RPD-75A - Meanwell

RPD-75 series
75W Dual Output Medical Type
MODEL:RPD-75A
OUTPUT FUNCTION TEST
NO
1
2
TEST ITEM
SPECIFICATION
TEST CONDITION
RIPPLE & NOISE
V1: 80
mVp-p (Max )
I/P: 230VAC
V1:
15
mVp-p (Max )
V2: 120
mVp-p (Max )
O/P:FULL LOAD
Ta:25℃
V2:
18
mVp-p (Max )
OUTPUT VOLTAGE
CH1: 4.75 V~-5.5 V
ADJUST RANGE
RESULT
VERDICT
I/P: 230
VAC
4.57V~
5.83
V / 230VAC
I/P: 115
VAC
4.57V~
5.83
V / 115VAC
P
O/P:MIN LOAD
Ta:25℃
3
4
5
6
OUTPUT VOLTAGE
V1: 2 %~ -2 % (Max)
I/P: 100 VAC / 264 VAC
V1:
0.1
%~
-0.1
%
TOLERANCE
V2: 6 %~ -6 % (Max)
O/P:FULL/ MIN
Ta:25℃
V2:
2.5
%~
-2.5
%
LINE REGULATION
LOAD REGULATION
CROSS REGULATION
LOAD
P
P
V1: 0.5 %~ -0.5 % (Max)
I/P: 100 VAC ~ 264 VAC
V1:
0
%~
0
%
V2: 1 %~ -1 % (Max)
O/P:FULL LOAD
Ta:25℃
V2:
0.24
%~
-0.24
%
V1: 1.5 %~ -1.5 % (Max)
I/P: 230 VAC
V1:
0
%~
0
%
V2: 3 %~ -3 % (Max)%
O/P:FULL ~MIN LOAD
Ta:25℃
V2:
0.8
%~
-0.8
%
V1: 1.5 %~ -1.5 % (Max)
I/P: 230 VAC
V1:
0
%~
0
%
V2: 3 %~ -3 % (Max)
O/P: Testing O/P 60%LOAD
V2:
2.1
%~
-2.1
%
P
P
Other O/P 40%LOAD
P
Change
Ta:25℃
7
SET UP TIME
230VAC:
500
ms (Max)
I/P: 230 VAC
230VAC/
152
ms
115 VAC:
500
ms (Max)
I/P: 115 VAC
115VAC/
135
ms
O/P:FULL LOAD
Ta:25℃
8
RISE TIME
230VAC:
30
ms (Max)
I/P: 230 VAC
230VAC/
4
ms
115VAC:
30
ms (Max)
I/P: 115 VAC
115VAC/
4
ms
O/P:FULL LOAD
Ta:25℃
9
HOLD UP TIME
230VAC:
90
ms (TYP)
I/P: 230 VAC
230VAC/
100.6
ms
115VAC:
20
ms (TYP)
I/P: 115 VAC
115VAC/
21.2
ms
<5
%
O/P:FULL LOAD
Ta:25℃
10
OVER/UNDERSHOOT TEST
I/P: 230 VAC
< +5%
TEST:
O/P:FULL LOAD
Ta:25℃
11
DYNAMIC LOAD
V1:
1000
mVp-p
I/P: 230 VAC
P
P
P
P
205
mVp-p
O/P:FULL /Min LOAD
P
90%DUTY/1KHZ
Ta:25℃
Model: RPD-75A
Test Report
1 / 5
RPD-75 series
75W Dual Output Medical Type
INPUT FUNCTION TEST
NO
1
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
INPUT VOLTAGE RANGE
90VAC~264 VAC
I/P:TESTING
52V~264V
VERDICT
O/P:FULL LOAD
Ta:25℃
I/P:
TEST: OK
LOW-LINE-3V= 87V
P
HIGH-LINE+15%=300 V
O/P:FULL/MIN LOAD
ON: 30 Sec . OFF: 30 Sec 10MIN
( AC POWER ON/OFF NO DAMAGE )
2
3
INPUT FREQUENCY RANGE
47HZ ~63 HZ
I/P: 90VAC ~ 264 VAC
NO DAMAGE OSC
O/P:FULL~MIN LOAD
Ta:25℃
77 % (TYP)
EFFICIENCY
TEST: OK
P
I/P: 230 VAC
77.08 %
O/P:FULL LOAD
Ta:25℃
4
INPUT CURRENT
P
230V/
1
A (TYP)
I/P: 230 VAC
I=
0.83
A/ 230 VAC
115V/
1.5
A (TYP)
I/P: 115 VAC
I=
1.42
A/ 115 VAC
P
O/P:FULL LOAD
Ta:25℃
5
INRUSH CURRENT
230V/
50
A (TYP)
I/P: 230 VAC
I=
37
A/ 230 VAC
115V/
25
A(TYP)
I/P: 115 VAC
I=
18
A/ 115 VAC
COLD START
6
LEAKAGE CURRENT
<
<
150
100
P
O/P:FULL LOAD
Ta:25℃
uA/ 264 VAC
I/P: 264 VAC
L-FG:
119
uA
for
N-FG:
119
uA
leakage current
O/P:Min LOAD
Ta:25℃
uA/264 VAC
I/P: 264 VAC
L-V+ :
84
uA
for
O/P:Min LOAD
Ta:25℃
L-V-:
84
uA
N-V+:
84
uA
N-V-:
84
uA
earth
touch
leakage current
P
P
PROTECTION FUNCTION TEST
NO
1
2
3
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
OVER LOAD PROTECTION
140%~ 180%
I/P: 230 VAC
165%/ 230 VAC
I/P: 115 VAC
164%/ 115 VAC
O/P:TESTING
Ta:25℃
Hiccup Mode
OVER VOLTAGE PROTECTION
SHORT PROTECTION
CH1: 5.75V~ 6.75V
I/P: 230 VAC
6.3V/ 230 VAC
I/P: 115 VAC
6.3V/ 115 VAC
O/P:MIN LOAD
Ta:25℃
Shunt down Re- power ON
SHORT EVERY OUTPUT
I/P: 264 VAC
NO DAMAGE
1 HOUR NO DAMAGE
O/P:FULL LOAD
Ta:25℃
Hiccup Mode
Model: RPD-75A
VERDICT
P
P
P
Test Report
2 / 5
RPD-75 series
75W Dual Output Medical Type
ENVIRONMENT TEST
NO
1
TEST ITEM
SPECIFICATION
TEMPERATURE RISE TEST
MODEL:RPD-75A WITH FAN
TEST CONDITION
RESULT
VERDICT
1. ROOM AMBIENT BURN-IN:1.5 HRS
I/P: 230VAC O/P: FULL LOAD Ta=29.6 ℃
2. HIGH AMBIENT BURN-IN:1 HRS
I/P: 230VAC O/P: FULL LOAD Ta=54.6 ℃
2
3
OVER LOAD BURN-IN TEST
LOW TEMPERATURE
ROOM AMBIENT HIGH AMBIENT
Ta= 29.6 ℃
Ta= 54.6 ℃
47.1℃
70.1℃
NO
Position
P/N
1
2
BD1
C6
US4K80 4A/800V SHI
100U/400V NCC 105℃
37.9℃
60.7℃
3
LF2
TR-719
47.3℃
69.9℃
4
C36
47U/63V RUB 105℃ YXF
41.3℃
64.2℃
5
Q1
STP9NK70ZFP 7.5A/700V
47.0℃
71.1℃
6
ZD1
P6KE250A LT
51.2℃
72.8℃
7
D1
HER308 3A/1KV REC
54.3℃
77.1℃
8
T1 COIL
TF-1522
61.0℃
83.3℃
STPS2045CT 20A/45V ST
58.4℃
79.4℃
9
D100
10
D200
SF10SC6 10A/60V SHI
64.8℃
83.3℃
11
C105
22000U/16V RUB 105℃
43.8℃
65.7℃
12
C205
1000U/16V RUB 105℃ ZLH
55.3℃
77.6℃
13
U1
NCP1230D165R2 ON
38.1℃
61.0℃
14
R205
150Ω/2W R/MO
62.9℃
85.8℃
TEST:OK
NO DAMAGE
I/P: 230 VAC
1 HOUR ( MIN )
O/P:V1=11A V2=5A
Ta:25℃
TURN ON AFTER
2 HOUR
TURN ON TEST
P
I/P: 230 VAC
O/P: 100﹪LOAD
P
TEST:OK
P
Ta= -20℃
4
AFTER 12 HOURS
I/P: 272 VAC
HIGH TEMPERATURE
IN CHAMBER ON
CONTROL 50 ℃
O/P:FULL LOAD
Ta=50 ℃
NO DAMAGE
+ 0.03 %(0~50℃)
I/P: 230 VAC
HIGH VOLTAGE
TURN ON TEST
5
TEMPERATURE
VIBRATION TEST
HUMIDITY=
P
95 %R.H
+ 0.003 %(0~50℃)
P
O/P:FULL LOAD
COEFFICIENT
6
TEST:OK
HIGH HUMIDITY
1 Carton & 1 Set
TEST:OK
(1) Waveform: Sine Wave
(2) Frequency:10~500Hz
(3) Sweep Time:10min/sweep cycle
P
