Agilent U1270 Series Handheld Digital Multimeters Shaped to fit, tailored to perform, built to last Data Sheet Features • Intelligent features: ZLOW*, Smart Ω*, Low Pass Filter, Qik-V** • Visual (Backlight Alert) and audible continuity indication in noisy environments • Ergonomic shape for better grip • IP 54 certified – water and dust resistant • Easy-to-operate large knob and buttons • 30,000-count dual display • AC + DC capability* • CAT III 1000 V, CAT IV 600 V safety rating • Easy maintenance – convenient fuse access • Easy connectivity to PC with optional IR-USB cable • Internal memory for data logging * U1272A only ** U1271A only Get a better grip on your DMM The U1270 series is shaped perfectly to fit in your hand, with or without gloves on. Its nonslip ergonomic shape enables you to carry your DMM and perform measurements on the go easily. Additionally, the controls are easy to operate due to the large knob and buttons. Improve productivity with intelligent features Designed for maximum efficiency and productivity in industrial settings, these DMMs offer convenient functions such as ZLOW to eliminate stray voltages, Smart Ω to minimize false readings due to leakage current and Qik-V to determine existence of AC and/or DC voltages. Continuity detection in noisy and dark places is made easy with the U1270 series’ loud beeper and Backlight Alert function which flashes the backlight to indicate continuity and improve safety. When it comes to Variable Frequency Drive (VFD) troubleshooting, the U1270 series has Low Pass Filter to handle the job with ease. Water and dust resistant The series’ tightly sealed design helps protect against water, dust and damage. Each DMM is IP 54 certified so that you can carry out test and measurement with confidence, even in harsh working conditions. Key functions Low Impedance (ZLOW) The U1272A is a dual input impedance digital multimeter. The DMM’s high input impedance is preferred in most electrical measurements because it would not load the circuit under test. However, to obtain accurate measurements on circuits that may contain stray voltages, the U1272A’s 2 kΩ low impedance function comes in handy. Stray voltages are usually found in non-energized electrical wiring adjacent to powered wires due to capacitive or inductive coupling between these wires. When a pair of test leads is placed between the open circuit and neutral conductor, the circuit is then complete and forms a voltage divider in conjunction with the input impedance of the multimeter. High input impedance multimeter is sensitive enough to measure voltage coupled into the disconnected conductor, thus giving an inaccurate indication of a live conductor. The low impedance function serves to eliminate false readings by dissipating the stray voltages, thus improves safety and measurement efficiency during voltage. Figure 1: U1272A helps you identify the presence of stray voltage on a disconnected wire running parallel with the wire powering up the VFD to an industrial motor. The image on the right shows the U1272A in low impedance mode. Low Pass Filter (LPF) The U1270 series offers a 1 kHz LPF or Low Pass Filter to provide accurate Variable Frequency Drive (VFD) output measurement. This function eliminates high frequency noise and harmonics. This ensures the efficiency of your motor filter as well. Figure 2: Comparison of voltage output from industrial motor VFD without and with Low Pass Filter functionality. Smart Ω The U1272A provides an additional 30 Ohm range for low resistance measurement. This Smart Ω function is available for ranges of 30 Ohm to 300 kOhm. It enhances measurement accuracy with offset compensation by removing residual DC voltages of up to 1000 mV induced by ground current and thermal EMF. This function also enables ‘live’ resistance measurement without isolating the measurement circuit. With this, you will be able to obtain leakage current using the secondary display. 2 Front and back panel description Dual display Backlight Alert flashes the backlight to provide visual indication in continuity test 30,000 counts resolution Large buttons Peak detect records transients as fast as 250µs Smart Ω** removes residual voltage of up to 1000 mV Low Pass Filter (LPF) removes unwanted high frequency signals Auto diode** automatically determines diode polarity Temperature measurement: J type thermocouple -200 to 1200 °C**, K type thermocouple -200 to 1732 °C ZLOW** provides both high and low impedance modes to eliminate stray voltages AC + DC** capability Large knob for ease of use Qik-V* allows AC+DC measurement to quickly check for energized voltage Front panel 3 * U1271A only ** U1272A only Hang hole Probe holder/storage IR-USB connectivity with optional IR-USB cable Slim, ergonomic design for better grip Easy fuse and battery access from battery cover beneath the stand Back panel 4 Choose between these two models Di U1271A U1272A 30,000 30,000 Yes Yes Basic Features Display resolution Auto/manual ranging Analog bar graph Yes Yes Backlight Yes Yes 20 kHz 100 kHz AC AC + DC AC bandwidth True RMS Measurements Voltage DC Range Accuracy 300 mV to 1000 V 0.05% + 2 cnts 30 mV to 1000 V 0.05% + 2 cnts Voltage AC Range Accuracy Bandwidth 300 mV to 1000 V 0.7% + 20 cnts 45 Hz to 20 kHz 30 mV to 1000 V 0.6% + 20 cnts 45 Hz to 100 kHz Current DC Range Accuracy 300 µA to 10 A 0.2% + 5 cnts 300 µA to 10 A 0.2% + 5 cnts Current AC Range Accuracy Bandwidth 300 µA to 10 A 0.9% + 25 cnts 45 Hz to 2 kHz 300 µA to 10 A 0.6% + 25 cnts 45 Hz to 2 kHz Resistance Range Accuracy 300 Ω to 100 MΩ 0.2 % + 5 cnts 30 Ω to 300 MΩ 0.2% + 5 cnts Frequency Range Accuracy 99.999 Hz to 999.99 kHz 0.005% + 5 cnts 99.999 Hz to 999.99 kHz 0.005% + 5 cnts Capacitance Range Accuracy 10 nF to 10 mF 1% + 2 cnts 10 nF to 10 mF 1% + 2 cnts Range Accuracy K: -200 to 1372 °C 1% + 1°C K: -200 to 1372 °C J: -200 to 1200 °C 1% + 1 °C Temperature Continuity with beeper Yes Yes Diode test Yes Yes Data Management Min/Max Recording Yes Yes Display Hold Yes Yes Peak Hold Yes Yes Manual Datalogging Yes Yes Null Yes Yes IR-USB IR-USB Yes Yes PC Connectivity % scale of 4-20 mA 5 Di U1271A U1272A Beep + Backlight Alert Yes Yes Low Pass Filter (LPF) Yes Yes Special Features ZLOW - Low impedance mode - Yes Smart Ω - Yes Yes - CAT III 1000 V, CAT IV 600 V CAT III 1000 V, CAT IV 600 V Yes Yes 4x AAA 4x AAA Qik-V Safety and Regulatory Over-voltage safety protection EN/IEC 61010-1:2001 compliance General Battery Operating temperature -20 °C to 55 °C, 0 to 80% R.H -20 °C to 55 °C, 0 to 80% R.H Standard Accessories Standard test leads, test probes with 19mm and 4-mm tips, K-type thermocouple and adapter, 4x AAA batteries, Certificate of Calibration, test report, Quick Start Guide Standard test leads, test probes with 19mm and 4-mm tips, K-type thermocouple and adapter, 4x AAA batteries, Certificate of Calibration, test report, Quick Start Guide 6 General Specifications Display Power consumption Battery Type Liquid crystal display (LCD) (with maximum reading of 33000 counts) 460 mVA maximum (with backlight enabled) • 4 × 1.5 V Alkaline battery (ANSI/NEDA 24A or IEC LR03), or • 4 × 1.5 V Zinc Chloride battery (ANSI/NEDA 24D or IEC R03) Battery Life • 300 hours typical (based on new Alkaline batteries for dc voltage measurement) • Low battery indicator will flash when the battery voltage drops below 4.4 V (approximately) Fuse • 10 × 35 mm 440 mA/1000 V 30 kA fast-acting fuse • 10 × 38 mm 11 A/1000 V 30 kA fast-acting fuse Operating Environment • Operating temperature from –20 to 55 °C, 0 to 80% RH • Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly to 50% RH at 55 °C • Altitude up to 2000 meters • Pollution degree II Storage Compliance Safety Compliance –40 to 70 °C, 0 to 80% RH • CAN/CSA-C22.2 No. 61010-1-04 • EN/IEC 61010-1:2001 • ANSI/UL 61010-1:2004 Measurement Category Electromagnetic Compatibility (EMC) Ingress Protection Rating Temperature Coefficient Common Mode Rejection Ratio (CMRR) Normal Mode Rejection Ration (NMRR) Dimensions (W x H x D) Weight CAT III 1000 V/ CAT IV 600 V Commercial limits compliance with EN61326-1 IP-54 0.05 × (specified accuracy) / °C (from –20 to 18 °C, or 28 to 55 °C) >120 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced) >60 dB at 50/60 Hz ± 0.1% 92 × 207 × 59 mm • U1271A: 518 grams (with batteries) • U1272A: 520 grams (with batteries) Warranty • Three years for product • Three months for product’s accessories Calibration Cycle One year Specification Assumptions • Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH. • AC V and AC μA/mA/A specifications are ac coupled, true RMS and are valid from 5% of range to 100% of range. • The crest factor may be up to 3.0 at full- scale except for the 1000 V range where it is 1.5 at full scale. • For non- sinusoidal waveforms, add (2% reading + 2% full scale) typical, for crest factors up to 3. • After ZLOW voltage measurements, wait at least 20 minutes for thermal impact to cool before proceeding with any other measurement. 7 Electrical specifications DC specifications for U1271A and U1272A Function Voltage [1] Range Accuracy ±(% of reading + Resolution counts of least significant digit) U1271A U1272A 30 mV 0.001 mV - 0.05 + 20 300 mV 0.01 mV 0.05 + 5 0.05 + 5 - 3V 0.0001 V 0.05 + 5 0.05 + 5 - 30 V 0.001 V 0.05 + 2 0.05 + 2 - 300 V 0.01 V 0.05 + 2 0.05 + 2 - 1000 V 0.1 V 0.05 + 2 0.05 + 2 - - 1 + 20 ZLOW (low input impedance) enabled, applicable for 1000 V range and resolution only. Resistance [2] Current [3] Diode Test [4] Test current/ Burden voltage 30 Ω 0.001 Ω - 0.2 + 10 0.65 mA 300 Ω 0.01 Ω 0.2 + 5 0.2 + 5 0.65 mA 3 kΩ 0.0001 kΩ 0.2 + 5 0.2 + 5 65 mA 30 kΩ 0.001 kΩ 0.2 + 5 0.2 + 5 6.5 mA 300 kΩ 0.01 kΩ 0.5 + 5 0.5 + 5 0.65 mA 3 MΩ 0.0001 MΩ 0.6 + 5 0.6 + 5 93 nA//10 MΩ 30 MΩ 0.001 MΩ 1.2 + 5 1.2 + 5 93 nA//10 MΩ 100 MΩ - 2.0 +10 - 93 nA//10 MΩ 93 nA//10 MΩ 300 MΩ 0.01 MΩ - 2.0 % + 10 @ <100 MΩ 8.0 % + 10 @ >100 MΩ 300 nS 0.01 nS 1 + 10 1 + 10 93 nA//10 MΩ 300 µA 0.01 µA 0.2 + 5 0.2 + 3 <0.04 V/ 100 Ω 3000 µA 0.1 µA 0.2 + 5 0.2 + 3 <0.4 V/ 100 Ω 30 mA 0.001 mA 0.2 + 5 0.2 + 3 <0.08 V/ 1 Ω 300 mA 0.01 mA 0.2 + 5 0.2 + 3 <1.00 V/ 1 Ω 3A 0.0001 A 0.3 +10 0.3 +10 <0.1 V/ 0.01 Ω 10 A 0.001 A 0.3 +10 0.3 +10 <0.3 V/ 0.01 Ω 3V 0.0001 V 0.5 + 5 0.5 + 5 Approximately 1 to 2 mA Auto 0.0001 V - 0.5 + 5 Approximately 0.1 to 0.3 mA 1. Notes for voltage specifications: • The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads). • For ZLOW measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode. 2. Notes for resistance specifications: • Overload protection: 1000 Vrms for short circuits with <0.3 A current. • Maximum open voltage is <+3.3 V • Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer than 1 ms. • The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads). • For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%. • The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead. • The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from -20 °C to 18 °C or 28 °C to 55 °C) 8 3. Notes for current specifications: • Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse • Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse • Specification for 300 mA range: 440 mA continuous. • Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the mea-sured time before proceeding with low current measurements. Notes for diode specifications: • Overload protection: 1000 Vrms for short circuits with <0.3 A current. • Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased di-ode or semiconductor junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V). • Open voltage for diode: <+3.3 V DC • Open voltage for Auto diode: <+2.5 V DC and > -1.0 V DC 4. AC specifications for U1271A Accuracy Function True RMS AC Voltage [1] Range Resolution 45 Hz to 65 Hz 30 Hz to 1 kHz 300 mV 0.01 mV 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40 3V 0.0001 V 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40 30 V 0.001 V 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40 300 V 0.01 V 0.7 + 20 1.0 + 25 2.0 + 25 - 1000 V 0.1 V 0.7 + 20 1.0 + 25 - - 1.0 + 25 @ <200 Hz 5.0 + 25 @ <440 Hz - - LPF (low pass filter) enabled, applicable for all voltage ranges and resolution Function True RMS AC Current [2] 1. 2. 1 kHz to 5 kHz to 5 kHz 20 kHz 0.7 + 20 Range Resolution Accuracy 45 Hz to 2 kHz Burden voltage/Shunt 300 µA 0.01 µA 0.9 + 25 <0.04 V/ 100 Ω 3000 µA 0.1 µA 0.9 + 25 <0.4 V/ 100 Ω 30 mA 0.001 mA 0.9 + 25 <0.08 V/ 1 Ω 300 mA 0.01 mA 0.9 + 25 <1.00 V/ 1 Ω 3A 0.0001 A 1.0 + 25 <0.1 V/ 0.01 Ω 10 A 0.001 A 1.0 + 25 <0.3 V/ 0.01 Ω Notes for voltage specifications: • Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current. • Input impedance: 10 MΩ (nominal) in parallel with <100 pF. Notes for current specifications: • Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse • Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse • Specification for 300 mA range: 440 mA continuous. • Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low current measurements. 9 AC specifications for U1272A Accuracy Function True RMS AC Voltage [1] Function True RMS AC Current [2] 20 kHz to 100 kHz Range Resolution 45 to 65 Hz 20 Hz to 1 kHz 30 mV 0.001 mV 0.6 + 20 0.7 + 25 1.0 + 25 1.0 + 40 3.5 + 40 300 mV 0.01 mV 0.6 + 20 0.7 + 25 1.0 + 25 1.0 + 40 3.5 + 40 3V 0.0001 V 0.6 + 20 1.0 + 25 1.5 + 25 2.0 + 40 3.5 + 40 30 V 0.001 V 0.6 + 20 1.0 + 25 1.5 + 25 2.0 + 40 3.5 + 40 300 V 0.01 V 0.6 + 20 1.0 + 25 1.5 + 25 2.0 + 40 - 1000 V 0.1 V 1 kHz to 5 kHz to 5 kHz 20 kHz 0.6 + 20 1.0 + 25 1.5 + 25 - - LPF (low pass filter) enabled, applicable for all voltage ranges and resolution 0.6 + 20 1.0 + 25 @ <200 Hz 5.0 + 25 @ <440 Hz - - - ZLOW 1000 V 2.0 + 40 2 + 40 @ <440 Hz - - - Accuracy 20 Hz 45 to to 65 Hz 2 kHz Burden voltage/ Shunt Range Resolution 300 µA 0.01 µA 0.6 + 25 0.9 + 25 <0.04 V/ 100 Ω 3000 µA 0.1 µA 0.6 + 25 0.9 + 25 <0.4 V/ 100 Ω 30 mA 0.001 mA 0.6 + 25 0.9 + 25 <0.08 V/ 1 Ω 300 mA 0.01 mA 0.6 + 25 0.9 + 25 <1.00 V/ 1 Ω 3A 0.0001 A 0.8 + 25 1.0 + 25 <0.1 V/ 0.01 Ω 10 A 0.001 A 0.8 + 25 1.0 + 25 <0.3 V/ 0.01 Ω 1. Notes for voltage specifications • Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current. • Input impedance: 10 MΩ (nominal) in parallel with <100 pF. • ZLOW impedance: 2 kΩ (nominal) • The input signal is lower than the product of 20,000,000 V×Hz. • For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10% of range. 2. Notes for current specifications • Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse • Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse • Specification for 300 mA range: 440 mA continuous. • Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low current measurements. 10 AC + DC specifications for U1272A Accuracy Function True RMS AC + DC Voltage [1] Function Range 20 kHz to 100 kHz 30 mV 0.001 mV 0.7 + 40 0.8 + 45 1.1 + 45 1.1 + 60 3.6 + 60 300 mV 0.01 mV 0.7 + 25 0.8 + 30 1.1 + 30 1.1 + 45 3.6 + 45 3V 0.0001 V 0.7 + 25 1.1 + 30 1.6 + 30 2.1 + 45 3.6 + 45 30 V 0.001 V 0.7 + 25 1.1 + 30 1.6 + 30 2.1 + 45 3.6 + 45 300 V 0.01 V 0.7 + 25 1.1 + 30 1.6 + 30 2.1 + 45 - 1000 V 0.1 V 0.7 + 25 1.1 + 30 1.6 + 30 - - Range 300 µA True RMS AC + DC Current [2] Resolution 45 Hz to 20 Hz to 1 kHz to 5 kHz to 65 Hz 2 kHz 5 kHz 20 kHz Accuracy Resolution 45 Hz to 20 Hz to 65 Hz 2 kHz Burden voltage/ Shunt 0.01 µA 0.8 + 30 1.1 + 30 <0.04 V/100 Ω 3000 µA 0.1 µA 0.8 + 30 1.1 + 30 <0.4 V/100 Ω 30 mA 0.001 mA 0.8 + 30 1.1 + 30 <0.08 V/1 Ω 300 mA 0.01 mA 0.8 + 30 1.1 + 30 <1.00 V/1 Ω 3A 0.0001 A 0.9 + 35 1.3 + 35 <0.1 V/0.01 Ω 10 A 0.001 A 0.9 + 35 1.3 + 35 <0.3 V/0.01 Ω 1. Notes for voltage specifications: • Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current. • Input impedance: 10 MΩ (nominal) in parallel with <100 pF. • For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10% of range. 2. Notes for current specifications: • Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse • Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse • Specification for 300 mA range: 440 mA continuous. • Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low current measurements. 11 Temperature specifications [1] - [6] Thermocouple type K J 1. Range Resolution -200 to 1372 °C Accuracy U1271A U1272A 0.1 °C 1% + 1 °C 1% + 1 °C -328 to 2502 °F 0.1 °F 1% + 1.8 °F 1% + 1.8 °F -200 to 1200 °C 0.1 °C - 1% + 1 °C -328 to 2192 °F 0.1 °F - 1% + 1.8 °F The specifications above is specified after 60 minutes of warm-up time. 2. The accuracy does not include the tolerance of the thermocouple probe. 3. Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages poses a shock hazard. 4. Ensure that the ambient temperature is stable within ±1 ºC and that the Null function is used to reduce the test lead’s thermal effect and temperature offset. Before using Null function, set the multimeter to measure temperature without ambient compensation (°C) and keep the thermocouple probe as close to the multimeter as possible (avoid contact with any surface that has a different temperature from the ambient temperature). 