ASM-SENSOR 3637-20

2000
3636-20 CLAMP LOGGER
3637-20 AC VOLTAGE LOGGER
Environmental Measuring
Instrument
Easy clamp current recording and
convenient, low-cost voltage recording
Extended recording of AC current and
voltage transients
● D a t a b a c k u p w h e n b a t t e r i e s a re e x h a u s t e d o r re p l a c e d
● I n s t a n t a n e o u s / a v e r a g e re c o rd i n g
( a v e r a g e o v e r re c o rd i n g i n t e r v a l )
● G r a p h i c d i s p l a y o f re c o rd e d d a t a t h ro u g h a c o m p u t e r
● Equipped with alarm output (3636-20)
http://www.hioki.co.jp/
ISO14001
JQA-E-90091
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
3636-20
CLAMP LOGGER
500A AC
3637-20
AC VOLTAGE LOGGER
600V AC
Analyze and Process Data on a
Personal Computer
The 3911-20 Communication Base is used to transfer data to a personal
computer. The 3911-20 accommodates up to 16,000 points of 16-channel data
or 32,000 points of 8-channel data. It captures data from multiple loggers, then
transfers data to the computer where it can be analyzed and managed.
3911-20
LOGGER
Functions
Infrared optical communication
Approximately 500 data points/second
RS-232C
Approximately 1000 data points/second
(Logger and 3911 must be in direct contact for
communication.)
Communication software included
with the 3911-20
Settings for communication software
Current time, recording interval, recording start,
recording method, comment, recording mode (3636-20/3637-20).
Alarm and channel selection (3636-20 only)
Compatible OS: Windows 95/98/NT4.0
● Graph display
● Measurement data tabulation
● Printing (data, graphs)
● Scaling
● File saving (in proprietary
format or text format)
Windows is a registered trademark of Microsoft Corporation
■ 3636-20/3637-20 Specifications
■ 3636-20/3637-20 Common Specifications
Display
* 50A or 500A range when using the 9650, and 500A range when using the 9651
: Measured value, recording status, recording interval, battery
condition, unit, recording in progress, preset active, average
value recording, maximum value, minimum value,
alarm (3636-20 only)
: Started manually or at preset time
Recording start
: Manual stop or until memory is full
Recording finish
: 1/2/5/10/15/20/30 seconds, 1/2/5/10/15/20/30/60 minutes
Recording interval
: One-time: Recording stops when memory becomes full.
Recording methods
Endless: Oldest data is overwritten when memory becomes full.
: Instantaneous value recording, average value recording
Recording modes
(average over the recording interval)
: Infrared optical communication (requires the 3911)
Interface
: LR03 (AAA) × 4 alkaline batteries
Power supply
: 0.1 VA
Maximum rated power
: About 1 year (with a recording interval of 1 minute and
Continuous use time
instantaneous value recording)
About 1 month (with average value recording)
: Approx. 57.5(W) × 86.5(H) × 30.0(D) mm, approx. 130 g
Dimensions/mass
: Indoors at an altitude of no more than 2000 m
Operating environment
: 0 to 50˚C, less than 80% rh (no condensation)
Ambient use conditions
Ambient storage conditions : -10 to 60˚C, less than 80% rh (no condensation)
: EMC EN61326-1 : 1997+A1 : 1998
Applicable standards
Safety EN61010-1 : 1993+A2 : 1995
Overvoltage category I
(anticipated overvoltage: 330V), pollution index 2 (3636-20)
Overvoltage category III
(anticipated overvoltage: 6000V), pollution index 2 (3637-20)
■ 9650/9651 Specifications
■ 3911-20 COMMUNICATION BASE Specifications
3636-20 CLAMP LOGGER
3637-20 AC VOLTAGE LOGGER
Input
AC current
AC voltage
Number of
input channels
2 channels
1 channel
Data recording
capacity
32,000 data points (1 ch)
16,000 data points (2 ch)
32,000 data points
Measurement
range
0.00 to 500.0 Arms.sin
(50A/500A, 2 ranges)
0.0 to 600.0 Vrms.sin
±1% rdg. ±5 dgt. (logger only)
±2.5% rdg. ±8 dgt. (logger + sensor)*
±1% rdg. ±5 dgt.
Accuracy
(50/60 Hz, 23±5˚C)
Alarm output
ON when set upper/lower limits exceeded
None
Measurement
system
True RMS calculation
True RMS calculation
LR03 (AAA) alkaline batteries × 4,
9632 connection cord (for alarm output)
LR03 (AAA) alkaline batteries ×4,
9639 connection cord (for input)
3911-20 Communication Base,
9650/9651 Clamp-on Sensor
3911-20 Communication Base
Accessories
Options
×
Rated primary
current
Rated secondary
current
9650 CLAMP ON SENSOR
9651 CLAMP ON SENSOR
AC100A
AC500A
AC100mA
AC500mA
±1.5% rdg. ±0.03%f.s.
