FUJI EDSX5-75

FCX – C SERIES
ABSOLUTE PRESSURE TRANSMITTER
DATA SHEET
FHH, FKH
The FCX –C absolute pressure transmitter accurately
measures absolute pressure and transmits proportional 4
to 20mA signal.
The transmitter utilizes the unique micromachined capacitive silicon sensor with state-of -the-art microprocessor technology to provide exceptional performance and functionality.
FEATURES
1. High accuracy
0.2% accuracy for all calibrated spans is the standard
feature for all AP models covering 13 to 3000kPa{0.13 to
30bar} high pressure range. Fuji's micro-capacitance silicon
sensor assures this feature for all suppressed calibration
ranges without additional adjustment.
2. Minimum inventory
Electronics unit, communication module, local indicators
and electronics housing are interchangeable among all
FCX –C models.
3. Replaceable Communication Module
Fuji micro-electronics manufacturing technology offers
replaceable communication module that makes FCX-AC
transmitter very unique design. In case of change in
communication protocol all that needs to be done is just
replace the module and the transmitter gets upgraded to
the new version.
4. Fuji/HART bilingual communication module
The communication module is "bilingual" to speak both
Fuji proprietary protocol and HART. Any HART compatible
devices can communicate with FCX-A/C series transmitters.
5. Application flexibility
Example features that render the FCX –C suitable for
almost any process applications includes:
— Analog indicator at either the electronics side or terminal side
— Full range of hazardous location approvals
— Built-in RFI filter and lightning arrester
— 4 12 -digits LCD meter
— The maximum span of each sensor can be converted to
in different units using below factors.
SPECIFICATIONS
Functional specifications
Type:
Model FHH:
Model FKH:
Service:
Span, range, and
4 to 20mA
4 to 20mA with digital signal
Liquid, gas, or vapour
overrange limit:
Span limit [kPa abs] {bar abs}
Type
F H 02
F
H 03
F
H 04
Min.
Max.
FKH
FHH
FHH/FKH
13
8.125
130
{0.13} {0.08125} {1.3}
50
31.25
500
{0.5}
{0.3125}
{5}
300
187.5
3000
{3}
{1.875}
{30}
Range limit
[kPa abs]
{bar abs}
0 to
{0 to
0 to
{0 to
0 to
{0 to
Overrange
limit
[MPa] {bar}
130
1.3}
500
5}
3000
30}
0.5
{5}
1.5
{15}
9
{90}
1MPa=103KPa=10bar=10.19716kgf/cm2=145.0377psi
1KPa=10mbar=101.9716mmH2O=4.01463inH2O
Output signal:
Model FHH: 4 to 20mA DC 2-wire
Model FKH: 4 to 20mA DC with digital signal superimposed on the 4 to 20mA signal.
Power supply:
Transmitter operates on 10.5V to 45V DC
at transmitter terminals.
10.5V to 32V DC for the units with optional
arrester
EDSX5-75
Date
Feb. 2, 1999
FHH, FKH
Load limitations: see figure below
[Ω]
Load resistance
2000
1533Ω
1500
1000
600Ω
Operating
area
500
24V
250
10
16.1
10.5V
Power voltage
45V
50
[V]
Note: For communication with FXW, min. of 250 Ω required.
Hazardous locations:
Authorities
BASEEFA
Factory
Mutual
CSA
SAA
Flameproof
Ex ds IIC T5, T6
Class I II III
Div. 1
Groups B thru. G
Class I II III
Div. 1
Groups C thru. G
Ex d II C T5, T6
IP 66/67
Intrinsic safety
EEx ia IIC T4, T5
Class I II III
Div. 1
Groups A thru. F
Class I II III
Div. 1
Groups A thru. G
Ex ia II C T5, T6
IP 66/67
Type N
Nonincendive
Ex N II T5
Class I II III
Div. 2
Groups A thru. G
Class I II III
Div. 2
Groups A thru. G
Ex n II C T5, T6
IP 66/67
Zero/span adjustment:
Model FHH: Zero is adjustable from an external adjustable screw.
The adjustable screw can also function
to adjust span when MODE SWITCH (located on the electronics unit) is in the span
mode. INHIBIT mode to disable the adjustable screw is also available.
Model FKH: Zero and span are adjustable either from
the HHC. Zero is also adjustable externally from the adjustable screw.
Damping:
Adjustable electrical damping
Model FHH: The time constant is adjustable to 0, 0.3,
1.2, 4.8, or 19.2 seconds.
Model FKH: The time constant is adjustable between
0 to 38.4 seconds. (9 steps)
Zero elevation/suppression:
Zero may be elevated within the specified
range limit of each sensor model.
Normal/reverse action:
Model FHH: Selectable by moving a jumper pin located
on the electronics unit.
Model FKH: Configurable from HHC.
Indication:
Analog indicator or 4 1 -digit LCD meter,
2
as specified.
Burnout direction:
Output hold
Output 21.6mA selectable
Output 3.8mA
Model FHH: Unless otherwise specified, the output is
in hold position.
