ETC PXI

E Series Multifunction DAQ –
333 kS/s, 16-Bit, 16 Analog Inputs
NI 6052E
NI 6052E
Driver Software
PCI-6052E
PXI-6052E
NI-DAQ
Windows 2000/NT/Me/9x
Mac OS
Analog Inputs
16 single-ended, 8 differential channels
333 kS/s sampling rate
333 kS/s stream-to-disk rate
16-bit resolution
Analog Output
Real-Time
See page 184.
Application Software
LabVIEW
Measurement Studio
ComponentWorks
VirtualBench
Measure
Lookout
2 channels,16-bit resolution
Digital I/O
8 (5 V/TTL) lines
Counter/Timers
Calibration Certificate Included
2 up/down, 24-bit resolution
See page 256.
Triggering
Analog and digital
Overview
Ordering Information
NI 6052E
PCI-6052E...........................................................777745-01
PXI-6052E ..............................................................777962-01
Includes NI-DAQ for Windows 2000/NT/Me/9x and Mac OS.
Extended warranty and
value added services ...........................................page 880
Recommended Configurations
DAQ Device
PCI-6052E
PXI-6052E
Accessory
SCB-68 (776844-01)
TBX-68 (777141-01)
Cable
SH6868-EP (184749-01)
SH6868-EP (184749-01)
See page 334 for accessory and cable information.
Analog
Bus
Inputs
PCI, PXI/CPCI 16 SE/8 DI
Resolution
16 bits
Sampling
Rate
333 kS/s
Input
Range
±0.05 to ±10 V
Analog
Outputs
2
Resolution
16 bits
Output
Rate
333 kS/s
Output
Range
±10 V
Digital
I/O
8
Counter/
Timers
2, 24-bit
Triggers
Analog and Digital
Table 1. NI 6052E Channel, Speed, and Resolution Specifications (see page 349 for detailed specifications)
National Instruments
Tel: (512) 794-0100 • Fax: (512) 683-9300 • [email protected] • ni.com
Measurements
Device
NI 6052E
EXPRESS
The NI 6052E DAQ devices use E Series
CODES
technology to deliver high-performance
and reliable data acquisition capabilities to
For information or to
meet a wide range of application
buy products online,
visit ni.com/catalog
requirements. You get up to 333 kS/s,
and enter:
16-bit performance on 16 single-ended
pci6052e
analog inputs. Depending on your type of
pxi6052e
hard drive, these devices can stream to disk
at rates up to 333 kS/s.
These NI 6052E DAQ devices feature
analog and digital triggering capability, as
well as two 24-bit, 20 MHz counter/timers, and eight digital I/O lines.
The NI 6052E DAQ devices also feature two 16-bit analog outputs.
See the E Series Multifunction DAQ Overview on page 306 for a
more detailed hardware overview.
319
NI 6052E
E Series Multifunction DAQ –
333 kS/s, 16-Bit, 16 Analog Inputs
Absolute Accuracy
Nominal Range (V)
Positive FS
% of Reading
Negative FS
24 Hrs
1 Year
Relative Accuracy
Noise + Quantization (µV)
Offset (µV)
Single Pt.
Averaged
Temp
Absolute Accuracy
Drift (%/°C)
at Full Scale (mV)
Resolution (µV)
Single Pt.
Averaged
10
-10
0.0354
0.0371
947
981
87.0
0.0006
4.747
1145
5
-5
0.0054
0.0071
476
491
43.5
0.0001
0.876
573
115
2.5
-2.5
0.0354
0.0371
241
245
21.7
0.0006
1.190
286
28.6
1
-1
0.0354
0.0371
99.2
98.1
8.7
0.0006
0.479
115
11.5
0.5
-0.5
0.0354
0.0371
52.1
56.2
5.0
0.0006
0.243
66.3
6.6
0.25
-0.25
0.0404
0.0421
28.6
32.8
3.0
0.0006
0.137
39.2
3.9
0.1
-0.1
0.0454
0.0471
14.4
22.4
2.1
0.0006
0.064
27.7
2.8
0.05
-0.05
0.0454
0.0471
9.7
19.9
1.9
0.0006
0.035
25.3
0.0001
1.232
57.3
2.5
10
0
0.0054
0.0071
476
491
43.5
573
57.3
5
0
0.0354
0.0371
241
245
21.7
0.0006
2
0
0.0354
0.0371
99.2
98.1
8.7
0.0006
2.119
286
28.6
0.850
115
1
0
0.0354
0.0371
52.1
56.2
5.0
0.0006
0.428
66.3
11.5
6.6
0.5
0
0.0404
0.0421
28.6
39.8
3.0
0.0006
0.242
48.2
3.9
0.2
0
0.0454
0.0471
14.4
22.4
2.1
0.0006
0.111
27.7
2.8
0.1
0
0.0454
0.0471
9.7
19.9
1.9
0.0006
0.059
25.3
2.5
Note: Accuracies are valid for measurements following an internal E Series Calibration. Averaged numbers assume dithering and averaging of 100 single-channel readings.
Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of external or factory-calibration temperature. One-year calibration
interval recommended. The Absolute Accuracy at Full Scale calculations were performed for a maximum range input voltage (for example, 10 V for the ±10 V range) after one year, assuming
100 pt averaging of data. See overview on page 312 for an example calculation of this type.
Table 2. NI 6052E Analog Input Accuracy Specifications
Nominal Range (V)
Absolute Accuracy
% of Reading
Positive FS Negative FS
Absolute
Temp
Accuracy at
Full Scale (mV)
24 Hrs
90 Days
1 Year
Offset (µV)
Drift (%/°C)
10
-10
0.0044
0.0052
0.0061
798
0.0001
1.405
10
0
0.0044
0.0052
0.0061
569
0.0001
1.176
Note: Temp Drift applies only if ambient is greater than ±10 °C of previous external calibration. See page 312 for example calculations.
See page 349 for more detailed
specifications.
Measurements
Table 3. NI 6052E Analog Output Accuracy Specifications
Figure 1. NI 6052E Hardware Block Diagram
320
National Instruments
Tel: (512) 794-0100 • Fax: (512) 683-9300 • [email protected] • ni.com
See page 310 in the E Series
Multifunction DAQ Overview
for I/O connector diagrams.