(where f.s. is rated primary current)
±1.5% rdg. ±0.03%f.s.
(where f.s. is rated primary current)
Frequency
response
40 Hz to 1 kHz, within ±8%
40 Hz to 1 kHz, within ±3%
Maximum
rated input
130A continuous (45 to 66 Hz)
600A continuous (45 to 66 Hz)
AC 300Vrms or less
(insulated conductor)
AC 600Vrms or less
(insulated conductor)
φ 15 mm or less
φ 46 mm or less
Approx. 3 m
Approx. 3 m
Accuracy
(50/60 Hz, 23±5˚C)
Circuit voltage
Measurable
conductor
diameter
Cord length
Dimensions/
mass
Approx. 46(W) × 135(H) × 21(D) mm, Approx. 77(W) × 151(H) × 42(D) mm,
approx. 200 g
approx. 340 g
: Max 16,000 data points × 16 ch / 32,000 data points × 8 ch
: Logger ⇔ 3911-20: by infrared optical communication (with
units in close contact)
3911-20 ¤ PC: by RS-232C connection
: LR03 (AAA) × 4 alkaline batteries
Power supply
:
0.2
VA
Maximum rated power
: Approx. 69(W) × 92(H) × 36(D) mm, approx. 150 g
Dimensions/mass
:
0
to
40.0˚C, max 80% rh (no condensation)
Ambient use conditions
Ambient storage conditions : -10.0 to 50.0˚C, max 80% rh (no condensation)
:
PC
communication
software
Accessories
Recording capacity
Communication type
9650
9651
9639
9632
CLAMP ON SENSOR
CLAMP ON SENSOR
CONNECTION CABLE
CONNECTION CABLE
■ Options
3636-20 CLAMP LOGGER
3637-20 AC VOLTAGE LOGGER
The 3636-20 CLAMP LOGGER cannot be used for measurement
by itself. A clamp-on sensor (sold separately) is required.
3911-20 COMMUNICATION BASE
9650 CLAMP ON SENSOR (100A)
9651 CLAMP ON SENSOR (500A)
9632 CONNECTION CABLE (included with the 3636-20, 1 m)
9639 CONNECTION CABLE (included with the 3637-20, 3 m)
9637 RS-232C CABLE (9 pin-to-9 pin, null modem, 1.8 m)
9638 RS-232C CABLE (9 pin-to-25 pin, null modem, 1.8 m)
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL+81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
ASM GmbH
Am Bleichbach 18 - 22
85452 Moosinning
All information correct as of Apr. 7, 2000. All specifications are subject to change without notice.
■Inter net HIOKI website h t t p : / / w w w . h i o k i . c o . j p /
F3636E1-04B-05P
Printed in Japan
2001
3638-20 LEAK LOGGER
DATA LOGGER
Simple Leak Current Recording with Clamp On Sensors. Easy Logging at Low Cost
Recording and Monitoring Leak Current
●Select from three types of clamp on leak current sensors
●Compact, light weight (130g), two-channel recording
●High-capacity recording (32,000 data points)
Recording interval: 15 steps from 1 second to 60 minutes
●Accurate true effective value calculation even for distorted waveforms
(100/1000 mA)
●Maximum or average value recording (max./ave. within recording interval)
●View recording data graphically on a PC
●Data is protected during battery discharge and replacement
Model
9657
Model 9658
Model
ISO14001
JQA-E-90091
9659
http://www.hioki.co.jp/
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
■ HIOKI 3638-20 LEAK LOGGER Specifications
Data Analysis and Processing on a PC
Sensor models
HIOKI 9657, 9658 and 9659 CLAMP ON LEAK SENSORS
Inputs & ranges
2 channels, 100.0 and 1000 mA (manual range selection)
Instrument accuracy
±1% rdg. ±5 dgt. at 50/60 Hz, 23 ±5˚C, with built-in filter for 50/60Hz
Data can be transferred to a PC using the HIOKI 3911-20
COMMUNICATION BASE. Up to 16,000 data points can
be collected from 16 channels, (or up to 32,000 data points
can be collected from 8 channels), and then transferred to the
PC for analysis and processing.
with 9657 or 9659 sensor ±2% rdg. ±10 dgt. (100 mA range)
±2% rdg. ±6 dgt. (1000 mA range)
±4.5% rdg. ±10 dgt. (100 mA range)
±4.5% rdg. ±6 dgt. (1000 mA range)
Combined accuracy
at 50/60 Hz, 23 ±5˚C
with 9658 sensor
Measurement period
1 or 0.2 s (one channel only)
Recording modes
Maximum or average value
Recording interval
1, 2, 5, 10, 15, 20 or 30 s, and 1, 2, 5, 10, 15, 20, 30 or 60 min.
