TRUMPOWER TM300-Q42

TRUMPower
300 WATT MEDICAL POWER SUPPLIES
D E SCR IPTION
The TM300 series comprising single and multiple
output models for 200 to 300 watts of continuous
output power is specially designed for medical
applications. They operate at 90 to 264VAC input
voltage without the need of a selector strap. All
auxiliary outputs are with magnetic amplifier to
keep regulation. The units are constructed on a
printed circuit board with a U-bracket for
mechanical
support
and
heat
sinking.
A
cover-and-fan assembly can be added during
manufacturing.
TM300 SERIES
( LV D )
RoHS
F EATU R E S
‧
‧
‧
‧
‧
‧
‧
‧
‧
EN61000-3-2 class A and D compliant
Pow er Factor 0.98 typical
O v e r v o l ta g e p r o t e c t i o n
Short-circuit protection
Pow er Fail Detect (PFD)
100% burn-in at full rated load
Remote sense on output #1 and output #2
Remote inhibit – TTL high disables output
C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
INPUT SPECIFICATIONS
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
E N V IR O N M ENTA L SP E C IF ICATI O N S
Input voltage :
90 to 264VAC
Input frequency:
47 to 63Hz
Input current :
4.7A (rms) for 115VAC
2.3A (rms) for 230VAC
Earth leakage current:
Saf et y Sta n da rd A ppr o va ls:
100µA max. @115VAC, 60Hz
220µA max. @ 230VAC, 50Hz
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
Cooling:
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and Noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
See rating chart
See rating chart
2% peak to peak maximum
Provided on output #1 only; set
at 115–140% of its nominal
output voltage
All outputs protected
to short circuit conditions
All outputs ±0.04%/ºC
maximum
Transient response:
Maximum excursion of 4%
or better on all models,
recovering to 1% of final value
within 500µs after a 25% step
load change
PFD signal:
TTL logic high for normal
operation and TTL logic low
upon loss of input power.
This signal appears at least
1ms prior to master output
dropping 5% below its nominal
value. This signal also
provides a minimum delay of
100ms after master output is
within regulation.
Requires an external TTL
high level signal to inhibit
outputs for standard models.
Remote inhibit:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
200/250/300 watts continuous
output power at 35 CFM forced
air cooling or 100/125/150 watts
at convention cooling
GENERAL SPECIFICATIONS
KHz ±10KHz
0.98 typical
70% minimum on all models
12 msec minimum at 110VAC
±0.2% maximum at full load
30 amps @ 115VAC or 60 amps
@230VAC, at 25ºC cold start
4000VAC from input to output
Withstand voltage:
1500VAC from input to ground
500VAC from output to ground
300,000 hours minimum at full load
MTBF:
at 25ºC ambient, calculated per
MIL-HDBK- 217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B
radiated
EN61000-3-2:
Harmonic distortion, Class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
Switching frequency:
Power factor:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
MEDICAL GRADE
TM300 SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Product No.
TM300-S3.3
TM300-S5.1
TM300-S12
TM300-S15
TM300-S24
TM300-S30
TM300-S48
TM300-Q403
TM300-Q40
TM300-Q41
TM300-Q42
TM300-Q45
TM300-Q46
TM300-Q47
TM300-Q48
TM300-Q49
TM300-Q410
TM300-Q411
TM300-Q412
TM300-Q413
TM300-Q414
Output # 1
Vnom. Imin. Imax.
3.3V
5.1V
12V
15V
24V
30V
48V
3.3V
5.1V
5.1V
5.1V
5.1V
5.1V
5.1V
5.1V
5.1V
24V
24V
24V
24V
24V
3.0A
60A
60A
3.0A
1.2A
25A
1.0A
20A
0.6A 12.5A
0.5A
10A
0.5A
6.3A
3.0A
35A
2.0A
35A
2.0A
35A
2.0A
35A
2.0A
35A
2.0A
35A
2.0A
35A
2.0A
35A
2.0A
35A
0.5A
6.3A
0.5A
6.3A
0.5A
6.3A
0.5A
6.3A
0.5A
6.3A
Tol.
Output # 2
Output # 3
Output # 4
Vnom. Imin. Imax. Tol. Vnom. Imin. Imax. Tol. Vnom. Imin. Imax.
3%
2%
2%
2%
2%
2%
2%
3%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
5.1V
12V
15V
12V
12V
12V
24V
24V
12V
12V
12V
12V
24V
24V
2.0A
1.0A
0.8A
1.0A
1.0A
1.0A
0.5A
0.5A
1.0A
1.0A
1.0A
1.0A
0.5A
0.5A
22A
10A
8A
10A
10A
10A
5A
5A
10A
10A
10A
10A
5A
5A
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
12V
12V
15V
12V
12V
12V
12V
5.1V
5.1V
5.1V
5.1V
12V
5.1V
12V
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
12V
5.1V
24V
12V
24V
15V
12V
15V
24V
12V
24V
12V
15V
12V
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
4A
4A
2.5A
4A
2.5A
4A
4A
4A
2.5A
4A
2.5A
4A
4A
4A
Tol.
Maximum
Output
Power (5)
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
200W
300W
300W
300W
300W
300W
300W
250W
300W
300W
300W
300W
300W
300W
300W
300W
300W
300W
300W
300W
300W
NOTES:
1.
Add suffix "C" for enclosed format with option fan control, e.g. TM300-Q45C.
2.
All outputs are floating. It can be connected externally for positive or negative
output.
3.
Output #1 & #2 can be adjusted within ±5% of their nominal voltage.
4.
Output #3 & #4 can be adjusted within ±15% of their nominal voltage.
5.
300 watts for "C" version with cover-and-fan assembly. 150 watts for U-bracket
version without moving air (maximum current of output #1 & #2 derated to 50%), or 300
watts with 35 CFM forced air provided by user.
6.
TM300-S5.1 is rated 200 watts with 35 CFM forced air cooling or 100 watts
convection cooled. TM300-Q403 is rated 250 watts with 35 CFM forced air
cooling (maximum current of output #1 & #2 derated to 50%) or 125 watts
convection coole).
7.
Single output models may be operated at no-load. At no-load, output voltage
tolerance increases to 10%.
MECHANICAL SPECIFICATIONS
Single Output Models
NOTES:
1.
Dimensions shown in inch [mm]
2.
Tolerance 0.02 [0.5] maximum
3.
Input connector P1 is Dinkle DT-35-B01W-03 screws are
M3, Nickel plated.
4.
Connector P4 mates with Molex housing 50-37-5103 and
pins 5263.
5.
Connectors P2, P3, P5 and P6: M3*0.5 screw connections
6.
Weight: 1.20 kgs. (2.64 lbs.) approx.
7.
DC fan on P4 is rated at 12V/0.1A.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
MEDICAL GRADE
TM300 SERIES
MECHANICAL SPECIFICATIONS
Multiple Output Models
NOTES:
1.
Dimensions shown in inch [mm]
2.
Tolerance 0.02 [0.5] maximum
3.
Input connector P1 is Dinkle DT-35-B01W-03 screws are
M3, Nickel plated.
4.
Connector P4 mates with Molex housing 50-37-5103 and
pins 5263.
5.
Connectors P2, P3, P5 and P6: M3*0.5 screw connections
6.
Output connector P7 mates with Molex housing 09-50-3041
and Molex 2878 series crimp terminal.
7.
Weight: 1.24 kgs. (2.73 lbs.) approx.
8.
DC fan on P4 is rated at 12V/0.1A.
PIN CHART
Single Output Models
Product No.
CONN
PIN
TM300-S5.1 TM300-S3.3
TM300-S12 TM300-S15
TM300-S24 TM300-S30
TM300-S48
`
P1 (AC)
P4
P2
1
2
3
Product No.
TM300-S5.1
TM300-S12
TM300-S24
TM300-S48
TM300-S3.3
TM300-S15
TM300-S30
1
2
3
4
+S (VO1)
-S (VO1)
PFD
SIGNAL GROUND
LIVE
+VO1
NEUTRAL GROUND
+VO1
(-VO1)
CONN
PIN
P3
P4
P5
5
6
7
8
9
10
INHIBIT
N.C.
N.C.
N.C.
0V (FAN)
FAN
P7
P6
-VO1
-VO1
1
2
3
4
VOID
VOID
VOID
VOID
2
3
4
+S (VO1)
-S (VO1)
PFD
Multiple Output Models
Product No.
CONN
PIN
TM300-Q403
TM300-Q41
TM300-Q45
TM300-Q47
TM300-Q49
TM300-Q411
TM300-Q413
TM300-Q40
TM300-Q42
TM300-Q46
TM300-Q48
TM300-Q410
TM300-Q412
TM300-Q414
P1 (AC)
1
P2
2
P4
P3
3
1
SIGNAL GROUND
LIVE
+VO1
NEUTRAL GROUND
-VO1
(-VO1)
CONN
P4
Product No.
PIN
TM300-Q403
TM300-Q41
TM300-Q45
TM300-Q47
TM300-Q49
TM300-Q411
TM300-Q413
TM300-Q40
TM300-Q42
TM300-Q46
TM300-Q48 INHIBIT
TM300-Q410
TM300-Q412
TM300-Q414
5
P5
6
7
8
9
10
N.C.
+S (VO2)
-S (VO2)
0V (FAN)
FAN
P7
P6
1
+VO2
-VO2
2
+VO3 -VO3
3
4
+VO4
-VO4
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com