Model FKH: Selectable from HHC.
}
2
Loop-check output:
Model FHH: Transmitter can output constant signal of
4mA, 12mA, or 20mA if MODE SWITCH
is set to the loop check mode.
Model FKH: Transmitter can be configured to provide
constant signal 3.8mA through 21.6mA
by HHC.
Temperature limit: Ambient: –40 to +85°C
(–20 to +80°C for LCD indicator)
(–40 to +60°C for arrester option)
For explosionproof units (flameproof or
intrinsic safety), ambient temperature
must be within the limits specified by
each standard.
Process: –40 to +85°C for silicone fill
sensor
Storage: –40 to +90°C
Humidity limit: O to 100% RH
Communication: (Model FKH only)
With HHC (Model FXW, consult Data
Sheet No. EDS8-47), following information can be remotely displayed or reconfigured.
Items
Tag No.
Model No.
Serial No.
Engineering unit
Range limit
Measuring range
Damping
Output mode
Burnout direction
Adjustment
Output adjust
Data
Self diagnoses
Printer
External switch lock
Transmitter display(*)
Display
Set
v
v
v
v
v
v
v
v
v
v
—
v
v
—
v
v
v
v
—
v
—
v
v
v
v
v
v
—
—
—
v
v
Note: (*) HHC's version must be more than 5.0 (or
FXW
1– 2), to use this function.
Performance specifications
Accuracy rating: (including linearity, hysteresis, and repeatability).
For spans greater than 1/10 of URL: ± 0.2% of span
For spans below 1/10 of URL (Model FKH only):
0.1 x URL
) % of span
span
Stability:
± 0.2% of upper range limit (URL) for 24
months
(In case of 6th digit code "3", "4")
Temperature effect:
Effect per 28°C change between the
limits of –40°C and +85°C
URL
Zero shift: ± (0.25 span )%/28°C
URL
Total effect: ± (0.25 + 0.25 span )%/28°C
± (0.1 + 0.1
Overrange effect: Zero shift, 0.3% of URL for any overrange
to maximum limit
Supply voltage effect:
Less than 0.05% of calibrated span per
10V
RFI effect:
Less than 0.2% of URL for the frequencies of 20 to 1000MHz and field strength
30V/m when electronics covers on.
(Classification: 2-abc: 0.2% span per
SAMA PMC 33.1)
Step response:
Time constant. 0.2 s
Dead time: about 0.3 s
(without electrical damping)
Mounting position effect:
Zero shift, less than 0.1kPa{1mbar} for a
10° tilt in any plane.
No effect on span. This error can be corrected by adjusting zero.
Dielectric strength:
500V AC, 50/60Hz 1 min., between circuit
and earth.
Insulation resistance:
More than 100MΩ at 500V DC.
Turn-on time:
4 sec
Internal resistance for external field indicator:
12Ω or less
Physical specifications
Electrical connections:
G1/2, 1/2-14NPT, Pg13.5, or M20 x 1.5
conduit, as specified.
Process connections:
1/2-14 NPT, 1/4-18NPT, Rc1/2 or Rc1/4
as specified.
Material
code
(7th digit Process cover
in "Code
symbols")
V
Diaphragm
Wetted
sensor body
Vent/drain
316 stainless 316L stainless 316 stainless 316 stainless
steel
steel
steel
steel
Process-wetted parts material:
Non-wetted parts material:
Electronics housing: Low copper die-cast
aluminum alloy (standard), finished
with epoxy/polyurethane double coating, as specified.
Fill fluid: Silicone oil
Mounting bracket: 304 stainless steel
Environmental protection:
IEC IP67 and NEMA 4X
Mounting:
On 60.5mm (JIS 50A or 2B) pipe using
mounting bracket, direct wall mounting,
or direct process mounting.
Mass{weight}:
Transmitter approximately 1.9kg without
options.
Add; 0.5kg for mounting bracket
0.8kg for indicator option
Optional features
Indicator:
A plug-in turnable analog indicator (1.5%
accuracy) can be housed in the electronics
compartment or in the terminal box of the
housing.
An optional 4 1 digits LCD meter is also
2
available.
Arrester:
A built-in arrester protects the electronics
from lightning surges.
Lightning surge immunity: 4KV (1.2×50µs)
Degreasing:
Process-wetted parts are cleaned, but the
fill fluid is standard silicone oil. Not for
use for oxygen or chlorine measurement.
NACE specification:
Metallic materials for all pressure boundary parts comply with NACE MR-01-75.
Customer tag:
A stainless steel tag for customer tag data
is wired to the transmitter.
ACCESSORIES
Hand held communicator:
(Model FXW, refer to Data Sheet No.EDS
8-47)
Communication module: (Standard for model FKH)
When using this module for model FHH,
remote setting function becomes available
Remark: When the communication module is
connected, the operation mode of external zero/span adjustable screw is
changed to zero adjustment only.
