FUJI EDSX3-137D

IN-SITU
ZIRCONIA OXYGEN ANALYZER
DATA SHEET
This oxygen analyzer is used to continuously measure
oxygen concentration in combustion exhaust gas of industrial
boilers or furnaces, and is ideally suited for combustion management and control.
The analyzer system is comprised of the detector and converter coupled together as a complete system. Detector setting configuration includes the detector flow guide tube and
detector sensor. The flow guide tube is inserted directly into
the gas and directs gas to the sensor for measurement. The
converter (ZKM) is comprised of the signal processor, input/
output and communications, display and system controls.
The converter is equipped with advanced functionality
such as performing the sensor diagnostics and sensor recovery function, so the detector can be used within long term
stability.
ZFK8, ZKM, ZTA
General-use detector
(ZFK8)
High-temperature detector
(ZTA)
FEATURES
1. Gas sampling device is unnecessary
For quick response, insert the detector directly into the
flue Gas sampling functions such as a gas aspirator and a
dehumidifier are not required.
2. Easy maintenance
The sensor equipped with the detector, has unit construction, it is easy to replace.
By separating the detector and the flow guide tube, filter
replacement is easy.
3. More reliable than sensor diagnosis, sensor recoverable function
Depending on the concentration of the measurement gas,
the power of the sensor might deteriorate. The equipment
includes sensor recovery function electronically, checking
the deterioration status of the sensor depletion.
Therefore, it has high reliability and long-lasting stability.
4. Safe and secure
System detects thermocouple break for heater control on
the sensor side. Safety functions of isolating power supply
to the detector or isolating power via external contact input
are also.
5. Easy operation
The operation and setting for the converter can be performed interactively, and available as English, Japanese or
Chinese for language display.
<IP66>
Converter (ZKM1)
<IP67>
Converter (ZKM2)
SPECIFICATIONS
General Specifications
Measuring object: Oxygen in noncombustible gas
Measuring method: Directly insert type zirconia system
Measuring range: 0 to 2 ··· setting range at option 2 in 50
vol% O2
(in 1 vol% O2 steps)
Repeatability: Within ±0.5%FS
Linearity: Within ±2%FS
Response time: Within 4 to 7 sec, for 90% (from calibration gas inlet)
Warmup time: More than 10 min
Analog output: 4 to 20mA DC (allowable load resistance
less than 500W) or 0 to 1V DC (output
resistance more than 100W)
Power supply: Rated voltage;
100 to 120V AC (operating voltage 90
to 132V AC)
200 to 240V AC (operating voltage
190 to 264V AC)
Rated frequency; 50/60Hz
Power consumption:
Maximum 240VA (Detector: approx.
200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx. 50VA,
Converter: approx. 20VA)
EDSX3-137d
Date
Apr. 1, 2011
ZFK8, ZKM, ZTA
Detector Specifications (ZFK)
Measured gas temperature:
Flow guide tube system; −20 to +600ºC
(for general-use, corrosive gas)
Ejector system; −20 to +1500ºC (for
high-temperature gas)
−20 to +800ºC (for general-use)
Measured gas pressure:
−3 to +3kPa (−306 to +306mmH2O)
Flow guide tube: With or without blow-down nozzle
Flange; JIS5K 65A FF
(JIS5K-80AFF for high particulate gas)
Insertion length; 0.3, 0.5, 0.75, 1m
Ejector (general-use):
Probe for guiding measured gas to
detector
Flange; JIS10K 65A RF
Insertion length; 0.5, 0.75, 1, 1.5m (according to customer's specification)
Operating temperature:
−10 to +60ºC for Primary detecting element
−5 to +100ºC for ejector section
125ºC or less at detector flange surface
with power applied
Storage temperature:
Sensing element: −20 to +70ºC
Ejector: −10 to +100ºC
Structure:
Dust/rain-proof structure(IEC IP66
equivalent)
Filter:
Alumina(filtering accuracy 50µm) and
quartz paper
Main materials of gas-contacting parts:
Detector; Zirconia, SUS316, platinum
Flow guide tube; SUS304 or SUS316
Ejector (general use); SUS316, SUS304
Ejector; (for high temperature) SiC,
SUS316, SUS304
Calibration gas inlet:
φ6mm tube join, φ1/4-inch tube join, or
ball valbe (as specified)
Reference air inlet (option):
φ6mm tube join or φ1/4-inch tube join (as
specified)
Detector mounting:
Horizontal plane ±45º, ambient surrounding air should be clean.
