INFRARED GAS ANALYZER DATA SHEET ZPA This gas analyzer (ZPA) is capable of measuring the concentration of NO,SO 2 ,CO 2 ,CO,CH 4 and O 2 components in sample gas. NO, SO 2 , CO 2 , CO and CH 4 are measured by non-dispersion infrared method (NDIR), while O 2 is measured by fuel cell or paramagnetic and zirconia method. Optimum use as an analyzer unit for measurement of combustion exhaust gas from refuse incinerator, boiler exhaust or gas from various industrial furnaces. FEATURES 1.High sensitivity Equipped with the newest upgraded mass flow sensor. 2.Easy maintenance Maintenance is easy due to simple measurement unit of single-beam system adapted. 3.Small and light The size is small 133x483x382mm (HxWxD). (This is for 19inch rack mounting type) and light (9kg). Unit is capable max. 5 component measurement in one analyzer. 4.Extensive functions Various optional functions are available such as auto calibration control, atmospheric pressure correction, high and low concentration alarm, remote range switch, and range identification signal,etc. SPECIFICATIONS Standard Specifications Principle of measurement: NO, SO 2 , CO 2 , CO, CH 4 ; Non-dispersion infrared-ray absorption method Single light source and single beams (single beam system) O 2 ;Fuel cell O 2 sensor (build-in) or paramagnetic O 2 sensor (build-in) or zirconia O 2 sensor (externally installed TYPE: ZFK7) Measurable gas components and measuring range: NO SO2 CO2 CO CH4 O2 Built in fuel cell O2 Built-in Paramagnetic O2 External Zirconia Minimum range 0 - 200ppm 0 - 200ppm 0 - 100ppm 0 - 200ppm 0 - 500ppm Maximum range 0 - 5000ppm 0 - 10vol% 0 - 100vol% 0 - 100vol% 0 - 100vol% 0 - 10vol% 0 - 25vol% 0 - 5vol% 0 - 100vol% None 100 - 95vol% 0 - 5vol% 0 - 25vol% •Max. 5 components measurement including O 2 . For reverse range O 2 measurement, infrared gas measurement is not available (single range O 2 only). •Measuring range ratio max. 1:10 (except O2 ) •Measuring ranges are changeable between the specified minimum and maximum range Settable one range or two ranges * In measurement range low range is called first range, high range is called second range. •For possible combinations of components and ranges, refer to Table1. Measured value indication: Digital indication in 4 digits (LCD panel with LED back light) •Instantaneous value of each component •Instantaneous value after O 2 correction (only in NO, SO 2 , CO measurement with O2 ) •Average value after O 2 correction (only in NO, SO 2 , CO measurement with O2 ) •O 2 average value EDS3-143 Date Nov. 2, 2011 ZPA Analog output signals: 4 to 20mA DC or 0 to 1V DC, isolated internally from circuit and ground. Output lines are non-isolated each other.; 12 outputs max. Allowable load 550Ω for 4 to 20mA DC Allowable load 100KΩ for 0 to 1V DC * Refer to Table2 for the channel No. of displayed values and analog output signals. Analog input signal: For signal input from externally installed O 2 sensor. Signal requirement; (1) Signal from Fuji’s Zirconia O 2 sensor (TYPE: ZFK7) (2) 0 to 1V DC from an O 2 sensor Input section is not isolated. This feature is effective when an O 2 sensor is not built in. * Externally installed O 2 sensor should be purchased separately. Digital output: (Option) 1c contact (24V DC/1A, resistive load) max.15 outputs Instrument error, calibration error, range identification, auto calibration status, solenoid valve drive for auto calibration, High/Low limit alarm contact output. * All relay contacts are isolated mutually and from the internal circuit. Digital input: (Option) Voltage contact (supply 12-24VDC (15mA Max.)) Max. 9 inputs Remote range change over, auto calibration remote start, remote hold, average value reset, Isolated from the internal circuit with photocoupler. Power supply: Voltage rating ;100V to 240V AC Allowable range ;85V to 264V AC Frequency ;50Hz/60Hz Power consumption ;100VA max. Operation conditions: Ambient temperature ; - 5˚C to 45˚C (40˚C max. when 2 optical system at 200V AC power source) Ambient humidity ;90% RH max., non-condensing Storage conditions: Ambient temperature ; -20˚C to 60˚C Ambient humidity ;100% RH max., non-condensing Dimensions (H × W × D): 133 x 483 x 382mm Mass: Approx. 9 kg Finish color: Front panel; Cool gray (PANTON 1C-F) Enclosure: Steel casing, for indoor use Material of gas-contacting parts: Gas inlet/outlet; SUS304 Sample cell; SUS304,chloroprene rubber Infrared-ray transmitting window; CaF2 Internal piping;Toaron,Teflon, Polypropylene Paramagnetic O 2 sensor cell : SUS316 Fuel cell O 2 sensor cell : ABS resin 2 Gas inlet/outlet: Rc1/4 or NPT1/4 internal thread Purge gas flow rate: 1L/min ( when required) Life time of fuel cell O 2 sensor: 2 years Standard Functions Output signal holding: Output signals are held unchanged during manual and auto calibrations by activation of holding (turning “ON” its setting). The values held are those just before start calibration mode or setting value. It is selectable. Indication of instantaneous values will not be held. Switch ranges: The switch ranges function is available in manual, auto, and remote modes. Only preset switch method is effective. Manual: Allows range to switch by key operation. Auto: Automatically switched from first range to second range when the measured value exceeds to 90%FS of first range. Automatically switched from second range to first range when the measured value drops to 80% or less first range. Remote: Voltage contact input (for measurable (Option) components) Allows range to switch via an external signal when remote