RPD-75 series 75W Dual Output Medical Type MODEL:RPD-75A OUTPUT FUNCTION TEST NO 1 2 TEST ITEM SPECIFICATION TEST CONDITION RIPPLE & NOISE V1: 80 mVp-p (Max ) I/P: 230VAC V1: 15 mVp-p (Max ) V2: 120 mVp-p (Max ) O/P:FULL LOAD Ta:25℃ V2: 18 mVp-p (Max ) OUTPUT VOLTAGE CH1: 4.75 V~-5.5 V ADJUST RANGE RESULT VERDICT I/P: 230 VAC 4.57V~ 5.83 V / 230VAC I/P: 115 VAC 4.57V~ 5.83 V / 115VAC P O/P:MIN LOAD Ta:25℃ 3 4 5 6 OUTPUT VOLTAGE V1: 2 %~ -2 % (Max) I/P: 100 VAC / 264 VAC V1: 0.1 %~ -0.1 % TOLERANCE V2: 6 %~ -6 % (Max) O/P:FULL/ MIN Ta:25℃ V2: 2.5 %~ -2.5 % LINE REGULATION LOAD REGULATION CROSS REGULATION LOAD P P V1: 0.5 %~ -0.5 % (Max) I/P: 100 VAC ~ 264 VAC V1: 0 %~ 0 % V2: 1 %~ -1 % (Max) O/P:FULL LOAD Ta:25℃ V2: 0.24 %~ -0.24 % V1: 1.5 %~ -1.5 % (Max) I/P: 230 VAC V1: 0 %~ 0 % V2: 3 %~ -3 % (Max)% O/P:FULL ~MIN LOAD Ta:25℃ V2: 0.8 %~ -0.8 % V1: 1.5 %~ -1.5 % (Max) I/P: 230 VAC V1: 0 %~ 0 % V2: 3 %~ -3 % (Max) O/P: Testing O/P 60%LOAD V2: 2.1 %~ -2.1 % P P Other O/P 40%LOAD P Change Ta:25℃ 7 SET UP TIME 230VAC: 500 ms (Max) I/P: 230 VAC 230VAC/ 152 ms 115 VAC: 500 ms (Max) I/P: 115 VAC 115VAC/ 135 ms O/P:FULL LOAD Ta:25℃ 8 RISE TIME 230VAC: 30 ms (Max) I/P: 230 VAC 230VAC/ 4 ms 115VAC: 30 ms (Max) I/P: 115 VAC 115VAC/ 4 ms O/P:FULL LOAD Ta:25℃ 9 HOLD UP TIME 230VAC: 90 ms (TYP) I/P: 230 VAC 230VAC/ 100.6 ms 115VAC: 20 ms (TYP) I/P: 115 VAC 115VAC/ 21.2 ms <5 % O/P:FULL LOAD Ta:25℃ 10 OVER/UNDERSHOOT TEST I/P: 230 VAC < +5% TEST: O/P:FULL LOAD Ta:25℃ 11 DYNAMIC LOAD V1: 1000 mVp-p I/P: 230 VAC P P P P 205 mVp-p O/P:FULL /Min LOAD P 90%DUTY/1KHZ Ta:25℃ Model: RPD-75A Test Report 1 / 5 RPD-75 series 75W Dual Output Medical Type INPUT FUNCTION TEST NO 1 TEST ITEM SPECIFICATION TEST CONDITION RESULT INPUT VOLTAGE RANGE 90VAC~264 VAC I/P:TESTING 52V~264V VERDICT O/P:FULL LOAD Ta:25℃ I/P: TEST: OK LOW-LINE-3V= 87V P HIGH-LINE+15%=300 V O/P:FULL/MIN LOAD ON: 30 Sec . OFF: 30 Sec 10MIN ( AC POWER ON/OFF NO DAMAGE ) 2 3 INPUT FREQUENCY RANGE 47HZ ~63 HZ I/P: 90VAC ~ 264 VAC NO DAMAGE OSC O/P:FULL~MIN LOAD Ta:25℃ 77 % (TYP) EFFICIENCY TEST: OK P I/P: 230 VAC 77.08 % O/P:FULL LOAD Ta:25℃ 4 INPUT CURRENT P 230V/ 1 A (TYP) I/P: 230 VAC I= 0.83 A/ 230 VAC 115V/ 1.5 A (TYP) I/P: 115 VAC I= 1.42 A/ 115 VAC P O/P:FULL LOAD Ta:25℃ 5 INRUSH CURRENT 230V/ 50 A (TYP) I/P: 230 VAC I= 37 A/ 230 VAC 115V/ 25 A(TYP) I/P: 115 VAC I= 18 A/ 115 VAC COLD START 6 LEAKAGE CURRENT < < 150 100 P O/P:FULL LOAD Ta:25℃ uA/ 264 VAC I/P: 264 VAC L-FG: 119 uA for N-FG: 119 uA leakage current O/P:Min LOAD Ta:25℃ uA/264 VAC I/P: 264 VAC L-V+ : 84 uA for O/P:Min LOAD Ta:25℃ L-V-: 84 