FUJI EDS3-212A

MULTI-CHANNEL ZIRCONIA OXYGEN ANALYZER
DATA SHEET
This oxygen analyzer measures at multiple points the
oxygen concentration included in the exhaust gas from a large
boiler, and simultaneously outputs the average value of the
concentrations.
Since the distribution of oxygen concentration in exhaust
gas is not uniform with a large boiler such as found in a power
plant, the concentration is measured at multiple points in a
flue. These values are averaged to provide an optimum
oxygen concentration signal which is necessary for
combustion control.
The converter (ZRN) will accept inputs from up to nine
direct-insertion type zirconia oxygen detectors (ZFK), and
carries out computations such as averaging of signals from
the detectors.
The detectors are available in general-use, anticorrosive
and high-temperature types for selection according to the
customer's intended use.
ZRN, ZFK
General-use detector
High-temperature detector
FEATURES
1. Up to nine detectors connectable
3, 6 or 9-channel configuration is selectable, and channels
are freely usable within the selected number. Addition of
channel unit allows expansion in units of 3 channels (9
channels max.).
2. Interactive setting of parameters
Various parameters for measurement and maintenance
are easily settable through interaction with the liquid
crystal display.
3. A variety of output signals
Available as analog output are 3 sets of average-value
outputs (average values from desired channel combination) and oxygen concentration output for each channel.
Also there are 4 contact outputs per channel and 7 contact
outputs common to all channels. All of the analog outputs
are isolated.
4. Full complement of maintenance functions
Auto calibration function and manual/auto blow-down
function are standard-equipped. A self-diagnosis function
for the main unit facilitates taking a remedy at the
occurrence of a fault.
5. Sampling device is unnecessary
Gas sampling devices such as gas aspirator and
dehumidifier are unnecessary since the detector is a
direct-insertion type. The detector is available in a hightemperature type for measurement of hot gases up to
1590°C and an anticorrosive type for measurement of
strongly corrosive gas from a garbage incinerator.
Converter
SPECIFICATIONS
General
Measuring object: Oxygen contained in noncombustible
gas
Measuring principle:
Direct-insertion zirconia system
Measuring range: 0 to 2 ……50 vol% O2, freely settable
(in 0.5% O2 steps)
Repeatability:
Within ± 0.5% of max. output signal
Linearity:
± 2% of full scale
Response time:
Within 7 sec for 90% response (from
calibration gas inlet)
Power supply:
100, 115 or 220VAC, 50/60 Hz
Power consumption:
Approx. 50VA + 50VA x No. of channels
(at steady state of general-use detector)
Approx. 50VA + 200VA x No. of channels
(at startup of general-use detector)
Warmup time:
Approx. 15 minutes
Oxygen detector (Type ZFK2,5), Ejector (Type ZTA)
Measuring detector:
For general-use: ZFK2
For corrosive gas: ZFK5
EDS3-212a
Date
Apr. 30, 1997
ZRN, ZFK
Measured gas temperature:
Flow guide tube system; –20 to +600°C
(for general-use, corrosive gas)
Ejector system; –20 to +1590 °C (for
high-temperature gas)
–20 to +800°C (for general-use)
Measured gas pressure:
–3 to +3kPa (–306 to +306mmH2O)
Flow guide tube: With or without blow-down nozzle
Flange; JIS5K 65A FF
(JIS5K-80AFF for high particulate gas)
Insertion length; 0.3, 0.5, 0.75, 1m
(0.8m for high particulate gas)
Ejector (general-use):
Probe for guiding measured gas to detector
Flange; JIS10K 65A RF
Insertion length; 0.5, 0.75, 1, 1.5m
(according to customer's specification)
Ambient temperature:
–20 to +60°C for cable section
–5 to +100°C for ejector section
125°C or less at detector flange surface
with power applied
Structure:
Dust/rain-proof structure(IEC IP55
equivalent)
Filter:
Alumina(filtering accuracy 50µm) and
quartz paper
