ASM-SENSOR 8835-01

MEMORY HiCORDER 8835-01
Recorders
High-visibility display, Compact body, Multi-channeled inputs
Field Measurement has never been easier
The MEMORY HiCORDER 8835-01 is a high-speed waveform recorder with
the special features of advanced performance of the basic “wave recording”
function for easy field measurement, easy-to-see color display, compact
dimensions of an A4-sized paper, and 4/8* channels for measurement. The
MEMORY HiCORDER 8835 -01 inherits all the functions of the MEMORY
HiCORDER 8835 and accommodates a total of 8* channels when used with
the input unit group to support a wide range of signals. The 8835-01 also comes
standardly equipped with 8x the memory of the previous unit, making long-term
recording possible.
* When using the 4ch ANALOG UNIT 8946, maximum input is 30V rms or 60V DC.
2
Compact 4ch/8ch* recorder
saves space with slim profile
1
*1 When using the 4ch ANALOG UNIT 8946, maximum input is 30V rms or 60V DC
- Features ● Conversion According to the Measurement, Plug-in
Input Function for a Maximum of 8 Channels*1
The 8835 -01 employs a plug-in unit system that can change
the measurement channels according to the measurement use.
Directly inputting physical signals through inserted conversion
amplifiers is also possible. A maximum of 8 channels*1 can be
used for measurement by mounting a 4-channel analog unit on
the recorder.
● High-visibility waveforms displayed on a 6.4-inch
color TFT liquid crystal display
The color display makes it easier to identify waveforms
and install the device. It enhances visibility and facilitates
operations.
● Compact and thin, occupying a space equivalent to
60% of an A4-size sheet of paper
Occupying desktop space equivalent to 60% of an A4-size
sheet of paper, the MEMORY HiCORDER 8835-01 is functionally
designed so as to permit operation on a flat bed.
● Highly Improved Basic Performance with 1MS/s,
12bit-A/D, 4MW
T he 8 8 3 5 - 0 1 employs a sa mpl i ng r at e of 1MS /s (1μs
cycle) a nd 12 -bit volt age -a x is r esolut ion for t he A / D
conver ter un it, wh ich digitizes measurement signa ls,
enabling accurate detection of signal waveforms.
With the 8835 -01, the standard memory capacity is 4MW.
● Converts to text file used with a Wave viewer (supplied
accessories, PC application software)
To open measurement data in PC applications such as Excel,
the data must be converted to text data in the CSV format.
The PC application software which comes standard in the
package enables easy operation.
● Connects to PCs and printers on a LAN network
Its use with PCs can be selected according to the usage.
It allows remote operation and data transfer via LA N
connection, GP-IB connection, or RS-232C connection.
● Function upgrade system to meet varied needs
The basic model provides several standard functions for users
who donʼt require functional complexity. Users requiring
a wider range of measurement functions can add functions
through the use of a function upgrade disk.
● On-screen help
To help the user get started or clarify operating steps, the
8835 -01 can display tips on-screen for many basic operations,
including key-button operations.
● CE Mark compliant
Complies with the EC directive determining safety standards
in Europe (within the EU).
A LAN-Connectable Recorder! Digitally Process
Test Data
With a conventional pen recorder, even if all test data is written on the
paper, usually only a small portion of the data is needed. However, to
look for just a small important part requires very extensive search of the
recording paper.
The MEMORY HiCORDER 8835 -01 stores
and manages all waveform measurement data
electronically. Furthermore, use of a LAN card
and the LAN COMMUNICATOR 9333 enables
high-speed data file transfer to PCs on a
network.
Example of using measurement
data in Excel
3
High-Speed Response for Capturing Transient Events
- Function Details -
Large memory capacity allows long-term
recording of high- speed data
A/D
conversion
Insulation
Function outline of memory recorder
Input voltage
The 8835 -01 can store a total of 4 mega-words, using
internal solid-state memory. This provides ample
capacity to store data for all 8 channels. The table at
right shows possible recording times, according to the
time axis setting and the number of channels in use.
A reduction in the number of channels prolongs the
recording time.
Write on memory
A/D
conversion
Insulation
Input voltage
Input voltage
Input waveform
Sampling period
Time
Time axis
100μs/DIV
200μs/DIV
500μs/DIV
1ms/DIV
2ms/DIV
5ms/DIV
10ms/DIV
20ms/DIV
50ms/DIV
100ms/DIV
200ms/DIV
500ms/DIV
1s/DIV
2s/DIV
5s/DIV
10s/DIV
30s/DIV
1min/DIV
2min/DIV
5min/DIV
■ Memory segmentation function
(an optional FUNCTION UP DISK 9540-01 is needed)
When using the memory recorder function, the data memory
can be divided into a maximum of 255 blocks. Data can be
written sequentially to the memory blocks, and the waveform
in a reference block and any other block can be superimposed
and compared.
Recording length using 8 channels
Segmentation
Segmentation into 3 blocks
Segmentation into 127 blocks
■ Zoom function
*In memory recorder function
To make the most of the large-capacity memory, it is possible
to display a compressed waveform simultaneously with a
magnified waveform. Since the 8835 -01 is capable of storing
a large amount of data, high-speed sampling is also possible
for waveforms with a long duration. Accordingly, while
observing the compressed image of the entire waveform, it is
also possible to observe the magnified details of desired parts.
Compressed display of a part of the entire waveform is also
possible.
■ Clock input for external sampling
*In memory recorder function
The sampling rate for the memory recorder can be controlled by
the timing of an external clock signal. This is useful for example to
collect data synchronized to the running cycle of an engine.
Maximum frequency 500kHz
Input voltage
Input waveform
Sampling controlled by external
clock signal
Thermal printer
Sampling period
Time
Output waveform
Sampling
period
1-channel setting
4 MW/channel, 40000 DIV
8-channel setting
500 kW/channel, 5000 DIV
1μs
2μs
5μs
10μs
20μs
50μs
100μs
200μs
500μs
1ms
2ms
5ms
10ms
20ms
50ms
100ms
300ms
0.6s
1.2s
3.0s
4s
8s
20 s
40 s
1 m 20 s
3 m 20 s
6 m 40 s
13 m 20 s
33 m 20 s
1 h 6 m 40 s
2 h 13 m 20 s
5 h 33 m 20 s
11 h 6 m 40 s
22 h 13 m 20 s
2 days 7 h 33 m 20 s
4 days 15 h 6 m 40 s
13 days 21 h 20 m
27 days 18 h 40 m
55 days 13 h 20 m
138 days 21 h 20 m
0.5 s
1s
2.5 s
5s
10 s
25 s
50 s
1 m 40 s
4 m 10 s
8 m 20 s
16 m 40 s
41 m 40 s
1 h 23 m 20 s
2 h 46 m 40 s
6 h 56 m 40 s
13 h 53 m 20 s
1 day 17 h 40 m
3 days 11 h 20 m
6 days 22 h 40 m
17 days 8 h 40 m
4
High-speed response and Effective value recorder
functions are useful in following signal variations
- Function Details -
Outline of recorder function
The input signal is converted to digital form and displayed
and printed in real time. The chart speed is a maximum of
20mm/s (in the 500ms/division range). Even with real-time
recording, the last 2000 divisions of the waveform can be
observed (by scrolling both horizontally and vertically) and
reprinted following measurement.