uA N-V+: 84 uA N-V-: 84 uA earth touch leakage current P P PROTECTION FUNCTION TEST NO 1 2 3 TEST ITEM SPECIFICATION TEST CONDITION RESULT OVER LOAD PROTECTION 140%~ 180% I/P: 230 VAC 165%/ 230 VAC I/P: 115 VAC 164%/ 115 VAC O/P:TESTING Ta:25℃ Hiccup Mode OVER VOLTAGE PROTECTION SHORT PROTECTION CH1: 5.75V~ 6.75V I/P: 230 VAC 6.3V/ 230 VAC I/P: 115 VAC 6.3V/ 115 VAC O/P:MIN LOAD Ta:25℃ Shunt down Re- power ON SHORT EVERY OUTPUT I/P: 264 VAC NO DAMAGE 1 HOUR NO DAMAGE O/P:FULL LOAD Ta:25℃ Hiccup Mode Model: RPD-75A VERDICT P P P Test Report 2 / 5 RPD-75 series 75W Dual Output Medical Type ENVIRONMENT TEST NO 1 TEST ITEM SPECIFICATION TEMPERATURE RISE TEST MODEL:RPD-75A WITH FAN TEST CONDITION RESULT VERDICT 1. ROOM AMBIENT BURN-IN:1.5 HRS I/P: 230VAC O/P: FULL LOAD Ta=29.6 ℃ 2. HIGH AMBIENT BURN-IN:1 HRS I/P: 230VAC O/P: FULL LOAD Ta=54.6 ℃ 2 3 OVER LOAD BURN-IN TEST LOW TEMPERATURE ROOM AMBIENT HIGH AMBIENT Ta= 29.6 ℃ Ta= 54.6 ℃ 47.1℃ 70.1℃ NO Position P/N 1 2 BD1 C6 US4K80 4A/800V SHI 100U/400V NCC 105℃ 37.9℃ 60.7℃ 3 LF2 TR-719 47.3℃ 69.9℃ 4 C36 47U/63V RUB 105℃ YXF 41.3℃ 64.2℃ 5 Q1 STP9NK70ZFP 7.5A/700V 47.0℃ 71.1℃ 6 ZD1 P6KE250A LT 51.2℃ 72.8℃ 7 D1 HER308 3A/1KV REC 54.3℃ 77.1℃ 8 T1 COIL TF-1522 61.0℃ 83.3℃ STPS2045CT 20A/45V ST 58.4℃ 79.4℃ 9 D100 10 D200 SF10SC6 10A/60V SHI 64.8℃ 83.3℃ 11 C105 22000U/16V RUB 105℃ 43.8℃ 65.7℃ 12 C205 1000U/16V RUB 105℃ ZLH 55.3℃ 77.6℃ 13 U1 NCP1230D165R2 ON 38.1℃ 61.0℃ 14 R205 150Ω/2W R/MO 62.9℃ 85.8℃ TEST:OK NO DAMAGE I/P: 230 VAC 1 HOUR ( MIN ) O/P:V1=11A V2=5A Ta:25℃ TURN ON AFTER 2 HOUR TURN ON TEST P I/P: 230 VAC O/P: 100﹪LOAD P TEST:OK P Ta= -20℃ 4 AFTER 12 HOURS I/P: 272 VAC HIGH TEMPERATURE IN CHAMBER ON CONTROL 50 ℃ O/P:FULL LOAD Ta=50 ℃ NO DAMAGE + 0.03 %(0~50℃) I/P: 230 VAC HIGH VOLTAGE TURN ON TEST 5 TEMPERATURE VIBRATION TEST HUMIDITY= P 95 %R.H + 0.003 %(0~50℃) P O/P:FULL LOAD COEFFICIENT 6 TEST:OK HIGH HUMIDITY 1 Carton & 1 Set TEST:OK (1) Waveform: Sine Wave (2) Frequency:10~500Hz (3) Sweep Time:10min/sweep cycle P (4) Acceleration:2G (5) Test Time:1 hour in each axis (X.Y.Z) (6) Ta:25℃ Model: RPD-75A Test Report 3 / 5 RPD-75 series 75W Dual Output Medical Type SAFETY TEST NO 1 2 3 TEST ITEM SPECIFICATION TEST CONDITION RESULT WITHSTAND VOLTAGE I/P-O/P: 4 KVAC/min I/P-O/P: 4.2 KVAC/min I/P-O/P: 2.01 mA I/P-FG: 1.5 KVAC/min I/P-FG: 1.8 KVAC/min I/P-FG: 1.29 mA O/P-FG: 1.5 KVAC/min O/P-FG: 1.8 KVAC/min Ta:25℃ O/P-FG: 1.34 mA I/P-O/P:500VDC>100MΩ I/P-O/P: 500 VDC I/P-O/P: 6 GΩ I/P-FG: 500VDC>100MΩ I/P-FG: 500 VDC O/P-FG:500VDC>100MΩ O/P-FG: 500 Ta:25℃ VDC I/P-FG: 13 GΩ O/P-FG: 14 GΩ ISOLATION RESISTANCE APPROVAL VERDICT P NO DAMAGE P NO DAMAGE TUV: Certificate NO : TA 50096088 P UL: File NO :E227340 E.M.C TEST NO 1 2 3 4 TEST ITEM SPECIFICATION TEST CONDITION RESULT HARMONIC EN61000-3-2 I/P: 230 VAC/50HZ PASS CLASS A O/P:FULL LOAD Ta:25℃ EN55022 I/P: 230 VAC (50HZ) PASS CLASS B O/P:FULL/50% LOAD Ta:25℃ Test by certified Lab CONDUCTION RADIATION E.S.D 5 E.F.T EN55022 I/P: 230 VAC (50HZ) PASS O/P:FULL LOAD Ta:25℃ Test by certified Lab EN61000-4-2 MEDICAL I/P: 230 VAC/50HZ CRITERIA A EN61000-4-4 MEDICAL SURGE P P O/P:FULL LOAD Ta:25℃ I/P: 230 VAC/50HZ P CRITERIA A O/P:FULL LOAD Ta:25℃ INPUT: 2KV 6 P CLASS B AIR:8KV / Contact:6KV VERDICT IEC61000-4-5 MEDICAL I/P: 230 VAC/50HZ P CRITERIA A O/P:FULL LOAD Ta:25℃ L-N :1KV P L,N-PE:2KV 7 Test by certified Lab & Test Report Prepare M.T.B.F & LIFE CYCLE CALCULATION NO 1 TEST ITEM SPECIFICATION CAPACITOR SUPPOSE C 105 IS THE MOST CRITICAL COMPONENT I/P: 230VAC O/P:FULL LOAD Ta= 25 ℃ LIFE TIME=997838HRS LIFE CYCLE I/P: 230VAC 2 3 TEST CONDITION O/P:FULL LOAD Ta= 50 ℃ RESULT VERDICT P LIFE TIME=265547HRS MTBF MIL-HDBK-217F NOTICES2 PARTS COUNT DMTBF/Accelerated Life Test TOTAL FAILURE RATE: 569.9K HRS Demonstration Mean Time Between Failure : Above 30,000 hours @ TA 40℃ P P Model: RPD-75A Test Report 4 / 5 RPD-75 series 75W Dual Output Medical Type COMPONENT STRESS TEST NO 1 TEST ITEM SPECIFICATION TEST CONDITION RESULT Power Transistor Q1 Rated I/P:High-Line +3V = 267 V (1) 692 V VERDICT ( D to S) or (C to E) Peak Voltage STP9NK70ZFP : 700V 7.5A O/P: (1)Full Load Turn on (2) 715 V P (2) Output Short Ta:25℃ 2 Diode Peak Voltage D100 Rated I/P:High-Line +3V = 267 V (1) 38.8 V STPS2045CT : 45V 20A O/P: (1)Full Load Turn on (2) 32.6 V (1) 50 V (2) 42 V (1) 630 V (2)Output Short Ta:25℃ P D200 Rated SF10SC6 : 60V 3 Clamp Diode Peak Voltage 10 A D1 Rated I/P:High-Line +3V = 267 V HER308 : 1000V 3A O/P: (1) Dynamic Load 90%Duty/1KHz P Ta:25℃ 4 Input Capacitor Voltage C5 Rated : 100u / 400V/ 105℃ I/P:High-Line +3V = 267 V (1) 384 V O/P: (1)Full Load Turn on /Off (2) 388 V (2) Min load Turn on /Off (3) 388 V P (3)Full Load /Min load Change Ta:25℃ 5 Control IC Voltage Test U1 Rated I/P:High-Line +3V = 267 V (1) 12.84 V 1230D165R2G : 18V O/P: (1)Full Load Turn on /Off (2) 11.69 V (2) Min load Turn on /Off (3) 12.84 V P (3)Full Load /Min load Change Ta:25℃ DATE 2006/9/5 2006/11/14 2007/5/7 SAMPLE TEST RESULT TESTER APPROVAL RD SAMPLE PASS VINCENT TSENG MAX LIN PASS VINCENT TSENG MAX LIN PASS VINCENT TSENG MAX LIN PRODUCT SAMPLE W0610A20 PRODUCT SAMPLE W0703A31 2003/12/12 A50-F023 Model: RPD-75A Test Report 5 / 5