Main materials of gas-contacting parts:
General-use detector; Zirconia, SUS316,
platinum
Anticorrosive detector; Zirconia, titanium, platinum
Flow guide tube; SUS304 or SUS316
Ejector (general use);SUS316, SUS304
Ejector; (for high temperature) SiC,
SUS316, SUS304
Calibration gas inlet:
Polypropylene or brass joint
Reference air inlet (option):
Rc1/8 or NPT1/8
Detector mounting:
Horizontal plane ± 45°, ambient surrounding air should be clean
Outer dimensions: (L x max. dia.) 210mm x 100mm
(detector)
Mass{weight}:
Detector; approx. 1.6kg
Ejector; approx. 15kg (insertion length
1m)
Flow guide tube (general-use, 1m);
approx. 5kg
Finish color:
Munsell N9.5 (silver) and SUS metallic
color
Ejector air inlet flow rate:
5 to 10N/min
Blow-down air inlet pressure:
200 to 300kPa (2.04 to 3.06kgf/cm2)
Ejector exhaust gas processing:
Within furnace, returned to flue
Heater temperature drop alarm output (ejector):
Alarm output when below +100 °C,
mechanical thermostat, 1N.O. (a) contact of 200V AC 2A
2
Oxygen converter (Type ZRN)
0 to 2 ……50 vol% O2, freely settable
(in 0.5% O2 steps)
Indication:
Oxygen concentration; 3-digit indication
by LED
Operation/setting display; 16 digits x 2
rows on LCD
Measurement channel status; 12 x 2
colors by LED
Oxygen concentration output signal per channel:
4 to 20mA DC (allowable load resistance
500 Ω or less)
or 0 to 1V (output resistance 100 Ω or
less), isolated output, linear
Average-value output signal (3 points):
4 to 20mA DC (allowable load resistance
500 Ω or less)
or 0 to 1V (output resistance 100 Ω or
less), isolated output, linear
Contact output signal:
(1) Independent per channel (4 points)
Unit alarm (contact closed during normal
measurement)
H/L alarm (contact closed when setpoint
exceeded)
Under calibration (contact closed during
calibration)
Under blow-down (contact closed during
blow-down)
(2) Common to all channels (7 points)
Unit alarm (contact closed when all
intended channels are under normal
measurement)
Under maintenance (contact closed in
maintenance mode, setting mode,
during auto calibration and during blowdown)
Zero valve (contact closed during zero
calibration)
Span valve (contact closed during span
calibration)
H/L alarm A (contact closed when average-value output A exceeds set- point)
H/L alarm B (contact closed when average-value output B exceeds setpoint)
H/L alarm C (contact closed when average-value output C exceeds setpoint)
(3) Contact specification:
1N.O.(a) 250V AC 2A, 24V DC 2A
Contact input signal:
Auto calibration start (auto calibration
starts when contact opens)
Calibration disable (calibration disabled
when contact closes)
Contact specification; isolated contact
input, ON at 1kΩ or less
Auto calibration: Zero and span automatically calibrated at
preset cycle; calibration gas supplied in
succession by driving external solenoid
valve
Calibration cycle; 0.1 to 99.9days
Calibration gas flow time; 1 to 9minutes
Auto calibration channel; freely settable
Output range:
Calibration gas concentration setting:
Settable range; 0.100 to 50.000% O2
(Zero: Up to 22.000% O2)
Density of recommended calibration gas
Zero: Less than 10%/residual N2 (more
than 0.15 vol.% O2) of full scale
Span: 80 to 100%/residual N2 of full scale
Blow-down:
A function for blowing out via compressed air the dust that has deposited in
the flow guide tube.
Blow-down can be performed for a predetermined time and at predetermined
intervals.
Blow-down cycle; 0.1 to 99.9hours
Blow-down time; 0.1 to 9.9minutes
Blow-down channel; freely settable
Output hold:
Output is held during maintenance,
setting, auto calibration and blow-down.
The hold function can also be released.
Transmission function (option):
Via RS-485 interface
Half-duplex bit serial transmission, startstop synchronization.
Remark: When connecting via an RS232C interface, then a 232 485 converter should be used.
