TRUMPOWER TMP110-Q46

TRUMPower
110 WATT MEDICAL POWER ADAPTERS
TMP110 SERIES
D E SCR IPTION
(MDD)
The TMP110 series of AC/DC switching power
adapters are for 110 watts of continuous
output power. They are enclosed in a 94V-1
rated polyphenylene-oxide case with an IEC
320 inlet to mate with interchangeable cord for
world-wide use. An on/off switch can be added
at request. All models meet CISPR 11 and FCC
class B emission limits, and are designed for
medical applications.
F EATU R E S
‧
‧
‧
‧
‧
‧
‧
‧
D u a l , t r i p l e o r q u a d r u p l e o u t p u ts
Optional output connectors
Optional on/off sw itch
Optional Power Fail Detect (PFD) signal
100% burn-in
W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
Input surge current protection
O v e r v o l ta g e p r o t e c t i o n
‧ Overcurrent protection
Saf et y Sta n da rd A ppr o va ls:
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
IN PUT S PE C IF ICAT ION S
E N V IR O N M ENTA L SP E C IF ICATI O N S
I n p u t v o l ta g e :
8 5 t o 2 6 4 VA C
Input frequency: 47 to 63 Hz
Input current:
3 . 2 0 A ( r m s ) f o r 11 5 VA C
1 . 8 0 A ( r m s ) f o r 2 3 0 VA C
L e a k a g e c u r r e n t : 9 0 µ A m a x . @ 11 5 VA C , 6 0 H z
1 5 0 µ A m a x . @ 2 3 0 VA C , 5 0 H z
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
OUTPUT SPECIFICATIONS
GENERAL SPECIFICATIONS
Output voltage/current:
See rating chart
Switching frequency: 20KHz to 250KHz, varies with load and line
Total output power:
110 watts maximum
Efficiency:
Ripple and noise:
1% peak to peak maximum
Overvoltage protection: Provided on output #1 only;
set at 112-132% of its nominal
output voltage
Overcurrent protection: All outputs protected to short
circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC
maximum
Transient response:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500us after a
25% step load change
PFD signal (optional):
TTL logic high for normal
operation and TTL logic low
upon loss of input power. This
signal appears at least 1ms
prior to +5V output dropping
5% below its nominal value.
This signal also provides a
minimum delay of 100ms after
+5V is within regulation.
Hold-up time:
Line regulation:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
70% minimum on single output models with
Vo ≥12V, 65% minimum on the others
12 msec minimum at 110VAC
±0.5% maximum at full load
Inrush current:
15 amps @ 115VAC, or 30 amps
@230VAC, at 25ºC cold start
Withstand voltage:
4000VAC from input to output 1500VAC
from input to ground
500VAC from output to ground
MTBF:
550,000 hours minimum at full load at 25ºC
ambient, calculated per MIL-HDBK-217F,
excluding DC fan.
EMC Performance (EN60601-1-2: 2001)
E N 5 5 0 11 :
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 :
Voltage dips, 30% reduction for 500ms, 60%
reduction for 100ms and >95% reduction for
10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
MEDICAL GRADE
TMP110 SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Product No.
Output #1
Output #2 (1)
Output #3
Vnom. Imin. Imax. Tol.
Vnom. Imin. Imax. Ipeak Tol
Vnom. Imin. Imax. Tol
Output
Output #4
Vnom. Imin. Imax. Tol. Power
TMP110-D23
+5V
0.8A 10A
5%
+12V
0A
5A
9.0A
5%
TMP110-T31
+5V
0.8A 10A
5%
+12V
0A
5A
9.0A
5%
-12V
0A
2A
5%
TMP110-T32
+5V
0.8A 10A
5%
+15V
0A
4A
7.5A
5%
-15V
0A
2A
5%
TMP110-Q40
+5V
0.8A 10A
5%
+12V
0A
5A
9.0A
5%
-12V
0A
2A
5%
-5V
0A
2A
5%
110W
TMP110-Q41
+5V
0.8A 10A
5%
+15V
0A
4A
7.5A
5%
-15V
0A
2A
5%
+24V
0A
2A
5%
110W
TMP110-Q42
+5V
0.8A 10A
5%
+12V
0A
5A
9.0A
5%
-12V
0A
2A
5%
+12V
0A
2A
5%
110W
TMP110-Q45
+5V
0.8A 10A
5%
+12V
0A
5A
9.0A
5%
-12V
0A
2A
5%
+24V
0A
2A
5%
110W
TMP110-Q45-1
+5V
2.0A 10A
5%
+12V
0A
5A
9.0A
5%
-12V
0A
2A
5%
+24V
1.5A
3A
10% 110W
TMP110-Q45-2
+5V
0.8A 10A
5%
+24V
0A
3A
5.0A
5%
-12V
0A
2A
5%
+12V
0A
2A
5%
110W
TMP110-Q46
+5V
0.8A 10A
5%
+15V
0A
4A
7.5A
5%
-15V
0A
2A
5%
-5V
0A
2A
5%
110W
Notes:
(N/A)
(N/A)
110W
(N/A)
110W
(N/A)
110W
(1) Peak output current with 10% maximum duty cycle for less than 60 seconds. Total peak power must not exceed 130 watts.
(2) The output #1 of model TMP110-Q45-1 needs a minimum current of 2A to support the other outputs at their maximum rated loads.
MECHANICAL SPECIFICATIONS
NOTES:
1.
Dimensions shown in inch [mm]
2.
Tolerance 0.02 [0.5] maximum
3.
Weight: 1.40 Kgs (3.08 Lbs.) approx.
4.
Standard output connector is an 8-pin
at 270º male DIN plug. For other
output connector options, please
consult factory.
5.
To order a model with on/off switch,
add suffix “S” to the product number.
e.g. TMP110-T31-S.
6.
To order a model with PFD signal,
add suffix “PF” to the product number,
e.g. TMP110-T31-PF.
PIN CHART
PIN
Product No.
TMP110-D23
TMP110-T31
TMP110-Q40
TMP110-Q42
TMP110-Q45-1
TMP110-Q46
TMP110-T32
TMP110-Q41
TMP110-Q45
TMP110-Q45-2
1
2
3
4
5
6
7
8
COMMON
RETURN
N.C.
OUTPUT
#1.
N.C.
OUTPUT
#2.
COMMON
RETURN
OUTPUT
#1
N.C. or
PFD
COMMON
RETURN
OUTPUT
#3
OUTPUT
#1.
N.C
OUTPUT
#2.
COMMON
RETURN
OUTPUT
#1
N.C. or
PFD
COMMON
RETURN
OUTPUT
#3
OUTPUT
#1.
OUTPUT
#4
OUTPUT
#2..
COMMON
RETURN
OUTPUT
#1
N.C. or
PFD.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com