Outer dimensions: (L × max. dia.) 210mm × 100mm (detector)
Mass (approx.) {weight}:
Detector; 1.6kg
Ejector; 15kg (insertion length 1m)
Flow guide tube (general-use, 1m); 5kg
Finish color:
Silver and SUS metallic color
Ejector air inlet flow rate:
5 to 10 L/min
Calibration gas flow:
1.5 to 2 L/min
Blowdown air inlet pressure:
200 to 300kPa {2 to 3 kgf/cm2}
Ejector exhaust gas processing:
Into furnace, returned to flue
Heater temperature drop alarm output (ejector):
Alarm output when below 100 ºC Mechanical thermostat
2
N.O. (1a) contact, 200V AC, 2A
Converter specification (ZKM)
Concentration value indication:
Digital indication in 4 digits
Contact output signal:
(1) Contact specification; 6 points, 1a 250V AC/3A or 30V DC/3A
(2) Contact function;
• Under maintenance
• Under blowdown Note3)
• Span calibration gas valve
• Zero calibration gas valve
• Instrument anomalies Note1)
• Alarm Note2)
Note1) The following Instrument errors (1) Thermocouples break (2) Sensor break (3) Temperature fault
(4) Calibration fault (5) Zero/span adjustment fault
(6) Output error turn the contact-ON
Note2) Alarm selects just one as mentioned below (1)
High (2) Low (3) Upper and Lower (4) High-high
(5) Low-low, it turns ON while operating.
Note3) Under blow down is available in case of option,
and it turns ON while operating.
Contact input signal:
(1) Contact specification; 3points (the following option)
ON; 0V (10mA or less), OFF; 5V
(2) Contact function;
• External hold
• Calculation reset
• Heater OFF
• Blow down (option)
• Inhibition of calibration
• Calibration start
• Range change
Calibration method:
(a) Manual calibration with key operation
(b) Auto. calibration (option)
Calibration cycle; 00 day 00 hour to
99 days 23 hours
(c) All calibration
Calibration gas: • Available range settings
Zero gas; 0.010 to 25.00% O2
Span gas: 0.010 to 50.00% O2
• Recommended calibration gas concentration
Zero gas; 0.25 to 2.0% O2
Span gas; 20.6 to 21.0% O2
(oxygen concentration in the air)
Blowdown:
A function for blowing out with compressed air dust that has deposited in
the flow guide tube. Blowdown can be
performed for a predetermined time and
(option)
at predetermined intervals.
Blowdown cycle; 00 hour 00 minute to
99 hours 59 minutes
Blowdown time; 0 minute 00 second
to 0 minutes 999
seconds
Output signal hold:
Output signal is held during calibration,
processing recoverable sensor, processing diagnosis of sensor, warm-up, PID
auto tuning, under set up maintenance
mode "available" and blowdown. The
hold function can also be released.
Valve and Flow meter (option):
Selects zero or span gas during manual
zero or span calibration. Mounted on the
side of the converter.
Communication function:
RS232C (MODBUS) standard specification
RS485 (MODBUS) (option)
Combustion efficiency display (option):
When you select this display, "rich mode
display" will be simultaneously displayed.
This function calculates and displays
combustion efficiency from oxygen
concentration and measured gas temperature.
Thermocouple (R) is required for temperature measurement.
Operating temperature:
−20 to +55ºC
Operating humidity:
95% RH or less, non condensing
Storage temperature:
−30 to +70ºC
Storage humidity: 95% RH or less, non condensing
Construction:
Dust-proof, rainproof construction
(corresponding to IP66 or IP67 of IEC)
Material:
Aluminum case
Outer dimensions (H x W x D):
170 X 159 X 70mm (IP66, Bench type)
220 X 230 X 95mm (IP67)
182 X 163.5 X 70.6mm (Bench type)
Mass {weight}:
IP66: Approx. 2kg (excluding cable and
detector)
IP67: Approx. 4.5kg (excluding cable and
detector)
Finish color:
IP66: Case: Silver
Cover: Pantone Cool Gray 1C-F
IP67: Munsell 6PB 3.5/10.5 (blue)
Cover: Silver (case)
Mounting method: Mounted flush on panel or on pipe
CODE SYMBOLS
(Detector)
4 5 6 7 8
ZFK 8 R
9 10 11 12 13
5 -
14 15 16
Description
- 1
Cal. gas inlet
For φ6mm tube (SUS)
For φ1/4 inch tube (SUS)
Ball valve
1
2
3
Power supply
100 to 120VAC 50/60Hz
200 to 240VAC 50/60Hz
1
3
Flow guide tube
flange
application
None
SUS304 general use
SUS304 general use
SUS304 general use
SUS304 general use
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 with blow-down nozzle
SUS316 with blow-down nozzle
SUS316 with blow-down nozzle
SUS316 with blow-down nozzle
SUS316 for high particulate
SUS316 for high particulate
SUS316 for high particulate
SUS316 for high particulate
SUS316 for high particulate with
cover
SUS316 for high particulate with
cover
SUS316 for high particulate with
cover
SUS316 for high particulate with
cover
Others
0Y 0
5A3
5A5
5A7
5A1
5B3
5B5
5B7
5B1
5C3
5C5
5C7
5C1
6D3
6D5
6D7
6D1
6E3
6E5
6E7
6E1
ZZZ
750mm
1000mm
Filter spec.