Self-diagnosis function:
Provided for detector temperature fault,
zero calibration fault, span calibration
fault, internal communication fault
Ambient temperature:
–10 to +50°C
Ambient humidity: 90% RH or less
Power supply:
100, 115 or 220V AC, 50/60Hz
Structure:
Dust-proof, rainproof structure ( corresponding to IP53 of IEC)
Material:
Made of steel
Outer dimensions (H x W x D):
221 X 482.6 X 380 (rack/panel mounting
type)
221 X 400 X 380mm (pipe stand mounting type)
Mass{weight}:
Approx. 18kg (excluding cable and detector)
Finish color:
Munsell 5Y 8/1
Mounting method: Mounted on pipe stand, mounted flush
on panel or on 19inch rack
3
ZRN, ZFK
CODE SYMBOLS
(Detector)
(Ejector)
4 5 6 7 8
ZFK
9 10 11 12 13
1 2 3 4 5 6 7 8
14
Description
4
Z T A
Application
General use
For corrosive gas (refuse incinerator)
2
5
1
Measured gas temperature
For high temperature (1590°C max.)
General use (800°C max.)
1
2
Kinds
Standard
R
Power supply
100/115VAC 50/60Hz
200/220VAC 50/60Hz
1
3
Flow guide tube
flange
application
None
SUS304 general use
SUS304 general use
SUS304 general use
SUS304 general use
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 with blow-down nozzle
SUS316 with blow-down nozzle
SUS316 with blow-down nozzle
SUS316 with blow-down nozzle
SUS316 for high particulate
SUS316 for high particulate with
cover
Otheres
0Y0
5A3
5A5
5A7
5A1
5B3
5B5
5B7
5B1
5C3
5C5
5C7
5C1
6D8
6E8
ZZZ
Insertion length [mm]
500
750
1000
1500
B
C
D
E
Cal. gas inlet
Polypropylene joint for φ6mm tube
Brass joint for φ1/4 inch tube
1
2
Power supply
100V/115V AC 50/60Hz
200V/220V AC 50/60Hz
1
3
length
(Converter)
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
800mm
800mm
Z RN
1 2 3 4 5 6 7 8
9
Description
1 -
No. of measurement channels
3 channels max.
6 channels max.
9 channels max.
3
6
9
Output signal
4 to 20mA DC
0 to 1V DC
B
E
Transmission function
None
with RS-485
0
1
Power supply
100V AC 50/60Hz
115V AC 50/60Hz
220V AC 50/60Hz
1
2
3
Protection cover
Without
With
Y
A
Description
1
1
2
Reference air inlet
None
Rc1/8
NPT1/8
Y
A
B
Z
None-standard spec.
Other none-standard items
(Exclusive special cable)(*1)
1 2 3 4 5 6 7 8
Z R ZN
9
Description
1 -
Connectable devices
For ZRN
N
Kinds
For R thermocouple
R
( Replacement Detector element)
4 5 6 7 8
9 10 11 12 13
4
0 Y 0 Y Y
ZFK
Description
Application
General use (TC516041)
For corrosive gas (refer to incinerator)
(TC516042)
2
5
Kinds
Standard
R
Cal. gas inlet
Polypropylene joint for φ6mm tube
Brass joint for φ1/4 inch tube
6
7
1
3
Power supply
100/115VAC 50/60Hz
200/220VAC 50/60Hz
Mounting method
Mounting flush on panel/on 19inch rack
Mounted on pipe stand (pipe stand provision
is excluded)
Conduit(*2)
None
None
None
None
None
None
None
None
None
None
None
None
6m
10m
15m
20m
YA
YB
YC
YD
YE
YF
YG
YH
YJ
YK
YL
YM
AA
BB
CC
DD
0
1
2
Cable length
6m
10m
15m
20m
30m
40m
50m
60m
70m
80m
90m
100m
6m
10m
15m
20m
Cable end treatment
None
One side (detector side)
Both sides
Notes: *
(1) One cable is used for one detector.
(2) Waterproof flexible type conduit.
