RPT-7503 - Meanwell

RPT-75 series
75W Triple Output Medical Type
MODEL:RPT-7503
OUTPUT FUNCTION TEST
NO
1
2
3
4
5
TEST ITEM
SPECICATION
TEST CONDITION
RIPPLE & NOISE
V1: 80
mVp-p (Max )
I/P: 230VAC
RESULT
V1:
10
mVp-p (Max )
VERDICT
V2: 120
mVp-p (Max )
29
mVp-p (Max )
mVp-p (Max )
O/P:FULL LOAD
Ta:25℃
V2:
V3: 120
V3:
35
mVp-p (Max )
OUTPUT VOLTAGE
V1: 4 %~ -4 % (Max)
I/P: 100 VAC / 264 VAC
V1:
2.5
%~
%
TOLERANCE
V2: 6 %~ -6 % (Max)
V2:
3
%~
%
V3: 10 %~ -6 % (Max)
O/P:FULL/ MIN
Ta:25℃
V3:
5.5
%~
%
V1: 1 %~ -1 % (Max)
I/P: 100 VAC ~ 264 VAC
V1:
0.36
%~
%
V2: 1 %~ -1 % (Max)
V2:
0.36
%~
%
V3: 1.5 %~ -1.5 % (Max)
O/P:FULL LOAD
Ta:25℃
V3:
1
%~
%
V1: 3 %~ -4 % (Max)
I/P: 230 VAC
V1:
1.7
%~
%
V2: 5 %~ -4 % (Max)
V2:
2.7
%~
%
V3: 6 %~ -6 % (Max)
O/P:FULL ~MIN LOAD
Ta:25℃
V3:
2
%~
%
V1: 3 %~ -4 % (Max)
I/P: 230 VAC
V1:
1.6
%~
%
V2: 5 %~ -4 % (Max)
O/P: Testing O/P 60%LOAD
V2:
2.9
%~
%
V3:
4.3
%~
%
LINE REGULATION
LOAD REGULATION
CROSS REGULATION
V3: 6 %~ -6 % (Max)
LOAD
Other O/P 40%LOAD
P
P
P
P
P
Change
Ta:25℃
6
SET UP TIME
230VAC:
500
ms (Max)
I/P: 230 VAC
230VAC/
146
ms
115 VAC:
500
ms (Max)
I/P: 115 VAC
115VAC/
130
ms
O/P:FULL LOAD
Ta:25℃
7
RISE TIME
230VAC:
30
ms (Max)
I/P: 230 VAC
230VAC/
4
ms
115VAC:
30
ms (Max)
I/P: 115 VAC
115VAC/
4
ms
O/P:FULL LOAD
Ta:25℃
8
HOLD UP TIME
230VAC:
90
ms (TYP)
I/P: 230 VAC
230VAC/
115
ms
115VAC:
20
ms (TYP)
I/P: 115 VAC
115VAC/
23.4
ms
<5
%
O/P:FULL LOAD
Ta:25℃
9
OVER/UNDERSHOOT TEST
I/P: 230 VAC
< +5%
TEST:
O/P:FULL LOAD
Ta:25℃
10
DYNAMIC LOAD
V1:
660
mVp-p
I/P: 230 VAC
P
P
P
P
246
mVp-p
O/P:FULL /Min LOAD
P
90%DUTY/1KHZ
Ta:25℃
Model: RPT-7503
Test Report
1 / 5
RPT-75 series
75W Triple Output Medical Type
INPUT FUNCTION TEST
NO
1
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
INPUT VOLTAGE RANGE
90VAC~264 VAC
I/P:TESTING
52V~264V
VERDICT
O/P:FULL LOAD
Ta:25℃
I/P:
TEST: OK
LOW-LINE-3V= 87V
P
HIGH-LINE+15%=300 V
O/P:FULL/MIN LOAD
ON: 30 Sec . OFF: 30 Sec 10MIN
( AC POWER ON/OFF NO DAMAGE )
2
3
INPUT FREQUENCY RANGE
47HZ ~63 HZ
I/P: 90VAC ~ 264 VAC
NO DAMAGE OSC
O/P:FULL~MIN LOAD
Ta:25℃
74% (TYP)
EFFICIENCY
I/P: 230 VAC
TEST: OK
P
73.6%
O/P:FULL LOAD
Ta:25℃
4
INPUT CURRENT
P
230V/
1
A (TYP)
I/P: 230 VAC
I=
0.75
A/ 230 VAC
115V/
1.5
A (TYP)
I/P: 115 VAC
I=
1.25
A/ 115 VAC
P
O/P:FULL LOAD
Ta:25℃
5
INRUSH CURRENT
230V/
50
A (TYP)
I/P: 230 VAC
I=
37
A/ 230 VAC
115V/
25
A(TYP)
I/P: 115 VAC
I=
18
A/ 115 VAC
COLD START
6
LEAKAGE CURRENT
<
<
150
100
P
O/P:FULL LOAD
Ta:25℃
uA/ 264 VAC
I/P: 264 VAC
L-FG:
119
uA
for
N-FG:
119
uA
leakage current
O/P:Min LOAD
Ta:25℃
uA/264 VAC
I/P: 264 VAC
L-V+ :
84
uA
for
O/P:Min LOAD
Ta:25℃
L-V-:
84
uA
N-V+:
84
uA
N-V-:
84
uA
earth
touch
leakage current
P
P
PROTECTION FUNCTION TEST
NO
1
2
3
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
OVER LOAD PROTECTION
140 %~ 180 %
I/P: 230 VAC
158 %/ 230 VAC
I/P: 115 VAC
155 %/ 115 VAC
O/P:TESTING
Ta:25℃
Hiccup Mode
OVER VOLTAGE PROTECTION
SHORT PROTECTION
CH1: 3.8V~ 4.45V
I/P: 230 VAC
4.3 V/ 230 VAC
I/P: 115 VAC
4.3 V/ 115 VAC
O/P:MIN LOAD
Ta:25℃
Shunt down Re- power ON
SHORT EVERY OUTPUT
I/P: 264 VAC
NO DAMAGE
1 HOUR NO DAMAGE
O/P:FULL LOAD
Ta:25℃
Hiccup Mode
Model: RPT-7503
VERDICT
P
P
P
Test Report
2 / 5
RPT-75 series
75W Triple Output Medical Type
ENVIRONMENT TEST
NO
1
TEST ITEM
SPECICATION
TEMPERATURE RISE TEST
MODEL: RPT-75A WITH FAN
TEST CONDITION
RESULT
VERDICT
1. ROOM AMBIENT BURN-IN:1HRS
I/P: 230VAC O/P: FULL LOAD Ta=31.3℃
2. HIGH AMBIENT BURN-IN:1.5HRS
I/P: 230VAC O/P: FULL LOAD Ta=52.5℃
Position
P/N
1
2
BD1
LF2
US4K80 4A/800V SHI
TR-719
52.8℃
70.7℃
3
C6
100U/400V NCC 105℃
38.6℃
59.9℃
4
C36
47U/50V RUB 105℃ YXF
42.8℃
64.0℃
5
T1COIL
TF-1524 JSI
66.6℃
89.9℃
6
D1
HER308 3A/1KV REC
55.7℃
78.6℃
7
ZD1
P6KE250A LT
39.4℃
63.9℃
8
Q1
STP9NK70ZFP 7.5A/700V
55.2℃
78.8℃
83.0℃
9
D100
STPS2045CT 20A/45V ST
58.5℃
10
D200
SF10SC6 10A/60V SHI
66.6℃
90.4℃
11
RG1
AN7905T
69.5℃
93.0℃
12
D300
SR360 3A/60V
61.9℃
82.7℃
13
C105
22000U/16V RUB 105℃
41.1℃
64.4℃
14
C205
1000U/16V RUB 105℃ ZLH
52.6℃
75.6℃
15
C305
470U/25V NCC 105℃ KY
56.6℃
78.4℃
16
R3
68KΩ/2W R/MO
57.4℃
78.4℃
NCP1230D165R2 ON
39.2℃
60.4℃
17
2
OVER LOAD BURN-IN TEST
ROOM AMBIENT HIGH AMBIENT
Ta=31.3 ℃
Ta=52.5 ℃
50.6℃
70.3℃
NO
U1
NO DAMAGE
I/P: 230 VAC
O/P:V1=10A V2=5A V3=1A
1 HOUR ( MIN )
P
TEST:OK
P
Ta:25℃
3
LOW TEMPERATURE
TURN ON AFTER
2 HOUR
TURN ON TEST
I/P: 230 VAC
O/P: 100﹪LOAD
TEST:OK
P
Ta= -20℃
4
AFTER 12 HOURS
I/P: 272 VAC
HIGH TEMPERATURE
IN CHAMBER ON
CONTROL 50 ℃
O/P:FULL LOAD
Ta= 50 ℃
NO DAMAGE
+ 0.03 %(0~50℃)
I/P: 230 VAC
HIGH VOLTAGE
TURN ON TEST
5
TEMPERATURE
VIBRATION TEST
HUMIDITY=
P
95 %R.H
+ 0.019 %(0~50℃)
P
O/P:FULL LOAD
COEFFICIENT
6
TEST:OK
HIGH HUMIDITY
1 Carton & 1 Set
TEST:OK
(1) Waveform: Sine Wave
(2) Frequency:10~500Hz
(3) Sweep Time:10min/sweep cycle
P
(4) Acceleration:2G
(5) Test Time:1 hour in each axis (X.Y.Z)
(6) Ta:25℃
Model: RPT-7503
Test Report
3 / 5
RPT-75 series
75W Triple Output Medical Type
SAFETY TEST
NO
1
2
3
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
WITHSTAND VOLTAGE
I/P-O/P: 4 KVAC/min
I/P-O/P: 4.2 KVAC/min
I/P-O/P:
2.14
mA
I/P-FG: 1.5 KVAC/min
I/P-FG: 1.8 KVAC/min
I/P-FG:
1.36
mA
O/P-FG: 1.5 KVAC/min
O/P-FG: 1.8 KVAC/min
Ta:25℃
O/P-FG:
1.57
mA
I/P-O/P:500VDC>100MΩ
I/P-O/P: 500
VDC
I/P-O/P:
9
GΩ
I/P-FG: 500VDC>100MΩ
I/P-FG:
500
VDC
O/P-FG:500VDC>100MΩ
O/P-FG: 500
Ta:25℃
VDC
I/P-FG:
9
GΩ
O/P-FG:
18
GΩ
ISOLATION RESISTANCE
APPROVAL
VERDICT
P
NO DAMAGE
P
NO DAMAGE
TUV: Certificate NO : TA 50096088
P
UL: File NO :E227340
E.M.C TEST
NO
1
2
3
4
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
HARMONIC
EN61000-3-2
I/P: 230 VAC/50HZ
PASS
CLASS A
O/P:FULL LOAD
Ta:25℃
EN55022
I/P: 230 VAC (50HZ)
PASS
CLASS B
O/P:FULL/50% LOAD
Ta:25℃
Test by certified Lab
CONDUCTION
RADIATION
E.S.D
5
E.F.T
EN55022
I/P: 230 VAC (50HZ)
PASS
O/P:FULL LOAD
Ta:25℃
Test by certified Lab
EN61000-4-2
MEDICAL
I/P: 230 VAC/50HZ
CRITERIA A
EN61000-4-4
MEDICAL
SURGE
P
P
O/P:FULL LOAD
Ta:25℃
I/P: 230 VAC/50HZ
P
CRITERIA A
O/P:FULL LOAD
Ta:25℃
INPUT: 2KV
6
P
CLASS B
AIR:8KV / Contact:6KV
VERDICT
IEC61000-4-5
MEDICAL
I/P: 230 VAC/50HZ
P
CRITERIA A
O/P:FULL LOAD
Ta:25℃
L-N :1KV
P
L,N-PE:2KV
7
Test by certified Lab & Test Report Prepare
M.T.B.F & LIFE CYCLE CALCULATION
NO
1
TEST ITEM
SPECICATION
CAPACITOR
RPT-75A WITH FAN : SUPPOSE C105 IS THE MOST CRITICAL COMPONENT
I/P: 230VAC O/P:FULL LOAD Ta= 25 ℃ LIFE TIME= 1419902 HRS
LIFE CYCLE
I/P: 230VAC
2
MTBF
TEST CONDITION
O/P:FULL LOAD Ta= 50 ℃
RESULT
VERDICT
P
LIFE TIME= 217049 HRS
MIL-HDBK-217F NOTICES2 PARTS COUNT
P
TOTAL FAILURE RATE: 521.2K HRS
Model: RPT-7503
Test Report
4 / 5
RPT-75 series
75W Triple Output Medical Type
COMPONENT STRESS TEST
NO
1
TEST ITEM
SPECICATION
TEST CONDITION
RESULT
Power Transistor
Q1 Rated
I/P:High-Line +3V = 267 V
(1)
646
V
VERDICT
( D to S) or (C to E) Peak Voltage
STP9NK70ZFP : 700V 7.5A
O/P: (1)Full Load Turn on
(2)
648
V
P
(2) Output Short
Ta:25℃
2
Diode Peak Voltage
D100 Rated
I/P:High-Line +3V = 267 V
(1)
26
V
STPS3045CT : 45V 30A
O/P: (1)Full Load Turn on
(2)
25
V
D200 Rated
(1)
37
V
STP2045CT : 45V 20 A
(2)
37
V
D300 Rated
(1)
108
V
HER306 : 600V 3A
(2)
111
V
(1)
596
V
(2)Output Short
Ta:25℃
3
Clamp Diode Peak Voltage
D1 Rated
I/P:High-Line +3V = 267 V
HER308 : 1000V 3A
O/P: (1) Dynamic Load
P
90%Duty/1KHz
P
Ta:25℃
4
Input Capacitor Voltage
C5 Rated
: 100u / 400V/ 105℃
I/P:High-Line +3V = 267 V
(1)
368
V
O/P: (1)Full Load Turn on /Off
(2)
372
V
(2) Min load Turn on /Off
(3)
372
V
P
(3)Full Load /Min load Change
Ta:25℃
5
Control IC Voltage Test
U1 Rated
I/P:High-Line +3V = 267 V
(1)
12.66
V
1230D165R2G : 18V
O/P: (1)Full Load Turn on /Off
(2)
10.75
V
(2) Min load Turn on /Off
(3)
12.66
V
P
(3)Full Load /Min load Change
Ta:25℃
DATE
2006/9/5
2006/11/14
2007/5/9
SAMPLE
TEST RESULT
TESTER
APPROVAL
RD SAMPLE
PASS
VINCENT TSENG
MAX LIN
PASS
VINCENT TSENG
MAX LIN
PASS
VINCENT TSENG
MAX LIN
PRODUCT SAMPLE
W0610A20
PRODUCT SAMPLE
W073A31
2003/12/12 A50-F023
Model: RPT-7503
Test Report
5 / 5