HP U1730

Agilent U1730 Series
Handheld LCR Meters
Take your expectations higher with
the latest LCR meters
Data Sheet
Agilent’s U1730 Series handheld LCR meters allow you to measure
at frequencies as high as 100 kHz—a capability typically found
only in benchtop meters. Get measurements done faster using the
one-touch automatic identification function button which displays
component type and more detailed component analysis such as
Z, ESR, and DCR. Ideal for testing on the go, these LCR meters
operate on a battery that lasts up to 16 hours. With the U1730
Series that is built for your convenience, you can perform quick
and basic LCR measurements at an affordable price.
Features
Key features
Frequency up to 100 kHz
Detailed component analysis
• 20,000 counts resolution
The test frequency now extends
as high as 100 kHz, providing more
flexibility to test a wider range of
components. A higher test frequency,
for example 100 kHz, is useful for
applications such as testing aluminum electrolytic capacitors used in
switching power supply circuits.
The handheld LCR meters allows
you to test various component types,
including secondary components
of Dissipation Factor (D), Quality
Factor (Q), and Angle Indication of
Impedance (θ). This new handheld
series also includes other functions that result in a more detailed
component analysis. For example, the
built-in Equivalent Series Resistance
(ESR) function helps you better
understand the inherent resistance
behavior typically found in capacitors
across selected frequencies. DCR is a
built-in DC resistance measurement
that eliminates the use of a separate
digital multimeter (DMM) for component test.
• 0.2% basic accuracy
• Wide LCR ranges with three to five
selectable test frequencies (up to
100 kHz for U1733C)
• Auto identification (Ai)
automatically determines and
displays component type and
measurements
• Detailed component analysis with
DCR, ESR, Z, D, Q, and θ functions
• Battery life of 16 hours/
AC-powered
• IR-to-USB connectivity for data
logging to PC
Automated identification
With Ai the testing and measuring
experience is easy; eliminating unnecessary trial and error time—with just
a single push of a button. This unique
feature automatically specifies L, C,
or R with parallel and series mode,
without the need to manually change
buttons.
Figure 1. Automate the recording of continuous readings when you hook the U1731C/U1732C/U1733C to a PC
2
Take a Closer Look
Visible and audible
tolerance mode for
component sorting
Secondary display
Maximum, Minimum and
Average values recording
20,000 counts resolution
Ai helps identify L, C,
and R components
automatically according
to level of impedance
Data Hold function
to freeze and save
measured values
ESR function for
capacitance series
resistance analysis
Allows manual or auto
ranging to be used
interchangeably
Auto calculation of Phase
Angle, Dissipation Factor,
and Quality Factor
To select desired test
frequency:
• U1731C: Up to 1 kHz
• U1732C: Up to 10 kHz
• U1733C: Up to 100 kHz
To select desired Z,
L, C, or R function
To enable DCR
measurement
(only in U1733C)
To enter rolerance range
of 1%, 5%, 10%, and 20%
Backlight function to ease
viewing in subdued lighting
(only in U1732C/U1733C)
To enter relative mode and
stores display reading as a
reference value
Figure 2. Front view of the U1733C
3
U1731C/U1732C/U1733C Electrical Specifications
Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than
80%. Please refer to the User Guide about the measuring mode specified for each range of L/C/R, series or parallel mode.
Measurements performed at the test socket and necessary Open and Short corrections must prior be done. The accuracy
is verified by design and specified type tests.
Impedance/Resistance
Accuracy = AZ + Offset
Range
1
2Ω
20 Ω
1
200 Ω1
Resolution
U1731C/U1732C/U1733C
U1732C/U1733C
U1733C
100 Hz
120 Hz
1 kHz
10 kHz
100 kHz
DCR
0.0001 Ω
0.7% + 50
0.7% + 50
0.7% + 50
0.7% + 50
1.0% + 50
0.7% + 50
0.001 Ω
0.7% + 8
0.7% + 8
0.7% + 8
0.7% + 8
0.7% + 8
0.7% + 8
0.01 Ω
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.2% + 3
2000 Ω
0.1 Ω
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.2% + 3
20 kΩ
0.001 kΩ
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.2% + 3
200 kΩ
0.01 kΩ
0.5% + 5
0.5% + 5
0.5% + 5
0.5% + 5
0.7% + 8
0.5% + 5
2000 kΩ
0.1 kΩ
0.5% + 5
0.5% + 5
0.5% + 5
0.7% + 5
NA
0.5% + 5
20 MΩ
0.001 MΩ
2.0% + 8
2.0% + 8
2.0% + 8
5.0% + 8
NA
2.0% + 8
200 MΩ
0.01 MΩ
6.0% + 80
6.0% + 80
6.0% + 80
NA
NA
6.0% + 80
1. This accuracy for the ranges of 2~ 200 Ω is specified after Math Null which is used to substrate the resistance of test leads and the contact
resistance.