4
SCOPE OF DELIVERY
Detector:
Detector main unit x 1, Viton O ring x 2,
mounting screw (M5mm x 16) x 6,
thermal sticker x 1, flow guide tube (as
specified) x 1, ceramic filter x 1, rainproof
cover (as specified) x 1
Converter:
Converter main unit x 1, mounting
bracket set, (according to specification) x
1 accessories (3A fuse x 3)
Ejector:
Ejector main unit x 1, insertion tube x 1,
M16mm nut, and washer x 4, packing x 1
Items to be prepared separately:
(1) Calibration gas unit
Code
used
TK7D5853C.
1 2 3 4 5
(2) Standard gas for calibration
Type ZBM NSH4-01 (up to 5% range)
ZBM NSJ4-01(over 5% range)
(3) Reducing valve for standard gas (type ZBD61003)
(4) For blow-down
• 2-way solenoid valve
ADK11-15A-02C-100/200VAC (supplied by CKD Co.,
Ltd.), connection port dia. Rc1/2
• Reducing valve AW4000-N04BG (supplied by SMC Co.,
Ltd.), connection port dia. Rc1/2
(5) Flowmeter (type ZBD42403)
1 to 10N/min for ejector
6
Panel/19inch
(3) (6) (9)
rack mount
Pipe mount
(3) (6) (9)
(Number of "ch")
DEVICE CONFIGURATION
The device to be combined differ according to the conditions of the gas to be measured. Select the devices to be combined
with reference to the following table.
Measured gas
DUST
Device configuration
Protection Note
cover
Application
Temperature Gas Flow
General-use
600°C or less 5 to 20m/s Less than 0.2g/mm3 —
Less than 10g/Nm3 —
(boiler)
Detector type
Converter Ejector
type
type
Fuel; gas, oil
ZFK2R
4-
5
A
ZRN
—
Fuel: coal
ZFK2R
4-
5
C
ZRN
—
with blow down
For corrosive 600°C or less 5 to 20m/s Less than 1g/Nm3
Less than 10g/Nm3
gas (refuse
—
Included low moisture
ZFK5R
4-
5
B
ZRN
—
—
Included low moisture
ZFK5R
4-
5
C
ZRN
—
ZFK5R
4-
6
D
ZRN
6
E
ZRN
with blow down
incinerator)
Less than 25g/Nm3
no
Included low moisture
with blow down
General-use
800°C or less Less than
(boiler)
Less than 25g/Nm3
yes
Included high moisture ZFK5R
with blow down
4-
Less than 1g/Nm3
—
SUS316 tube
ZFK2R
4-OYO
ZRN
ZTA2
ZFK2R
4-OYO
ZRN
ZTA1
with blow down
1m/s
1590°C or
Less than
less
1m/s
Less than 1g/Nm3
—
SIC tube
with blow down
Note (1) Dust volume is approximate value.
(2) Instrument quality air or bottled air is available as reference air by selecting detector with reference air inlet.
CAUTIONS
• If combustible gas (CO, H2 etc.) exists in the measured
gas, error will occur due to burning at the sensor section.
The inclusion of corrosive gas (Si vapor, alkaline metal, P,
Pb etc.) will shorten the life of the sensor.
• When the measured gas temperature is high (300°C or
higher), the flange should be separated from the furnace
wall in order to bring the detector flange surface
temperature below the specified value(125°C). The flow
guide should be attached in the direction in which the gas
flow to the detector decreases.
•
•
When much dust is included in the gas, the flow guide
tube should be attached at an inclination so that the flow
goes from above to below. And the flow guide should be
attached in the direction in which the gas flow to the
detector decreases.
In the case of a refuse incinerator, automatic blowback of
the flow guide should not be performed (to prevent
corrosion of the flow guide tube due to drainage). Blowdown should be performed manually when change in the
indication has become very little with the furnace
stopped.
5
ZRN, ZFK
CONFIGURATION
Flow guide tube system
Rc 1/4
(upstream blow port
should be selected)
Blow-down 200 to 300kPa
{ 2 to 3 kgf/cm2 }
Reducing valve
(unnecessary when supply air pressure is 200 to 300 kPa
{ 2 to 3 kgf/cm2 } )
Solenoid valve
for below 1
Air supply
Detector
ZFK2
ZFK5
Rainproof cover
1
Flow guide tube
or 10/ 8 PTFE pipe
or copper pipe
15A SGP
equivalent tube
(not supply)
2
3
4
5
Gas temp.