Recording Time of the Recorder
Virtual recording
The 8835 -01 supports a high-speed recording function in the
memory with no need for recording paper. Although realtime recording on the recording paper is not possible in the
high-speed range of the recorder function (10ms to 200ms/
division), the waveforms are stored in the memory and can
therefore be monitored on the screen. The last 2000 divisions
of the waveform are retained in the memory before the
measurement is completed. If the recording length is not set
to “continuous”, the printer can also be operated, allowing
waveforms to be printed out later.
Time axis
10ms*2/DIV
20ms*2/DIV
50ms*2/DIV
100ms*2/DIV
200ms*2/DIV
500ms/DIV
1 s/DIV
2 s/DIV
5 s/DIV
10 s/DIV
30 s/DIV
1 min/DIV
2 min/DIV
5 min/DIV
10 min/DIV
30 min/DIV
1 hr/DIV
Chart speed
Sampling
period
20mm/s
1μs
10μs
100μs
20mm/s
1μs, 10μs
100μs, 1ms
10mm/s
5mm/s
2mm/s
1mm/s
20mm/min
10mm/min
5mm/min
2mm/min
1mm/min
20mm/hr
10mm/hr
1μ, 10μ, 100μs
1ms, 10ms
1μs
10μs
100μs
1ms
10ms
100ms
Recording time for approx.
1 roll of recording paper (30m)*1
20 s
40 s
1 m 40 s
3 m 20 s
6 m 40 s
24 m 45 s
49 m 30 s
1 h 39 m 00 s
4 h 7 m 30 s
8 h 15 m
24 h 45 m
2 days 1 h 30 m
4 days 3 h 00 m
10 days 7 h 30 m
20 days 15 h
61 days 21 h
123 days 18 h
*1 Based on 2970 divisions, assuming that about 30 cm of the paper length will not be used.
*2 Real-time recording on the recording paper is not possible.
Continuous X-Y recorder function
This function allows two signals converted to digital form to
be combined in an x-y plot and stored in memory. Any of the
four analog channels can be used for an x-y plot, and up to
three x-y plots can be combined. The x-y plot can be viewed
on the screen in real time, and the recording time is unlimited.
The x-y plot can also be reprinted.
Effective value (RMS-value) recorder function
This function is designed exclusively for use on 50/60Hz
power supply lines and DC. High-speed sampling is applied to
calculate the rms value from the waveform data, and the result
is recorded as a graph.
Note: Using fixed 200μs sampling, data for two waveforms are captured for
calculating the rms value. This process is repeated 20 times per second,
resulting in high-speed response that is 10 times faster than that of a digital
tester or similar (using a 2-second update rate).
5
Easily interfaced with a PC or a waveform
comparator
Support for connection to PCs via Ethernet
■ Connect HiCORDER to departmental LAN (using TCP/IP communication protocol)
8835 -01 can be connected to Ethernet, a standard network
protocol in the Internet age (using the optional LAN CARD and LAN
COMMUNICATOR 9333) . For those who frequently analyse
measurement data on PCs, this function offers a good match.
Also, connection to PCs using RS-232C connection or GP-IB
connection is possible (using the optional RS-232C CARD 9557
and GP-IB CARD 9558). 8835 -01 data can be sent to PCs or be
remotely controlled from PCs.
Ethernet
* Because LAN card, GP-IB card, and RS-232C card all use the same PC card slot of
8835 -01, when one of them is inserted into the PC card slot, then any memory card cannot
be used at the same time.
Copy or remote operations of HiCORDER data
files from client PCs running 9333 (software)
■ Save data to network server
Color image output to the
network printer
(using TCP/IP communication protocol)
Offline data exchange with PCs
The supplied waveform viewer (PC application) can
convert saved waveform data to text files (CSV format).
For data storage, FD/PC card (supplied as standard) can
be used. This allows easy offline data exchange with
PCs.
Ethernet
PC running 9333 (software)
Copy measurement data files obtained from
HiCORDER to PC on network
Memory / main unit
* In addition to HiCORDER's read/write native file (binary format), data can
also be saved to text files (CSV format) which can be opened by PC spreadsheet
applications, or waveform bitmap files (BMP format). However, because data
saved in text files cannot be read by HiCORDER, it is recommended that text
data conversion be performed on PCs.
Memory / main unit
Floppy disk
Memory / main unit
Flash ATA card
Data compatibility with the model 8730/8731
The MEMORY HiCORDER 8835 - 01 is judgment
area and waveform data compatible with HIOKI WAVE
COMPARATORs 8730/8731.
The 8835 - 01 lends itself to detailed analysis and
printing of waveform judgment data on production lines.
Waveform judgment screen on the 8835-01
(* Waveform data is binary data within 500 DIV.)
(* The waveform judgment function of the 8835-01 must be upgraded with the
FUNCTION UP DISK 9540-01.)
Waveform judgment screen on the 8731
6
- Function Details -
Trigger functions for monitoring of all four channels
In all of the functions, including the memory recorder and
recorder, all eight analog input channels and sixteen logic input
channels can be used for trigger input. In addition to a level
trigger, which compares voltages based on a reference value,
the 8835 -01 supports the following triggers.
● Window trigger that compares voltages based on two reference values
● Voltage-drop detection trigger that detects voltage drops in commercial
power lines
● RMS-value level trigger that compares signals based on the RMS-value level
● Period detection trigger that measures periods and detects all deviating
periods
● Pattern trigger that compares signals based on the logic signal ON/OFF
pattern
Variable (span adjustment), vernier (fine adjustment)
W hen sensors a re used to measu re a nd record noise,
temperature, acceleration or other physical quantities, precise
calibration is important. This is facilitated by the vernier
function that allows fine adjustment of amplitude. The variable
function lets the user numerically specify the measurement
span, such as 1 - 5V or 4 - 20mA. This is useful for matching
the range of instrumentation to the full span of the recording
paper. A scaling function for converting measurement results is
also available.
Manipulation using the cursor
Use of the two cursors on the screen enables the user to read
the time difference and potential difference.
Signal outputs, control inputs, DC input
The results of waveform decisions, parameter decisions, and
triggers are output as open collector signals. The 8835-01 is
also provided with signal inputs for remote control of the start,
stop, and print buttons.
The 8835 -01 has a dual AC/DC power-supply specification,
and an external battery can be used by means of the DC
POWER ADAPTER 9439, in addition to normal AC power
supply. This allows vehicle-mounted applications, where
an AC power supply is not available. If both supplies are
connected, the AC power supply takes precedence, but if
the AC power fails, the unit automatically switches to DC
operation.