range switch input is received. When the contact input terminals for each component are input voltage, the first range is selected, and it is switched to the second range when the terminals are open. * These range value are settable between original first range and second range. Optional Functions Remote output holding: Output signal is held at the last value or preset value by voltage input the remote output holding input terminals. Holding is maintained while the voltage input the terminals. Indication of instantaneous values are not held. Range identification signal: The present measuring range is identified by a contact position. The contact output terminals for each component close when the first range is selected, and when the second range is selected, the terminals are open. Auto calibration: Auto calibration is carried out periodically at the preset cycle. When a standard gas cylinder for calibration and a solenoid valve for opening / closing the gas flow line are prepared externally by the customer, calibration will be carried out with the solenoid valve drive contacts for zero calibration and each span calibration turned on/off sequentially at the set auto calibration timing. Auto calibration cycle setting: Auto calibration cycle is set. Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day). Gas flow time setting: The time for flowing each calibration gas in auto calibration is set. Settable within 60 to 900 seconds (in increments of 1 second) Auto calibration remote start: Auto calibration starts by opening the auto calibration remote start input terminal after short circuiting for 1.5 sec or longer. Auto calibration starts when contacts open. Auto zero calibration: Auto zero calibration is carried out periodically at the preset cycle. This cycle is independent on “Auto calibration” cycle. When zero calibration gas and solenoid valve for opening/closing the calibration gas flow line are prepared externally by the customer, zero calibration will be carried out with the solenoid valve drive contact for zero calibration turned on/off at the set auto zero calibration timing. Auto zero calibration cycle setting: Auto zero calibration cycle is set. Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day) Gas flow time setting: The timing for flowing zero gas in auto zero calibration is set. Settable within 60 to 900 seconds (in increments of 1 second) High/low limit alarm: Alarm contact output turns on when measurement value reaches the preset high or low limit alarm value. Contacts close when the instantaneous value of each channel exceeds the high alarm limit value or falls below the low alarm limit value. Instrument error contact output: Contacts turn on at occurrence of analyzer error No. 1, 2, 3 or 10. Calibration error contact output: Contacts turn on at occurrence of manual or auto calibration error (any of errors No. 4 to 9). Auto calibration status contact outputs: Contacts turn on during auto calibration. O 2 correction: Correction of measured NO, SO 2 and CO gas concentrations into values at reference O 2 concentration Correction formula: C= 21-On 21-Os x Cs C :Sample gas concentration after O 2 correction C S :Measured concentration of sample gas O S :Measured O 2 concentration (Limit setting: 1 to 20% O 2 ) O n :Reference O 2 concentration (value changeable by setting.0 to 19% O 2 ) Average value after O 2 correction and O 2 average value calculation: The result of O 2 correction or instantaneous O 2 value can be output as an average value over the preset period of time. Used for averaging is the moving average method in which sampling is carried out at intervals of 30 seconds. (Output is updated every 30 seconds. It is the average value in the determined period of time just before the latest updating.) Averaging time is settable within 1 to 59 minutes (in increments of 1 minute) or 1 to 4 hours (in increments of 1 hour). Average value resetting: The above-mentioned output of average value is started from the initial state by opening the average value resetting input terminals after short circuiting for 1.5 sec or longer. Output is reset by input voltage and restarted by opening Communication function: RS-485 (9pins D-sub connector) Half-duplex bit serial Start-stop synchronization ModbusTM protocol Contents :Read/ Write parameters Read measurement concentration and instrument status Remark : W hen connecting via RS 232C interface, an RS-232C ↔ RS-485 converter should be used. Atmospheric pressure correction: Measure atmospheric pressure and calculate compensation (for use, be sure to relieve the exhaust gas from analyzer to the atmosphere) After atmospheric pressure correction; Zero point:No influenced Span point:The change is 0.5%%measured value or less relating to the change of the atmospheric pressure 1%. Correction range: 700hPa-1050hPa 3 ZPA Performance ±0.5% of full scale 1% of full scale prior to atmospheric pressure correction (option) Zero drift: ±2% of full scale/week In the case of Auto zero calibration use for 500ppm or less range. Span drift: ±2% of full scale/week Response time (for 90% FS response) : 1 to 15 sec electrical response. Within 10-30 seconds including replacement time of sampling gas. Gas replacement time depends on the number of measuring components,and measuring range. Interference from other gases: Repeatability: Linearity: Interference component NO CO2 CO CH4 SO2 analyzer analyzer analyzer analyzer analyzer CO 1000ppm ≤1.0%FS CO2 15% H2O saturation at 20°C H2O saturation at 2°C CH4 1000ppm –– ≤1.0%FS ≤1.0%FS ≤1.0%FS –– *1) *2) ≤1.0%FS ≤1.0%FS ≤1.0%FS ≤1.0%FS *2) ≤1.0%FS ≤1.0%FS ≤1.0%FS –– ≤2.0%FS ≤1.0%FS ≤1.0%FS –– –– –– ≤2.0%FS ≤2.0%FS –– ≤20ppm –– Corrosive component: 1 ppm or less Standard gas for calibration: 1))Infrared-ray measurable component, standard O 2 Zero gas ;Dry N2 Span gas ; Each sample gas having concentration 90 to 100% of its measuring range (recommended). In case a zirconia O2 analyzer is installed externally and calibration is carried out on the same calibration gas line: Zero gas ;Dry air or atmospheric air (Do not use with CO2 measurement) Span gas ;For other than O2 measurement, each sample gas having concentration 90 to 100% of its measuring range For O2 measurement, O2 gas of 1 to 2 vol%/remains N2 gas 2))Reverse range O2 measurement Zero gas ;100% O2 Span gas ;For O2 measurement, O2 gas of 95.0%~95.5%/remains N2 gas *For reverse range O 2 measurement, infrared measurable component is not measurable. *1) 0-200ppm range ≤ 2.0%FS *2) 0-500ppm range or less ≤ 2.0%FS Installation Requirements EC Directive Compliance The product conforms to the requirements of the Low Voltage Directive 2006/95/EC and EMC directive 89/336/EEC (as amended by Directive 92/31/EEC), both as amended by Directive 93/68/EEC. It conforms to following standards for product safety and electromagnetic compatibility ; EN61010-1 : 2010Safety requirements for electrical equipment for measurement, control and laboratory use. “Installation Category II” “Pollution Degree 2” EN61326-1 : 2006 Electrical equipment for measurement, control and laboratory use — EMC requirements. Standard Requirements for Sample Gas Flow rate: Temperature: Pressure: Dust: Mist: Moisture: 4 0.5L / min ±0.2L / min 0 to 50˚C 10 kPa or less (Gas outlet side should be open to the atmospheric air.) 100 µg/Nm3 or less in particle size of 0.3 µm or less Unallowable Below a level where saturation occurs at room temperature (condensation unallowable). Below the level where saturation occurs at 2˚C for CO measurement in 0 to 200 ppm range, NO measurement, and SO 2 measurement. •Indoor use (Select a place where the equipment does not receive direct sunlight, draft/rain or radiation from hot substances. If such a place cannot be found, a roof or cover should be prepared for protection.) •Avoid a place where unit receives heavy vibration •Select a place where atmospheric air is clean Principle diagram of NDIR type measurement (For NO, SO2, CO2, CO, CH4) Front expansion chamber Gas inlet Infrared-ray light source Gas outlet Rear expansion chamber Detector Motor Mass flow sensor Measuring cell Chopper Preamplifier Signal processing and calculation block Indication Output Principle diagram of fuel cell type measurment (For O2) Resistor Thermistor Electrolyte Negative electrode Diaphragm Positive electrode Principle diagram of paramagnetic type measurment (For O2) Permanent magnet Mirror Dumb-bell Gas inlet Measuring cell Magnetic field Gas outlet Preamplifier Permanent magnet Photo diode Light emitting diode Indication Signal processing and calculation block Output 5 ZPA Examples of sampling system configuration including gas analyzer To measure low moisture content (saturated at room temperature or lower) sample gas (CO, CO2, CH4) 9 Mist filter Sample gas inlet 1 Flow meter 10 Pump 3 Two-way Solenoid valve 7 7 7 7 Gas analyzer (ZPA) Solenoid valve Drain pot 5 11 Membrane filter Demister O2 analyzer 6 Ball valve 6 Ball valve (In the case of zirconia type) Exhaust (atmospheric pressure) Drain Reference gas for zero calibration 8 Reference gas for span calibration To measure high moisture content sample gas, NO, SO2, or CO (0 to 200 ppm range) 13 1 Mist filter 3 Pump Sample gas inlet 3 Solenoid valve 9 10 7 Pump Atmospheric (Air) *1) 2 4 Safety drain trap Flow meter Two-way Solenoid valve 12 7 7 7 Electronic cooler 5 6 NOX/NO converter (for NOX measurement) Solenoid valve Drain pot Ball valve Drain Membrane filter Gas analyzer (ZPA) O2 analyzer Reference gas for zero calibration 8 Reference gas for span calibration (In the case of zirconia type) Exhaust (atmospheric pressure) *1) Be sure to use a dehumidifier such as electronic cooler for NO, SO2, and CO analyzers of 0 to 200 ppm range (=2ºC saturation or lower). List of sampling devices (example) No. Device name Fuji’s type 1 Mist filter ZBBK1V03-0 2 Safety drain trap ZBH51603 3 Pump ZBG80 4 Electoric cooler ZBC91004 5 Drain pot ZBH13003 (Length 255mm) 6 Ball valve ZBFB1 7 Two-way solenoid valve 8 Standard gas for calibration ZBM 9 Flow meter ZBD42203 10 Membrane filter ZBBM2V03-0 11 Demister ZBH35003 12 NO2/NO converter ZDL02001 13 Y04-0 (Codes in to be selected depending on application) Three-way solenoid valve Note) The above is a typical configuration example. As configuration may differ depending on measuring objects, please consult us. 6 CODE SYMBOLS standard multi components Description Digit 4 <Specification/Structure> Horizontal type (Terminal block for power supply) Horizontal type (Power inlet, with lock) 5 <Mounting> 19inch rack mounting type EIA conformity 6 <Measurable component (NDIR)> 1st component 2nd component 3rd component 4th component None NO SO2 CO2 CO CH4 NO SO2 NO CO CO2 CO CH4 CO CO2 CH4 NO SO2 CO CO CH4 CO2 NO SO2 CO2 CO Others 7 <Measurable component (O2)> None External O2 analyzer External zirconia O2 sensor (ZFK7) Built-in fuel cell O2 sensor Built-in paramagnetic O2 sensor 8 <Revision code> 9 <Measuring range (NDIR)>1st component, 1st range 10 <Measuring range (NDIR)>1st component, 2nd range 11 <Measuring range (NDIR)>2nd component, 