600°C max.
6
Power supply
Detector 1
Rc 1/4
(upstream blow port should be selected)
Blow-down 200 to 300 kPa
{ 2 to 3 kgf/cm2 )
Detector
ZFK2
ZFK5
Solenoid valve
for belw 9
15A SGP equivalent
tube (not upply)
Reducing valve
Rainpproof
cover
1
Flow guide tube
3
4
5
6
Gas temp.
600°C max.
Detector 9
BL9 BL-C
BL1 BL-C
2
4 cores
Cell/
thermocouple
output
2 cores
Heater
I1
I1
TC1
TC1
H1
power supply
H1-C
4 cores
Cell/
I9
thermocouple
output
2 cores
Heater
power supply
I9
TC9
TC9
H9
H9-C
Sample gas outlet
Heater
~ 3ch
Main
board
~ 3ch
board
~ 9ch
Zirconia type multichannel oxygen
analyzer ZRN
Heater
board
~ 3ch
Heater
~ 3ch
I/O board
Main
board
~ 9ch
SV
ZV COM
V1
COM
board
~ 9ch
V9
AC
Power supply
AC
COM
Power supply
Heater
board
~ 9ch
6
5
4
7
8
15
18
1
2
Reducing valve
(unnecessary when suppy air
pressure is 50 to 100kPa
{0.51 to 1.02 kgf/cm2 }
Calibration gas unit
6/ 4mm or 1/ 4inch
PTFE tube for
calibration gas (to be
prepared separately)
Gas temperature.
600°C max.
Flow guide tube for
high particulate
(*1) Air suplly
Reducing
valve
Flow meter
(ZBD4)
(*) Standard gas or
instrumentation air
supply is usable for
span point.
ZBD61
Flow meter
(ZBD4)
Reducing valve
(1 to 1.5R/min)
(*2) Ret. air
(1 to 1.5R/min)
to Detector
Sample gas outlet
Gas temperature.
600°C max.
Flow guide tube for high
particulate with cover
Note:
*1 Standard gas or instrumentation air is usable to span point.
*2 Instrument quality air or bottled air available as reference air instead of ambient air.
6
(*) Standard gas for
span point (ZBM)
Standard gas for
zero point (ZBM)
0.45 to 1.0% O2 /N2
Ejector system
Rc 1/4
Blow-down 200 to 300kPa
{ 2 to 3 kgf/cm2 }
Solenoid valve
for blow 1
Reducing valve
(unnecessary when supply air pressure is 200 to 300 bkPa
{ 2 to 3 kgf/cm2 }
Air supply
Blow inlet
15A SGP equivalent tube
or 10/ 8 copper pipe
(not supply)
1
Air supply for blow
2
Reducing valve
3
Ejectoy
inlet
1 2
4
6/
4
4 coppre pipe (not supply)
5
5
6
Power supply
Heater temp.
drop alarm
Power supply
Detector 1
(5 to 10R/min)
Flowmeter 1
(5 to 10R/min)
Flowmeter 9
Solenoid valve
for blow 9
15A SGP equivalennt tube or
10/ 8 copper pipe (not supply)
Rc 1/4
Blow-down 200 to 300 kPa
{ 2 to 3 kgf/cm2 )
Reducing valve
1
3
4
1 2
4
5
5
6
Power supply
Heater temp.
drop alarm
Detector 9
BL9 BL-C
BL1 BL-C
2
4 cores
Cell/
thermocouple
output
2 cores
Heater
I1
I1
TC1
TC1
H1
power supply
H1-C
4 cores
Cell/
I9
thermocouple
output
2 cores
Heater
power supply
I9
TC9
TC9
H9
H9-C
Heater
~ 3ch
Main
board
~ 3ch
board
~ 9ch
Zirconia type multichannel oxygen
analyzer ZRN
Heater
board
~ 3ch
Heater
~ 3ch
I/O board
Main
board
~ 9ch
SV
ZV COM
V1
COM
board
~ 9ch
V9
AC
Power supply
AC
COM
Power supply
Heater
board
~ 9ch
6
5
4
7
8
15
18
1
2
Reducing valve
(unnecessary when suppy air
pressure is 50 to 100kPa
{0.51 to 1.02 kgf/cm2 }
Calibration gas unit
6/ 4mm or 1/ 4inch
PTFE tube for
calibration gas (to be
prepared separately)
(*1) Air suplly
Reducing
valve
Flow meter
(ZBD4)
(*) Standard gas or
instrumentation air
supply is usable for
span point.