DC input
(The DC POWER ADAPTER 9439
must be used)
External trigger input
External control
connector
Trigger signal output
Judgment outputs
* When using the F/V UNIT 8940 with 12 V DC power, the printer can only be
used for up to 2 channels.
7
Upgrading provides sophisticated functions
Additional functions provided by the FUNCTION UP DISK 9540-01 (with the 8835, use the FUNCTION
UP DISK 9540)
Recorder and memory functions
If an abnormal event is detected by triggers during the realtime recording of signals using the recorder function, it
is stored in memory by the high-speed sampling memory
recorder. The recorder function works independently and is
therefore continuous. These functions are useful when the
user wants to record normal waveforms as well as abnormal
waveforms.
FFT analysis functions
The single-channel FFT function is used in spectrum analysis.
The two-channel FFT function analyzes transfer functions.
The octave analysis function is used in acoustic analysis. The
signal source for FFT analysis is a section obtained from the
waveforms captured in the memory recorder
(the required number of pieces of data for FFT analysis are 1000, 2000, 5000
and 10000).
Waveform judgment functions
*In memory recorder functions and FFT analysis functions
The 8835 -01 can monitor a measured waveform for a preset
judgment area. Unlike a voltage level only comparison such as a
trigger setting, this function makes it possible to easily prepare
a standard judgment area making it an excellent system for
comparisons both in level direction and in time axis direction.
Patented: Registration number 2028013
The 8835 -01 is waveform data and area data compatible with
the WAVE COMPARATOR 8730 and 8731.
(* Wavefor m data is bina r y data within 50 0 DIV. The 873 0 and 8731
comparators do not provide an FFT analysis function.)
Simultaneous computation on eight channels
It is possible to simultaneously compute four different types
of waveforms, each of which has been captured on one of the
eight channels in memory recorder mode. The results of four
basic arithmetic computations, differentiation, or integral are
displayed in a waveform.
(The waveform computation requires a FUNCTION UP DISK. 9540 -01 With
the 8835, the waveform computation requires a FUNCTION UP DISK 9540.)
For parameter calculations that calculate numerical values such
as the maximum and minimum values, up to eight waveforms
can be operated simultaneously on four channels.
(The parameter operations are a standard function.)
Waveform judgment screen on the 8835-01
Waveform judgment screen on the
8731
8
Example Printouts
(Actual size)
Sample recording in memory
recorder function (full width)
The recording width can also be divided into two
or four parts, and X-Y plots are possible.
Sample recording
(two sections)
Sample recording in X-Y format
* Example when used with 1-4 channels set
Synthesized X-Y waveforms can be output where
amplitude data for each input channel is plotted
on the vertical and horizontal axes, based on the
waveform data obtained using the memory recorder.
The recording size is 100 mm × 100 mm.
Although the sample print out is in
Japanese, the actual print out appears
in English. (Two languages selectable)
Sample recording (four sections)
Full-width recordings and recordings divided
into two or four parts are also possible.
Logging output
* Example when used with 1-4 channels set
Instantaneous voltage values are printed for each sampling.
Although the sample print out is in Japanese, the actual print out appears in English. (Two languages selectable)
PC card slot for type-III cards
A PC card-type GP-IB or
RS-232C interface is used for
this slot as well.
Waveform monitor (color TFT LCD)
Floppy-disk drive
External control
terminal
DC power inlet
Connect the optional DC
POWER ADAPTER 8439
AC power inlet
Logic probes input terminals
Analog input unit mount
Thermal printer: 110 mm in width, 8 dots/mm
Basic specifications 8835 -01 (use 2ch input type modules) 8835 -01 (use 4ch ANALOG UNIT 8946)
Input type/number
of channels
Plug-in input modules
Max. 4 analog chʼs + 16 logic chʼs
(Isolated analog channels, isolated input and
frame, logic has common GND)
Plug-in input modules
Max. 8 analog chʼs + 16 logic chʼs
(Isolated analog channels, isolated input and
frame, logic has common GND)
Measurement
functions
MEM (high-speed recording)
REC (real-time recording)
RMS (50/60Hz, or DC only)
REC & MEM (Additional functions provided by the optional FUNCTION UP DISK 9540-01)
FFT (Additional functions provided by the optional FUNCTION UP DISK 9540-01)
Maximum
sampling rate
1 MS/second (1μs, all channels simultaneously)
External sampling (500kS/second, 2μs)
Memory capacity
Data storage media
Interfaces (option)
(12 analog bits + 4 logic bits) × 4 Mwords/channel (1 channel used) to (12 analog
bits + 4 logic bits) × 500 kwords/channel (8 channels used)
PC Card Type III slot × 1: up to 1GB (Flash ATA)
3.5" Floppy disk drive × 1: 1.44MB, 1.2MB, 720KB, MS-DOS format
File format: Binary, text, BMP
Mini-jack 3.5 mm in dia.: Trigger input/output
Terminal board: External start, stop, print input/sampling input,
decision output
Note: Use one of the following: RS-232C CARD 9557, GP-IB CARD 9558, LAN CARD
(HIOKI-tested)
Compliance
standard
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Power
requirements
100 to 120V AC or 200 to 240V AC (50/60 Hz)
10 to 28V DC (use the DC POWER ADAPTER 9439 : option)
Power consumption
120VA, max. for 100V AC (approx. 60VA with the printer off)
70VA, max. for 12V DC (approx. 30VA with the printer off)
Power consumption
(when using two units of 8940)
100μs to 5min/DIV, 20 ranges or external sampling, time axis
resolution 100 points/DIV, time axis zoom: ×2 to ×10 in 3 stages,
compression: 1/2 to 1/2,000 in 10 stages
Sampling rate
1/100 of time axis ranges (minimum sampling period 1μs)
External sampling Max. 500kS/s (minimum sampling period 2μs)
Settable in 1 DIV steps, 20 to 40,000 DIV*1
Recording length *1 Depending on the number of channels in use.
Pre-trigger
Can record data from before the trigger point, 0 to 100% or -95% of
recording length; 15 settings
Other functions
waveform processing*2, waveform parameter processing, waveform
averaging*2, memory segmentation (up to 255 segments)*2, logging
(numerical printout), X-Y waveform plot, voltage axis zoom ×2 to ×
10, 3 settings, compression 1/2, zoom, variable display, graph
superimposition, waveform judgment function*2
*2 Additional functions provided by the optional FUNCTION UP DISK 9540-01
Recorder functions (time axis waveform and X-Y format)
10ms to 1 hour/DIV with 17 ranges, time axis resolution 100 points/
DIV, time axis compression: 1/2 to 1/50 in 5 stages
Time axis
At 10ms to 200ms/DIV, printing in real time is not possible, but waveform data are stored in
memory and can be monitored on screen. Data are stored for 2,000 divisions before the end
of measurement. At recording length settings other than "Continuous", the printer can be
used simultaneously, for follow-up printing of waveforms.