Notes:
a. For the ranges of 20 MΩ and 200 MΩ, the R.H. is specified for < 60%
b. Resistance is specified to Q < 10 and D > 0.1, otherwise the accuracy is (AZ + Offset) x √ 1 + Q2
c. Equivalence Series Resistance (ESR) measurement is determined by impedance measurement and range. The maximum
display is up to 199.99 kΩ and the accuracy is (AZ + Offset) x √ 1 + Q2
Capacitance
Accuracy = AC + Offset
Range
Resolution
U1731C/U1732C/U1733C
U1732C/U1733C
U1733C
100 Hz
120 Hz
1 kHz
10 kHz
100 kHz
20 mF
0.001 mF
0.5% + 8
0.5% + 8
NA
NA
NA
2000 μF
0.1 μF
0.5% + 5
0.5% + 5
0.5% + 8
NA
NA
200 μF
0.01 μF
0.3% + 3
0.3% + 3
0.5% + 5
0.5% + 8
NA
20 μF
0.001 μF
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
5.0% + 10
2000 nF
0.1 nF
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.7% + 10
200 nF
0.01 nF
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 3
0.7% + 10
20 nF
0.001 nF
0.5% + 5
0.5% + 5
0.2% + 3
0.5% + 3
0.7% + 10
2000 pF1
0.1 pF
0.5% + 10
0.5% + 10
0.5% + 5
0.5% + 3
2.0% + 10
200 pF1
0.01 pF
NA
NA
0.5% + 10
0.8% + 10
2.0% + 10
0.001 pF
NA
NA
NA
1.0% + 20
2.5% + 10
20 pF
1
1. This accuracy for the ranges of 20 pF~2000 pF is specified after Math Null which is used to substrate the stray capacitances for test leads.
Notes:
a. The accuracy for the ceramic capacitor will be influenced depending on the dielectric constant (K) of the material used to make
the ceramic capacitor. For related influence factors, please refer to the Component dependency factors section in the Impedance
Measurement Handbook, downloadable for free at http://www.agilent.com/find/lcrmeters
4
U1731C/U1732C/U1733C Electrical Specifications
Inductance
Accuracy = AL + Offset
Range
Resolution
U1731C/U1732C/U1733C
U1732C/U1733C
U1733C
100 Hz
120 Hz
1 kHz
10 kHz
100 kHz
20 μH
0.001 μH
NA
NA
NA
1.0% + 5
2.5% + 20
200 μH
0.01μH
NA
NA
1.0% + 5
0.7% + 3
2.5% + 20
2000 μH
0.1 μH
0.7% + 10
0.7% + 10
0.5% + 3
0.5% + 3
0.8% + 20
20 mH
0.001 mH
0.5% + 3
0.5% + 3
0.2% + 3
0.3% + 3
0.8% + 10
200 mH
0.01 mH
0.5% + 3
0.5% + 3
0.2% + 3
0.2% + 3
1.0% + 10
2000 mH
0.1 mH
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
1.0% + 10
20 H
0.001 H
0.2% + 3
0.2% + 3
0.5% + 5
1.0% + 5
2.0% + 10
200 H
0.01 H
0.7% + 5
0.7% + 5
1.0% + 5
2.0% + 8
NA
2000 H
0.1 H
1.0% + 5
1.0% + 5
2.0% + 8
NA
NA
Phase Angle of Impedance
Range
Resolution
Accuracy (θe)
Condition
–180° ~180°
0.1°/1°
(AZ + Offset/Zx) x180/π
D < 1 or Q > 1
Impedance
Zx
AZ
Offset
θe
1999.9 Ω
19999
0.2%
3
±0.12 °
199.9 Ω
1999
0.2%
3
±0.20 °
19.9 Ω
199
0.2%
3
±0.98 °
1.9 Ω
19
0.2%
3
±9.16 °
Notes:
a. Specifications are applicable to all models (U1731C, U1732C, and U1733C) unless specified
b. The “AZ” and Offset are the accuracy specified at impedance
c. The “π” is approximately 3.14159
Dissipation/Quality Factor
Function
Range
Accuracy (De)
Condition
Z
0.001~999
AZ + Offset/Zx x 100% + 3
D < 1 or Q > 1
L
0.001~999
AL + Offset/Lx x 100% + 3
D < 1 or Q > 1
C
0.001~999
AC + Offset/Cx x 100% + 3
D < 1 or Q > 1
Capacitance
Cx
AC
Offset
De
88.88 μF
8888
0.2%
3
0.203% + 3
Notes:
1. Specifications are applicable to all models (U1731C, U1732C, and U1733C) unless specified
2. The “AZ, AL, AC” and Offset are the accuracy specified at Impedance, Inductance, and Capacitance, respectively
3. The Zx, Lx, and Cx are the display count of the reading. For example, the Cx is 8888 as if the capacitance is 88.88 μF for the range of
200 μF.
4. The Quality Factor is the reciprocal of Dissipation Factor
5
U1731C/U1732C/U1733C Electrical Specifications
Test Signal
Test signal level
Test frequency
Model
Selection
Level
Accuracy
Frequency
Accuracy
U1731C/U1732C/U1733C
100 Hz
0.74 Vrms
0.05 Vrms
100 Hz
0.01%
120 Hz
0.74 Vrms
0.05 Vrms
120.481 Hz
0.01%
1 kHz
0.74 Vrms
0.05 Vrms
1 kHz
0.01%
10 kHz
0.70 Vrms
0.05 Vrms
10 kHz
0.01%
U1732C/1733C
U1733C
100 kHz
0.70 Vrms
0.05 Vrms
100 kHz
0.01%
DCR
+1.235 V
0.05 V
NA
NA
Source Impedance of Impedance/Resistance Measurement
Typical source impedance
Range
U1731C/U1732C/U1733C
U1732C/U1733C
U1733C
100 Hz
120 Hz
1 kHz
10 kHz
100 kHz
DCR
2Ω
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
20 Ω
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
200 Ω
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
2000 Ω
1 kΩ
1 kΩ
1 kΩ
1 kΩ
1 kΩ
1 kΩ
20 kΩ
10 kΩ
10 kΩ
10 kΩ
10 kΩ
1 kΩ
10 kΩ
200 kΩ
100 kΩ
100 kΩ
100 kΩ
10 kΩ
1 kΩ
100 kΩ
2000 kΩ
100 kΩ
100 kΩ
100 kΩ
10 kΩ
NA
100 kΩ
20 MΩ
100 kΩ
100 kΩ
100 kΩ
100 kΩ
NA
100 kΩ
200 MΩ
100 kΩ
100 kΩ
100 kΩ
NA
NA
100 kΩ
Source Impedance of Capacitance Measurement
Typical source impedance
Range
20 mF
U1732C/U1733C
U1733C
100 Hz
U1731C/U1732C/U1733C
120 Hz
1 kHz
10 kHz
100 kHz
100 Ω
100 Ω
NA
NA
NA
2000 μF
100 Ω
100 Ω
100 Ω
NA
NA
200 μF
100 Ω
100 Ω
100 Ω
100 Ω
NA
20 μF
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
2000 nF
1 kΩ
1 kΩ
100 Ω
100 Ω
100 Ω
200 nF
10 kΩ
10 kΩ
1 kΩ
100 Ω
100 Ω
20 nF
100 kΩ
100 kΩ
10 kΩ
1 kΩ
100 Ω
2000 pF
100 kΩ
100 kΩ
100 kΩ
10 kΩ
1 kΩ
200 pF
NA
NA
100 kΩ
10 kΩ
1 kΩ
20 pF
NA
NA
NA
100 kΩ
1 kΩ
6
U1731C/U1732C/U1733C Electrical Specifications
Source Impedance of Inductance Measurement
Typical source impedance
Range
U1731C/U1732C/U1733C
U1732C/U1733C
U1733C
100 Hz
120 Hz
1 kHz
10 kHz
100 kHz
20 μH
NA
NA
NA
100 Ω
100 Ω
200 μH
NA
NA
100 Ω
100 Ω
100 Ω
2000 μH
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
20 mH
100 Ω
100 Ω
100 Ω
100 Ω
100 Ω
200 mH
100 Ω
100 Ω
100 Ω
1 kΩ
1 kΩ
2000 mH
100 Ω
100 Ω
1 kΩ
10 kΩ
1 kΩ
20 H
1 kΩ
1 kΩ
10 kΩ
10 kΩ
1 kΩ
200 H
10 kΩ
10 kΩ
100 kΩ
100 kΩ
NA
2000 H
100 kΩ
100 kΩ
100 kΩ
NA
NA
7
General Specifications
Parameter
U1731C
U1732C
U1733C
Measurements
Z/L/C/R/D/Q/θ/ESR
Z/L/C/R/D/Q/θ/ESR
Z/L/C/R/D/Q/θ/ESR/DCR
Display
Primary display: Maximum display 19,999 counts
Secondary display: Maximum display 999 counts
Automatic polarity indication
Test frequency (Accuracy =
± 0.1% of actual test frequency)
100 Hz, 120 Hz, 1 kHz
100 Hz, 120 Hz, 1 kHz, 10 kHz
100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz
Backlight
No
Yes
Yes
Test signal level
Selection
Test signal level
Test frequency
100 Hz
0.74 Vrms
100 Hz
120 Hz
0.74 Vrms
120.481 Hz
1 kHz
0.74 Vrms
1 kHz
10 kHz
0.74 Vrms
10 kHz
100 kHz2
0.74 Vrms
100 kHz
+1.235 V
NA
1
2
DCR
Tolerance mode
1%, 5%, 10%, 20%
Ranging mode
Auto and manual
Measurement rate
1 time/second, nominal
Response time
Approximately 1 second/DUT (Device Under Test)
Auto power-off
~0-99 mins without operation
Power supply
Single standard 9 V battery (alkaline or carbon-zinc) or optional power adaptor
Power consumption
225 mVA maximum without backlight
Input protection fuse
Resettable over-current protection
Battery life
16 hours based on alkaline battery
] will appear when voltage drops below ~7.2 V
Low battery indicator
[
Operating temperature
–10 to 55 °C, 0 to 80% R.H.
Storage temperature
–20 to 70 °C, 0 to 80% R.H. without battery
Temperature coefficient
0.1 × (specified accuracy)/°C (from –10 to 18 °C or 28 to 55 °C)
Relative humidity
Maximum 80% R.H. for temperature up to 30 °C decreasing linearly to 50% R.H. at 55 °C
Weight
337 grams with battery
Dimensions (H x W x D)
184 mm x 87 mm x 41 mm
Safety and EMC Compliance
In compliance with EN61010-1 (IEC61010-1:2001) for low voltage directive and Pollution Degree II
Environment. Susceptibility and Emissions (EMC): Commercial Limits per EN61326-1
Calibration
One-year calibration cycle recommended
Warranty
• 3 years for main unit
• 3 months for standard shipped accessories
1. Only applicable for U1732C/U1733C
2. Only applicable for U1733C
8
Ordering Information
Standard shipped items
Standard U1731C, U1732C, and U1733C ordering include:
•
•
•
•
Quick Start Guide
Certificate of Calibration (CoC)
Alligator clip leads
9 V alkaline battery
Recommended accessories
U1731P
Combo Kit
Includes one U1731 Series handheld and four accessories:
• U5491A soft carrying case
• U1173A IR-USB cable
• U1780A AC adaptor
• U1782A SMD tweezer
U1732P
Combo Kit
Includes one U1732 Series handheld and four accessories:
• U5491A soft carrying case
• U1173A IR-USB cable
• U1780A AC adaptor
• U1782A SMD tweezer
U1733P
Combo Kit
Includes one U1733 Series handheld and four accessories:
• U5491A soft carrying case
• U1173A IR-USB cable
• U1780A AC adaptor
• U1782A SMD tweezer
U1174A
Soft carrying case
U5481A
IR-to-USB cable
U1782A
SMB tweezer
U1780A
Power adaptor and cord (according to country)
U1781A
Alligator clip leads
9
www.agilent.com
www.agilent.com/find/handheld-tools
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.axiestandard.org
AdvancedTCA® Extensions for
Instrumentation and Test (AXIe) is
an open standard that extends the
AdvancedTCA® for general purpose
and semiconductor test. Agilent
is a founding member of the AXIe
consortium.
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
Agilent Advantage Services is committed to your success throughout
your equipment’s lifetime. We share
measurement and service expertise
to help you create the products that
change our world. To keep you competitive, we continually invest in tools
and processes that speed up calibration and repair, reduce your cost of
ownership, and move us ahead of
your development curve.
www.agilent.com/find/advantageservices
www.lxistandard.org
LAN eXtensions for Instruments puts
the power of Ethernet and the Web
inside your test systems. Agilent
is a founding member of the LXI
consortium.
http://www.pxisa.org
PCI eXtensions for Instrumentation
(PXI) modular instrumentation
delivers a rugged, PC-based highperformance measurement and
automation system.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s
measurement expertise and product
breadth, combined with channel
partner convenience.
www.agilent.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
(11) 4197 3500
01800 5064 800
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 375 8100
www.agilent.com/quality
Europe & Middle East
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
Ireland
Israel
Italy
Netherlands
Spain
Sweden
United Kingdom
49 (0) 7031 464 6333
1890 924 204
972-3-9288-504/544
39 02 92 60 8484
31 (0) 20 547 2111
34 (91) 631 3300
0200-88 22 55
44 (0) 118 9276201
For other unlisted Countries:
www.agilent.com/find/contactus
Revised: October 14, 2010
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2011
Printed in USA, April 20, 2011
5990-7778EN