PEN RECORDER PR8111, PR8112

PEN RECORDER PR8111, PR8112
Recorders
Portable, easy-to-use pen recorder
built for the field
Easily portable
Compact size
Support for three power sources
Can be powered with dry-cell batteries.
Outdoor-ready
Ships with a drip-proof cover.
Pen-based
Records data reliably.
Easy enough for anyone to use
Features simple operation.
PR8111: 1 pen, PR8112: 2 pens
2
Take reliable measurements. Anywhere.
Wherever you take them, the PR8111 and PR8112 will yield reliable measurement data
that can be stored for extended periods of time.
n Measure up to 250V with high 1mV sensitivity
The PR8111/PR8112 can accommodate a variety of input signals thanks to a broad series of measurement ranges.
Measurement ranges: 17 ranges
±1, 2.5, 5, 10, 25, 50, 100, 250, 500 mV, ± 1, 2.5, 5, 10, 25, 50, 100, 250 V
n Support for three power sources to allow use anywhere
The PR8111/PR8112 can be powered using its AC adapter, dry-cell batteries*1, or a DC power supply*2.
*1: Requires six D size Alkaline batteries
Continuous operating time: Approx. 50 h (using 1 pen) or approx. 25 h (using 2 pens)
With the front cover closed
*2: A connection cable is available by special order.
Please contact your HIOKI representative.
n Simple dust- and drip-proof design for outdoor use
• A front cover protects measurement hardware from wind and dust
(and keeps recording paper from blowing away in the wind).
• A drip-proof cover protects the instrument and recorded data from water droplets and dust
(affording peace of mind in the event of a sudden rain shower).
n Instant confirmation of measurement operation
by watching pen movements
With the drip-proof cover on
The ability to view measurements being taken, even from a distance, allows
you to immediately detect any malfunction.
n Reliable storage of important data
• Pen-based operation provides a reliable record of measurement data.
• Recording paper can be stored as-is for extended periods of time.
So simple, it just works.
With the drip-proof cover on
You need an instrument you can get out to the field easily
and start using right away. The PR8111/PR8112 delivers.
n Excellent portability
A more compact design means
even better portability.
Compact
Lightweight
(The PR8111/PR8112 is approximately half
the size of previous HIOKI models by volume.)
Don’t be fooled by their smaller footprint—
these instruments can use the same recording
paper as their predecessors.
Recording width: 150 mm
[ PR8112 ]
[ Previous HIOKI model: EPR-3500s ]
n Simple, intuitive operation
n Find anomalies quickly
• Simple,
• You can check recorded waveforms in the field.
• Recordings can easily be compared to past data.
intuitive controls
• Settings that are visible
at a glance
Start using it
right away
Find problems
immediately
Abnormal
Operating panel
(2-pen model)
Normal
3
PR8111/PR8112 Features
PR8111 PR8112
1 pen
2 pens
Pictured: PR8112
(2-pen model)
The days of big, heavy pen recorders are over.
Introducing a compact, lightweight instrument that’s designed for simple ease of use.
Compact and lightweight
Simple operation
Easily portable for
outstanding mobility
Dark gray color so
dirt doesn’t stand out
Intuitive knobs and switches
Easy-to-understand operation:
Each knob or switch controls
a single function.
No need to worry about how
to operate the instrument.
Power supply LED
Provides a color-based
indication of remaining
battery life.
Green: Ample remaining
battery life
Orange: Limited remaining
battery life
Red: Replace batteries.
Inputs
(rear)
Support for Johnson,
banana, and crimp-type
terminals
Recording hardware
Front cover (standard)
Protects the recording hardware from wind and dust (removable).
Channel 2 POWER switch (rear)
Check waveforms and make
notes in the field.
Recording paper can be
stored as-is for extended
periods of time.
Recording paper width:
150 mm
Allows you to turn off power to channel 2 (pen 2) to save
energy when performing measurements with just one
channel. Available on the PR8112 (2-pen model) only.
(Can use paper designed for
use with previous models.)
Applications
Record output from
measuring instruments
Record analog output from temperature/
humidity, pH, and other environmental
measuring instruments, insulation
testers, and sensors
Maintenance
Experiment
Print example (Actual size)
Environment
n Specifications
Input unit (product guaranteed for one year)
Operating method
Input
Measurement
ranges
Recording
accuracy
Dead zone
Inter-range error
Input impedance
Recording unit
Self-balancing
±0.25% of effective recording width
2 MΩ±1 %
Max. allowable input 250 V DC (V range) or 30 V DC (mV range)
Max. rated voltage between
300 V DC/AC
terminals and ground
3-stage LED display (green, orange, red)
(when operating on dry-cell battery power)
Battery check
Accuracy is guaranteed even when red indicator is lit.
Operating temperature
0 ℃ to 40 ℃, 40 % to 80 % rh (non-condensating)
and humidity
(1) AC ADAPTER 9418-15: 100 to 240 V AC (50/60 Hz)
(2) D Size alkaline battery × 6
Power supply
Max. rated power
Dimensions
Mass
PR8111: 1 pen, PR8112: 2 pens
Recording method Disposable felt pens
Recording width 150 mm
No. of pens
DC Voltage (isolated input)
±1, 2.5, 5, 10, 25, 50, 100, 250, 500 mV,
±1, 2.5, 5, 10, 25, 50, 100, 250 V
17ranges
±0.5% of effective recording width (excluding contraction and
expansion of recording paper)
(Including linearity in reference range; reference range of 250 mV)
±0.2% of effective recording width
Pen interval
5 mm±1 mm
500 mm/s or greater (using AC adapter)
Pen speed
10, 20, 30, 60, 120, 180, 300, 600 mm/m
10, 20, 30, 60, 120, 180, 300, 600 mm/h 16 ranges
Accuracy: ±0.25%
Paper feed
speed
Zero point
movement range
Chart feed
Can be adjusted throughout each range’s full area of
movement.
300 mm/min, slow-up method
Pen-lifting mechanism Gang lift
Vertical
Fanfold plain paper SE-10Z-2 (length: 15 m)
Recording paper
Roll plain paper SE-10 (length: 20 m)
Orientation
(When used with the AC adapter, the adapter takes precedence.)
[Continuous use] (based on in-house testing conditions)
PR8111: Approx 50h, PR8112: Approx 25h
(3) DC power supply: 10 to 27V DC (cable available by special order)
4VA (AC ADAPTER, DC POWER) or 3VA (dry-cell batteries)
292 mm (11.5 in) W×177 mm (6.97 in) H×182 mm (7.17 in) D
PR8111: 3.9 kg (137.6oz) , PR8112: 4.4 kg (155.2oz) (instrument only)
PR8111: 4.8 kg (169.3oz) , PR8112: 5.3 kg (186.9oz)
(with dry-cell batteries, recording paper, and pen(s))
Storage temperature 20 ℃ to 55 ℃, 10 % to 80 % rh (non-condensating)
and humidity
Safety: EN61010-1
Applicable
EMC: EN61326-1 CLASS A, EN61000-3-2, EN61000-3-3
standards
PR8111 (1-pen model)
PR8112 (2-pen model)
Accessories
FELT PEN PR8111: P-1201A (Red)
PR8112: P-1201A (Red), P-1202A (Green)
RECORDING PAPER × 1 (fanfold)
AC ADAPTER 9418-15 × 1, Drip-proof cover × 1, Instruction manual × 1
n OPTION
RECORDING PAPER SE-10Z-2 (fanfold)
Width:170 mm (effective recording width: 150 mm) ×15 m,
Set of 10
The PR8111/PR8112 uses the same recording paper
and felt pens as previous HIOKI models
(the EPR-3000 series and EPR-3500 series).
RECORDING PAPER SE-10 (roll)
Width:170 mm (effective recording width: 150 mm) ×20 m,
Set of 10
FELT PENS
P-1201A (Red)
P-1202A (Green)
P-1203A (Blue)
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
http://www.hioki.co.jp / E-mail: [email protected]
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
http://www.hiokiusa.com / E-mail: [email protected]
1608-1610,Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-63910090/63910092 FAX +86-21-63910360
http://www.hioki.cn / E-mail: [email protected]
Beijing Office : TEL +86-10-84418761/84418762
Guangzhou Office : TEL +86-20-38392673/2676
DISTRIBUTED BY
HIOKI SINGAPORE PTE. LTD. :
33 Ubi Avenue 3, #03-02 Vertex, Singapore 408868
TEL +65-6634-7677 FAX +65-6634-7477
E-mail: [email protected]
HIOKI INDIA PRIVATE LIMITED :
Khandela House, 24 Gulmohar Colony Indore 452 018 (M.P.), India
TEL +91-731-4223901, 4223902 FAX +91-731-4223903
http://www.hioki.in / E-mail: [email protected]
All information correct as of Apr. 18, 2011. All specifications are subject to change without notice.
PR8111E1-14B