ZBD61
Flow meter
(ZBD4)
Reducing valve
(1 to 1.5R/min)
(*2) Ret. air
(1 to 1.5R/min)
(*) Standard gas for
span point (ZBM)
Standard gas for
zero point (ZBM)
0.45 to 10% O2 /N2
to Detector
Note:
*1 Standard gas or instrumentation air is usable to span point.
*2 Instrument quality air or bottled air available as reference air instead of ambient air.
7
ZRN, ZFK
Detector 2 to 8
2
Detector 1 ZFK2/5
3
4
5
6
7
8
Detector 9
1
Heater
1
Heater
2
2
+
3
Sensor
output
Thermocoule
–
–
4
+
6
5
Sensor
output
3
Thermocoule
5
4
6
Zirconia type multi-channel
oxygen analyzer ZRN
Input/output
board
AC power
supply
1
21
2
22
3
23
4
24
5
25
6
26
power
supply
7
27
AC
8
28
AC
9
29
E
10
30
E
11
31
Heatre board
1 to 3ch
Main board
1 to 3ch
Heatre board
*
7 to 9ch
Main board Main board
4 to 6ch * 7 to 9ch
*
21
1
21
1
21
1
21
1
21
1
21
2
22
2
22
2
22
2
22
2
22
2
22
3
23
3
23
3
23
3
23
3
23
3
23
4
24
4
24
4
24
4
24
4
24
4
24
5
25
5
25
5
25
5
25
5
25
5
25
6
26
6
26
6
26
6
26
6
26
6
26
7
27
7
27
7
27
7
27
7
27
7
27
8
28
8
28
8
28
8
28
8
28
8
28
9
29
9
29
9
29
9
29
9
29
9
29
10
30
10
30
10
30
10
30
10
30
10
30
11
31
11
31
11
31
11
31
11
31
11
31
1
SV for
blow 3
AC
power
supply
Heatre board
*
4 to 6ch
SV for
blow 2
(2) Use a cable of 1.25mm2 or larger for the main power
supply (AC power supply of ZRN)
SV for
blow 8
SV for
blow 5
SV for
blow 1
Remarks
(1) Grounding equal to 3rd class grounding structure of
Japanese standard (grounding resistance 100Ω or
less) must be conected.
SV for
blow 9
SV for
blow 6
SV for
blow 4
(3) Shieled cable of 0.5mm2 or larger shouuld be
used for the input/output cable.
SV for
blow 7
(4) Use a compensating conductor 0.5mm2 or
lage for the detector thermocouple input
(for R or K themocoule).
Note: *Board is provided for channel 3, 6 and 9 in case
of specified.
Calibration gas unit
1
2
3
ZV:SV for zero gas
SV:SV for span gas
V1 to V9:SVs for
detectors 1 to 9
8
4
5
6
ZV
SV
7
8
V1
10
11 12
V2 V3
V4
9
13
14
V5 V6
15
16
17
18
V7
V8 V9
19
20
OUTLINE DIAGRAM
(Unit:mm)
Detector (ZFK2)
80
Proof cover (to order)
Approx. 130
67
Approx. 65
45.4
Approx. 118
36
5
26
Filter
EXTERNAL CONNECTION DIAGRAM
85
Heater
Thermocouple
Element output
60
°
A
Approx. 25
1
3
2
4
Black / White/ red / White/
2-core wire
6- 6
5
6
yellow / blue
4-core wire
Exclusive-special cable
Ref. air inlet (to order)
NPT1/8 or Rc1/8
Calibration gas inlet (to order)
6/ 4 PTEF tube or brass
PP for
for 1/4 inch tube.