1st range 12 <Measuring range (NDIR)>2nd component, 2nd range 13 <Measuring range (NDIR)>3rd component, 1st range 14 <Measuring range (NDIR)>3rd component, 2nd range 15 <Measuring range (NDIR)>4th component, 1st range 16 <Measuring range (NDIR)>4th component, 2nd range 17 <Measuring range (O2)> None 0-5/10% 0-5/25% 0-10/25% 0-5% 0-10% 0-25% 0-50% 0-100% 100-95% Others 18 <Gas connection> Rc1/4 NPT1/4 19 <Output> DC0-1V DC4-20mA DC0-1V+Communication function DC4-20mA+Communication function 20 <Indication/power supply cable> Japanese (PSE) English (UL) English (CEE) Chinese (CCC) 21 <O2 correction and O2 correction average output> None O2 correction O2 correction average O2 correction and O2 correction average 22 <Optional function (DIO)> FAULT A. Cal. H/L Alarm RangeID/Remote range None 1 2 3 4 5 6 7 8 note note1 ZPA B 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 1 Digit A D B note2 note3 note4 note4 note4 note4 note4 note4 note4 note4 note5 Y P A D B E F G J K L N T V Z Y 1 2 3 4 1 Y A B C L M V P R S Z 1 2 A B C D note6 note6 note6 note6 note7 J E U C Y A B C note8 Y A B C D E F G H 7 ZPA 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 1 Digit Y 1 A B note9 A C D Z note10 RANGE CODE Range None 0~100ppm 0~200ppm 0~250ppm 0~300ppm 0~500ppm 0~1000ppm 0~2000ppm 0~2500ppm 0~3000ppm 0~5000ppm ZRE note Description Digit 23 <Pressure compensation> None Pressure compensation 24 <Un> ppm, Vol% mg/m3, g/m3 25 <Adjustment> For standard For heat treatment furnace For converter Others O2 measurement range Code Y B C D S E F G U T H Range 0~1% 0~2% 0~3% 0~5% 0~10% 0~20% 0~25% 0~40% 0~50% 0~70% 0~100% Others Code J K Q L M N V W P X R Z Measurement Range Fuel cell O2 Paramagnet Zirconia O2 range code sensor ic O2 sensor sensor (external) (built - in) (built - in) A 0~5/10 vol% 0~5/25 vol% B 0~10/25 vol% C 0~5 vol% L 0~10 vol% M 0~25 vol% V 0~50 vol% P 0~100 vol% R 100~95 vol% S note1)When “D” is specified at 4th digit, Power cable is supplied in the scope of supply. Cable specification should be specified at the 20th digit. note2)When only O2 measurement is necessary, "Y" should be specified at 6th digit. note3)When "1" is specified at 7th digit, O 2 pt sensor signal has to be set as 0-1V DC linear corresponding to full scale. External zirconia O2 sensor and external O2 analyzer are not included in the scope of supply, and has to be separately ordered. note4)Refer to Table 1 for possible combination of measuring components and ranges in the data sheet. When "Y" is specified at 6th digit, "Y" should be specified at 9th to 16th digit. For fuel cell O 2 sensor, range is 0-10% or more. note5)Select the type of voltage rating, plug type and applicable standard of the power cable by 20th digit. Select a power cable for using at the location of end-user. When “A” is specified at 4th digit, Power cable will not be included in scope of delivery. When English is specified for display language, "E" should be selected at 20th digit. note6)O2 correction is calculated only for NO, SO 2 and CO. note7)When 5 components measurement is specified, "H" must not be specified at 22nd digit. When 4 components measurement is specified and “H” is specified at 22nd digit, 3 points is maximum for alarm output function. note8)When "B" is specified at 24th digit, measuring range should be specified by ppm range code. In this case NO, SO 2 and CO measuring range are corresponding range in mg/m 3. Please refer to the table shown below for the corresponding range code based on "mg/m 3". note9)When A to D is specified at 25th digit, the analyzer will be adjusted and delivered with the following gasses. Standard "A": balance gas N2. For heat treatment furnace "C": balance gas 30% H 2/remamins N2. For converter "D": balance gas CO, CO 2. When other adjustment is required, please specify "Z". When "Z" is specified, please attach a list of gas composition contained in the measuring gas. Corresponding mg/m3 Range code C D S E F G U T H 8 Unit : ppm 0-200ppm 0-250ppm 0-300ppm 0-500ppm 0-1,000ppm 0-2,000ppm 0-2,500ppm 0-3,000ppm 0-5,000ppm Corresponding range in mg/m3 NO SO2 CO 0-260mg/m3 0-250mg/m3 0-570mg/m3 0-325mg/m3 0-700mg/m3 0-300mg/m3 0-400mg/m3 0-850mg/m3 0-375mg/m3 0-650mg/m3 0-1,400mg/m3 0-600mg/m3 0-1,300mg/m3 0-2,800mg/m3 0-1,250mg/m3 0-2,600mg/m3 0-5,600mg/m3 0-2,500mg/m3 0-3,300mg/m3 0-7,100mg/m3 0-3,000mg/m3 0-4,000mg/m3 0-8,500mg/m3 0-3,750mg/m3 0-6,600mg/m3 0-14.00g/m3 0-6,250mg/m3 The conversion formula "ppm" unit into "mg/m3" unit. NO (mg/m3) = 1.34 × NO (ppm) SO2 (mg/m3) = 2.86 × SO2 (ppm) CO (mg/m3) = 1.25 × CO (ppm) Table 1 Measurable component and range - availability check table Procedure of range selection On one component analyzer: First determine 1st range, then select 2nd range from the right of your determination range in following tables. More than two components: The procedure is same as one component. Difference is 2nd range in the tables. The 2nd range in the tables for two and more components is maximum available range. 2nd range is selectable from 1st range to 2nd range(max) on the table. 1-component analyzer : CO 1st range 0 - 200ppm 0 - 250ppm 0 - 300ppm 0 - 500ppm 0 - 1000ppm 0 - 2000ppm 0 - 2500ppm 0 - 3000ppm 0 - 5000ppm 0 - 1% 0 - 2% 0 - 3% 0 - 5% 0 - 10% 0 - 20% 0 - 25% 0 - 40% 0 - 50% 0 - 70% 0 - 100% 1-component analyzer : NO 2nd range None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1% None, 0 - 2500ppm,3000ppm,5000ppm,1%,2% None, 0 - 3000ppm,5000ppm,1%,2% None, 0 - 5000ppm,1%,2% None, 0 - 1%,2%,3%,5% None, 0 - 2%,3%,5%,10% None, 0 - 3%,5%,10%,20% None, 0 - 5%,10%,20%,25% None, 0 - 10%,20%,25%,40%,50% None, 0 - 20%,25%,40%,50%,70%,100% None, 0 - 25%,40%,50%,70%,100% None, 0 - 40%,50%,70%,100% None, 0 - 50%,70%,100% None, 0 - 70%,100% None, 0 - 100% None 1-component analyzer : CO2 1st range 0 - 100ppm 0 - 200ppm 0 - 250ppm 0 - 300ppm 0 - 500ppm 0 - 1000ppm 0 - 2000ppm 0 - 2500ppm 0 - 3000ppm 0 - 5000ppm 0 - 1% 0 - 2% 0 - 3% 0 - 5% 0 - 10% 0 - 20% 0 - 25% 0 - 40% 0 - 50% 0 - 70% 0 - 100% 2nd range None, 0 - 200ppm,250ppm,300ppm,500ppm,1000ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1% None, 0 - 2500ppm,3000ppm,5000ppm,1%,2% None, 0 - 3000ppm,5000ppm,1%,2% None, 0 - 5000ppm,1%,2% None, 0 - 1%,2%,3%,5% None, 0 - 2%,3%,5%,10% None, 0 - 3%,5%,10%,20% None, 0 - 5%,10%,20%,25% None, 0 - 10%,20%,25%,40%,50% None, 0 - 20%,25%,40%,50%,70%,100% None, 0 - 25%,40%,50%,70%,100% None, 0 - 40%,50%,70%,100% None, 0 - 50%,70%,100% None, 0 - 70%,100% None, 0 - 100% None 2-component analyzer : NO/SO2 1-component : NO 1st range 2nd range (max.) 0 - 200ppm 0 - 2000ppm 0 - 250ppm 0 - 2500ppm 0 - 300ppm 0 - 2500ppm 0 - 500ppm 0 - 5000ppm 0 - 1000ppm 0 - 5000ppm 0 - 2000ppm 0 - 5000ppm 0 - 2500ppm 0 - 5000ppm 0 - 3000ppm 0 - 5000ppm 0 - 5000ppm None 1st range 0 - 200ppm 0 - 250ppm 0 - 300ppm 0 - 500ppm 0 - 1000ppm 0 - 2000ppm 0 - 2500ppm 0 - 3000ppm 0 - 5000ppm 2nd range None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm None, 0 - 2500ppm,3000ppm,5000ppm None, 0 - 3000ppm,5000ppm None, 0 - 5000ppm None 1-component analyzer : SO2 1st range 0 - 200ppm 0 - 250ppm 0 - 300ppm 0 - 500ppm 0 - 1000ppm 0 - 2000ppm 0 - 2500ppm 0 - 3000ppm 0 - 5000ppm 0 - 1% 0 - 2% 0 - 3% 0 - 5% 0 - 10% 2nd range None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1% None, 0 - 2500ppm,3000ppm,5000ppm,1%,2% None, 0 - 3000ppm,5000ppm,1%,2% None, 0 - 5000ppm,1%,2% None, 0 - 1%,2%,3%,5% None, 0 - 2%,3%,5%,10% None, 0 - 3%,5%,10% None, 0 - 10% None, 0 - 10% None 1-component analyzer : CH4 1st range 0 - 500ppm 0 - 1000ppm 0 - 2000ppm 0 - 2500ppm 0 - 3000ppm 0 - 5000ppm 0 - 1% 0 - 2% 0 - 3% 0 - 5% 0 - 10% 0 - 20% 0 - 25% 0 - 40% 0 - 50% 0 - 70% 0 - 100% 2nd range None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1% None, 0 - 2500ppm,3000ppm,5000ppm,1%,2% None, 0 - 3000ppm,5000ppm,1%,2% None, 0 - 5000ppm,1%,2% None, 0 - 1%,2%,3%,5% None, 0 - 2%,3%,5%,10% None, 0 - 3%,5%,10%,20% None, 0 - 5%,10%,20%,25% None, 0 - 10%,20%,25%,40%,50% None, 0 - 20%,25%,40%,50%,70%,100% None, 0 - 25%,40%,50%,70%,100% None, 0 - 40%,50%,70%,100% None, 0 - 50%,70%,100% None, 0 - 70%,100% None, 0 - 100% None 2-component analyzer : NO/CO 2-component : SO2 1st range 2nd range (max.) 0 - 200ppm 0 - 2000ppm 0 - 250ppm 0 - 2500ppm 0 - 300ppm 0 - 2500ppm 0 - 500ppm 0 - 5000ppm 0 - 1000ppm 0 - 5000ppm 0 - 2000ppm 0 - 5000ppm 0 - 2500ppm 0 - 5000ppm 0 - 3000ppm 0 - 5000ppm 0 - 5000ppm None • The 2nd component should be selected as shown in the right table. 1-component : NO 1st range 2nd range (max.) 0 - 200ppm 0 - 2000ppm 0 - 250ppm 0 - 2500ppm 0 - 300ppm 0 - 2500ppm 0 - 500ppm 0 - 5000ppm 0 - 1000ppm 0 - 5000ppm 0 - 2000ppm 0 - 5000ppm 0 - 2500ppm 0 - 5000ppm 0 - 3000ppm 0 - 5000ppm 0 - 5000ppm None 2-component : CO 1st range 2nd range (max.) 0 - 200ppm 0 - 2000ppm 0 - 250ppm 0 - 2500ppm 0 - 300ppm 0 - 2500ppm 0 - 500ppm 0 - 5000ppm 0 - 1000ppm 0 - 5000ppm 0 - 2000ppm 0 - 5000ppm 0 - 2500ppm 0 - 5000ppm 0 - 3000ppm 0 - 5000ppm 0 - 5000ppm None • The 2nd component should be selected as shown in the right table. 9 ZPA 2-component analyzer: CO2/CO 2-component: CO 1-component: CO2 1st range 2nd range (max.) 1st range/2nd range (max.) 0-100ppm 0-1000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm 0-200ppm 0-2000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%, 0-250ppm 0-2500ppm 0-3% 0-300ppm 0-500ppm 0-500ppm 0-5000ppm 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%, 0-3% 0-1000ppm 0-5000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-2000ppm 0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-1000ppm 0-1% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-2000ppm 0-1% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-2000ppm 0-2% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-2500ppm 0-1% 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-2500ppm 0-2% 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-3000ppm 0-1% 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-3000ppm 0-2% 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/25%, 0-10/100%,0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-5000ppm 0-3% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-1% 0-5% 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-5% 0-5000ppm 0-5% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-1% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-20% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-3% 0-25% 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-5% 0-50% 0-10% 0-100% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-20% 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-25% 0-40% 0-50% 0-70% 0-100% None 2-component analyzer: CH4/CO 2-component: CO 1-component: CH4 1st range 2nd range (max.) 1st range/2nd range (max.) 0-500ppm 0-5000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm 0-1000ppm 0-5000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm 0-1000ppm 0-1% 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm 0-2000ppm 0-5000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20% 0-2500ppm 0-5000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% 0-3000ppm/1%, 0-3000ppm 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25% 0-2000ppm 0-1% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20% 0-2500ppm 0-1% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25% 0-3000ppm 0-2000ppm 0-2% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20% 0-2500ppm 0-2% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25% 0-3000ppm 0-5000ppm 0-1% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-5000ppm 0-3% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-5000ppm 0-5% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5% 0-2/5%, 0-3/25%, 0-5/25%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-1% 0-5% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-1% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/1%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-20% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-3% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-3% 0-25% 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-5% 0-25% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-5% 0-50% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/5%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-10% 0-50% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-10% 0-100% 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-20% 0-50% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-25% 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-40% 0-20% 0-100% 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-25% 0-40% 0-50% 0-70% 0-100% None 10 2-component analyzer: CO2/CH4 2-component: CH4 1-component: CO2 1st range 2nd range (max.) 1st range/2nd range (max.) 0-100ppm 0-1000ppm 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm 0-200ppm 0-2000ppm 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/1%, 0-1% 0-250ppm 0-2500ppm 0-300ppm 0-2500ppm 0-500ppm 0-2500ppm 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10% 0-500ppm 0-5000ppm 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10% 0-1000ppm 0-2500ppm 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20% 0-1000ppm 0-5000ppm 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3% 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20% 0-1000ppm 0-1% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20% 0-2000ppm 0-2500ppm 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20% 0-2000ppm 0-5000ppm 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20% 0-2000ppm 0-2% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20% 0-2500ppm 0-5000ppm 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25% 0-2500ppm 0-2% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25% 0-3000ppm 0-2% 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5% 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25% 0-5000ppm 0-3% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-5000ppm 0-5% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-1% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50% 0-2% 0-20% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-10% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-3% 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-3% 0-25% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-5% 0-20% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-5% 0-50% 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-10% 0-20% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-10% 0-50% 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-20% 0-50/100%, 0-70/100%, 0-100% 0-25% 0-40% 0-10% 0-100% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-20% 0-100% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-25% 0-100% 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-40% 0-70/100%, 0-100% 0-50% 0-70% 0-100% None 3-component analyzer: NO/SO2/CO >>> Combination of 1st component NO and 2nd component SO2 / 3rd component CO 2-component: SO2 3-component: CO 1-component: NO 1st range 2nd range (max.) 1st range 2nd range (max.) 1st range/2nd range (max.) 0-200ppm 0-2000ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm 0-200ppm 0-2000ppm 0-250ppm 0-2500ppm 0-250ppm 0-2500ppm 0-300ppm 0-300ppm 0-2500ppm 0-500ppm 0-2500ppm 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-5000ppm 0-500ppm 0-5000ppm 0-1000ppm 0-1000ppm 0-5000ppm 0-2500ppm None 0-2000ppm 0-5000ppm 0-1000ppm 0-5000ppm 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm 0-2500ppm 0-5000ppm 0-2000ppm 0-3000ppm 0-5000ppm 0-2500ppm 0-5000ppm None 0-3000ppm 0-5000ppm None 3-component analyzer: CO2/CO/CH4 >>> Combination of 1st component CO2 / 2nd component CO and 3rd component CH4 1-component: CO2 2-component: CO 3-component: CH4 1st range 2nd range (max.) 1st range/2nd range (max.) 1st range 2nd range (max.) 0-5000ppm 0-5% 0-5000ppm 0-3% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-1% 0-5% 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-5% 0-5000ppm 0-5% 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-1% 0-10% 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-2% 0-20% 0-1% 0-10% 0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-3% 0-25% 0-2% 0-20% 0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-5% 0-10% 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-10% 0-20% 0-2% 0-10% 0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3% 0-25% 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-20% 0-25% 0-5% 0-50% 0-25% 0-40% 0-10% 0-100% 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-40% 0-50% 0-20% 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-25% 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100% 0-50% 0-70% 0-40% 0-50% 0-70% 0-100% None 0-70% 0-100% 0-100% None Availability of product Product available only when CO analyzer max. measuring range is 50% or less Product available Product available only when CO analyzer measuring range is 0 to 1000ppm or more. Product available only when CO analyzer measuring range is 0 to 5000ppm or more. Product available only when CO analyzer measuring range is more than 5000ppm or CO2 analyzer range is more than 2%. 11 ZPA 4-component analyzer: NO/SO2/CO2/CO >>> Combination of 1st component NO /4th component CO and component 2nd component SO2/3rd component CO2 1-component: NO 1st range 2nd range (max.) 0-200ppm 0-2000ppm 0-250ppm 0-2500ppm 0-300ppm 0-2500ppm 0-500ppm 0-2000ppm 0-1000ppm 0-2000ppm 0-2000ppm None 0-500ppm 0-5000ppm 0-1000ppm 0-5000ppm 0-2000ppm 0-5000ppm 0-2500ppm 0-5000ppm 0-3000ppm 0-5000ppm 0-5000ppm None 4-component: CO 1st range/2nd range (max.) 0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None 2-component analyzer: SO2 3-component analyzer: CO2 1st range 2nd range (max.) 1st range/2nd range (max.) 0-200ppm 0-2000ppm 0-250ppm 0-2500ppm 0-300ppm 0-2500ppm 0-500ppm 0-5000ppm 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%/None 0-1000ppm 0-5000ppm 0-2000ppm 0-5000ppm 0-2500ppm 0-5000ppm 0-3000ppm 0-5000ppm 0-5000ppm None Table 2 Channel (Ch) No. and display/output contents comparison table Code symbol 6th digit 7th digit 21st digit Y 1 to 3 Y P Y Y A Y Y D Y Y B Y Y E Y Y F Y Y G Y Y J Y Y K Y Y L Y Y N Y Y T Y Y V Y Y P 1 to 3 Y A 1 to 3 Y D 1 to 3 Y B 1 to 3 Y E 1 to 3 Y F 1 to 3 Y G 1 to 3 Y J 1 to 3 Y K 1 to 3 Y L 1 to 3 Y N 1 to 3 Y T 1 to 3 Y V 1 to 3 Y P 1 to 3 A* A 1 to 3 A* B 1 to 3 A* F 1 to 3 A* G 1 to 3 A* J 1 to 3 A* N 1 to 3 A* V 1 to 3 A* P 1 to 3 C * A 1 to 3 C * B 1 to 3 C * F 1 to 3 C * G 1 to 3 C * J N 1 to 3 1 to 3 C * C * V 1 to 3 C * Display/output contents Ch1:O2 Ch1:NO Ch1:SO2 Ch1:CO2 Ch1:CO Ch1:CH4 Ch1:NO, Ch2:SO2 Ch1:NO, Ch2:CO Ch1:CO2, Ch2:CO Ch1:CH4, Ch2:CO Ch1:CO2, Ch2:CH4 Ch1:NO, Ch2:SO2, Ch3:CO Ch1:CO2, Ch2:CO, Ch3:CH4 Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO Ch1:NO, Ch2:O2 Ch1:SO2, Ch2:O2 Ch1:CO2, Ch2:O2 Ch1:CO, Ch2:O2 Ch1:CH4, Ch2:O2 Ch1:NO, Ch2:SO2, Ch3:O2 Ch1:NO, Ch2:CO, Ch3:O2 Ch1:CO2, Ch2:CO, Ch3:O2 Ch1:CH4, Ch2:CO, Ch3:O2 Ch1:CO2, Ch2:CH4, Ch3:O2 Ch1:NO, Ch2:SO2, Ch3:CO, Ch4:O2 Ch1:CO2, Ch2:CO, Ch3:CH4, Ch4:O2 Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2 Ch1:NOx, Ch2:O2, Ch3:corrected NOx Ch1:SO2, Ch2:O2, Ch3:corrected SO2 Ch1:CO, Ch2:O2, Ch3:corrected CO Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2 Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO Ch1:NOx, Ch2:O2, Ch3:corrected NOx, Ch4:corrected NOx average Ch1:SO2, Ch2:O2, Ch3:corrected SO2, Ch4:corrected SO2 average Ch1:CO, Ch2:O2, Ch3:corrected CO, Ch4corrected CO average Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2, Ch6:corrected NOx average, Ch7:corrected SO2 average Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO, Ch6:corrected NOx average, Ch7:corrected CO average Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO, Ch5:corrected CO average Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO, Ch8:corrected NOx average, Ch9:corrected SO2 average, Ch10:corrected CO average Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO, Ch9:corrected NOx average, Ch10:corrected SO2 average2, Ch11:corrected CO average * When the 21st digit code is A or C, the component of the NO analyzer is displayed as NOx. 12 OUTLINE DIAGRAMS (Unit : mm) Refer to appendix. ((TC drawing) 429 57.2 <FRONT VIEW> 132.5 7 <TOP VIEW> 415 463 483 Power Switch Analog Output Connctor (A/O) Digital Input/output Connector (DIO1 to 3) Purge Gas Inlet Rc1/4 or NPT1/4 Communication Connector (RS485) Sample Gas Inlet Rc1/4 or NPT1/4 Analog Input Connector (A/I) Sample Gas Outlet Rc1/4 or NPT1/4 Fuse <REAR VIEW> (M4 Terminal Block Type) 124.5 Power Inlet(100V to 240V AC 50/60Hz) <SIDE VIEW> 24 2 25 380 396 16 (M4 Terminal Block Type) (Power Inlet Type(100V to 240V AC 50/60Hz)) 13 ZPA Mounting method The analyzer weight should be supported at the bottom of the case. 19-inch rack mounting type 57.2 <Rack dimensions> 450 or more 465 Mounting diagram Guide rails Guide rails EXTERNAL CONNECTION <Analog output> A/O connector 13 1 <RS485 communication signal> 1 14 2 25 14 15 3 D-sub 25pins female 16 4 * In standard, displayed Channel No. and Analog Output No. are same. 17 5 18 6 <Screw terminal (M4)> P N L 19 7 20 8 100 to 240V AC 21 9 22 10 <Power inlet> 23 11 24 12 POWER INLET 100-240V AC 25 13 14 AO1+ AO1AO2+ AO2AO3+ AO3AO4+ AO4AO5+ AO5AO6+ AO6AO7+ AO7AO8+ AO8AO9+ AO9AO10+ AO10AO11+ AO11AO12+ AO12NC 5 (GND) 1 1 6 RT×D+ 2 9 6 7 RT×D- 3 D-sub 9pins female 8 4 9 5 <Analog input> A/I connector (O2 signal input) AI- AI+ <Digital I/O> DIO 1 to 3 connector (option) 13 DIO1 DIO2 DIO3 connector connector connector 1 1 14 25 2 14 15 3 D-sub 25pins female 16 4 * DIO 1 to 3 are all as same connector. 17 5 18 6 19 7 20 Contents of digital input signal DI1 Remote hold DI2 Average value reset DI3 A. cal. start DI4 A. zero. cal. start DI5 Remote range Ch1 DI6 Remote range Ch2 DI7 Remote range Ch3 DI8 Remote range Ch4 DI9 Remote range Ch5 8 21 9 22 10 23 11 NC com NO NC com NO NC com NO NC com NO NC com NO DI1+ DI1DI2+ DI2DI3+ DI3- DI4+ DI4DI5+ DI5DI6+ DI6- DI7+ DI7DI8+ DI8DI9+ DI9- DO1 DO6 DO11 DO2 DO7 DO12 DO3 DO8 DO13 DO4 DO9 DO14 DO5 DO10 DO15 Digital input OFF : 0V ON : 12 to 24V DC Digital output max. contact load rating 24V DC/1A 24 12 25 13 Contents of digital output signal Independent on the number of component 1-component analyzer 22th digit A, C B, E D, F, G, H B, D, E, F, G, H B, D, E, F, G, H DO1 Instrument error Instrument error Instrument error Instrument error Instrument error DO2 Calibration error Calibration error Calibration error Calibration error Calibration error DO3 A.cal.status (A.cal.status) (A.cal.status) (A.cal.status) DO4 For zero gas (For zero gas) (For zero gas) (For zero gas) For span gas Ch1 (For span gas Ch1) (For span gas Ch1) (For span gas Ch1) (For span gas Ch2) (For span gas Ch2) DO5 2-component analyzer 3-component analyzer DO6 (Alarm1) (Alarm1) DO7 (Alarm2) (Alarm2) (For span gas Ch3) DO8 (Alarm3) (Alarm3) (Range identification Ch1) DO9 (Alarm4) (Alarm4) DO10 (Alarm5) (Alarm5) (Range identification Ch1) (Range identification Ch2) Range identification Ch1 (Range identification Ch2) (Range identification Ch3) DO11 (Alarm1) (Alarm1) (Alarm1) DO12 (Alarm2) (Alarm2) (Alarm2) DO13 (Alarm3) (Alarm3) (Alarm3) DO14 (Alarm4) (Alarm4) (Alarm4) DO15 (Alarm5) (Alarm5) (Alarm5) 4-component analyzer The items in the parentheses may not be available depending on the selected type on 22th digit. The normal open side (NO) of digital output is close when the function is active without range ID. In case of range ID, normal open (NO) side is close with First range. The normal close (NC) side is close with Second range. 5-component analyzer 22th digit B, E D, F G H B, E D, F G DO1 Instrument error Instrument error Instrument error Instrument error Instrument error Instrument error Instrument error DO2 Calibration error Calibration error Calibration error Calibration error Calibration error Calibration error Calibration error DO3 A.cal.status A.cal.status A.cal.status A.cal.status A.cal.status DO4 For zero gas For zero gas For zero gas For zero gas For zero gas DO5 For span gas Ch1 For span gas Ch1 For span gas Ch1 For span gas Ch1 For span gas Ch1 DO6 For span gas Ch2 For span gas Ch2 For span gas Ch2 For span gas Ch2 Range identification Ch1 For span gas Ch2 DO7 For span gas Ch3 Range identification Ch1 For span gas Ch3 For span gas Ch3 For span gas Ch3 Range identification Ch2 For span gas Ch3 DO8 For span gas Ch4 Range identification Ch2 For span gas Ch4 For span gas Ch4 For span gas Ch4 Range identification Ch3 For span gas Ch4 DO9 Range identification Ch3 Range identification Ch1 For span gas Ch5 Range identification Ch4 For span gas Ch5 DO10 Range identification Ch4 Range identification Ch2 Range identification Ch5 DO11 (Alarm1) (Alarm1) (Alarm1) (Alarm1) (Alarm1) Range identification Ch1 DO12 (Alarm2) (Alarm2) Range identification Ch1 (Alarm2) (Alarm2) (Alarm2) Range identification Ch2 DO13 (Alarm3) (Alarm3) Range identification Ch2 (Alarm3) (Alarm3) (Alarm3) Range identification Ch3 DO14 (Alarm4) (Alarm4) Range identification Ch3 Range identification Ch3 (Alarm4) (Alarm4) Range identification Ch4 DO15 (Alarm5) (Alarm5) Range identification Ch4 Range identification Ch4 (Alarm5) (Alarm5) Range identification Ch5 SCOPE OF DELIVERY ORDERING INFORMATION •Gas analyzer … 1 unit •Replacement fuse (250V, 2A AC, delay type) … 2 pcs •Instruction manual … 1 copy •Connector for I/O connection … 1 set •Power cable(standard inlet type 2m) … 1 pc 1. Code symbols 2.Application and composition of sample gas 15 ZPA Gas inlet/outlet size: Rc1/4 or NPT1/4 Power supply: Rated voltage Exclusive Zirconia O2 Sensor (to be purchased separately) For O2 correction, the gas analyzer ZPA can accept linearized 0 to 1V DC signal coming from analyzer calibrated 0 to 25% O2 full scale. If the analyzer is not available, Fuji can supply exclusive Zirconia O2 sensor Model ZFK. Measuring method: Zirconia system Measurable component and measuring range: Measurable component Range O 2 Oxygen 0 to 25vol% Repeatability: Within ± 0.5% of full scale Linearity: Within ± 1% of full scale Zero drift: Within ± 1% of full scale/week Span drift: Within ± 2% of full scale/week Response time: Approx. 20 seconds (for 90% response) Measured gas flow rate: 0.5 ± 0.25L / min Remark:The Zirconia system, due to its principle, may produce a measuring error due to relative concentration versus the combustible O2 gas concentration. Also, a corrosive gas (SO2 of 250 ppm or more, etc.) may affect the life of the sensor. OUTLINE DIAGRAM ;100 to 115V AC or 200 to 240V AC Rated frequency ;50Hz/60Hz Max. rated power;215VA (during power ON) 65VA (during steadystate operation) Enclosure: Steel casing, for indoor application Indication: Temperature indication (LED) Temperature alarm output: Contact output 1a contact, Contact capacity 220V, 1A AC (resistive load) Outer dimensions (H x W x D): 141 x 170 x 190mm Mass {weight}: Approx. 3kg Finish color: Munsell 5Y 7/1 CODE SYMBOLS 1 2 3 4 5 6 7 8 Z F K 7 Y Y 9 10 11 12 13 4 - Measuring method Zirconia method 7YY Power supply 100 to 115V AC 50/60Hz(Standard) 200 to 240V AC 50/60Hz(Standard) 200 to 240V AC 50/60Hz(CE mark) 9 B C (Unit:mm) Description Y 0 Y Y Gas inlet/outlet size Rc 1/4 NPT 1/4 1 8 170 4 152 OXYGEN ANALYZER US 1 AC250V T3.15A 2 3 4 5 6 7 E F 110 131 141 TEMP. L N AC INPUT OUT ALM 9TH DIGIT 1: RC1/4 8: NPT1/4 ø5 EXTERNAL CONNECTION DIAGRAM 1 2 3 4 + AC power supply E 5 6 INLET OUTLET 7 Output Temperature to analyzer alarm output Caution on Safety *Before using this product, be sure to read its instruction manual in advance. International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fujielectric.com/products/instruments/ Information in this catalog is subject to change without notice. Printed in Japan