Standard
Instruction manual language
Japanese
English
Chinese
J
E
C
1
2
Specification name plate
Standard (100 to 120V AC 50/60Hz)
Standard (200 to 240V AC 50/60Hz)
(Replacement Detector element)
Power supply
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC
as detailed within the technical construction file
number TZ734575. The applicable standards used to
demonstrate compliance are :
500mm
Reference air inlet
Non
For φ6mm tube (SUS)
For φ1/4 inch tube (SUS)
Y
A
B
1
Overvoltage category
; II power supply input
; I relay interfaces
(IEC1010-1)
External overcurrent protective device
; 10A
Equipment interfaces are safety
separated (SELV)
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
Protection cover
Without
With
Y
A
Electrical Safety:
length
Code symbols
100 to 120V AC
ZFK8YY15-0Y0YY-0YY
200 to 240V AC
ZFK8YY35-0Y0YY-0YY
EN 55011 : 1992 CLASSA Conducted and Radiated emissions
EN 50082-1 : 1992
Radiated immunity, ESD and FBT
ZFK, ZKM
3
ZFK8, ZKM, ZTA
(Converter)
(Ejector)
1 2 3 4 5 6 7 8
Z KM
1 2 3 4 5 6 7 8
9 10 11 12
1 -
Description
1
Construction
IP66
IP67
Bench type
1
2
3
1
Mounting bracket
None (Specify "None" when the bench
type is selected)
Mounting on panel surface
Pipe mounting
Y
1
2
Optional Functions
None
Combustion efficiency display function Note4)
Blowdown
Auto calibration
Combustion efficiency indication
+ Blowdown Note4)
Combustion efficiency indication
+ Auto calibration Note4)
Blowdown + Auto calibration
Combustion efficiency indication + Blowdown
+ Auto calibration Note4)
Y
1
2
3
4
5
6
7
Display language
Japanese
English
Chinese
J
E
C
Y
1
2
Option
None (Specify "None" when the bench
type or the auto calibration is selected)
With valve
With valve + flowmeter
Note4) When you select this display, rich mode
will be a simultaneous display.
(Exclusive-special cable)
1 2 3 4 5 6 7 8
Z R Z KR
9
Description
1 -
Connectable devices
For ZKM
K
Conduit length
None
None
None
None
None
None
None
None
None
None
None
None
6m
10m
Note5
15m
20m
YA
YB
YC
YD
YE
YF
YG
YH
YJ
YK
YL
YM
AA
BB
CC
DD
0
1
2
Insertion length [mm]
500
750
1000
1500
1
3
5
Power supply
100V/115V AC 50/60Hz
200V/220V AC 50/60Hz
230VAC 50/60Hz
SCOPE OF DELIVERY
Detector:
Detector main unit × 1, Viton O ring ×
1, mounting screw (M5mm × 16) × 6,
thermal sticker × 1, flow guide tube (as
specified) × 1, ceramic filter × 1, rainproof cover (as specified) × 1, Instruction
manual × 1
Converter:
Converter main unit × 1, mounting
bracket set, (as specified) × 1
Accessories (AC250V 500mA T fuse ×
2, AC250V 2.5A T fuse × 2),
Instruction manual × 1
Ejector:
Ejector main unit × 1, insertion tube × 1,
M16mm nut, and washer × 4, packing ×
1
Items to be prepared separately:
(1) Standard gas for calibration
Type ZBM NSH4-01 (up to 5% O2 range)
Type ZBM NSJ4-01 (over 5% O2 range)
(2) Reduction valve for standard gas (type ZBD61003)
(3) Flowmeter
Type; ZBD42203, 0.2 to 2L/min (for calibrating gas)
Type; ZBD42403, 1 to 10L/min (for ejector)
CAUTIONS
Types
For R thermocouple
R
Measured gas temperature
For high temperatures (+1500°C max.)
General-use (+800°C max.)