Sampling rate
1μs to 100ms; 6 settings (selectable from 1/100 or less of time axis)
Recording length
Settable in 1 DIV steps, 20 to 2,000 DIV*3, or "Continuous"*4
At X-Y format: only continuous for X-Y plotting
GP-IB, RS-232C, LAN
Environmental conditions Operation: +5°C (41°F) to +40°C (104°F), 35% to 80% rh
(no condensation)
Storage: -10°C (14°F) to +50°C (122°F), 20% to 90% rh
(when using two units of 8936)
Time axis
4 Mwords total
Clock and setting conditions: battery life of at least 10 years
Backup functions Waveform data: battery life of at least 1 hour after system power is
(at 25°C)
turned OFF (at 2 minutes after power-on)
External control
connectors
Memory functions
170VA, max. for 100V AC (approx. 110VA with the printer off)
80VA, max. for 12V DC (approx. 50VA with the printer off)
Note: When using the F/V UNIT 8940 with 12V DC power, the printer can only be used for
up to 2 channels.
Dimensions and
mass
285mm (11.22in) W × 220mm (8.66in) H × 132mm (5.20in) D, 4.5kg
Supplied
accessories
Instruction Manual × 1, Power cord × 1, Printer paper ×1, Protective
cover ×1, Roll paper attachment ×2, PC card protector ×1, Application
Disk (Wave Viewer Wv, Communication Commands table) ×1
(158.73oz.) (main unit only)
X-Y sampling period 100μs; fixed (dot), 100μs to 25ms (line)
X-Y axis resolution 40dots/DIV (display), 80dots (horizontal) × 80 dots (vertical)/DIV (printer)
Waveform memory
Store data for most recent 2,000 DIV in memory. Backward
scrolling and re-printing available.
Other functions
logging (numerical printout), additional recording (recording is resumed
without overwriting previous data), voltage axis magnification ×2 to ×10;
3 settings, compression 1/2; 1 setting, variable display.
RMS Recorder Function (for 50/60 Hz and DC)
Time axis
5s to 1 hr/DIV; 9 settings, time axis compression 1/2 to 50; 5 settings
Sampling rate
200μs fixed (20 rms datas/s)
RMS calculation accuracy ±3% f.s.
Settable in 1 DIV steps, 20 to 2,000 DIV*5, or "Continuous"
Recording length *5 Measure all channels.
Waveform memory
Store data for most recent 2,000 DIV in memory. Backward
scrolling and re-printing available.
Other functions
logging (numerical printout), additional recording (recording is resumed
without overwriting previous data), voltage axis magnification ×2 to ×10;
3 settings, compression 1/2; 1 setting, variable display.
Print/display section
Display
6.4 inch TFT color LCD, with English/Japanese selector (640 × 480 dots)
Recording paper
110 mm (4.33 in) × 30 m (98.4 ft), thermal paper roll
Recording width
10 divisions for full scale, 1 DIV = 10 mm (0.39 in) (80 dots)
Paper feed density
10 rows/mm (250 rows/in) * 20 rows/mm (500 rows/in with the memory
Recording speed
recorder's smooth print function
Max. 25 mm/s (0.98 in/s)
Auxiliary Functions
General
Trigger functions
Trigger sources
Trigger types
(analog)
CH1 to CH8 (analog), CHA to CHD (logic), external, timer, manual
(either ON or OFF for each source), logical AND/OR of sources
Level: Digital setting of voltage. Triggered when set value is
exceeded in UP or DOWN direction.
Window: When entering or exiting a level range defined by upper
or lower limit
Voltage drop: Only for AC power lines. Triggered when the peak
voltage falls below setting value
RMS level: Only for DC and AC power lines. Triggered when rms
value crosses set value in UP or DOWN direction
Period: When rising or falling edge of set voltage does not fall
within cycle range
Scaling
Vernier function
Pattern trigger: 1, 0, or × (disregard), logical product (AND) or
logical sum (OR) set for 4 channels
Trigger filter
OFF, setting range 0.1 to 10.0 DIV in 9 stages (MEM, REC & MEM
function), ON/OFF (REC function)
(logic)
(analog/logic)
Other functions
Pre-trigger function to capture pre- and post-trigger waveform,
trigger output (active Low with φ 3.5mm mini-jack and open
collector 5 voltage output), Start & stop trigger in REC function
Printing of settings including input range, trigger time, etc, cursor
measurement, scaling, free comment input, screen hard copy,
registration of setting conditions (eight conditions), start condition
retention, auto setup, auto saving, remote control, auto ranging,
view function, online help, key lock, list printing, etc.
Scaling: Translation of amplitude gradation only
Variable: Arbitrary setting of the upper and lower limit of the
waveform display range
Allows precision adjustment of input voltage.
Average value, effective (rms) value, peak to peak value, maximum
Waveform parameter value, time to maximum value, minimum value, time to minimum
calculation
value, period, frequency, rise time, fall time, area value, X-Y area
(in MEM function)
value, and standard deviation.
■ PC Software Specifications
Wave Viewer (Wv) Software (Application disk CD-R, bundled accessory)
Level setting resolution Equivalent to 0.25% when full scale is set to 10 divisions
Trigger types
*3 Measure all channels.
*4 When time 10 ms − 200 ms/DIV and printer is ON, continuous is not available.
