zero and span adjustment
Note: For operation of the “3” type transmitter (“3” at the 8th digit
of product code), a hand-held communicator is required to
have a version 1.6 or higher, but a communicator before
version 1.6 can be operated with memory data updated.
(Refer to the instruction manual of transmitter.)
3
FBK···3
SYSTEM BLOCK DIAGRAM
Host system
Master station
Control room
Star coupler
Field
Transmitter
Field equipment
(max. 8 units)
Transmission distance: 1.5 km max.
(between master station and field equipment)
1:1 system
Control room
1 to 5V DC
4 to 20mA DC
Optical fiber type
electric converter
(Type; PSN)
Field
Transmitter
Transmission distance: 4 km max.
4
CODE SYMBOLS
1 2 3 4 5 6 7 8
F BK 0
3
9 10111213
F
1415
Description
Process cover (4th digit)
Process connection Cable lead-in port
Rc1/4
G1/2
1/4-18NPT
1/2-14NPT
S
T
Measuring span (6th digit)
1.6...........16 kPa abs
3.25.......130 kPa abs
12.5.......500 kPa abs
75........3000 kPa abs
1
2
3
4
Material (7th digit)
Process cover
SCS14
SCS14
SCS14
SCS14
V
H
M
T
Seal diaphragm
SUS316L
Hastelloy-C
Monel
Tantalum
Other wetted parts
SUS316
Hastelloy-C
Monel
Tantalum Note) Note)
Not available
when 6th digit is “1”.
Indicator (9th digit)
Not provided
Digital, % indication
Digital, real scale indication
A
L
P
Explosion-proof (10th digit)
Non-explosion proof
Intrinsic safety, JIS
A
G
Side vent/drain and fixture (11th digit)
Side vent/drain
None
None
Yes
Yes
A
C
D
F
Fixture
None
Yes (stainless steel)
None
Yes (stainless steel)
Treatment and Fill-fluid (13th digit)
Treatment
None
Degreasing
NACE specification
Y
G
N
Fill-fluid
Silicon oil
Silicon oil
Silicon oil Note)
Note)
Unavailable when 7th digit is “T”.
O-ring material (for process cover)(14th digit)
Viton
A
C
D
E
Bolt/nut (15th digit)
NACE bolt/nut (ASTM A193 B7M/A194 2HM)
NACE bolt/nut (ASTM A320 L7M/A194 2HM)
SUS304/SUS304.... (For general use)
5
FBK···3
OUTLINE DIAGRAM
(Unit : mm)
P
241
80
FBK S 0
T
108
70
V
H
M 3–
T
Mounting pipe
JIS 50A (2B)
A
C
U-bolt
(M8)
70
φ102
225
8
315
54
15
Indicator (option)
Gas vent, drain port
D
E
Details of “B”
View from arrow P
6
8
15 3
41.3
Pressure connection
27
φ7
φ15
A
4–7/16-20UNF
Thread depth 13
(for oval flange)
37 27
47
Type
2.5
B
Optical cable lead-in port
φ18.5
Details of “A”
FBKS
FBKT
D
Rc1/4
1/4–18NPT
E
G1/2
1/2–14NPT
SCOPE OF DELIVERY
Instrument body and pipe fixture (as specified)
ITEM TO BE PREPARED SEPARATELY
Oval flange:
To be used as a flange of connecting pipe
port.
For details, refer to the DATA SHEET of
oval flange (EDS6-10).
ORDERING INFORMATION
1.
2.
3.
4.
Model type
Measuring range
Indication scale for real scale specification
Others
7
FBK···3
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan