RPS-160 160W Single Output Medical Type series Features : Universal AC input / Full range Built-in active PFC function Protections: Short circuit / Overload / Over voltage / Over temperature 5"x3" compact size Free air convection for 110W and 160W with 20.5 CFM forced air With power good and fail signal output Built-in remote sense function for 5~15V No load power consumption under 0.75W by PS-ON control (Optional) Standby 5V@1A with Fan/0.6A without Fan (Optional) 3 years warranty RPS G 160 12 G: With 5Vsb Blank: Without 5Vsb EN60601-1 IEC60601-1 SPECIFICATION MODEL RPS -160-5 RPS -160-12 RPS -160-15 RPS -160-24 5V 12V 15V 24V 48V RATED CURRENT (20.5CFM) 30A 12.9A 10.3A 6.5A 3.25A CURRENT RANGE (convection) 0 ~ 20A 0 ~ 9.1A 0 ~ 7.3A 0 ~ 4.6A 0 ~ 2.3A CURRENT RANGE (20.5CFM) 0 ~ 30A 0 ~ 12.9A 0 ~ 10.3A 0 ~ 6.5A 0 ~ 3.25A RATED POWER (convection) Note.7 103W 112.2W 112.5W 113.4W 113.4W RATED POWER (20.5CFM) Note.8 155W 159.8W 159.5W 161W 161W RIPPLE & NOISE (max.) Note.2 100mVp-p 100mVp-p 120mVp-p 150mVp-p 250mVp-p VOLTAGE ADJ. RANGE CH1:10.8 ~ 13.2V 3.0% CH1:13.5 ~ 16.5V 3.0% CH1:22 ~ 27V 2.0% CH1:43.2 ~ 52.8V 2.0% DC VOLTAGE OUTPUT VOLTAGE TOLERANCE Note.3 LINE REGULATION 0.5% 0.5% 0.5% 0.5% 0.5% LOAD REGULATION 1.0% 1.0% 1.0% 1.0% 1.0% 1200ms, 30ms/230VAC SETUP, RISE TIME VOLTAGE RANGE Note.6 90 ~ 264VAC 47 ~ 63Hz POWER FACTOR (Typ.) PF>0.93/230VAC EFFICIENCY (Typ.) 85% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 2A/115VAC 1.1A/230VAC COLD START 35A/115VAC 70A/230VAC 13.8 ~ 16.2V 5.75 ~ 6.75V 17.25 ~ 20.25V 27.6 ~ 32.4V 88% 55.2 ~ 64.8V Protection type : Shut down o/p voltage, re-power on to recover 110 (5V),105 (12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor) 95 (5V),90 (12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor) Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down Protection type : (TSW2)Shut down o/p voltage, re-power on to recover PS-ON INPUT SIGNAL (OPTIONAL) Power on: PS-ON = "Hi" or ">2V" ; Power off: PS-ON = "Low" or "<0.5V" PF>1ms POWER GOOD / POWER FAIL 500ms>PG>10ms 5 ~ 15V REMOTE SENSE SAFETY STANDARDS WITHSTAND VOLTAGE -20 ~ +70 (Refer to output load derating curve) 20 ~ 90% RH non-condensing -40 ~ +85 , 10 ~ 95% RH 0.03%/ (0 ~ 50 ) 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes UL60601-1, TUV EN60601-1 approved I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B HARMONIC CURRENT Compliance to EN61000-3-2,-3 EMS IMMUNITY MTBF NOTE 87% OVERLOAD ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT VIBRATION OTHERS 87% Earth leakage current <300uA / 264VAC, Patient leakage current <100uA/264VAC WORKING TEMP. WORKING HUMIDITY (Note 4) 87% 105 ~ 135% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed OVER TEMPERATURE EMC PF>0.98/115VAC at full load LEAKAGE CURRENT PROTECTION SAFETY & 127 ~ 370VDC FREQUENCY RANGE OVER VOLTAGE FUNCTION 2500ms, 30ms/115VAC at full load 16ms/230VAC/115VAC at full load HOLD UP TIME (Typ.) INPUT CH1:4.5 ~ 5.5V 4.0% RPS -160-48 DIMENSION PACKING Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A 230.5Khrs min. MIL-HDBK-217F (25 127*76.2*34.6mm (L*W*H) 0.32Kg; 36pcs/12.5Kg/0.79CUFT ) 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. 5. HS1,HS2 & HS3 can not be shorted. 6. Derating may be needed under low input voltages. Please check the derating curve for more details. 7. The rated power includes 5Vsb @ 0.6A. 8. The rated power includes 5Vsb @ 1A. RPS-160 160W Single Output Medical Type Mechanical Specification Unit:mm FAN AC Input Connector (CN1) : JST B3P-VH or equivalent 5cm Pin No. 1 2 3 5 20.5CFM min. SVR1 CN3 1 2 3 4 66.2 76.2 HS1 1 2 3 Pin No. 1,2,3,4 5,6,7,8 1 2 3 4 5 6 7 8 HS3 FS2 Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent 117 Assignment +V -V Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Pin No. Status Mating Housing Terminal 1 2 3 4 PG GND -S +S JST XHP or equivalent JST SXH-001T-P0.6 or equivalent CN901 LED 4- 5 Mating Housing Power Good Connector(CN3):JST B4B-XH or equivalent CN2 CN901 Assignment AC/N No Pin AC/L DC Output Connector (CN2) : JST B8P-VH or equivalent HS2 FS1 CN1 series 1234 3.5 5VSB Connector(CN901) : JST B-XH or equivalent 127 Pin No. 1 2,4 3 PS/ON 34.6 5mA(max.) GND Assignment Mating Housing Terminal JST XHP or equivalent JST SXH-001T or equivalent PS/ON GND 5VSB 3 max. HS1,HS2,HS3 can not be shorted Block Diagram fosc :100KHz(5V) 70KHz(12~48V) +5VSB DETERCTION CIRCUIT RECTIFIERS & FILTER POWER SWITCHING PWM CONTROL RECTIFIERS & PFC EMI FILTER I/P +5VSB +S RECTIFIERS & FILTER POWER SWITCHING +V -V -S PG PG DETECTION CIRCUIT FG CONTROL O.L.P. O.V.P. PS/ON CONTROL Derating Curve PS/ON Output Derating VS Input Voltage Ta=25 100 100 With FAN 90 80 80 70 Without FAN LOAD (%) LOAD (%) 60 40 20 -20 0 10 20 30 40 50 AMBIENT TEMPERATURE ( ) 60 70 (HORIZONTAL) 60 50 40 90 95 100 115 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz File Name:RPS-160-SPEC 2008-09-23