B
C
D
E
Communication function
RS-232C
RS-485
1
2
Description
1
1
2
Output signal
4 to 20mA DC
0 to 1V DC
Other
B
E
Z
Cable length
6m
10m
15m
20m
30m
40m
50m
60m
70m
80m
90m
100m
6m
10m
15m
20m
Cable end treatment
None
One side (detector side)
Both sides
Note5) For connection between detector and
converter, the conduit to be used should
be rainproof flexible type.
4
Z T A
• If combustible gas (CO, H2 etc.) exists in the measured
gas, error will occur due to burning at the sensor section.
The inclusion of corrosive gas (Si vapor, alkaline metal, P,
Pb etc.) will shorten the life of the sensor.
• When the measured gas temperature is high (+300°C
or higher), the flange should be separated from the
furnace wall in order to bring the detector flange surface
temperature below the specified value +125°C). The flow
guide should be attached in the direction in which the gas
flow to the detector decreases.
• When much dust is included in the gas, the flow guide
tube should be attached at an inclination so that the flow
goes from below to above. And the flow guide tube
should be attached in the direction in which the gas flow
to the detector decreases.
• In the case of a refuse incinerator, automatic blow down
of the flow guide should not be performed (to prevent
corrosion of the flow guide tube due to drainage). Blowdown should be performed manually when change in
the indication has become very little with the furnace
stopped.
CONFIGURATION
Flow guide tube system (with valve)
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Solenoid valve
(not supplied)
Reduction valve
(unnecessary when supply air pressure is
200 to 300kPa)
Supply air
15ASGP
tube equivalent
or φ10/φ8 PTFE
pipe or copper pipe
(not supplied)
15ASGP
tube equivalent
or φ10/φ8 PTFE
pipe or copper pipe
(not supplied)
Rainproof cover
Flow guide tube
Detector
(ZFK8)
AC power
supply
Gas temperature.
600ºC max.
6
1
3
5
2
4
*2
Ref. air
Blue
Sample gas outlet
Red
Yellow
Rainproof flexible
conduit.
(Max. 20m)
White
Black
White
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Sample gas outlet
Converter (ZKM)
1
Gas temperature.
600ºC max.
Flow guide tube for high
particulate with cover
2
3
4
+ O2 −
+ TC1 −
O2 input
Thermocouple
input
5
6
AO
+
7
φ6mm or φ1/4 inch tube
for calibration gas
(not supplied)
Thermocouple
for combustion
control
or
FAULT
10
11
12
13
TM-1
14
ZERO
GAS
BLOW
HEATER
Blow Calibrating
Fault
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
+ TC2 −
DI1
DI2
DI3 DICOM
15
17
18
19
16
9
ZV
Analog output
4 to 20mA DC
or 0 to 1V DC
*4
8
−
20
SPAN
GAS
MAINTE
COM
21
22
23
24
25
SV
26
AC power
supply
PE
*3
N AC L
27
28
1
2
RS232C
or
RS485
(option)
TM-2
3
(Option)
Flowmeter
Valve
Reduction valve
(ZBD6)
MEASURE
ZERO
Standard gas
for zero point (ZBM)
SPAN
Reduction valve
(ZBD6)
Flowmeter
*2 Ref. air
to Detector
(0.2 to 0.5L/min)
(not supplied)
Note: *1 Standard gas or instrumentation air
can be used in place of span gas.
*2 Instrument quality air or bottled air is
*1
available as reference air instead of
Span gas (ZBM)
ambient air.
*3 Protective earth.
*1 Air supply
*4 Connect the shield of a exclusive cable
with the ground terminal in the converter.
Reduction valve
(unnecessary when
supply air pressure
is 50 to 100kPa)
5
ZFK8, ZKM, ZTA
Flow guide tube system
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Reduction valve
(unnecessary when supply air pressure is
200 to 300kPa)
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
or φ10/φ8 PTFE
pipe or copper pipe
(not supplied)
15ASGP
tube equivalent
or φ10/φ8 PTFE
pipe or copper pipe
(not supplied)
Rainproof cover
Flow guide tube
Detector
(ZFK8)
AC power
supply
Gas temperature.
600ºC max.
6
1
5
3
2
4
*2
Ref. air
Blue
Sample gas outlet
Yellow
Red
Rainproof flexible
conduit.
(Max. 20m)
White
Black
White
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Sample gas outlet
Converter (ZKM)
1
Gas temperature.
600ºC max.
