MEANWELL RPSG-160-5

RPS-160
160W Single Output Medical Type
series
Features :
Universal AC input / Full range
Built-in active PFC function
Protections: Short circuit / Overload / Over voltage / Over temperature
5"x3" compact size
Free air convection for 110W and 160W with 20.5 CFM forced air
With power good and fail signal output
Built-in remote sense function for 5~15V
No load power consumption under 0.75W by PS-ON control (Optional)
Standby 5V@1A with Fan/0.6A without Fan (Optional)
3 years warranty
RPS G
160 12
G: With 5Vsb
Blank: Without 5Vsb
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
RPS -160-5
RPS -160-12
RPS -160-15
RPS -160-24
5V
12V
15V
24V
48V
RATED CURRENT (20.5CFM) 30A
12.9A
10.3A
6.5A
3.25A
CURRENT RANGE (convection) 0 ~ 20A
0 ~ 9.1A
0 ~ 7.3A
0 ~ 4.6A
0 ~ 2.3A
CURRENT RANGE (20.5CFM) 0 ~ 30A
0 ~ 12.9A
0 ~ 10.3A
0 ~ 6.5A
0 ~ 3.25A
RATED POWER (convection) Note.7 103W
112.2W
112.5W
113.4W
113.4W
RATED POWER (20.5CFM) Note.8 155W
159.8W
159.5W
161W
161W
RIPPLE & NOISE (max.) Note.2 100mVp-p
100mVp-p
120mVp-p
150mVp-p
250mVp-p
VOLTAGE ADJ. RANGE
CH1:10.8 ~ 13.2V
3.0%
CH1:13.5 ~ 16.5V
3.0%
CH1:22 ~ 27V
2.0%
CH1:43.2 ~ 52.8V
2.0%
DC VOLTAGE
OUTPUT
VOLTAGE TOLERANCE Note.3
LINE REGULATION
0.5%
0.5%
0.5%
0.5%
0.5%
LOAD REGULATION
1.0%
1.0%
1.0%
1.0%
1.0%
1200ms, 30ms/230VAC
SETUP, RISE TIME
VOLTAGE RANGE
Note.6 90 ~ 264VAC
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.93/230VAC
EFFICIENCY (Typ.)
85%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
2A/115VAC
1.1A/230VAC
COLD START 35A/115VAC
70A/230VAC
13.8 ~ 16.2V
5.75 ~ 6.75V
17.25 ~ 20.25V
27.6 ~ 32.4V
88%
55.2 ~ 64.8V
Protection type : Shut down o/p voltage, re-power on to recover
110 (5V),105 (12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor)
95 (5V),90
(12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor)
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover
PS-ON INPUT SIGNAL (OPTIONAL) Power on: PS-ON = "Hi" or ">2V" ; Power off: PS-ON = "Low" or "<0.5V"
PF>1ms
POWER GOOD / POWER FAIL 500ms>PG>10ms
5 ~ 15V
REMOTE SENSE
SAFETY STANDARDS
WITHSTAND VOLTAGE
-20 ~ +70
(Refer to output load derating curve)
20 ~ 90% RH non-condensing
-40 ~ +85
, 10 ~ 95% RH
0.03%/ (0 ~ 50 )
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
UL60601-1, TUV EN60601-1 approved
I/P-O/P:4KVAC
I/P-FG:1.5KVAC
O/P-FG:0.5KVAC
ISOLATION RESISTANCE
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
HARMONIC CURRENT
Compliance to EN61000-3-2,-3
EMS IMMUNITY
MTBF
NOTE
87%
OVERLOAD
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
OTHERS
87%
Earth leakage current <300uA / 264VAC, Patient leakage current <100uA/264VAC
WORKING TEMP.
WORKING HUMIDITY
(Note 4)
87%
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER TEMPERATURE
EMC
PF>0.98/115VAC at full load
LEAKAGE CURRENT
PROTECTION
SAFETY &
127 ~ 370VDC
FREQUENCY RANGE
OVER VOLTAGE
FUNCTION
2500ms, 30ms/115VAC at full load
16ms/230VAC/115VAC at full load
HOLD UP TIME (Typ.)
INPUT
CH1:4.5 ~ 5.5V
4.0%
RPS -160-48
DIMENSION
PACKING
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
230.5Khrs min. MIL-HDBK-217F (25
127*76.2*34.6mm (L*W*H)
0.32Kg; 36pcs/12.5Kg/0.79CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. HS1,HS2 & HS3 can not be shorted.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. The rated power includes 5Vsb @ 0.6A.
8. The rated power includes 5Vsb @ 1A.
RPS-160
160W Single Output Medical Type
Mechanical Specification
Unit:mm
FAN
AC Input Connector (CN1) : JST B3P-VH or equivalent
5cm
Pin No.
1
2
3
5
20.5CFM min.
SVR1
CN3
1
2
3
4
66.2
76.2
HS1
1
2
3
Pin No.
1,2,3,4
5,6,7,8
1
2
3
4
5
6
7
8
HS3
FS2
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
117
Assignment
+V
-V
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Pin No.
Status
Mating Housing
Terminal
1
2
3
4
PG
GND
-S
+S
JST XHP
or equivalent
JST SXH-001T-P0.6
or equivalent
CN901
LED
4-
5
Mating Housing
Power Good Connector(CN3):JST B4B-XH or equivalent
CN2
CN901
Assignment
AC/N
No Pin
AC/L
DC Output Connector (CN2) : JST B8P-VH or equivalent
HS2
FS1
CN1
series
1234
3.5
5VSB Connector(CN901) : JST B-XH or equivalent
127
Pin No.
1
2,4
3
PS/ON
34.6
5mA(max.)
GND
Assignment
Mating Housing
Terminal
JST XHP
or equivalent
JST SXH-001T
or equivalent
PS/ON
GND
5VSB
3 max.
HS1,HS2,HS3 can not be shorted
Block Diagram
fosc :100KHz(5V)
70KHz(12~48V)
+5VSB
DETERCTION
CIRCUIT
RECTIFIERS
&
FILTER
POWER SWITCHING
PWM CONTROL
RECTIFIERS
&
PFC
EMI
FILTER
I/P
+5VSB
+S
RECTIFIERS
&
FILTER
POWER
SWITCHING
+V
-V
-S
PG
PG
DETECTION
CIRCUIT
FG
CONTROL
O.L.P.
O.V.P.
PS/ON
CONTROL
Derating Curve
PS/ON
Output Derating VS Input Voltage
Ta=25
100
100
With FAN
90
80
80
70
Without FAN
LOAD (%)
LOAD (%)
60
40
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60
70
(HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-160-SPEC 2008-09-23