Recording data
quantity
32,000 data points (single channel), 16,000 data points (for dual-channel
recording)
Recording start / end
Manual, or at preset time / Manual, or when memory full
Recording methods
Single recording, Endless recording
Max./Min. display
Maximum and Minimum values are displayed
Alarm output
On when input crosses preset upper and lower limits (open-collector output)
Backup
Data backup is provided
Interface
Serial infrared transfer
3911-20
3911-20
3638-20
(Infrared transfer)
RS-232C Cable
Settings that can be made from communications software:
Power supply,
AAA-size alkaline batteries (LR03 × 4), 0.1 VA
consumption
Continuous operating One month (with power saver on, 1s measurement period)
Ten days (with power saver off, 0.2s measurement period)
period
Approx. 57.5(W) × 86.5(H) × 30.0(D) mm (less projections), approx. 130g
Size & weight
(including batteries)
current time, recording interval, recording start, recording methods,
comment, recording mode, alarm, channel selection
■ HIOKI 3911-20 COMMUNICATION BASE
Specification
Communication
method
Up to 16,000 data points × 16 channels,
Up to 32,000 data points × 8 channels
Infrared between Leak Logger and 3911-20,
RS-232C serial between 3911-20 and PC
Recording capacity
Power supply
Four AAA size alkaline batteries (LR03 × 4)
Operating temperature &
humidity
0 to 50˚C, 80% RH or less (noncondensing)
Size & weight
Approx. 69(W) × 92(H) × 36(D) mm, approx. 150 g
Supplied accessories
AAA-size alkaline batteries (LR03 × 4), Model 9632 CONNECTION CABLE
(for alarm output)
Supplied accessories PC communications software
Supported
operating systems
Windows 9x, NT 4.0
■ Clamp On Current Leak Sensor Specifications
9657
9658
9659
Up to 40 mm dia.
Up to 12 × 30 mm
Up to 30 × 150 mm
60A Continuous at 45 to 65 Hz
10A Continuous at 45 to 65 Hz
Model
Measurable conductor size
1.0A AC
Rated primary current
Maximum allowable input
Amplitude accuracy
Residual current
±1.0% rdg. ±12 µV
±3.5% rdg. ±12 µV
±1.0% rdg. ±12 µV
5 mA (at 100A AC in)
1 mA (at 10A AC in)
30 mA (at 500A AC in)
5 mA equiv. With 400A/m (AC), 7.5 mA max.
External magnetic field effect
Approx. 3 m
Cable length
Voltage to ground
Insulation withstand voltage
300 Vrms AC
150 Vrms AC
460 Vrms AC
3.7 kV (for 1 min.)
2.3 kV (for 1 min.)
2.2 kV (for 1 min.)
0 to 50˚C, 80% RH or less (non-condensating)
Operating temp. & humidity
Size & weight
100A Continuous at 45 to 65 Hz
25 mV/A AC
Output voltage
Approx. 74W × 145H × 42D mm, 340 g
Approx. 65W × 52H × 18D mm, 50 g
NOTES:
1. Voltage to ground is that voltage between a power line and earth ground in a groundingdependent electric circuit, or the voltage between one power line and any other power line
in a grounding-independent circuit.
2. Coordination of Insulation: determination of the safe and appropriate characteristics of
electrical insulation of wiring and connected devices according to the operating voltage.
3. Model 9658 is suitable for coordination of insulation with circuits of up to 150V voltage
to ground, up to 240V for 3P3W lines, and up to 120/240V for 3P4W lines.
4. Models 9657 and 9659 are suitable for coordination of insulation with circuits of up to
300V voltage to ground, up to 500V for 3P3W lines, and up to 277/480V for 3P4W lines.
Approx. 358W × 108H × 48D mm, 2.5 kg
COMMON ELECTRIC CIRCUITS and COORDINATION of INSULATION
3P-4W (3-phase,
4-wire) Voltage
120/208 V
120/240 V
230/400 V
277/480 V
3P-3W (3-phase,
3-wire) Voltage
Voltage to Ground
CATIII Recommended
Impulse Withstand
Voltage
240 V
150 V
2500 V
500 V
300 V
4000 V
(From EN61010-2-032, Annexes J, Table J.1)
Model 3638-20 LEAK LOGGER
■ Options
Model 3911-20 COMMUNICATION BASE
Model 9657 CLAMP ON LEAK SENSOR
Model 9658 CLAMP ON LEAK SENSOR
Model 9659 CLAMP ON LEAK SENSOR
Model 9637 RS-232C CABLE (9-pin to 9-pin crossover, 1.8 m)
Model 9638 RS-232C CABLE (9-pin to 25-pin crossover, 1.8 m)
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
ASM GmbH
Am Bleichbach 18 - 22
85452 Moosinning
All information correct as of Apr. 20, 2001. All specifications are subject to change without notice.
■ Inter net HIOKI website h t t p : / / w w w . h i o k i . c o . j p /
363820E2-14M-05P
Printed in Japan