Flow guide tube for high
particulate with cover
2
3
4
+ O2 −
+ TC1 −
O2 input
Thermocouple
input
5
6
AO
+
7
or
Thermocouple
for combustion
control
φ6mm or φ1/4 inch tube
for calibration gas
(not supplied)
10
11
12
13
TM-1
14
ZERO
GAS
BLOW
HEATER
Blow Calibrating
Fault
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
+ TC2 −
DI1
DI2
DI3 DICOM
15
17
18
19
16
9
ZV
FAULT
Analog output
4 to 20mA DC
or 0 to 1V DC
*4
8
−
20
COM
21
22
23
AC power
supply
SPAN
GAS
MAINTE
24
25
PE
*3
N AC L
SV
26
27
28
1
2
RS232C
or
RS485
(option)
TM-2
3
AC power
supply
Solenoid valve
(not supplied)
Joint
Reduction valve
(ZBD6)
Standard gas
for zero point (ZBM)
Flowmeter
(ZBD4)
(1 to 1.5L/min)
Reduction valve
(unnecessary when supply air pressure is 50 to 100kPa)
*2 Ref. air
*1 Air supply
to Detector
(0.2 to 0.5L/min)
(not supplied)
Solenoid valve
(not supplied)
Note: *1 Standard gas or instrumentation air can be used in place of span gas.
*2 Instrument quality air or bottled air is available as reference air instead
of ambient air.
*3 Protective earth.
*4 Connect the shield of a exclusive cable with the ground terminal in
the converter.
6
Reduction valve
(ZBD6)
Span point
*1
Standard gas for
span point (ZBM)
Ejector system (with valve)
Ejector (ZTA)
Flowmeter
(ZBD)
Ejector (5 to 10L/min)
Reduction valve
(unnecessary when supply air pressure
is 200 to 300kPa )
Joint
Supply air
Power supply
1
2
4
5
Gas temperature.
1500ºC max.
Copper pipe φ6/φ4mm
(not supplied)
Heater temperature drop
Alarm
Solenoid valve
(not supplied)
Copper pipe
φ10/φ8mm
(not supplied)
Copper pipe φ10/φ6mm
(not supplied)
Detector (ZFK8)
6
1
5
AC power
supply
3
2
4
White
Rainproof flexible
conduit.
(Max. 20m)
Black
White
Bule
Yellow
Red
1
2
3
4
+ O2 −
+ TC1 −
O2 input
Thermocouple
input
5
6
AO
+
7
FAULT
Analog output
4 to 20mA DC
or 0 to 1V DC
Thermocouple
for combustion
control
φ6mm or φ1/4 inch tube
for calibration gas
(not supplied)
10
11
12
13
TM-1
14
ZERO
GAS
BLOW
HEATER
Blow Calibrating
Fault
Heater
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
+ TC2 −
DI1
DI2
DI3 DICOM
15
17
18
19
16
9
ZV
or
*4
8
−
20
SPAN
GAS
MAINTE
COM
21
22
23
24
25
SV
26
AC power
supply
*3
PE
N AC L
27
28
1
2
RS232C
or
RS485
(option)
TM-2
3
(Option)
Flowmeter
Valve
AC power
supply
Reduction valve
(ZBD6)
MEASURE
ZERO
Standard gas
for zero point (ZBM)
SPAN
Reduction valve
(ZBD6)
*1
Span gas (ZBM)
Flowmeter
*2 Ref. air
*1 Air supply
to Detector
(0.2 to 0.5L/min)
(not supplied)
Reduction valve
(unnecessary when
supply air pressure
is 50 to 100kPa)
Note: *1 Standard gas or instrumentation air
can be used in place of span gas.
*2 Instrument quality air or bottled air is
available as reference air instead of
ambient air.
*3 Protective earth.
*4 Connect the shield of a exclusive cable
with the ground terminal in the converter.
7
ZFK8, ZKM, ZTA
Ejector system
Ejector (ZTA)
Flowmeter
(ZBD)
Ejector (5 to 10L/min)
Reduction valve
(unnecessary when supply air pressure
is 200 to 300kPa )
Joint
Supply air
1
2
4
5
Gas temperature.
1500ºC max.
Power supply
Copper pipe φ6/φ4mm
(not supplied)
Heater temperature drop
Alarm
Solenoid valve
(not supplied)
Copper pipe
φ10/φ8mm
(not supplied)
Copper pipe φ10/φ6mm
(not supplied)
Detector (ZFK8)
6
1
5
AC power
supply
3
2
4
White
Rainproof flexible
conduit.
