FUJI EDSX6-101

FCX – AX SERIES REMOTE SEAL TYPE
DIFFERENTIAL PRESSURE TRANSMITTER
FHD, FKD...3
DATA SHEET
The FCX – AX differential pressure transmitter accurately
measures differential pressure, liquid level or gauge pressure
and transmits a proportional 4 to 20mA signal. The transmitter utilizes a unique micromachined capacitance silicon
sensor with state-of-the-art microprocessor technology to
provide exceptional performance and functionality. Totally
welded construction of the seals assures excellent reliability
in high temperature and highly corrosive process conditions.
FEATURES
1. High accuracy
0.2% accuracy for all calibrated spans is a standard
feature for all DP models covering 3.2kPa{32mbar} range
to 500kPa{5bar} high differential pressure range. 0.1%
accuracy is available as option. Fuji's micro-capacitance
silicon sensor assures this accuracy for all elevated or
suppressed calibration ranges without additional
adjustment.
2. Minimum environmental influence
The "Advanced Floating Cell" design which protects the
pressure sensor against changes in temperature, static
pressure, and overpressure substantially reduces total
measurement error in actual field applications.
3. Replaceable Communication Module
Fuji micro-electronics manufacturing technology offers
replaceable communication module that makes FCX –
AX transmitter very unique in design. In case of change
in communication protocol, all that needs to be done is
just to replace the module and the transmitter gets
upgraded to the new version.
4. Fuji/HART bilingual communication module
The communication module is "bilingual" to speak both
Fuji proprietary protocol and HART. Any HART compatible
devices can communicate with FCX – AX series transmitters.
5. Application flexibility
Various options that render the FCX – AX suitable for
almost any process applications include:
– Analog indicator at either the electronics side or
terminal side
– Full range of hazardous area approvals
– Built-in RFI filter and lightning arrester
– 4 12 -digits LCD meter
– Stainless steel electronics housing
– Wide selection of materials
– High temperature, high vacuum seals
6. Programmable output Linearization Function
In addition to Linear and Square Root, output signal
can be freely programmable.
(Up to 14 compensated points at approximation.)
(Available for amplifier unit from version 24 and FXW(HHC) version
5.3.)
7. Burnout current flexibility (Under Scale: 3.2 to 3.8mA,
Over Scale: 20.8 to 21.6mA)
Burnout signal level is adjustable using Model FXW hand
Held Communicator (HHC) to comply with NAMUR
NE43.
8. Dry calibration without reference pressure
Thanks to the best combination of unique construction
of mechanical parts (Sensor unit) and high performance
electronics circuit (Electronics unit), reliability of dry
calibration without reference pressure is at equal level
as wet calibration.
SPECIFICATIONS
Functional specifications
Type:
Model FHD: 4 to 20mA
Model FKD: 4 to 20mA with digital signal
Service:
Liquid, gas, or vapour
Static pressure, span, and range limit:
Span limit [kPa]
(m bar}
Type
FHD
F D
3
F D
5
F D
6
}
{
Up tp
flange rating {
{
Range limit
[kPa] {m bar}
Max.
Min.
Static pressure
FKD
FHD/FKD
3.2
0.32
32 } { 3.2 } {
13
1.3
130 } { 13 } {
50
5
500 } { 50 } {
32
320 }
130
1300 }
500
5000 }
+/–
{ +/–
+/–
{ +/–
+/–
{ +/–
32
320 }
130
1300 }
500
5000 }
Remark : To minimize environmental influence, span should be greater
than 1/40 of the max. span in most applications.
– Lower limit of static pressure (vacuum limit),
Silicone fill sensor: See Fig. 1
Fluorinated fill sensor: Atmospheric pressure
– The maximum span of each sensor can be converted
to different units using factors as below.
1MPa=103kPa=10bar=10.19716kgf/cm2=
145.0377psi
1kPa=10mbar=101.976mmH2O=4.01463H2O
Overrage limit: To maximum static pressure limit
Output signal:
Model FHD: 4 to 20mA DC 2-wire, linear signal
Model FKD: 4 to 20mA DC (linear or square root) with
digital signal superimposed on the 4 to
20mA signal
(Available for amplifier unit from version 24 and FXW(HHC) version
5.3.)
EDSX6-101
Date
Mar. 15, 1999
FHD, FKD...3
Power supply:
Transmitter operates on 10.5V to 45V DC
at transmitter terminals.
10.5V to 32V DC for the units with optional
arrester.
Load limitations: see figure below
[Ω]
Load resistance
2000
1533Ω
1500
1000
600Ω
Operating
area
500
24V
250
10
16.1
10.5V
Power voltage
45V
50
[V]
Burnout direction: If self-diagnostic detect transmitter failure, the analog signal will be driven to either “Output Hold”, “Output Overscale”
or “Output Underscale” modes.
Model FHD: Unless otherwise specified in the order,
the transmitter will be shipped in “Output
Hold” mode.
(Output signal just before failure happens
is maintained.)
Model FKD: Selectable from HHC
“Output Hold”:
Output signal is hold as the value just before failure happens.
“Output Overscale”:
Approx. 21.6mA
(Adjustable within the range 20.8mA to
21.6mA from HHC)
“Output Underscale”:
Approx. 3.8mA
(Adjustable within the range 3.2mA to
3.8mA from HHC)
Note: For communication with HHC (Model: FXW), min. of 250Ω required.
3.2
Hazardous locations: (Approval pending)
Authorities
BASEEFA
Factory
Mutual
CSA
RIIS
SAA
Flameproof
Ex ds IIC T5, T6
Class I II III
Div. 1
Groups B thru. G
Class I II III
Div. 1
Groups C thru. G
Exds IIB + H2 T4
Ex d II C T5, T6
IP 66/67
Intrinsic safety
EEx ia IIC T4, T5
Class I II III
Div. 1
Groups A thru. F
Class I II III
Div. 1
Groups A thru. G
—
Ex ia II C T5, T6
IP 66/67
Type N
Nonincendive
Ex N II T5
Class I II III
Div. 2
Groups A thru. G
Class I II III
Div. 2
Groups A thru. G
—
Ex n II C T5, T6
IP 66/67
Zero/span adjustment:
Model FHD: Zero is adjustable from the external adjustment screw.
The adjustment screw can also function to
adjust span when MODE SWITCH (located on the electronics unit) is in the span
mode. INHIBIT mode to disable the adjustment screw is also available.
Model FKD: Zero and span are adjustable from the
HHC. Zero is also adjustable externally
from the adjustment screw.
Damping:
Adjustable electrical damping.
Model FHD: The time constant is adjustable to 0, 0.3,
1.2, 4.8, or 19.2 seconds.
Model FKD: The time constant is adjustable between 0
to 38.4 seconds.
Zero elevation/suppression:
–100% to +100% of URL
Normal/reverse action:
Model FHD: Selectable by moving a jumper pin located
on the electronics unit.
Model FKD: Selectable from HHC
Indication:
Analog indicator or 4 12 -digit LCD meter, as
specified.
3.8 4
20 20.8
Over scale
Burnout
0
Normal operating range
Under scale
Probable under range
Burnout
Loop-check output:
Model FHD: Transmitter can output constant signal of
4mA, 12mA, or 20mA if MODE SWITCH
is set to the loop check mode.
Model FKD: Transmitter can be configured to provide
constant signal 3.8mA through 21.6mA by
HHC.
Temperature limit:
Ambient: – 40 to + 85°C
(– 20 to + 80°C for LCD indicator)
(– 40 to + 60°C for arrester option)
(– 10 to + 60°C for fluorinated oil fill transmitter)
(– 10 to + 85°C for silicone oil "H", "S", "K")
(+ 20 to + 85°C for silicone oil "J", "T")
For explosionproof units (flameproof or intrinsic safety),
ambient temperature must be within the limits specified in each standard.
Process:
Fill fluid
Code in the
13th digit of
"Code symbols"
Fluorinated oil
W, A and D
– 20 to 120°C
Silicone oil
H
– 15 to 250°C
J
Process
temperature
Lower limit
of
static press.
Atmospheric
pressure
85 to 300°C
Y and G
– 40 to 120°C
S
– 15 to 250°C
T
85 to 300°C
K
– 15 to 200°C
Storage: – 40 to +90°C
2
21.6
Probable over range
2.7kPa abs
{20mmHg
abs}
0.13kPa abs
{1mmHg abs}
or more
Humidity limit:
0 to 100% RH
Communication: (Model FKD only)
With HHC (Model FXW, consult Data
Sheet No. EDS8-47), following information can be remotely displayed or reconfigured.
Items
Static pressure effect:
Zero shift; 0.2% of URL for flange rating
pressure
Double the zero shift for material code.
"H", "F", "G", "K", "L", "M", "T", "P" and "R"
+0
Span shift: – 0.2 –0.1 % of calibrated span
for flange rating pressure
Display
Set
Tag No.
v
v
Model No.
v
v
Serial No.
v
—
Engineering unit
v
v
Range limit
v
—
Measuring range
v
v
Damping
v
v
Output mode
v
v
Burnout direction
v
v
Adjustment
v
v
Output adjust
—
v
Data
v
—
Self diagnoses
v
—
Printer
—
—
Range code
2
1.7 s
Overrange effect: Zero shift; 0.3% of URL for flange rating
pressure
Double the effects for material code.
"H", "F", "G", "K", "L", "M", "T", "P" and "R"
Supply voltage effect:
Less than 0.05% fo calibrated span per
10V
RFI effect:
Less than 0.2% of URL for the frequencies of 20 to 1000MHz and field strength
30 V/m when electronics covers on.
(Classification: 2-abc: 0.2% span per
SAMA PMC 33.1)
Step response: (without electrical damping)
External switch lock
v
v
"3"
Transmitter display(*)
v
v
"5"
Linearise
(**)
v
v
Rerange
(**)
v
v
Notes: (*) HHC's version must be more than 5.0
(or FXW
1 – 2 ), to use this
function.
(**) HHC's version must be more than 5.3,
and Amplifier unit version 24.
Programmable output linearization function:
In smart version, output signal can be
characterized with “14 points linear approximation function” from HHC.
Accuracy rating: (including linearity, hysteresis, and repeatability)
(Standard)
For spans greater than 1/10 of URL: 0.2% of span
For spans below 1/10 of URL (Model FKD only):
0.1 × URL
± 0.1+0.1
% of span
Span
(Option)
For spans greater than 1/10 of URL: 0.1% of span
For spans below 1/10 of URL (Model FKD only):
0.1 × URL
± 0.05+0.05
% of span
Span
Linearity:
0.1% of calibrated span
Stability:
± 0.2% of upper range limit (URL) for 24
months
Temperature effect (*):
Effects per 28°C change between the limits of – 40°C and +85°C
(Standard)
Zero shift: ± 0.35% of URL
Total effect: ± 0.5% of URL
(Option)
Zero shift: ± 0.25% of URL
Total effect: ± 0.275% of URL
(
Dead time
s
Approx. 0.3 s
"6"
1.7 s
Note: * Capillary length: 1.5m
Dielectric strength:
500V AC, 50/60Hz 1 min., between circuit
and earth.
Insulation resistance:
More than 100MΩ/500V DC.
Turn-on time:
4 sec.
Internal resistance for external field indicator:
12Ω or less
Physical specifications
Performance specifications
(
Time constant (*)
)
)
Note: * Excluding effect by temperature difference
between the seals.
Electrical connections:
G1/2, 1/2-14 NPT, Pg13.5, or M20 × 1.5
conduit, as specified.
Process connections:
JIS, ANSI, or DIN raised face flanges.
JIS: 10K80A, 10K100A, 30K80A, or
30K100A
ANSI: 150LB 3", 150LB 4", 300LB 3", or
300LB 4"
DIN: PN40 DN80 or PN16 DN100
See OUTLINE DIAGRAM for detailed dimensions.
Diaphragm extension:
0, 50, 100, 150, or 200mm as specified.
(See model code. Extended diaphragm is
available only with 316L stainless steel or
Hastelloy-C diaphragm)
Process-wetted parts material:
Diaphragm: 316L stainless steel, HastelloyC,
Monel, Tantalum, Titanium or
Zirconium
Flange face:316 stainless steel, HastelloyC lining
Monel lining, or Tantalum lining
Extension: 3 1 6 s t a i n l e s s s t e e l o r
Hastelloy-C
3
FHD, FKD...3
Non-wetted parts material:
Electronics housing: Low copper die-cast
aluminum alloy (standard), finished
with polyester coating, or 316 stainless
steel (SCS14 per JIS G5121), as specified.
Capillary: In case of 13th code "Y, W, G, A,
D", PVC armored stainless steel.
In case of 13th code "H, J, S, T, K",
stainless steel armored stainless steel.
Mounting flange: 304 stainless steel or
carbon steel
Fill fluid: Silicone oil (standard) or fluorinated oil (Daifloil)
Mounting bracket: Carbon steel with epoxy coating or 304 stainless steel, as
specified
Environmental protection:
IEC IP67 and NEMA 4X
Mounting:
On 60.5mm (JIS 50A) pipe using mounting bracket, direct wall mounting
Mass {weight}:
Transmitter approximately 15kg without
options.
Add; 0.5kg for mounting bracket
0.8kg for indicator option
4.5kg for stainless steel housing
option
1.5kg per 50mm extension of diaphragm
Optional features
Indicator:
A plug-in analog indicator (1.5% accuracy)
can be housed in the electronics compartment or in the terminal box of the housing.
An optional 4 12 digits LCD meter is also
available.
Arrester:
A built-in arrester protects the electronics
from lightning surges.
Lightning surge immunity:
4KV (1.2 × 50µs)
Oxygen service: Special cleaning procedures are followed
throughout the process to maintain all process wetted parts oil-free.
The fill fluid is fluorinated oil.
Chlorine service: Oil-free procedures as above. Includes
fluorinated oil for fill.
Degreasing:
Process-wetted parts are cleaned, but the
fill fluid is standard silicone oil. Not for use
on oxygen or chlorine measurement.
Vacuum service: Special silicone oil and filling procedure
are applied.
See below figure.
Customer tag:
A stainless steel tag for customer tag data
is wired to the transmitter.
Coating of cell: Cell's surface is finished with epoxy/polyurethane double coating. Specify if environment is extremely corrosive.
ACCESSORIES
Hand-held communicator:
(Model FXW, refer to Data Sheet No. EDS
8-47)
Communication module: (standard for model FKD)
By adding communication module, remote setting function becomes available
for model FHD.
Remark: When the communication module is connected, the operation mode of external
zero/span adjustment screw is limited to
zero adjustment only.
[kPa abs]
{mbar abs}
Silicone (Code Y,G)
[kPa abs]
{mbar abs}
Silicone (Code S)
Silicone (Code T)
Sealing liquid:Silicone (Code K)
101
{1010}
53.3
{533}
Operating pressure
Operating pressure
101
{1010}
Operating arear
13.3
{133}
Non-operating
area
0.27
{2.7}
0.13
{1.3}
2.7
{27}
–40 –15
60 85 120
Process temperature[°C ]
200 230 250
Fig. 1 Relation between process temperature and
operating pressure
4
Operating arear
300
–15
100
Process temperature[°C ]
200
Fig. 2 Relation between process temperature and
operating pressure
CODE SYMBOLS
1 2 3 4 5 6 7 8
3
9 10 11 12 13
14 15
Description
0
Type
4 to 20mA, Output type
4 to 20mA with digital signal, Output type
FHD
FKD
Conduit connection
G 1/2
1/2-14NPT
Pg 13.5
M20 × 1.5
S
T
V
W
Flange
Mounting flange
0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F
G
H
J
K
304 stainless
steel
Carbon steel
None
(wafer type)
P
Q
Flange size and rating
JIS 10K 80A
JIS 10K 100A
JIS 30K 80A
JIS 30K 100A
ANSI/JPI 150LB 3"
ANSI/JPI 150LB 4"
ANSI/JPI 300LB 3"
ANSI/JPI 300LB 4"
DIN PN16/40 DN80
DIN PN16 DN100
JIS 10K 80A
JIS 10K 100A
JIS 30K 80A
JIS 30K 100A
ANSI/JPI 150LB 3"
ANSI/JPI 150LB 4"
ANSI/JPI 300LB 3"
ANSI/JPI 300LB 4"
DIN PN16/40 DN80
DIN PN16 DN100
3 inch wafer
4 inch wafer
Span limit (*1) [kPa]{m bar}
FHD/FKD
3.2/0.32…32/32
{32/3.2…320/320}
13/1.3…130/130
{130/13…1300/1300}
50/5…500/500
{500/50…5000/5000}
3
5
6
Material/diaphragm extension
Diaphragm
Flange face
W
A
B
C
D
316L stainless
steel
316 stainless
steel
H
F
G
K
L
Hastelloy-C
Hastelloy-C
M
T
P
R
Monel
Tantalum
Titanium
Zirconium
Monel
Tantalum
Titanium
Zirconium
Diaph. extension [mm]
0
50
100
150
200
0
50
100
150
200
0
0
0
0
}
(*2)
Notes: * (1) 100: 1 turn down is possible for model FKD, but should be used at a span greater than
1/40 of the maximum span for better performance.
(2) In case of 7th digit code "A", "B", "C", "D" and 13th digit code. "S", "T", "K", 5th digit
code "1", "3", "5", "7", "B", "D", "E", "F", "H", "Q" is available.
5
FHD, FKD...3
1 2 3 4 5 6 7 8
F HD
9 10 11 12 13
14 15
3
0
Description
0
Indicator and arrester
Indicator
None
Analog, 0 to 100% linear scale
Analog, 0 to 100% sq. root scale
Analog, custom scale
Analog, double scale
None
Analog, 0 to 100% linear scale
Analog, 0 to 100% sq. root scale
Analog, custom scale
Analog, double scale
Digital, 0 to 100%
Digital, custom scale
Digital, 0 to 100%
Digital, custom scale
A
B
C
D
J
E
F
G
H
K
L
P
Q
S
Arrester
None
None
None
None
None
Yes
Yes
Yes
Yes
Yes
None
None (Model FKD only)
Yes
Yes (Model FKD only)
Approvals for hazardous locations (Approval pending)
None (for ordinary locations)
JIS, Flameproof (Conduit seal)
(Available for 4th digit code "S")
JIS, Flameproof (Cable gland seal)
(Available for 4th digit code "S")
FM, Flameproof (or explosionproof)
(Available for 4th digit code "T")
CSA, Flameproof (or explosionproof) (Available for 4th digit code "T")
BASEEFA, Flameproof (Conduit seal)
BASEEFA, Flameproof (Cable gland seal) (Conduit connection G1/2 only)
FM, Intrinsic safety and nonincendive
CSA, Intrinsic safety and nonincendive
CENELEC, Intrinsic safety
CENELEC, Intrinsic safety and BASEEFA, Type N
SAA Flameproof (Conduit seal)(Available for 4th digit cord ("S,T,W")
SAA Intrinsic safety (Available for 4th digit cord ("S,T,W")
SAA Type–N (non-sparking)(Available for 4th digit cord ("S,T,W")
A
B
C
D
E
M
N
H
J
K
P
R
T
Q
Capillary and mounting bracket
Capillary
Mounting bracket
1.5 m
Carbon steel
3
Carbon steel
5
Carbon steel
6
Carbon steel
7
(*1)
Carbon steel
8
(*1)
Carbon steel
10 (*1)
Carbon steel
1.5
Stainless steel
3
Stainless steel
5
Stainless steel
6
Stainless steel
7
(*1)
Stainless steel
8
(*1)
Stainless steel
10 (*1)
Stainless steel
A
B
G
C
H
J
K
D
E
L
F
M
N
P
Stainless steel parts (*2)
Stainless steel tag plate
None
Yes
None
Yes
None
Yes
None
Yes
Y
B
C
E
M
N
P
Q
Stainless steel elec. housing
None
None
Yes
Yes
None
None
Yes
Yes
Special applications and fill fluid
Treatment
Fill fluid
None (standard)
Silicone oil
None (standard)
Fluorinated oil
Degreasing
Silicone oil
Oxygen service
Fluorinated oil (7th digit code "W", "A", "B", "C" and "D")
Chlorine service
Fluorinated oil (7th digit code "H", "F", "G", "K", "L"and"T")
High temp. 250°C
Silicone oil
High temp. 300°C
Silicone oil
High temp. and vacuum (250°C) Silicone oil
(*3)
High temp. and vacuum (300°C) Silicone oil 7th digit code "W", "A", "B", "C", and "D"
High temp. and high vacuum
Silicone oil
Y
W
G
A
D
H
J
S
T
K
}
Y
C
}
Teflon membrane
None
Yes (Available for 5th digit code "0", "2", "4", "6", "8", "A", "C", "E", "G", "J", "P" and 7th digit code
"W", "H", "M", "T", "P", "R".)
Notes: * (1) Available for 13th digit code "Y, W, G, A, D".
Inquire about in case of 13th other code.
(2) Not applicable to carbon steel flange material.
(3) Treatment; None
6
Coating of cell
None
None
None
None
Yes
Yes
Yes
Yes
OUTLINE DIAGRAM (Unit:mm)
Fastener
(In case of digit code
“B,C,M,N“)
Zero/span ADJ. screw
In case of 4th
digit code
“S,V,W,X“
U - Bolt M8
Mtg.pipe
JIS 50A (2")
70
50
Up
110
43.5
φ80
104
2–M8 12
Down
88
28 38
“A“
Conduit conn.
No flange is
attahed in
case of 5th
digit code
“P,Q“
233
Cable gland (option)
see note
4 Holes
9 dia.
Inner of gasket
“B“
Conduit
conn.
φH
φG
φF
φE
φD
P
N Holes
h dia
58
t
t
270
Earth terminal
(both side)
see table 1
51
71
(27)
L
Detail “A“
(Conduit conn.)
Detail “B“
φ13
φ15
17
M
261
K
φ80
With field ind .
(option)
J
G1/2
See table 1
S
CONNECTION
DIAGRAM
+ – CK+ CK–
S
S
T
H
D V
F
K
W
X
H.P.
L.P.
+–
54
70
120
φD
φE
φF
φG
φH
4TH DIGIT
CODE
t
P
N - φh
Flange
150
126
100
73
38
116
8–19
JIS–10K–80A
210 175
151
103
96
38
141
8–19
JIS–10K–100A
210 170
126
100
73
48
116
8–23
JIS–30K–80A
185
240 195
151
103
96
52
141
8–25
JIS–30K–100A
191
152.5
126
100
73
44
116
4–20
ANSI/JPI–150LB-3"
229
190.5
CONDUIT CONN.
J
K
3
EARTH TERMINAL
M
S
G1/2
17
8
T
1/2-14NPT
16
5
V
Pg13.5
W
M20 × 1.5
8
16
M4
No.8-32UNC
4.5
M4
5
M4
TABLE 1
7th digit
code
L
Mass.approx.
(kg)
11th digit S 11th digit S
(m)
code
(m) code
151
103
96
44
141
8–20
ANSI/JPI–150LB-4"
W,H,M,T,P,R
0
14 to 19.5
A,D
1.5
H,M
210 168
126
100
73
49
116
8–23
ANSI/JPI–300LB-3"
A,F
50
15 to 30.5
B,E
3
J,N
8
254 200
151
103
96
52
141
8–23
ANSI/JPI–300LB-4"
B,G
100
15.5 to 31
G,L
5
K,P
10
200 160
126
100
73
44
116
8–18
DIN
PN40
DN80
C,K
150
16 to 31.5
C,F
6
220 180
151
103
96
40
141
8–18
DIN
PN16
DN100
D,L
200
16.5 to 32
7
Note) : Cable gland is supplied in case of flameproof packing type.
φ11 cable is suitable.
7
FHD, FKD...3
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TN510412. The applicable standards used to demonstrate
compliance are :EMI (Emission)
EN50081-1 : 1992
Test item
Frequency range
Basic standard
Applicable Electromagnetic Radiation
Disturbance
30-1000MHz
EN55022
Class B
EMS (Immunity)
EN50082-1 : 1992
No.
Test item
Test
specification
Basic
standard
Performance
criteria
1
Electrostatic
discharge
8kV (Air)
IEC 801-2:1984
B
2
Radio-frequency 27-500MHz
electromagnetic 3V/m
IEC 801-3:1984
field.
(Unmodulated)
A
3
Fast transients
common mode
0.5kV, 5/50
(Tr/Th) ns
5kHz Rep.
IEC 801-4:1988
B
"LVD - The transmitter is not covered by the requirements
of the LVD standard."
Head office
11-2 Osaki 1-chome, Shinagawa-ku, Tokyo, 141-0032 Japan
Phone: 81-3-5435-7111
http://www.fujielectric.co.jp/eng/sg/KEISOKU/welcome.htm
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202
Fax: 81-42-585-6187, 6189
Information in this catalog is subject to change without notice.
Printed in Japan