Functions
• Simple display of waveform file
• Text conversion: convert binary data file to text format, with
selectable space or tab separators in addition to CSV, and specifiable section, thinning available
• Display format settings: scroll functions, enlarge/reduce display, display channel settings
• Others: voltage value trace function, jump to cursor/trigger
position function
Compatible PC operating systems Windows 95/98/Me, Windows NT 4.0 (SP3 or later), 2000, XP
9
10
■ Options (sold separately)
FUNCTION UP DISK 9540 -01
Additional functions to the MEMORY HiCORDER 8835-01
REC & MEM functions Additional functions provided by the FUNCTION UP DISK 9540-01
Time axis
(REC)
500ms to 1hour/DIV; 17 settings, 1 DIV = 100 samples, time axis
compression 1/2 to 1/50, 5 settings
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 300 g (10.6 oz) Accessories: None
VOLTAGE/TEMP UNIT 8937
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up
time and zero-adjust; accuracy guaranteed for 1 year)
Measurement functions Number of channels: 2, for voltage measurement/temperature measurement with thermocouple
Voltage input: metallic BNC connector (input impedance 1MΩ, input capacitance 50pF),
Input connectors thermocouple input: terminal connector (input impedance min. 5.1MΩ), Max. rated
voltage to earth: 30Vrms or 60V DC (with input isolated from the unit, the maximum voltage
that can be applied between input channel and chassis and between input channels without damage)
Note: Sampling period 1/100 of time axis range at MEM function
Time axis
(MEM)
100μs to 5 minutes/DIV; 20 settings, 1 DIV = 100 samples, time axis
zoom ×2 to ×10; 3 settings, compression 1/2 to 1/2000, 10 settings
Voltage measurement
range
1mV to 5V/DIV, 12 ranges, full scale: 10DIV, low-pass filter: 5Hz/
500Hz/ 5kHz/ 100kHz, Measurement resolution: 1/160 of measurement
range (using 12-bit A/D conversion; installed in the 8835-01)
Recording length
REC: Settable in 1-division steps, 20 to 1,000 DIV, or continuous
MEM: Settable in 1-division steps, 20 to 2,000 DIV
Temperature
measurement range
20°C to 200°C/DIV, 4 ranges, full scale: 10DIV, low-pass filter:
5Hz/500Hz, Measurement resolution:1/160 of measurement range
Trigger source
REC: timer trigger, or OFF
MEM: CH1 to CH8 (analog), logic A to D, or external trigger
K: -200 to 1350°C, E: -200 to 800°C, J: -200 to 1100°C, T: -200 to 400°C,
N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 300 to 1800°C,
Reference junction compensation: internal/ external (switchable)
Other functions
Only real-time waveform is output when printer output is started,
reprinting of stored REC waveform data (last 1000 DIV), Additional
recording function (recording is resumed without overwriting previous
data), variable display
Thermocouple
range
Highest sampling rate
Voltage input: 1MS/s, Temperature measurement: 4kS/s
Accuracy
Voltage input: DC amplitude ±0.4% of full scale, zero position ±0.15% of full
scale, Temperature measurement (K, E, J, T, N): ±0.1% of full scale ±1°C,
±0.1% of full scale ±2°C (-200 to 0°C), (R, S): ±0.1% of full scale ±3°C, (B):
±0.1% of full scale ±4°C (400 to 1800°C) , Reference junction compensation
accuracy: ±0.1% of full scale ±1.5 °C (internal reference junction compensation)
Frequency
characteristics
Voltage input: DC to 400 kHz +1/−3dB
Temperature measurement: DC to 1kHz +1/−3dB
Input coupling
DC, GND, AC
Max. allowable input
30Vrms or 60V DC (the maximum voltage that can be applied across input pins without damage)
Note: Sampling period 1/100 of time axis range (min. 1μs)
FFT functions Additional functions provided by the FUNCTION UP DISK 9540-01
Analysis mode
Storage waveform, Linear spectrum, RMS spectrum, Power
spectrum, Cross-power spectrum, Auto-correlation function,
Histogram, Transfer function, Cross-correlation function, Unitimpulse response, Coherence function, Octave analysis
Analysis channels 1 or 2 selected channels out of all analog channels
(using 12-bit A/D conversion; installed in the 8835-01)
(simultaneous sampling in 2 channels)
Frequency range
133mHz to 400kHz, External, (resolution 1/400, 1/800, 1/2000, 1/4000)
Number of
sampling points
1000, 2000, 5000, 10000 points
Analysis data
Selected from: Newly loaded data / MEM function waveform
data / MEM waveform of REC & MEM function
Windows
Rectangular, Hanning, Exponential
STRAIN UNIT 8939
Time axis / frequency axis simple averaging, exponential averaging,
peak hold
Measurement
functions
Number of channels: 2, for distortion measurement (electronic
Input connectors
Via conversion cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to
earth: 30Vrms or 60V DC (with input isolated from the unit, the maximum voltage that can
Suitable
transducer
Strain gauge converter, bridge impedance: 120Ω to 1kΩ, bridge
voltage 2 ±0.05V
Measurement
range
50με to 2000με/DIV, 6 ranges, full scale: 10DIV, low-pass filter:
10Hz/30Hz/300Hz/3kHz
Averaging
function
Additional features Additional functions provided by the FUNCTION UP DISK 9540-01
Waveform processing
calculations
(MEM function)
(Maximum possible calculation up to 1000 DIV; accuracy is within the tolerance of
the input module.)
Four arithmetic operations, absolute value, exponentiation, common
logarithm, square root, moving average, differentiation once and
twice, integration once and twice, parallel displacement along the
time axis. 8 arbitrary operational equation.
(MEM function)
(FFT function)
Type: Area judgment using reference waveform for time axis
waveform, X-Y plot, or FFT display. Parameter judgment for
waveform parameter processing. Judgment output: pass/fail
output, open-collector 5V voltage output
Others
Waveform averaging; memory segmentation (up to 255 segments)
Waveform judgment
function
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 250 g (8.8 oz) Accessories: Conversion cable × 2
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up
time and auto-balance; accuracy guaranteed for 1 year)
auto-balancing, balance adjustment range within ±10000με)
be applied between input channel and chassis and between input channels without damage)
Measurement resolution 1/160 of measurement range (using 12-bit A/D conversion; installed in 8835-01)
Highest sampling rate
1MS/s (2-channel simultaneous sampling)
Accuracy After auto-balancing DC amplitude: ±(0.5% of full scale +2με), zero position: ±0.5% of full scale
Frequency characteristics DC to 20 kHz +1/−3dB
Max. allowable input
10V DC + AC peak (the maximum voltage that can be applied across input pins without damage)
CONVERSION CABLE 9318 (to connect 9270 to 9272, 9277 to 9279 and 8940)
CONVERSION CABLE 9319 (to connect 3273, 3273-50 and 8940)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 300 g (10.6 oz) Accessories: None
■ Input unit specifications (sold separately)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 290 g (10.2 oz) Accessories: None
ANALOG UNIT 8936
Input connectors
Measurement
functions
Number of channels: 2, for voltage input based frequency measurement, integration,
pulse duty ratio, current (with optional clamp-on sensor), and voltage measurement
Input connectors
Metallic BNC connector (input impedance 1MΩ, input capacitance 60pF), sensor
connector (dedicated connector for clamp-on sensor via conversion cable, common
ground with recorder), Max. rated voltage to earth: 30Vrms or 60V DC (with
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
time and zero-adjust; accuracy guaranteed for 1 year)
Measurement functions Number of channels: 2, for voltage measurement
Isolated BNC connector (input impedance 1MΩ, input capacitance 30pF), Max. rated
voltage to earth: 370V AC, DC (with input isolated from the unit, the maximum voltage that
10mV to 50V/DIV, 12 ranges, full scale: 10DIV, AC voltage for
possible measurement/display using the memory function: 280V
rms, low-pass filter: 5Hz/500Hz/5kHz/100kHz
Measurement
range
Measurement resolution 1/160 of measurement range (using 12-bit A/D conversion; installed in the 8835-01)
