MPSIND HBU60-106

HBU60 SERIES
60W Open frame type medical power supplies for health care applications
Description:
The HBU60 series of compact, open frame constructed, AC/DC
switching mode power supplies provide 60 Watts of continuous
output power .They are suited for use in health care applications.
All models meet FCC Part-18 class B and CISPR-11 EN55011 class
B emission limits and are designed to comply with UL/c-UL(UL
60601-1:2ndEdition) ,TUV/T-mark(EN 60601-1:2ndEdition) and new
CE requirements. All units are 100% burned in and tested.
Features:
Wide Operating voltage 90 to 264 VAC,47 to 63 Hz
Single output
Input connector mates with molex housing 09-50-3031/35977-0300
and molex 2478/ 35922 series crimp terminal.
Output connector mates with molex housing 09-50-3061/35977-0600
and molex 2478/ 35922 series crimp terminal.
Input surge current, over voltage and over load protection
Size: 2"x4”x1”
Class I
2 year warranty
Safety approvals :
CB
Electrical characteristics:
Sym.
Parameter
fin
Po
Vo
Io
Iil
Iih
Safety Approvals Input
Voltage Range
Operate Voltage Range
Input Frequency
Output Power Range
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
I rl
Low Line Inrush Current
Vin
I rh
Eff
REG-i
REG -o
OVP
OCP
Ttr
Thold
Ts
Vp-p
Ilk
TC
High Line Inrush Current
Efficiency
Line Regulation
Load Regulation
Over Voltage Protection
Over Current Protection
Time of Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise (Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
Test Conditions
Io=Full load,
Io=Full load,
Io=Full load,
Vin=115VAC
Io=Full load,
Vin=230VAC
Io=Full load,
Io=Full load
Vin=230VAC
Vin=115VAC
Vin=230VAC
25 C,Cool start,
IEC 60601-1
Min. Typ. Max. Unit
100
Vin=90 to 264 VAC
EN 60601-1
240
264
90
47
63
60
0
See rating Chart
See rating Chart
1.62
0.72
26
30
25 C,Cool start,
43
47
Vin=230VAC
85
VAC
VAC
Hz
W
V
A
A
A
A
A
Io=Full load to Half Load, Vin=100VAC
12
16
Io=Full load, Vin=110VAC
Io=Full load, Vin=100VAC
0.3
1
2
Full load, Vin=90VAC
1
Io=Full load, Vin=240VAC
0.1 0.3
All output
-0.04
0.04
%
%
%
%
%
mS
mS
S
%
mA
%/ C
Min. Typ. Max.
Unit
40
70
C
1
5
132
150
4
112
110
Environmental :
Sym.
Parameter
T e s t Conditions
Toper
Operating Temperature
Tstg
Storage Temperature
-40
85
C
Ho
Operating Humidity
0
95
%
Hr
Storage Humidity
0
75
%
MTBF
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd
Derate linearly from 100% load at 40 C to 50% load at 70 C
0
0.1M
Hrs
HBU60 SERIES
60W Open frame type medical power supplies for health care applications
Safety specifications:
Sym.
Parameter
Dielectric Withstanding Voltage
Vps for
Primary to secondary
Dielectric
Voltage
Vpg for Primary Withstanding
to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Min. Typ. Max.
Unit
Primary to secondary
5656
VDC
Primary to ground
2828
VDC
50
B
B
M
CLASS
CLASS
T e s t Conditions
Test Voltage=500VDC
Vin=220VAC
Vin=110VAC
Output voltage and current rating chart (Single output) :
Model Number
Output Voltage
Output Current
Total Regulation
Maximum
Output Power
HBU60-105
012 ~ 13 VDC
5.00 ~ 4.61 A
5%
60W
HBU60-106
013 ~ 16 VDC
4.61 ~ 3.75 A
5%
60W
HBU60-107
16 ~ 21 VDC
3.75 ~ 2.85 A
5%
60W
HBU60-108
21 ~ 27 VDC
2.85 ~ 2.22 A
3%
60W
HBU60-109
27 ~ 33 VDC
2.22 ~ 1.81 A
3%
60W
HBU60-110
33 ~ 40 VDC
1.81 ~ 1.50 A
3%
60W
Mechanical specifications :
Pin Char t
PIN
MODEL
HBU60-1XX
1
2
3
4
5
6
OUT OUT OUT RTN RTN RTN
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 140gs approx.
3. Input connector mates with Molex housing
09-50-3031/35977-0300 and Molex 2478/ 35922
series crimp terminal.
4. Output connector mates with Molex housing
09-50-3061/35977-0600 and Molex 2478/ 35922
series crimp terminal.