5. When measuring temperature with respect to any temperature calibrator, try to set both the calibrator and multimeter with an external reference (without internal ambient compensation). If both the calibrator and multimeter are set with internal reference (with internal ambient compensation), some deviations may show between the readings of the calibrator and multimeter, due to differences in ambient compensation between the calibrator and multimeter. Keeping the multimeter close to the output terminal of calibrator will help reduce the deviation. 6. The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175. Capacitance specifications [7][8] Accuracy Range Resolution U1271A U1272A 10 nF 0.001 nF 1% + 5 1% + 5 100 nF 0.01 nF 1% + 2 1% + 2 1000 nF 0.1 nF 1% + 2 1% + 2 10 µF 0.001 µF 1% + 2 1% + 2 100 µF 0.01 µF 1% + 2 1% + 2 1000 µF 0.1 µF 1% + 2 1% + 2 10 mF 0.001 mF 1% + 2 1% + 2 7. Overload protection: 1000 Vrms for short circuits with <0.3 A current. 8. The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads). 12 Frequency specifications [1][2] Range Resolution Accuracy 99.999 Hz 0.001 Hz 0.02% + 5 999.99 Hz 0.01 Hz 0.005% + 5 9.9999 kHz 0.1 Hz 0.005% + 5 99.999 kHz 1 Hz 0.005% + 5 999.99 kHz 0.01 kHz 0.005% + 5 >1 MHz 0.1 kHz 0.005% + 5 @ <1 MHz Minimum input frequency 0.5 Hz 1. Overload protection: 1000 V; input signal is <20,000,000 V × Hz (product of voltage and frequency). 2. The frequency measurement is susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors. Turning on the low pass filter may help you to filter out the noise and achieve a stable reading. Duty Cycle [3] Mode Range Accuracy at full scale DC Coupling 99.99% 0.3 % per kHz + 0.3 % AC Coupling 99.99% 0.3 % per kHz + 0.3 % Pulse Width [4] Range Resolution Accuracy at full scale 999.99 ms 0.01 ms (duty cycle accuracy/frequency) + 0.01 ms 2000.0 ms 0.1 ms (duty cycle accuracy/frequency) + 0.1 ms 3. Notes for duty cycle specifications: • The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the dc 3 V range. For ac couplings, the duty cycle range can be measured within the range of 10% to 90% for signal frequencies >20 Hz. • The range of the duty cycle is determined by the frequency of the signal: {10 μs × frequency × 100%} to {[1 – (10 μs × frequency)] × 100%}. • The pulse width (positive or negative) must be >10 μs. The range of the pulse width is determined by the frequency of the signal. 4. Notes for pulse width specifications: • The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the dc 3 V range. • The pulse width (positive or negative) must be >10 μs. The range of the pulse width is determined by the frequency of the signal. 13 Frequency sensitivity for voltage measurements [1][2][3] Minimum sensitivity (RMS sine wave) Input range 15 Hz to 100 kHz 0.5 Hz to 200 kHz Up to 1 MHz Trigger level for dc coupling 0.5 Hz to 200 kHz U1271A U1272A 30 mV 3 mV 3 mV - - 5 mV 300 mV 6 mV 8 mV 40 mV 10 mV 15 mV 3V 0.12 V 0.2 V 0.4 V 0.15 V 0.15 V 30 V 0.6 V 0.8 V 2.6 V 1.5 V 1.5 V 300 V 6V 8 V @ <100 kHz - 9 V @ <100 kHz 9 V @ <100 kHz 1000 V 50 V 50 V @ <100 kHz - 90 V @ <100 kHz 90 V @ <100 kHz 1. Maximum input for specified accuracy, refer to “AC specifications” on page 9. 2. 30 mV range applicable for U1272A only. 3. 200 kHz to 1 MHz range applicable for U1272A only. Frequency sensitivity for current measurements [4] Input range Minimum sensitivity (RMS sine wave) 2 Hz to 30 kHz 300 µA 100 µA 3000 µA 70 µA 30 mA 1.2 mA 300 mA 12 mA 3A 0.12 A 10 A 1.2 A 4. Maximum input for specified accuracy, refer to “AC specifications” on page 9. Peak hold Signal width Accuracy for DC Voltage and Current Single event >1 ms Specified accuracy + 400 Repetitive >250 µs Specified accuracy + 1000 14 Decibel (dB) for U1272A [1] - [3] dB base Reference Default reference 1 mW (dBm) 1 to 9999 Ω 50 Ω 1 V(dBV) 1V 1V 1. The reading of dBm is indicated in decibels of power above or below 1 mW, or decibels of voltage above or below 1 V. The formula is calculated according to the voltage measurement and specified reference impedance. Its accuracy is depended on the accuracy of the voltage measurement. See Decibel (dBV) accuracy table below. 2. Auto-ranging mode is used. 3. The bandwidth is according to voltage measurement. Decibel (dBV) accuracy dBV range Range Accuracy 20 Hz to 1 Hz to 5 kHz to 20 Hz to 45 to 65 Hz 1 kHz 5 kHz 20 kHz 100 kHz Minimum Maximum -56.48 -30.46 300 mV -36.48 -10.46 0.06 0.07 0.09 0.1 0.32 3V -16.48 +9.54 0.06 0.09 0.14 0.19 0.32 30 mV 0.06 0.07 0.09 0.1 0.32 30 V +3.52 +29.54 0.06 0.09 0.14 0.19 0.32 300 V +23.52 +49.54 0.06 0.09 0.14 0.19 - 1000 V +33.98 +60 0.06 0.09 0.14 - - Measurement rate (approximate) Function ACV Times/second U1271A U1272A 7 7 DCV 7 7 Ω 14 14 Ω with offset compensation - 3 Diode 14 14 Auto Diode - 3 Capacitance 4 (<100 µF) 4 (<100 µF) DCI 7 7 ACI 7 7 Temperature 7 7 Frequency 2 (>10 Hz) 2 (>10 Hz) Duty cycle 1 (>10 Hz) 1 (>10 Hz) Pulse width 1 (>10 Hz) 1 (>10 Hz) 15 Ordering Information U1271A U1272A Standard Shipped Accessories Standard test leads, test probes with 19-mm and 4-mm tips, K-type thermocouple and adapter, 4x AAA batteries, Certificate of Calibration, test report, Quick Start Guide Optional Accessories Measuring Accessories (non-temperature) U1160A Standard test lead kit Includes two test leads (red and black), alligator clips, fine-tip test probes, SMT grabbers and mini grabber (black). • • • • • Test leads: CAT III 1000 V, 15 A Alligator clips: CAT III 1000 V, 10 A Fine-tip test probes: CAT II 300 V, 3 A SMT grabbers: CAT II 300 V, 3 A Mini grabber: CAT II 300 V, 3 A U1161A Extended test lead kit U1162A Alligator clips Includes two test leads (red and black), two test probes, medium-sized alligator clips and 4-mm banana plugs. • One pair of insulated alligator clips (red and black). Recommended for use with Agilent standard test leads. • Rated CAT III 1000 V, 10 A. • • • • Test leads: CAT III 1000 V, CAT IV 600 V, 15 A Test probes: CAT III 1000 V, 15 A Medium-sized alligator clips: CAT III 600 V, 10 A 4-mm banana plugs: CAT II 600 V, 10 A U1163A SMT grabbers • One pair of SMT grabbers (red and black). Recommended for use with Agilent standard test leads. • Rated CAT II 300 V, 3 A. U1164A Fine-tip test probes U1165A Test probe leads • One pair of fine-tip test probes (red and black). Recommended for use with Agilent standard test leads. • Rated CAT II 300 V, 3 A. • Rated CAT III 1000 V, 15 A 16 Measuring Accessories (non-temperature) U1168A Standard test lead kit U1169A Test probe leads U1583B AC current clamp Includes two test leads (red and black), 19-mm and 4-mm test probes, alligator clips, fine-tip test probes, SMT grabbers and mini grabber (black). Includes two test leads (red and black), and a pair each of 19-mm and 4-mm test probes. • Dual range: 40 A and 400 A • Rated CAT III 600 V • BNC-to-banana-plug adapter provided for use with DMMs • Test leads: CAT III 1000 V, CAT IV 600 V, 15 A • Test probes (19-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A • Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A (highly recommended for CAT IV environment) • Alligator clips: CAT III 1000 V, 10 A • Fine-tip test probes: CAT II 300 V, 3 A • SMT grabber: CAT II 300 V, 3 A • Mini grabber: CAT II 300 V, 3 A • Test leads: CAT III 1000 V, CAT IV 600 V, 15 A • Test probes (19-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A • Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A (highly recommended for CAT IV environment) Measuring Accessories (temperature) U1180A Thermocouple adapter+lead kit, J and K types U1181A Immersion temperature probe U1182A Industrial surface temperature probe • Type-K T/C for use in oil and other liquids • Measurement range: –50 °C to 700 ºC • Includes adapter U1184A for connection to DMM • Type-K T/C for use on still surfaces • Measurement range: –50 ºC to 400 ºC • Includes adapter U1184A for connection to DMM U1183A Air temperature probe U1184A Temperature probe adapter U1185A J-type thermocouple and adapter • Type-K T/C for use in air and non-caustic gas • Measurement range: –50 ºC to 800 ºC • Includes adapter U1184A for connection to DMM • Mini-connector-to-banana-plug adapter for use with DMM • T/C adapter J/K-type • T/C bead J-type: –20 °C to 200 °C Includes thermocouple adapter, thermocouple bead J-type and thermocouple bead K-type. • T/C adapter J/K-type • T/C bead J-type: –20 °C to 200 °C • T/C bead K-type: –20 °C to 200 °C U1186A K-type thermocouple and adapter • T/C adapter J/K-type • T/C bead J-type: –20 °C to 200 °C 17 Cable U1173A IR-to-USB cable U1174A Soft carrying case • For remote control and data logging to PC • Max. baud rate: 19,200 bits per second The convenient way to carry your DMM and essential accessories • Dimension: 9” (H) x 5” (W) x 3” (D) Hanging Kit U1171A Magnetic hanging kit For fastening of DMM to a steel surface so both hands are free. Probe Clip Light U1176A LED Probe Clip Light • 3 inches in length • To be clipped onto test probes to increase visibility • Comes with one AAA battery 18 www.agilent.com www.agilent.com/find/ handhelddmm For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. www.lxistandard.org LXI is the LAN-based successor to GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium. Agilent Advantage Services is committed to your success throughout your equipment’s lifetime. We share measurement and service expertise to help you create the products that change our world. To keep you competitive, we continually invest in tools and processes that speed up calibration and repair, reduce your cost of ownership, and move us ahead of your development curve. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent’s measurement expertise and product breadth, combined with channel partner convenience. www.agilent.com/find/advantageservices www.agilent.com/quality www.agilent.com/find/contactus Americas Canada Brazil Latin America Mexico United States (877) 894 4414 (11) 4197 3500 305 269 7500 01800 5064 800 (800) 829 4444 Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Thailand 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008 Europe & Middle East Austria 43 (0) 1 360 277 1571 Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 €/minute Germany 49 (0) 7031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 Switzerland 0800 80 53 53 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: July 8, 2010 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2010 Printed in USA, October 25, 2010 5990-6425EN