Highest sampling rate
1MS/s (simultaneous sampling in 2 channels)
Accuracy
DC amplitude: ±0.4% of full scale, zero position: ±0.1% of full scale (after zero adjustment)
DC, GND, AC
Max. allowable input
400V DC (the maximum voltage that can be applied across input pins without damage)
maximum voltage that can be applied across input pins without damage)
FFT ANALOG UNIT 8938
from the 8940, max. 4 sensors total)
Measurement
range
Anti-aliasing filter
Integrated filter for suppressing aliasing distortion caused by FFT
processing (automatic cutoff frequency setting/OFF)
Other functions
Other specifications same as the ANALOG UNIT 8936
DC amplitude accuracy: ±0.4% of full scale, zero position ±0.15%
of full scale (current measurement accuracy dependent on sensor accuracy/
characteristics)
Frequency characteristics: DC to 400kHz ±3dB
Full scale: 10DIV, Max. allowable input: 30Vrms or 60V DC (the
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
time and zero-adjust; accuracy guaranteed for 1 year)
Measurement functions Number of channels: 2, for voltage measurement
Frequency: DC to 100kHz, with 0.1Hz to 10kHz/DIV, 11 ranges, 10 (r/
min) to 1k (r/min)/DIV, 5ranges, P50Hz (40 to 60Hz), P60Hz (50 to 70Hz)
* Power line frequency measurement requires the DIFFERENTIAL PROBE 9322 or PT 9303 ,
Accuracy: ±0.2% of full scale (except 10kHz/DIV range), ±0.7% of full
scale (10kHz/DIV range), ±0.032Hz (P50Hz, P60Hz range)
Integration: DC to 90kHz, with 10counts to 1Mcounts/DIV, 11 ranges
Pulse duty ratio: 10Hz to 100kHz, with 100% of full scale, 1 range,
Accuracy: ±1% of full scale (10Hz to 10kHz)
Threshold: −10 to +10V (settable in 0.2V steps)
Full scale: 10DIV, Max. allowable input: 30Vrms or 60V DC (the
Voltage: 1mV to 5V/DIV, 12 ranges
Current: 10mA to 200A/DIV, 10 ranges, using current sensor (powered
Frequency characteristics DC to 400kHz ±3dB, with AC coupling: 7Hz to 400kHz ±3dB
Input coupling
input isolated from the unit, the maximum voltage that can be applied between input
channel and chassis and between input channels without damage)
Compatible current sensors 9270, 9271, 9272, 9277, 9278, 9279, 3273, 3273 -50
can be applied between input channel and chassis and between input channels without damage)
Measurement
range
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
time and zero-adjust; accuracy guaranteed for 1 year)
F/V UNIT 8940
maximum voltage that can be applied across input pins without damage)
Measurement resolution 1/160 of measurement range (installed in 8835-01, excluding current range when using 9279)
1MS/s (simultaneous sampling in 2 channels), (frequency/duty ratio
Highest sampling rate measurement: 1.125μs cycle)
Other functions
Voltage input pull-up: ON (10kΩ)/OFF, input coupling: DC, GND, AC
(voltage/current), DC (others), low-pass filter: 5Hz/500Hz/5kHz/100kHz
* The 8940 can be used with the 8835-01. but the 8835, main unit only, current probe cannot be used. The 8940 can be used with the 8835
standard models later than Ver. 2.10, 9540 install models later than Ver.5.10 can be used.
11
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 310 g (10.9 oz) Accessories: None
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up
time and zero-adjust; accuracy guaranteed for 1 year)
4ch ANALOG UNIT 8946
Measurement functions Number of channels: 4, for voltage measurement
Input connectors
Metallic BNC connector (input impedance 1MΩ, input capacitance 15pF),
Max. rated voltage to earth: 30Vrms or 60V DC (with input isolated
Measurement
range
20mV to 5V/DIV, 8 ranges, full scale: 10DIV, low-pass filter,
5Hz/500Hz/5kHz/50kHz, input coupling: DC, GND
from the unit, the maximum voltage that can be applied between input channel
and chassis and between input channels without damage)
Measurement resolution 1/160 of measurement range (using 12-bit A/D conversion; installed in the 8835-01)
Highest sampling rate
1MS/s (simultaneous sampling in 4 channels)
Accuracy
DC amplitude: ±0.5% of full scale, zero position: ±0.15% of
full scale (after zero adjustment)
Frequency characteristics DC to 100kHz ±3dB
Max. allowable input
30Vrms or 60V DC (the maximum voltage that can be applied across input pins without damage)
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,
approx. 310 g (10.9 oz) Accessories: None
CHARGE UNIT 8947
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up
time and zero-adjust; accuracy guaranteed for 1 year)
Measurement functions Number of channels: 2, for acceleration measurement
Voltage input/integrated preamplifier input: metallic BNC connector
Input connectors
Suitable
transducer
Measurement
range
Charge input
(miniature connector)
Internal pre-amp input
(BNC connector)
Measurement
range
Voltage input (BNC connector)
(for voltage input: input impedance 1MΩ, input capacitance 200pF or less)
Charge input: miniature connector (#10-32 UNF)
Max. rated voltage to earth: 30Vrms or 60V DC (with input isolated
from the unit, the maximum voltage that can be applied between input channel and
chassis and between input channels without damage)
Charge input: Charge-output type piezoelectric acceleration pick-up sensor
Internal preamp input: Acceleration pick-up sensor with an internal preamp
100m (m/s2)/DIV to 20k (m/s2)/DIV, 12 ranges × 6 types, charge input sensitivity:
0.1 to 10 pC/(m/s2), integrated pre-amplifier input: 0.1 to 10 mV/(m/s2),
amplitude accuracy: ±2% of full scale, frequency characteristics: 1 to 50kHz,
+1/−3dB, low-pass filter: 500Hz/5kHz, pre-amplifier drive power source:
2mA ±20%, +15V ±5%, maximum input charge: ±500pC (high-sensitivity setting,
6 ranges), ±50000pC (low-sensitivity setting, 6 ranges)
1mV to 5V/DIV, 12 ranges, DC amplitude accuracy: ±0.4% of full scale, frequency
characteristics: DC to 400kHz, +1/−3 dB, low-pass filter: 5Hz/500Hz/5kHz/100kHz,
input coupling: DC, GND, AC, Max. allowable input: 30Vrms or 60V DC
1/160 to 1/64 of measurement range (depending on measurement
Measurement resolution sensitivity; installed in the 8835-01)
Highest sampling rate
1MS/s (simultaneous sampling in 2 channels)
Integrated filter for suppressing aliasing distortion caused by FFT
Anti-aliasing filter processing (automatic cutoff frequency setting/OFF)
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 30 cm
(0.98 ft), approx. 150 g (5.3 oz)
LOGIC PROBE 9320
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh; accuracy guaranteed for 1 year)
Function
Detection of voltage signal or relay contact signal for High/Low state recording
Input
4 channels (common ground between unit and channels), digital/contact input,
switchable (contact input can detect open-collector signals), input impedance:
1MΩ (with digital input, 0 to +5V), 500kΩ or more (with digital input, +5 to
+50V), pull-up resistance: 2kΩ (contact input: internally pulled up to +5V)
Digital input threshold 1.4V/2.5V/4.0V
Contact input
detection resistance
1.5kΩ or higher (open) and 500Ω or lower (short), 3.5kΩ or higher (open) and
1.5kΩ or lower (short), 25kΩ or higher (open) and 8kΩ or lower (short)
Response speed
Detectable pulse width 500ns or lower
Max. allowable input
0 to +50V DC (the maximum voltage that can be applied across input pins without damage)
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 1 m (3.28 ft),
approx. 320 g (11.3 oz)
LOGIC PROBE 9321
Function
Input
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh; accuracy guaranteed for 1 year)
Detection of AC or DC relay drive signal for High/Low state recording
Can also be used for power line interruption detection
4 channels (isolated between unit and channels), HIGH/LOW range switching
Input impedance: 100kΩ or higher (HIGH range), 30kΩ or higher (LOW range)
LAN COMMUNICATOR 9333
Distribution media
One CD-R
Operating
environment
Computer equipped with Pentium (133 MHz) or better CPU, running
under Windows 95/98/Me or Windows NT 4.0/ 2000/XP operating
system, with network adapter installed and configured to use TCP/IP
protocol, and at least 64 MB of memory.
HiCORDER side
Standard LAN connector, LAN card
Output (H)
detection
170 to 250V AC, ±DC (70 to 250V ) (HIGH range)
60 to 150V AC, ±DC (20 to 150V) (LOW range)
Communications
Ethernet, TCP/IP
Output (L)
detection
0 to 30V AC, ±DC (0 to 43V) (HIGH range)
0 to 10V AC, ±DC (0 to 15V) (LOW range)
Remote control
Response time
Rising edge 1ms max., falling edge 3ms max. (with HIGH range at
200V DC, LOW range at 100V DC)
Maximum allowable 250Vrms (HIGH range), 150Vrms (LOW range) (the maximum voltage
that can be applied across input pins without damage)
input voltage
Cable length and mass: Main unit cable 1.3 m (4.27 ft), input section cable 46 cm
(1.51 ft), approx. 350 g (12.3 oz)
DIFFERENTIAL PROBE 9322
Function
Waveform data
acquisition
Waveform viewer
Remote control of MEMORY HiCORDER (by sending key codes and receiving
print reports, print images from the screen, receive
waveform data in same format as waveform files from the MEMORY
HiCORDER (binary only)
Accept auto-saves from the MEMORY HiCORDER, same format
as auto-save files of MEMORY HiCORDER (binary only), print
automatically with a MEMORY HiCORDER from a PC. The
MEMORY HiCORDER's print key launches printouts on the PC
images on screen) ,
Simple display of waveform files, conversion to CSV format, Scroll
function, enlarge/reduce display, display CH settings.
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh, after 30 minutes of warm-up time; accuracy guaranteed for 1 year)
For high-voltage floating measurement, power line surge noise
detection, RMS rectified output measurement
DC mode
For waveform monitor output, frequency characteristics: DC to 10MHz
(±3dB), amplitude accuracy: ±1% of full scale (at max. 1000V DC), ±3%
of full scale (at max. 2000V DC) (full scale: 2000V DC)
AC mode
For detection of power line surge noise, frequency characteristics: 1kHz
to 10MHz ±3dB
RMS mode
DC/AC voltage RMS output detection, frequency characteristics: DC, 40Hz to
100kHz, response speed: 200ms or less (400V AC), accuracy: ±1% of full scale
(DC, 40Hz to 1kHz), ±4% of full scale (1kHz to 100kHz) (full scale: 1000V AC)
Input
Input type: balanced differential input, input impedance/capacitance:
H-L 9MΩ/10pF, H/L-unit 4.5MΩ/20pF, Max. rated voltage to earth:
when using grabber clip 1500V AC/DC (CAT II ), 600V AC/DC (CAT III),
when using alligator clip: 1000V AC/DC (CAT II), 600V AC/DC (CAT III)
Maximum allowable 2000V DC, 1000V AC (CAT II)
600V AC/DC (CAT III)
input voltage
Output
Voltage divider for 1/1000 of input, BNC connectors (output switchable for
Power source
Use of the AC ADAPTER 9418 -15 (12V DC)
3 modes DC, AC, RMS)
WAVE PROCESSOR 9335
Distribution media
One CD-R
Operating
environment
Computer equipped with Pentium (133 MHz) or better CPU and at least
32 MB of memory, and running under Windows 95/98/Me, Windows
NT 4.0/2000/XP, or Windows Vista 32-bit type (recommended system:
Pentium (200 MHz) or better with at least 64 MB of memory)
Display functions
Waveform display/X-Y display/digital value display/cursor function/
scroll function/maximum number of channels (32 channels analog, 32
channels logic)/gauge display (time, voltage axes)/graphical display
File loading
Readable data formats (.MEM, .REC, .RMS, .POW)
Maximum loadable file size: Maximum file size that can be saved by
a given device (file size may be limited depending on the computer
configuration)
Data conversion
Conversion to CSV format, tab delimited, space delimited/data culling
(simple)/convert for specified channel/batch conversion of multiple files
Print functions
Print formatting (1 up, 2-to-16 up, 2-to-16 rows, X-Y 1-to-4 up) /preview/
hard copy functions usable on any printer supported by operating system
Other
Parameter calculation/search/clipboard copy/launching of other
applications
12
Composition of options
Note: Product names appearing herein are trademarks or registered trademarks of various companies.
Logic signal measurement
Input modules
Install by inserting into the main unit. Can be replaced by user.
Note: Input cables are not supplied. Please purchase the appropriate cable
for the intended application.
ANALOG UNIT 8936
VOLTAGE/TEMP UNIT 8937
FFT ANALOG UNIT 8938
STRAIN UNIT 8939
F/V UNIT 8940
4ch ANALOG UNIT 8946 (The 8946 can be used with the
LOGIC PROBE 9320
LOGIC PROBE 9321
4-channel type, for voltage/contact
signal ON/OFF detection
8835-01, but cannot be used with the 8835.)
4 isolated channels, ON/OFF detection
of AC/DC voltage
Current measurement * The 3274, 3275, and 3276 cannot be used for the F/V UNIT 8940
CHARGE UNIT 8947
F/V UNIT 8940
Voltage measurement
9279 Not CE certified
3273-50
9277
CONVERSION CABLE 9318 (to connect 9277 - 9279 and 8940) 9278 CONVERSION CABLE 9319 (to connect 3273-50 and 8940)
UNIVERSAL CLAMP ON CT 9277 CLAMP ON PROBE 3273-50
DC to 50MHz wideband response,
Observe waveforms from DC to distorted
mA-class current up to 30A rms
AC. DC to 100kHz response, input 20A /
output 2V AC
CONVERSION CABLE 9319
to connect 3273-50 and 8940
PROBE 9322
CONNECTION CORD DIFFERENTIAL
For up to 2kV DC or 1kV AC, AC adapter
9197
9418-15 required for operation.
UNIVERSAL CLAMP ON CT 9278
Observe waveforms from DC to distorted
AC. DC to 100kHz response, input 200A /
output 2V AC
For high voltage (up to 500V)
AC ADAPTER 9418-15
For powering Differential probe 9322,
100 to 240V AC, 12V/2.5 A
CONNECTION CORD
9198
For low voltage (up to 300V)
POWER CORD
9324
Use to the 9322 from the logic terminal.
* Note that usage restrictions apply.
POWER CORD
9325
Observe waveforms from DC to distorted
AC. DC to 20kHz response, input 500A /
output 2V AC
MEMORY HiCORDER 8835-01 (main unit)
* The MEMORY HiCORDER 8835-01 cannot operate alone.
F/V UNIT 8940
CLAMP ON SENSOR 9271
5Hz to 50kHz response, up to 200A rms,
2V AC output
Insulation transformer, 400V or
200V AC input, 10V AC output,
for AC power line measurement.
Input from 10 to 500A, 40Hz to 3kHz for
0.2V AC output. BNC terminal
1Hz to 100kHz response, selectable 20 and CLAMP ON PROBE 9132-10
Input from 20 to 1000A, 40Hz to 1kHz for 0.2V
AC output. BNC terminal, Not CE certified.
200A rms ranges, 2V AC output
CONVERSION CABLE 9318
Not CE certified
CLAMP ON LEAK SENSOR 9657-10
to connect 9270/9271/9272 and 8940
TRIGGER CORD 9305
Supplied with
PC Card
adapter
CONVERSION ADAPTER 9199
Banana-to-BNC, use to connect to
insulation-BNC terminal on Input section
PC Card Precaution
Communication
3274
3275
3272
3273-50
3276
3269
CLAMP ON PROBE 3273-50
CLAMP ON PROBE 3276
CLAMP ON PROBE 3274
POWER SUPPLY 3272
CLAMP ON PROBE 3275
POWER SUPPLY 3269
DC to 50MHz wideband response,
mA-class current up to 30A rms
DC to 10MHz wideband response,
mA-class current up to 150A rms
DC to 2MHz wideband response,
mA-class current up to 500A rms
DC to 100MHz wideband response,
mA-class current up to 30A rms
Connect and power up to one Clamp-On Probes
to use in combination with voltage input modules
(up to two PROBEs restrictively)
Connect and power up to four Clamp-On Probes
to use in combination with voltage input modules
For leak current only, 10A for rated current,
40Hz to 5kHz, output rate 100mV/A AC,
BNC terminal
φ 3.5mm (0.14in) mini-plug, 1.5m
(4.92 ft) length
Printing
Model 9555-10 is required to power the ClampOn CTs, except for connecting Clamp-On CT
Model 9277/78/79 to Model 8940, and for
signal output, Model 9217 Connection Cord is
required to connect for input modules.
to connect 9277/9278/9279 and 8940
9555-10
CLAMP ON PROBE 9018-10
CLAMP ON SENSOR 9272-10
CONNECTION CORD 9165
Metal BNC connectors at both
ends, and connects to metal
BNC connectors
9272-10
9270, 9271
Not CE certified
SENSOR UNIT 9555-10
CONVERSION CABLE 9318 (to connect 9270 - 9272 and 8940)
Model 9555-10 is required to power the Clamp-On
CLAMP ON SENSOR 9270
Sensors, except for connecting Clamp-On Sensor
5Hz to 50kHz response, up to 20A rms, 2V
Model 9270, 9271, 9272-10 to Model 8940, and
for signal output, Model 9217 Connection Cord is
AC output
required to connect for input modules.
PT 9303
Not CE certified
SENSOR UNIT 9555-10
CONVERSION CABLE 9318
Current measurement
Not CE certified
Insulated BNC connectors at
both ends, and connects to
insulated BNC connectors
Cord has insulated BNC connectors
at both ends, and connects to the
9555-10 and input moduls.
UNIVERSAL CLAMP ON CT 9279
Use to the 9322 from the F/V UNIT 8940.
* Note that usage restrictions apply.
CONNECTION CORD 9217
CONNECTION CORD 9217
U s e o n l y P C C a rd s s o l d
by HIOKI . Compatibility
and performance are not
guaranteed for PC cards made
by other manufacturers. You
may be unable to read from or
save data to such cards.
PC CARD 128M 9726
(128 MB capacity)
PC CARD 256M 9727
(256 MB capacity)
PC CARD 512M 9728
(512 MB capacity)
PC CARD 1G 9729
(1 GB capacity)
FUNCTION UP DISK 9540 -01
For the model 8835-01, use the
FUNCTION UP DISK 9540 with the
the model 8835.
DC POWER ADAPTER 9439 CARRYING CASE 9388
Supplies operating power in the range
10 to 28V DC.
With casters for convenient
transportation.
RECORDING PAPER 9221
110 mm (4.33 in) × 30 m (98.43 ft),
10 rolls set
Not CE certified
Not CE certified
PAPER WINDER 220H
Paper width 70mm (2.75 in) to
220mm (8.66 in), using specialpurpose AC adapter
RS-232C CARD 9557 GP-IB CARD 9558
LAN CARD
PCMCIA-compliant
May not operate depending on the
environment., 10BASE-T
PCMCIA-compliant, cord
length: 2m (6.6ft)
LAN CABLE 9642
LAN COMMUNICATOR 9333 WAVE PROCESSOR 9335
Application software to create a LAN
Supplied with cross conversion connection with Windows 95/98/Me, or
cable, straight Ethernet cable,
Windows NT 4.0/2000/XP.
length: 5m (16.4ft)
Data conversion, print functions, waveform
display, compatible with Windows 95/98/Me,
Windows NT 4.0/2000/XP, and Windows
Vista 32-bit type.
■ Combination example: 8835-01 (normal choice of the input unit, up to 200V AC direct input.)
Model number × quantity
Input cable
Main unit
8835-01×1
Memory 4MW
standard
2ch
8936×1
9198×2
4ch
8936×2
9198×4
■ Combination example: 8835-01 (select an automotive input unit that can handle up to 30 V AC.)
Model number × quantity
Input cable
Main unit
8835-01×1
Memory 4MW
standard
4ch
8946×1
9198×4
8ch
8946×2
9198×8
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: [email protected]
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: [email protected]
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: [email protected]
DISTRIBUTED BY
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: [email protected]
Guangzhou Office :
Room 303, Profit Plaza, No.76, West Huangpu Road
Guangzhou, P.R.China POSTCODE: 510623
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: [email protected]
All information correct as of Sep. 6, 2007. All specifications are subject to change without notice.
8835-01E11-79B-00P