(Max. 20m)
Black
White
Bule
Yellow
Red
1
2
3
4
+ O2 −
+ TC1 −
O2 input
Thermocouple
input
5
6
AO
+
7
or
Thermocouple
for combustion
control
φ6mm or φ1/4 inch tube
for calibration gas
(not supplied)
10
11
12
13
TM-1
14
ZERO
GAS
BLOW
HEATER
Blow Calibrating
Fault
contact gas contact
contact
output
output
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
+ TC2 −
DI1
DI2
DI3 DICOM
15
17
18
19
16
9
ZV
FAULT
Analog output
4 to 20mA DC
or 0 to 1V DC
*4
8
−
20
SPAN
GAS
MAINTE
COM
21
23
22
24
25
AC power
supply
*3
PE
N AC L
SV
26
27
28
1
2
RS232C
or
RS485
(option)
TM-2
3
AC power
supply
(1 to 1.5L/min)
Joint
Solenoid valve
(not supplied)
Reduction
valve (ZBD6)
Standard gas
for zero point (ZBM)
Flowmeter
(ZBD4)
Reduction valve (unnecessary when
supply air pressure is 50 to 100kPa)
*2 Ref. air
to Detector
*1 Air supply
(0.2 to 0.5L/min)
(not supplied)
Note: *1 Standard gas or instrumentation air can be used in place of span gas.
*2 Instrument quality air or bottled air is available as reference air instead
of ambient air.
*3 Protective earth.
*4 Connect the shield of a exclusive cable with the ground terminal in
the converter.
8
Solenoid valve
(not supplied)
Reduction
valve (ZBD6)
Span gas
*1 Standard gas for
span point (ZBM)
DEVICE CONFIGURATION
The device to be combined differ according to the conditions of the gas to be measured. Select the devices to be combined
with reference to the following table.
Measured gas
Application
Temperature Gas Flow
General-use
600ºC or
(boiler)
less
For corrosive 600ºC or
gas (refuse
DUST
Device configuration
Protection Note
cover
5 to 20m/s Less than 0.2g/Nm3 —
Less than 10g/Nm3 —
5 to 20m/s Less than 1g/Nm3
Less than 10g/Nm3
less
—
—
incinerator)
Less than 25g/Nm3 no
Less than 25g/Nm3 yes
General-use
800ºC or
Less than Less than 1g/Nm3
—
(boiler)
less
1500ºC or
1m/s
Less than Less than 1g/Nm3
—
less
1m/s
Detector type
Converter Ejector
type
type
Fuel; gas, oil
ZFK8R
5- A 5
-1
Fuel: coal
ZFK8R
5- C 5
-1
with blow down
Included low moisture
ZFK8R
5- B 5
-2
Included low moisture
ZFK8R
5- C 5
ZFK8R
with blow down
Included high moisture ZFK8R
with blow down
SUS316 tube
ZFK8R
with blow down
ZFK8R
SIC tube
with blow down
Included low moisture
ZKM
ZKM
—
—
-2
ZKM
ZKM
—
—
5- D 6
-2
ZKM
—
5- E 6
-2
ZKM
—
5-0Y0
-1
ZKM
ZTA2
5-0Y0
-1
ZKM
ZTA1
with blow down
Note (1) Dust volume is approximate value.
(2) Instrument quality air or bottled air is available as reference air by selecting detector with reference air inlet.
9
ZFK8, ZKM, ZTA
OUTLINE DIAGRAM
(Unit:mm)
Detector (ZFK8)
Approx.132
φ 80
Approx.62
Approx.130
φ 67
5
6-φ 6
Temp.Proof cover
(To order)
Terminal box
φ 85
Heater
°
60
Filter
Approx.30
1
3
2
4
Black / White/ red / White/
2-core wire
Ground-wire
Screw: M4
Calibration gas inlet
Ref. Air inlet (to order) (To order)
SUS316,for φ 6/φ 4 tube SUS316,for φ 6/φ 4 tube
or 1/4 inch tube
or 1/4 inch tube
Exclusive
Cable gland
Approx.130
5
Temp.proof cover
(to order)
Terminal box
°
60
125.5
φ 26
φ 79
6-φ 6
φ 85
Approx.62
φ 67
Filter
Calibration gas inlet
Rc1/4 INT.THD.
10
Approx.100
ZFK8R 3
Ball valve
Ref. air inlet (to order)
SUS316,for φ 6/φ 4 tube
or 1/4 inch tube
Element output
Ground-wire
screw: M4
Exclusive cable gland
Valve handle
5
6
yellow / blue
4-core wire
Exclusive-special cable
Approx.132
φ 80
Thermocouple
125.5
φ 26
φ 79
EXTERNAL CONNECTION DIAGRAM
Sensor unit (ZFK8YY)
(132.7)
3.5
20
(?17.5)
(88.4)
4-φ 3.5(Mount for detector)
TK4J4664R0
3
5
S-
H
4
2
45°
T+
S+ H
6
1
made in japan
40
45°
φ 35
T-
Flow guide tube
155
Approx. L
12
Approx. 20
130
67
34
Gas inlet
20
ZFK8R
5-5A3
5
7
1
6-M5
detector side
Gas outlet
8–15
MTG. holes
Oxygen detector
1
Code 11th
3
5
L (m)
0.3
0.5
0.75 1.0
MASS
2.7
Approx.(kg)
3.3
4.1
7
4.8
Z
L=
(to order)
Flow guide tube (with blow-down nozzle)
155
Approx. L
18
Approx. 20
130
67
60.5
67
3
Gas inlet
6-M5
detector side
ZFK8R
5-5C3
5
7
1
Gas outlet
Oxygen detector
Blow down air inlet
Code 11th
3
L (m)
0.3
Mass
3.0
Approx.(kg)
5
7
1
Z
0.5 0.75 1.0
3.8
4.8
5.7
L=
(to order)
8–15
MTG. holes
4-Rc1/4
with plug int.thd.
11
ZFK8, ZKM, ZTA
Flow guide tube (for high particulate)
14
5
180
4 - Rc 1/4
with plug int thd.
Blow down air inlet
Select an upper port to
avoid the condensation
in the piping
8 - 19
MTG. hole
67
29.
5
Flange: JIS 5K80A FF
ZFK MTG. position
185
Packing
5-6D3
5
7
1
800
50
ZFK8R
Gas outlet
30
Code 11th
3
L (m)
0.3
Mass
4.5
Approx.(kg)
5
7
1
Z
0.5 0.75 1.0
5.6
7.0
8.3
L=
(to order)
1
Z
Gas flow
Tube (50A SCH40)
5
14
180
Flow guide tube (for high particulate with cover)
4 - Rc 1/4
with plug int thd.
Blow down air inlet
8 - 19
MTG. holes
Select an upper port to
avoid the condensation
in the piping
67
29
.5
Flange: JIS 5K80A FF
(175)
ZFK MTG. position
50
Packing
ZFK8R
5-6E3
5
7
1
800
(790)
Gas outlet
12
Code 11th
3
L (m)
0.3
Mass
7.1
Approx.(kg)
Gas flow
Protection tube (65A SCH40)
30
(38)
Tube (50A SCH40)
5
7
0.5 0.75 1.0
L=
9.0 11.4 13.6 (to order)
Flow guide tube (for corrosive gas)
155
12
Approx. 40
Approx. L
130
3
67
67
60.5
Gas inlet
ZFK8R
6-M5
detector side
Gas outlet
5-5B3
5
7
1
1
Code 11th
3
L (m)
0.3
0.5 0.75
1.0
MASS
3.3
Approx.(kg)
4.5 6.1
7.6
5
7
Z
L=
(to order)
8- 15
MTG. holes
Oxygen detector
Ejector (ZTA)
4-M16 bolt
Viewed from P direction
42.7
30
8
170
P
Gas inlet
217
25
105
190
Approx. L
140
(JIS 10K65ARF)
L [mm] 500 750 1000 1500
40
Detector
(ZFK8)
Ejector air
outlet
(Rc1/4)
35
EXTERNAL CONNECTION DAIAGRAM
1
2
AC
power supply
AC100/110V
AC200/220V
AC230V
3
4
5
Blow -down air inlet
(Rc1/4)
Ejector air inlet
(Rc1/4)
40
50
45
Cable gland
(A15C)
Heater temp.drop alarm
13
ZFK8, ZKM, ZTA
Converter (ZKM1)
<IP66>
Mounting hole
2-ø9
100
2.5
2.5
75
100
10
72
Valve (option)
170
ESC
ENTER
S
Z
Oxygen Analyzer
POWER
270
MODE
55
O
N
M E A S
U bolts M8
(option)
ZKM
Oxygen Analyzer
10
Calibration gas outlet
(for F6 tube)
35
153
Mounting for
2B pipe
For contact Input/output
62.5
Earth terminal (M4)
For sensor (coupling)
For F23.5/F15.8 tube
Zero gas inlet
(for F6 tube)
Mtg. Plate
For communication
For output
Span gas inlet
(for F6 tube)
For power supply
Converter (ZKM2)
<IP67>
Mounting hole 4-F9
U bolts M8 (option)
Bracket
324
72
195
59
20
Valve (option)
498
55
312
220
POWER
N
ENTER
S
ESC
O
MODE
Z
M E A S.
Mounting for
2B pipe
Earth terminal (M4)
230
14
5.7
Calibration gas outlet
(for F6 tube)
95
95
3.2
39
U bolts M8 (option)
95
For sensor (coupling)
For F23.5/F15.8 tube
Zero gas inlet
(for F6 tube)
For sensor (coupling)
For F23.5/F15.8 tube
40
For contact Input/output
(G1/2)
For communication
(G1/2)
14
40
40
40
For power supply
(G1/2)
For output
(G1/2)
Span gas inlet
(for F6 tube)
Converter (ZKM1)
<IP66>
2.5
2.5
75
Mounting hole
2-ø9
100
100
10
72
Valve (option)
170
ESC
ENTER
N
S
POWER
230
270
MODE
O
Oxygen Analyzer
Z
M E A S
U bolts M8
(option)
Flowmeter (option)
ZKM
Oxygen Analyzer
10
Calibration gas outlet
(for F6 tube)
153
Mounting for
2B pipe
(125)
For contact Input/output
62.5
Earth terminal (M4)
For sensor (coupling)
For F23.5/F15.8 tube
Mtg. Plate
Zero gas inlet
(for F6 tube)
For communication
Span gas inlet
(for F6 tube)
For output
For power supply
Converter (ZKM2)
<IP67>
Mounting hole 4-F9
U bolts M8 (option)
Bracket
324
72
195
59
20
Valve (option)
230
498
312
220
POWER
N
ENTER
S
ESC
O
MODE
Z
M E A S
Mounting for
2B pipe
Flowmeter
(option)
5.7
Earth terminal
(M4)
230
14
Calibration gas outlet
(for F6 tube)
95
95
95
39
3.2
U bolts M8 (option)
180
For sensor (coupling)
For F23.5/F15.8 tube
Zero gas inlet
(for F6 tube)
For sensor (coupling)
For F23.5/F15.8 tube
40
For contact Input/output
(G1/2)
For communication
(G1/2)
40
40
40
Span gas inlet
(for F6 tube)
For power supply
(G1/2)
For output
(G1/2)
15
ZFK8, ZKM, ZTA
Converter (ZKM3)
<Bench type>
163.5
10
70.6
143.5
Oxygen Analyzer
50/60Hz 60VA
OXYGEN TRANSM ITTER
Made in Japan
Fuji Electric systems Co.,Ltd.
TYPE
OUTPUT DC
POWER AC
SER. NO.
MFD
172
ESC
ENTER
注 意
POWER
182
MODE
POWER
FUSE T 0.5A
変換器へ供給する電源電圧は、接続される
検出器の電源電圧に合わせてください。
HEATER
CAUTION
FUSE T 2.5A
Power supply must be selected
the same voltage of detector.
POWER
ON
OFF
+ O2 − + TC1− + AO− FAULT
BLOW ZV HEATER
1
2
3
4
5
6
7
8
9
10 11 12 13 14
15 16 17 18 19 20 21 22 23 24 25 26 27 28
ALARM MAINAC
COM SV
+ TC2 − DI1 DI2 DI3 DIN
L
TENANCE
COM
4-φ4.5 Mounting hole
O2 sensor
input
EXTERNAL TERMINAL (TM1) / Terminal M3
O2 sensor
thermocouple
input
*2
Exclusive
cable
Analog
output
4-20mA DC
or 0-1V DC
1
+
O2
2
3
−
+
4
TC1
−
5
+
AO
6
Fault
contact output
250V AC/3A
30V DC/3A
7
8
9
+
15
TC2
−
DI1
16
17
Thermocouple
input type R
(not supplied)
DI2
18
11
12
FAULT
BLOW
ZERO
GAS
14
21
20
22
Alarm
contact output
250V AC/3A
30V DC/3A
Contact input
Short circuit : on (Current 5mA)
Open : off (Voltage 5V)
HEATER
SPAN
GAS
MAINTE
DI3 DICOM
19
13
ZV
ALARM
Ground
terminal
M4
10
−
or
*1
O2 sensor
heater power
output
Blow contact
output
250V AC/3A
30V DC/3A
23
24
COM
25
Maintenance
contact output
250V AC/3A
30V DC/3A
SV
26
Calibrating gas
contact output
250V AC/3A
30V DC/3A
N
27
AC
L
28
AC
Power supply
input
100 to120V AC
200 to 240V AC
COMMUNICATION TERMINAL (TM2) /INSERTION TERMINAL
Terminal number
Remarks
1
2
3
RS232C
TXD
RXD
GND
Standard
RS485
TRX+
TRX–
GND
Option
Note 1) The heater power supply is the same
as the converter power supply.
Note 2) Be sure to connect the shield of the
cable to the ground in the main body.
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan