XPPOWER PBM25US05

AC-DC Medical
25 Watts
THE XPERTS IN POWER
PBM25 Series
Small Size 2.50” x 4.25” x 0.96”
•
•
Low Leakage Current
•
Medical Safety Agency Approvals
•
Both Line and Neutral Fused
•
Meets Class B, EN55011 Conducted
•
Fixed Switching Frequency
Specification
Input
General
Input Voltage
Input Frequency
Input Current
•
•
•
Inrush Current
•
Earth
Leakage Current
•
85-264 VAC
47-63 Hz
0.7 A (rms) at 115 VAC,
0.4 A (rms) at 230 VAC
10 A max at 115 VAC,
25 A max at 230 VAC
90 µA max at 115 VAC 60 Hz
150 µA max at 230 VAC 50 Hz
Output
Efficiency
•
Isolation
•
Switching Frequency •
MTBF
•
Size
•
Weight
Output Voltage
Output Power
Minimum Load
•
•
•
Start Up Delay
Start Up Rise TIme
Hold up Time
Initial Set Accuracy
Line Regulation
Total Regulation
Transient Response
•
•
•
•
•
•
•
Ripple & Noise
Overvoltage
Protection
•
•
Overcurrent
Protection
Short Circuit
Protection
Temperature
Coefficient
•
5.1 to 28 VDC, See Table
25/28 Watts, See Table
No minimum for single outputs models
See Table for multi outputs models
3 s max
12 ms
10 ms minimum at 110 VAC
±1%
±0.5% max at full load
See Table
4% max deviation, 500 µs recovery
time for a 25% load change
±1% max pk-pk (See Note 1)
Provided on output 1 only; set at
112-132% of its nominal output
voltage
125-200% with auto recovery
•
Trip & restart with auto recovery
•
±0.04%/°C
•
68% minimum for 5 V & multi output
models, 70% minimum all others
4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
42 KHz ±5 KHz
500,000 hrs per min MIL-HDBK-217F
4.25 x 2.50 x 0.96”
107.95 x 63.50 x 24.38 mm
Single output models 163 g,
Multi output models 175 g
Environmental
Operating
Temperature
•
Storage Temperature
Cooling
Relative Humidity
Operating Altitude
Shock
Vibration
•
•
•
•
•
•
0 °C to +70 °C,
derate linearly from 100% load at
+50°C to 50% load at +70 °C
-40 °C to +85 °C
Convection cooled
5-95%, non-condensing
3000m
20 G in 3 directions, 3 times each
0.15mm over 10-55Hz and 1
octave/min. 3 axes for 30 min
EMC & Safety
EMC
•
ESD
Radiated Immunity
EFT/Burst
Surge
Safety Approvals
•
•
•
•
•
EN55011 (CISPR11) Class B & FCC
Class B conducted
EN61000-4-2 Level 2 Perf Criteria A
EN61000-4-3 3 V/m Perf Criteria A
EN61000-4-4 Level 3 Perf Criteria A
EN61000-4-5 Level 3 Perf Criteria A
UL2601-1, CSA C22.2 No 601.1,
EN60601-1, IEC 601-1
•
•
www.xp-medical.com
[email protected]
OUTPUT VOLTAGE & CURRENT RATINGS
Output Power
Maximum
28 W
28 W
28 W
28 W
28 W
25 W
25 W
25 W
25 W
25 W
Notes
•
Vnom
5.1 V
12.0 V
15.0 V
24.0 V
28.0 V
5.1 V
5.1 V
5.1 V
5.1 V
5.1 V
Output 1
Imin
Imax
0.0 A
5.5 A
0.0 A
2.3 A
0.0 A
1.9 A
0.0 A
1.2 A
0.0 A
1.0 A
0.4 A
2.5 A
0.4 A
2.5 A
0.4 A
2.5 A
0.4 A
2.5 A
0.4 A
2.5 A
Reg.(2)
2%
1%
1%
1%
1%
3%
3%
3%
3%
3%
PBM25
Vnom
Output 2
Imin
Imax
+12
+24
+12
+15
+24
0.2
0.1
0.2
0.1
0.1
V
V
V
V
V
A
A
A
A
A
1.5
0.8
1.5
1.0
0.8
A
A
A
A
A
Reg.(2)
Vnom
5%
5%
5%
5%
5%
-12 V
-15 V
-12 V
Output 3
Imin
Imax
0.05 A
0.05 A
0.05 A
0.2 A
0.2 A
0.2 A
Reg.(2)
4%
4%
4%
Model
Number
PBM25US05
PBM25US12*
PBM25US15*
PBM25US24*
PBM25US28*
PBM25UD01*
PBM25UD02*
PBM25UT01*
PBM25UT02*
PBM25UT03*
1. Ripple and noise measured with 20 MHz bandwidth with 10 µF electrolytic in parallel with 0.1 µF ceramic.
2. Regulation includes line and load regulation and set point accuracy. All multiple output models may be operated at no-load,
but output regulation will increase to 10% on outputs V2 & V3.
* Also available from Farnell InOne, see pages 350 & 351.
Mechanical Details - Single Output Models
PIN CONNECTIONS
Pin
Function
1
Return
2
Return
3
Output 1
4
Output 1
Notes:
1. Dimensions shown in inches (mm).
2. Tolerance 0.03 (0.76) maximum.
3. Input connector P1 mates with Molex housing
09-50-3031 and Molex 2878 series crimp terminal.
4. Output connector P2 mates with Molex housing
09-50-3041 and Molex 2878 series crimp terminal.
5. Weight 163 g.
6. Covers available, order part number PBM25 COVER*,
or add suffix ‘-C’ to have cover fitted (78.8 x 12.0 x 34.8 mm).
7. Mating connector kit available, order part number
PBM25 CONKT-S*.
Mechanical Details - Multi Output Models
PIN CONNECTIONS
Pin
Dual Output
Triple Output
1
Output 2
Output 2
2
Output 1
Output 1
3
Output 1
Output 1
4 Common Return Common Return
5 Common Return Common Return
6
No Connection
Output 3
Notes:
1. Dimensions shown in inches (mm).
2. Tolerance 0.03 (0.76) maximum.
3. Input connector P1 mates with Molex housing
09-50-3031 and Molex 2878 series crimp terminal.
4. Output connector P2 mates with Molex housing
09-50-3061 and Molex 2878 series crimp terminal.
5. Weight 175 g.
6. Covers available, order part number PBM25 COVER*,
or add suffix ‘-C’ to have cover fitted (78.8 x 120.0 x 34.8 mm).
7. Mating connector kit available, order part number
PBM25 CONKT-D/T*.
www.xp-medical.com
[email protected]
Rev 3.1