Exclusive
cable gland
Flow guide tube
155
Approx. L
12
Approx. 20
130
67
20
34
Gas inlet
Oxygen detector
3
5
L (m)
0.3
0.5
MASS
2.7
Approx (kg)
1
7
ZFK2R
4-5C3
5
7
1
Z
8–15
MTG. holes
0.75 1.0
4.1
3.3
4-5A3
5
7
1
6-M5
detector side
Gas outlet
Cord 11th
ZFK2R
4.8
L=
(to order)
Flow guide tube (with blow-down nozzle)
155
Approx. L
18
Approx. 20
130
67
60.5
67
3
Gas inlet
6-M5
detector side
Gas outlet
Oxygen detector
Blow down air inlet
Cord 11th
3
L (m)
0.3
Mass
3.0
Approx (kg)
5
7
1
Z
0.5 0.75 1.0
3.8
4.8
5.7
Select an upper port to
avoid the condensation
in the piping
8–15
MTG. holes
4-Rc1/4
with plug int.thd.
L=
(to order)
9
ZRN, ZFK
5
14
4 - Rc 1/4
with plug int thd.
180
Flow guide tube (for high particulate)
Blow down air inlet
Select an upper port to
avoid the condensation
in the piping
8 - 19
MTG. hole
67
29.
5
Flange: JIS 5K80A FF
ZFK MTG. position
185
Packing
4 - 6 D8
Gas outlet
800
50
ZFK2R
30
Gas flow
Tube (50A SCH40)
5
14
4 - Rc 1/4
with plug int thd.
180
Flow guide tube (for high particulate with cover)
Blow down air inlet
8 - 19
MTG. holes
Select an upper port to
avoid the condensation
in the piping
67
29
.5
Flange: JIS 5K80A FF
(175)
ZFK MTG. position
50
Packing
800
(790)
Gas outlet
Gas flow
10
Protection tube (65A SCH40)
(38)
Tube (50A SCH40)
30
ZFK2R
4 - 6 E8
Detector (ZFK5)
80
67
Proof cover (to order)
Approx. 130
Approx. 65
45.4
Approx. 118
36
5
26
Filter
EXTERNAL CONNECTION DIAGRAM
Thermocouple
Element output
60
°
Heater
Approx. 35
1
3
2
4
Black / White/ red / White/
2-core wire
5
6
yellow / blue
4-core wire
Exclusive-special cable
Ref. air inlet (to order)
NPT1/8 or Rc1/8
6- 6
Calibration gas inlet (to order)
6/ 4 PTEF tube or brass
PP for
for 1/4 inch tube.
Exclusive cable gland
Flow guide tube
155
12
Approx. 40
Approx. L
130
3
67
67
60.5
Gas inlet
4-5B3
5
7
1
ZFK5R
4-5C3
5
7
1
6-M5
detctor side
Gas outlet
Oxygen detector
ZFK5R
Cord 11th
3
L (m)
0.3
MASS
3.3
Approx (kg)
5
1
7
0.5 0.75
4.5 6.1
Z
8- 15
MTG. holes
1.0
7.6
L=
(to order)
Flow guide tube (with blow - down nozzle)
155
18
Approx. 20
Approx. L
130
3
67
60.5
67
Gas inlet
6-M5
detctor side
Gas outlet
Blow down air inlet
8- 15
MTG. holes
Oxygen detector
Cord 11th
3
L (m)
0.3
MASS
3.3
Approx (kg)
5
7
0.5 0.75
4.2 5.7
1
Z
1.0
7.4
L=
(to order)
Select an upper port to
avoid the condensation
4- Rc 1/4
in the piping
with plug int. thd.
11
ZRN, ZFK
5
14
180
Flow guide tube (for high particulate)
4 - Rc 1/4
with plug int thd.
Blow down air inlet
8 - 19
MTG. hole
Select an upper port to
avoid the condensation
in the piping
67
29.
5
Flange: JIS 5K80A FF
ZFK MTG. position
185
Packing
4 - 6 D8
Gas outlet
800
50
ZFK5R
30
Gas flow
Tube (50A SCH40)
5
14
4 - Rc 1/4
with plug int thd.
