MPSIND MPU63-110

MPU63 SERIES
63W Desk Top Switching Power Supplies For Medical Equipment.
Description:
The MPU63 series of AC/DC switching mode power supplies
provide 63 Watts of continuous output power. All supplies are
UL94V-1 min compliant. They are suited for use in hospital
instrument and many other applications. All models meet FCC
Part-18 class B and CISPR-11 EN55011 class B emission Limits
and are designed to comply with UL/c-UL(UL 60601-1:2ndEdition)
,TUV/T-mark (EN 60601-1:2ndEdition), and new CE requirements.
All units are 100% burned in and tested.
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single Output
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Output Voltage Protection(Crowbar Design)
Energy Star 2.0, Efficiency level V
Class I
3 year warranty
Safety Approvals :
EN 60601-1
S&E
IEC 60601-1
PS
E
RT
PS
E
Agree to apply for the PSE
if order on hand
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Vin
Safety Approvals Input
Voltage Range
fin
Po
Vo
Io
Operate Voltage Range
Input Frequency
Output Power Range
Output Voltage Range
Output Current Range
Vin=90 to 264VAC
Iil
Iih
Input Current (Low Line)
Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
I rl
Low Line Inrush Current
I rh
High Line Inrush Current
Io=Full load, 25 C , Cool start,
Vin= 115VAC
Io=Full load, 25 C , Cool start,
Vin= 230VAC
Io=Full Load, Vin=230VAC
Eff
REG-i
REG -o
OVP
OCP
Ttr
Thold
Ts
Vp-p
Ilk
Pno
TC
Efficiency
Line Regulation
Load Regulation
Over Voltage Protection
Min. Typ. Max. Unit
100
240
VAC
90
264
63
VAC
Hz
W
V
A
47
63
0
See rating chart
See rating chart
77
Io=Full Load
Vin=230VAC
1.62
0.72
A
A
28
38
A
45
55
A
85
88
0.5
3
1
7
132
%
%
%
112
Over Current Protection
110
150
Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
4
Hold-Up Time
Io=Full Load, Vin=110VAC
16
Io=Full Load, Vi n= 10 0V AC
Start Up Time
0.3
1
2
Ripple & Noise (Peak to Peak) Full Load, Vin=100VAC
0.5
1
Safety Ground Leakage C urrent Io = Fu ll Load, Vin=240VAC
0.1
No-Load Power Consumption
No load,Vin=230VAC
0.3 0.4 0.5
Temperature Coefficient
-0.04
All output
0.04
%
%
mS
mS
S
%
mA
W
%/ C
Environmental :
Sym.
Toper
Parameter
T e s t Conditions
Operating Temperature
Min. Typ. Max.
Unit
50
70
C
Storage Temperature
-40
85
C
Operating Humidity
0
95
%
0
75
Tstg
Ho
Hr
Storage Humidity
MTBF
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd
Derate linearly from 100% load at 50 C to 50% load at 70 C
0
0.1M
%
Hrs
MPU63 SERIES
63W Desk Top Switching Power Supplies For Medical Equipment.
Safety Specifications:
Parameter
Sym.
Test Conditions
Dielectric Withstanding Voltage
for Primary to secondary
Dielectric Withstanding Voltage
Vpg for
Primary to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Vps
Min. Typ. Max.
Unit
Primary to secondary
5656
VDC
Primary to ground
2828
VDC
50
B
B
M
CLASS
CLASS
Test Voltage = 500VDC
Vin=220VAC
Vin=110VAC
Output Voltage And Current Rating Chart (Single Output) :
Model Number
MPU63-102
MPU63-103
MPU63-104
MPU63-105
MPU63-106
MPU63-107
MPU63-108
MPU63-109
MPU63-110
*
*
*
*
*
*
Output Voltage
5
7
9
12
15
18
24
30
36
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
Output Current
9.00
7.85
6.44
5.25
4.20
3.50
2.62
2.10
1.75
Total Regulation ①
A
A
A
A
A
A
A
A
A
7%
7%
5%
5%
5%
5%
3%
3%
3%
M aximum
Output Power
45W
55W
58W
63W
63W
63W
63W
63W
63W
MPU63-103~109 had been approved by PSE .
Mark “
①
*
” means approved by CEC level V and the output cable must be using as below.
MPU63-102~103 are required to use AWG#16 / 5C/4FT output cable.
MPU63-104~105 are required to use AWG#16 /2C/ 4FT output cable.
MPU63-106~107 are required to use AWG#16 /2C/ 4FT output cable.
MPU63-108~110 are required to use AWG#18 /2C/ 6FT output cable.
The regulation and efficiency will be changed by modified output cable.
Mechanical Specifications:
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 510-560gs approx.
3. Optional output connector:
See page Appendix.