3
FHH, FKH
CODE SYMBOLS
1 2 3 4 5 6 7 8
F
H
0
9 10 11 12 13
1
14 15
Description
0
Type
4 to 20 mA, Output type
4 to 20 mA with digital signal, Outout type
H
K
Connections
S
T
V
W
Process conn.
Conduit conn.
1/2-14NPT
G1/2
1/2-14NPT
Pg 13.5
M20 x 1.5
Span limit
[kPa abs] {bar abs}
13/8.125…130 {0.13/0.08125…1.3}
50/31.25…500 {0.5/0.3125…5}
300/187.5…3000 {3/1.875…30}
2
3
4
Material
Process cover
316 stainless steel
V
Diaphragm
316L stainless steel
Indicator and arrester
Indicator
None
Analog, 0 to 100% linear scale
Analog, custom scale
Analog, double scale
None
Analog, 0 to 100% linear scale
Analog, custom scale
Analog, double scale
Digital, 0 to 100%
Digital, custom scale
Digital, 0 to 100%
Digital, custom scale
A
B
D
J
E
F
H
K
L
P
Q
S
Wetted cell body
316 stainless steel
Arrester
None
None
None
None
Yes
Yes
Yes
Yes
None
None (Model FKH only)
Yes
Yes (Model FKH only)
Approvals for hazardous locations
None (for ordinary locations)
FM, Flameproof (or explosionproof) ( Available for 4th digit code "T")
CSA, Flameproof (or explosionproof) ( Available for 4th digit code "T")
BASEEFA, Flameproof (Conduit seal)
BASEEFA, Flameproof (Cable gland seal) (Conduit connection G 1/2 only)
FM, Intrinsic safety and Nonincendive
CSA, Intrinsic safety and Nonincendive
CENELEC, Intrinsic safety
CENELEC, Intrinsic safety and BASEEFA, Type N
SAA Flameproof (Conduit seal) ( Available for 4th digit code "S, T, W")
SAA Intrinsic safety ( Available for 4th digit code "S, T, W")
SAA Type – N (non-speaking) ( Available for 4th digit code "S, T, W")
A
D
E
M
N
H
J
K
P
R
T
Q
Mounting bracket
None
Yes (stainless steel)
A
C
Optional specification
Stainless tag
None
Yes
Y
B
Special applications and fill fluid
Y
G
N
Y
A
B
C
4
Treatment
Filled liquid
None (standard)
Degreasing
NACE specification
Silicon oil
Silicon oil
Silicon oil
Process adaptor
None (1/2 -14NPT)
Rc1/4
Rc1/2
1/4-18NPT
OUTLINE DIAGRAM (Unit:mm)
Fastener
(Flameproof EX. only)
2 holes
9 dia.
In case of 4th digit code
"S, V, W"
U-bolt M8
170
110
50
108
f80
UP
88
DOWN
Zero/span adj. screw
MTG. PIPE
JIS 50A (2B)
"B"
Conduit conn.
196
Earth terminal
(both side)
See table 1
51
"C"
Conduit conn.
(27)
71
Details of "A"
Details of "B"
(Conduit Conn.)
(Press. Conn.)
Details of "C"
φ13
f80
φ15
214
F
φ18
G
(27)
16
E
17
70
D
1/2-14NPT
"A" Press. Conn.
See table 1
85
"D" Adapter (option)
See table 1
Conduit conn.
E
F
D
4th of Code
symbols
Earth
terminal
S
G1/2
17 8
M4
T
1/2-14NPT
16 5
No. 8-32UNC
V
Pg13.5
W
M20x1.5
8 4.5
16 5
G1/2
See table 1
CONNECTION DIAGRAMS
+
–
CK+
CK–
M4
+
M4
–
Table
<Optional stainless steel tag>
S
T
F
H V
K
W
75
H
V1
20
Note *: Cable gland is supplied in case of flameproof packing type.
φ11 cable is suitable.
5
FHH, FKH
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TN510412. The applicable standards used to demonstrate
compliance are :EMI (Emission)
EN50081-1 : 1992
Test item
Frequency range
Basic standard
Applicable Electromagnetic Radiation
Disturbance
30-1000MHz
EN55022
Class B
EMS (Immunity)
EN50082-1 : 1992
No.
Test item
Test
specification
Basic
standard
Performance
criteria
1
Electrostatic
discharge
8kV (Air)
IEC 801-2:1984
B
2
Radio-frequency 27-500MHz
electromagnetic 3V/m
IEC 801-3:1984
field.
(Unmodulated)
A
3
Fast transients
common mode
0.5kV, 5/50
(Tr/Th) ns
5kHz Rep.
IEC 801-4:1988
B
"LVD - The transmitter is not covered by the requirements
of the LVD standard."
Head office
11-2 Osaki 1-chome, Shinagawa-ku, Tokyo, 141-0032 Japan
Phone: 81-3-5435-7111
http://www. fujielectric.co.jp/eng/sg/KEISOKU/welcome.htm
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo 191-8502 Japan
Phone: 81-42-585-6201, 6202
Fax: 81-42-585-6187, 6189
Information in this catalog is subject to change without notice.
Printed in Japan