180
Flow guide tube (for high particulate with cover)
Blow down air inlet
8 - 19
MTG. holes
Select an upper port to
avoid the condensation
in the piping
67
29
.5
Flange: JIS 5K80A FF
(175)
ZFK MTG. position
50
Packing
(800)
790
Gas outlet
Gas flow
12
Protection tube (65A SCH40)
(38)
Tube (50A SCH40)
30
ZFK5R
4 - 6 E8
OUTLINE DIAGRAM
(Unit:mm)
26.5
482.6
207.5
37.2
465.1
209.2
205
7
146.1
7
ZRN
Supprot
482.6
15
400
364
17- 23 with
rubber bush
Panel cutout dimensions
465.1
216.5
146.1
393
386.4
34
391.6
4-M6 or
7mm
Panel flush mounting/19 inch rack mounting type
400
15
364
209.2
221
ZRN
Pipe stand
(additional specification)
17- 23 with
rubber buch
159
60
Mounting hole 4-M6
1000
76.3
4-
130
18
386.4
220
350
Pipe stand type
13
ZRN, AFK
EXTERNAL CONNECTION DIAGRAM
I/O board
Average-value
analog outut
External transmission
(RS-458)
AV3
1
21
AV3
AV2
2
22
AV2
AV1
3
23
AV1
DXA
4
24
DXA
5
25
FG (for external tranmission)
26
Contact imput common
7
27
(UA) common
8
28
Maintenace
9
29
Valve-maintenance common
10
30
H/L alarm common
11
31
Average-value output 2 H/L alarm
Calibration disable
Contact
input
6
Calibration start
Unit alarm (UA)
Span valve
Contact
output
Contact
output
Heater board Ch3/Ch4 to Ch6/Ch7 to Ch9
Zero valve
Average-value output 3
H/L alarm
Average-value output 1
H/L alarm
Contact
output
H/L alarm 3,6,9
1
21
H/L alarm common
H/L alarm 1,4,7
2
22
H/L alarm 2,5,8
Contact
output
Blow valve 3,6,9
3
23
Blow valve common
Blow valve 1,4,7
4
24
Blow valve 2,5,8
Calibration gas
valve 3,6,9
Calibration gas
valve 1,4,7
AC power supply
5
25
Calibration gas valve
common
6
26
7
27
3,6,9
8
28
2,5,8
9
29
1,4,7
10
30
11
31
Contact
output
Detector
heater
Detector ground
Analog
output
Detector
input
Heater common
Detector ground
Powe supply
Main board Ch1 to Ch3/Ch4 to Ch6/Ch7 to Ch9
Contact
output
Calibration gas
valve 2,5,8
AC power supply
AC
Unit alarm 3,6,9
1
21
Unit alarm common
Unit alarm 1,4,7
Oxygern concentration output 3,6,9
Oxygern concentration output 2,5,8
Oxygern concentration output 1,4,7
Thermocouple
signal 3,6,9
2
22
Unit alarm 2,5,8
3
23
Oxygen concentration output 3,6,9
4
24
Oxygen concentration output 2,5,8
5
25
Oxygen concentration output 1,4,7
6
26
Thermocouple signal 3,6,9
Zirconia signal 3,6,9
7
27
Zirconia signal 3,6,9
Thermocouple
signal 2,5,8
8
28
Thermocouple signal 2,5,8
Zirconia signal 2,5,8
9
29
Zirconia signal 2,5,8
Thermocouple
signal 1,4,7
Zirconia signal 1,4,7
10
30
Thermocouple signal 1,4,7
11
31
Zirconia signal 1,4,7
AC
AC power supply
E
E
Analog output
Ejector (ZTA)
4-M16 bolt
190
30
42.7
Approx. L
170
P
217
Gas inlet
L [mm] 500 750 1000 1500
40
Detector
(ZFK2)
Viewed from P direction
8
25
105
140
(JIS 10K65ARF)
35
Blow -down air inlet
(Rc1/4)
Ejector air inlet
(Rc1/4)
50
45
Cable gland
(A15C)
EXTERNAL CONNECTION DAIAGAM
1
Ejector air
outlet
(Rc1/4)
40
2
3
AC
power supply
4
5
Heater temp.drop alam
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan