ASM-SENSOR 8206-10

MICRO HiCORDER 8205-10, 8206-10
Recorders
Direct Measurement of Current up to 1000A !
New and improved
Successor to the long-popular Model 8202 & 8203 MICRO HiCORDERs
True RMS Voltage and Current Recorders
with Multimeter Operability !
The MICRO HiCORDERs 8205 -10 and 8206 -10 incorporate the latest technology in HIOKI’s
best selling MICRO HiCORDER 8200 series. Maintaining the operational simplicity of an analog multimeter, these data recorders offer easy to use features such as;
l A fast LCD level meter vs. a stylus to quickly display the measured level
l True RMS rectification system to accurately read distorted waveforms
l Climate-resistant recording paper is easy to read and lasts longer!
l Sampling rate of 100 samples/second with selectable chart speeds
There are two models to choose from. The 8205-10 has one channel for recording either AC/DC
voltage or AC current, while the 8206-10 has two channels for recording of AC voltage and AC
current simultaneously, allowing it to be used for recording power line fluctuations. Packaged in a
rugged case, the HIOKI MICRO HiCORDERs work where you work !
The Standard for Recorders
Depending on your application, select either the 8205-10 for general applications or the 8206-10 for power lines.
: For Recording AC Voltage, DC Voltage, or AC Current on a Single Channel
The 8205 -10 can be used for basic recording in a wide range of applications, with
features such as a wide AC/DC measurement range of 0.1V to 500V, an analog level
monitor function, and a special clamp-on probe for recording large currents.
●Range:AC/DC 0.1V to 500V (AC 10A to 500A when using the 9651, AC 10A to 1000A when using
the 9668) ●Input Channels: One, for voltage or current
●Paper Feed Speed: 20 cm/min to 2 cm / hr ●Sampling Frequency: 10ms
Easy viewing of the
input level with a fullrange level monitor.
For DC measurement,
the zero position can
be changed in steps
that are -20% of the
range, which is convenient for tasks such as
recording both polarities.
Example of a level display for the DC 10V range
Measurements are possible
from -9V to 11V by moving the
zero position.
0%
Standard position
11V
•Range and mode setting data is
printed every 10 divisions.
•Voltage or current values can be
read directly from the vertical axis
scale.
•Time is printed every 5 divisions.
•Interval printing of average values
on a gr id with points ever y 2
divisions.
-40%
Intermediate position
-80%
Minimum position
0V
-9V
: For 2-Channel Special Purpose Recording of AC Voltage and Current from Power Lines
The 8206 -10 provides simultaneous voltage and current measurement (a frequent requirement
For voltage measurements,
of power line management) along with a zoom function that allows 2X recording centered
the display can be
around the range value. Providing ranges that are expressly intended use with power lines, the
expanded from 25% to
instrument is capable of recording voltages of up to 600V.
-35% around the center
●Range: AC 100V, 200V, or 500V (AC 10A to 500A when using the 9651, AC 10A to 1000A when
using the 9668) ●Input Channels: Two, for simultaneous recording of voltage and current
●Paper Feed Speed: 60 cm/hr to 2 cm/hr ●Sampling Frequency: 10ms
of the range. The
center va lue ca n be
changed in steps that
are -10% of the range, which is convenient when
it is necessary to adjust the measurement position
for electric circuits with different rated voltages.
Example of a level display for the 200V range
M e a su re m e nts a re p o s s i b l e
from 90V to 290V by changing
the center value.
290V
0%
Standard position
•Range and mode setting data is
printed every 10 divisions.
•Voltage or current values can be
read directly from the vertical
axis scale.
•Time is printed every 5 divisions.
•Interval printing of average values
on a gr id with points ever y 2
divisions.
-20%
Minimum position
200V
90V
+20%
Maximum position
Simple Operation, Rugged Design and Quiet !
The 8206-10 has two recording channels,
one for voltage and one for current.
(clamp on sensor sold separately)
Operational simplicity with
multimeter ease-of-use.
Intuitive analog dials are used to set the input
range and paper feed speed. This makes the
recorders as easy to operate as a multimeter.
With 2-channel recording, a single instrument
can simultaneously record both voltage and
current. This is ideal for management and
testing of power lines.
A sensor allow measurement up to
1000A.
(clamp on sensor sold separately)
Sensitized recording paper
provides clean, easy-to-read
printouts.
■ Paper feed speed and recording time
*With a time axis resolution of 80 points / DIV (1DIV=10mm)
8205-10
8206-10
9235,9236 -01 Sensitizedpaper
15m (49ft 2.5in) roll
20 cm (7.9 in)/m
−
−
1 hr 15 m
Paper feed speed Paper feed speed
6 cm (2.36 in)/m
60 cm (23.8 in)/m 60 cm (23.8 in)/hr 4 hr 10 m
1 day 1 hr
−
60 cm (23.8 in)/hr 3 days 3 hr
−
6 cm (2.36 in)/hr 10 days 10 hr
10 cm (3.94 in)/hr 10 cm (3.94 in)/hr 2 cm (0.79 in)/hr
2 cm (0.79 in)/hr 6 days 6 hr
31 days 6 hr
Adaptable for Special Applications
The recorder provides selectable current ranges.
Used together with a special clamp on sensor,
this makes it possible to read values directly and
eliminates the need for scaling conversion. In
addition to 100A and 500A sensors, a sensor has
been added to allow measurement up to 1000A.
Sensitized recording paper is used, providing clean,
quiet recording. Further, HIOKI's RECORDING
PAPER 9236 -01 and provides superior climate
resistance and is much less susceptible to
discoloration due to chemical exposure than
ordinary thermal recording paper, providing much
longer print life.
Standard/Average Print Modes
As a special-order option, the DC power supply connector
can be converted to a signal input terminal. Although this
makes it impossible to use a DC power supply, it allows
recording on paper to be turned on or off externally, or to
input a contact signal and record the occurrence of events
on recording paper. Please contact your nearest HIOKI
sales representive for details.
• Special order modification (disables DC power supply)
• External CHART ON / OFF control
• Event recording
Example of Instantaneous Power Outage Recording
Also runs on DC Power !
AC/DC power supply
selector switch
Power switch
DC power inlet
AC power inlet
Standard Mode
Average Mode
For printing, you can select standard mode or average mode.
l Standard mode prints values interpolated from the maximum
and minimum values sampled every 10 ms. This is effective
for recording widths of steep input signal changes, such as
instantaneous power outages.
l Average mode prints the average value of the data during
the print interval. This makes it possible to print something
that is smooth and easy to see, at times when there are violent
fluctuations.
* Switch to average mode by holding down the paper feed button when turning on
the power.
50 Hz
60 Hz
This example shows recording of a
simulated instantaneous power outage by
the 8206, for AC 100V with frequencies of
50Hz or 60Hz and 0V durations of 0.5, 1, 2,
4, or 8 cycles.
Even for a 0.5 cycle instantaneous power
failure, the high speed response can
catch the line abnormality.
In addition to a universal power
AC supply of 100 to 240V, the
recorders can operate on 9.5 to
14V DC power supplies. This is
useful outdoors and in other field
applications.
* Runs off either AC or DC. DC cannot be
used as a backup power supply.
■ External appearance and dimensions (for both the 8205-10 and 8206-10)
8205-10 and 8206-10 Common Specifications
Recording method 60mm (2.36in) amplitude (1 DIV = 10mm/0.39in), heat-sensitive recording
Sampling rate
Level meter and scaling values by LCD bar graph, plus other
setting information
122 mm
(4.8 in)
Display
100S/s (sampling period : 10ms fixed)
Recording time axis accuracy Within ±0.5%.
Environmental conditions Operation: +5°C (41°F) to +40°C (104°F), 35% to 80% rh
Storage: -10°C (14°F) to +50°C (122°F), within 80% rh
(non-condensating)
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
250 mm (9.4 in)
100 to 240V AC (auto selecting) at 50/60 Hz, or 9.5 to 14V DC
93.5 mm
(3.66 in)
Applicable
standards
Power
Maximum power consumption 30 VA maximum (AC or DC operation)
Dimensions and mass 250mm (9.4in) × 122mm (4.8in) × 93.5mm (3.66in), 1.2kg (42.33oz)
RECORDING PAPER 9235 × 1, Roll paper holder × 2, Power cord × 1,
Included
CONNECTION CORD 9257 × 1, CARRYING CASE 9344 × 1
Accessories
8205 -10 (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh ; accuracy guaranteed for 1 year)
Input channels
8206 -10 (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh ; accuracy guaranteed for 1 year)
One channel, AC or DC voltage, or one channel AC current.
One channel AC voltage and one channel AC current.
(Simultaneous recording not available, one channel may be either type, input is voltage isolated.)
(Simultaneous recording based on two channel alternating sampling, for commercial power lines, input is voltage isolated.)
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, or 500V AC/DC. In DC mode, the zero 100, 200, or 500V AC, Max. input voltage: 600V rms. Magnified display from
position can be set in steps of 20% of the range.
+25% to -35% of the range. Center position of the magnified display can be set
Voltage
Max. input voltage: 500V rms. Accuracy: ±2% of the range. (ACV/ 45Hz to 66Hz) in 10% steps from +20% to -20%. Accuracy: ±2% of the range. (45Hz to 66Hz)
measurement ranges Frequency characteristic: 20Hz to 30kHz/ +3 dB to -3 dB
Frequency characteristic: 30 Hz to 30 kHz/ +0.5 dB to -3 dB
* Resolution = 400points/ range
Max. rated voltage to earth: 500V rms (the maximum voltage that can be applied be- Max. rated voltage to earth: 600V rms (the maximum voltage that can be applied between input channel and chassis without damage)
tween input channel and chassis without damage)
Combination accuracy with the clamp-on sensor (% of full scale value)
10, 20, 50, or 100A AC (with CLAMP ON SENSOR 9650).
10, 20, 50, 100, 200, or 500A AC (with CLAMP ON SENSOR 9651).
Current
measurement ranges 10, 20, 50, 100, 200, 500 or 1000A AC (with CLAMP ON SENSOR 9668).
* Resolution = 400points/ range Bar graph displays from 0% to 110% of range.
Frequency characteristic is determined by the clamp on sensor.
Range
10A
20A
50A
100A
200A
500A
1000A
9650
±3.8%
±3.65%
±3.56%
±3.53%
-
-
-
9651
±5.0%
±4.25%
±3.8%
±3.65%
±3.58%
±3.53%
-
9668
±8.0%
±6.5%
±5.6%
±5.3%
±5.15%
±5.06%
±6.03%
True RMS.
Rectification system
Input resistance
Voltage input: approx. 1MΩ. Clamp input: 1Ω ±10%.
Ordering Information
Options
Supplied as standard
CONNECTION CORD
9326
MICRO HiCORDER 8205 -10 (1-channel recording)
MICRO HiCORDER 8206 -10 (2-channel simultaneous recording)
(1.7m/ 5.58ft, for 8205 and
8205-10 voltage input)
CONNECTION CORD
9257
The CLAMP ON SENSOR 9650,9651 and 9668 for current measurement are not
supplied with the MICRO HiCORDERs 8205 -10 and 8206 -10. To measure current,
an optional CLAMP ON SENSOR must be purchased separately.
(1.3m/ 4.27ft, voltage input,
Supplied as standard)
RECORDING PAPER 9236 -01
(Climate-resistant, 15m/ 49ft roll, 10
rolls/ set)
CARRYING CASE 9344
Holds optional accessories
(Supplied as standard)
RECORDING PAPER 9235
(15m (49.22ft) roll, 10 rolls/
sets; one roll only, provided
with recorder)
■ Option Specifications (sold separately)
CLAMP ON SENSOR 9650
Individual Specifications
CLAMP ON SENSOR 9651
9650
CLAMP ON SENSOR 9668
9668
Primary: 500A AC, Secondary: 500mA AC
Primary: 1000A AC, Secondary: 1000mA AC
Frequency characteristic ±8% or better from 40Hz to 1kHz
±3% or better from 40Hz to 1kHz
±3% or better from 40Hz to 1kHz
Application circuit voltage 300V rms AC or less (insulated conductor)
600V rms AC or less (insulated conductor)
Rated current
Secondary current
amplitude accuracy
Primary: 100A AC, Secondary: 100mA AC
9651
Combined accuracy with recorder depends on recorder Combined accuracy with recorder depends on recorder Combined accuracy with recorder depends on recorder
accuracy.
accuracy.
accuracy.
Maximum input current 130A continuous at 45 to 66Hz
600A continuous at 45 to 66Hz
1000A continuous at 45 to 66Hz
600V rms AC or less (insulated conductor)
ø 55mm (2.17 in) or less,
ø 46mm (1.81 in) or less
80mm (3.15 in) × 20mm (0.79 in) busbar
Approx. W 46mm (1.81 in) × H 135mm (5.31 in) × D 21mm (0.83 in) Approx. W 77mm (3.03 in) × H 151mm (5.94 in) × D 42mm (1.65 in) Approx. W 46mm (1.81 in) × H 135mm (5.31 in) × D 21mm (0.83 in)
Cable length: 3m (9.84 ft), mass: 230g (8.1 oz)
Cable length: 3m (9.84 ft), mass: 360g (12.7 oz)
Cable length: 3m (9.84 ft), mass: 230g (8.1 oz)
Diameter of measurable ø 15mm (0.59 in) or less
conductor
Dimensions and mass
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: [email protected]
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: [email protected]
Shanghai Representative Office :
1904, Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
E-mail: [email protected]
All information correct as of Mar. 5, 2007. All specifications are subject to change without notice.
8205-10E7-73B-00P
Printed in Japan