(4) Acceleration:2G
(5) Test Time:1 hour in each axis (X.Y.Z)
(6) Ta:25℃
Model: RPD-75A
Test Report
3 / 5
RPD-75 series
75W Dual Output Medical Type
SAFETY TEST
NO
1
2
3
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
WITHSTAND VOLTAGE
I/P-O/P: 4 KVAC/min
I/P-O/P: 4.2 KVAC/min
I/P-O/P:
2.01
mA
I/P-FG: 1.5 KVAC/min
I/P-FG: 1.8 KVAC/min
I/P-FG:
1.29
mA
O/P-FG: 1.5 KVAC/min
O/P-FG: 1.8 KVAC/min
Ta:25℃
O/P-FG:
1.34
mA
I/P-O/P:500VDC>100MΩ
I/P-O/P: 500
VDC
I/P-O/P:
6
GΩ
I/P-FG: 500VDC>100MΩ
I/P-FG:
500
VDC
O/P-FG:500VDC>100MΩ
O/P-FG: 500
Ta:25℃
VDC
I/P-FG:
13
GΩ
O/P-FG:
14
GΩ
ISOLATION RESISTANCE
APPROVAL
VERDICT
P
NO DAMAGE
P
NO DAMAGE
TUV: Certificate NO : TA 50096088
P
UL: File NO :E227340
E.M.C TEST
NO
1
2
3
4
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
HARMONIC
EN61000-3-2
I/P: 230 VAC/50HZ
PASS
CLASS A
O/P:FULL LOAD
Ta:25℃
EN55022
I/P: 230 VAC (50HZ)
PASS
CLASS B
O/P:FULL/50% LOAD
Ta:25℃
Test by certified Lab
CONDUCTION
RADIATION
E.S.D
5
E.F.T
EN55022
I/P: 230 VAC (50HZ)
PASS
O/P:FULL LOAD
Ta:25℃
Test by certified Lab
EN61000-4-2
MEDICAL
I/P: 230 VAC/50HZ
CRITERIA A
EN61000-4-4
MEDICAL
SURGE
P
P
O/P:FULL LOAD
Ta:25℃
I/P: 230 VAC/50HZ
P
CRITERIA A
O/P:FULL LOAD
Ta:25℃
INPUT: 2KV
6
P
CLASS B
AIR:8KV / Contact:6KV
VERDICT
IEC61000-4-5
MEDICAL
I/P: 230 VAC/50HZ
P
CRITERIA A
O/P:FULL LOAD
Ta:25℃
L-N :1KV
P
L,N-PE:2KV
7
Test by certified Lab & Test Report Prepare
M.T.B.F & LIFE CYCLE CALCULATION
NO
1
TEST ITEM
SPECIFICATION
CAPACITOR
SUPPOSE C 105 IS THE MOST CRITICAL COMPONENT
I/P: 230VAC O/P:FULL LOAD Ta= 25 ℃ LIFE TIME=997838HRS
LIFE CYCLE
I/P: 230VAC
2
3
TEST CONDITION
O/P:FULL LOAD Ta= 50 ℃
RESULT
VERDICT
P
LIFE TIME=265547HRS
MTBF
MIL-HDBK-217F NOTICES2 PARTS COUNT
DMTBF/Accelerated Life Test
TOTAL FAILURE RATE: 569.9K HRS
Demonstration Mean Time Between Failure : Above 30,000 hours @ TA 40℃
P
P
Model: RPD-75A
Test Report
4 / 5
RPD-75 series
75W Dual Output Medical Type
COMPONENT STRESS TEST
NO
1
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
Power Transistor
Q1 Rated
I/P:High-Line +3V = 267 V
(1)
692
V
VERDICT
( D to S) or (C to E) Peak Voltage
STP9NK70ZFP : 700V 7.5A
O/P: (1)Full Load Turn on
(2)
715
V
P
(2) Output Short
Ta:25℃
2
Diode Peak Voltage
D100 Rated
I/P:High-Line +3V = 267 V
(1)
38.8
V
STPS2045CT : 45V 20A
O/P: (1)Full Load Turn on
(2)
32.6
V
(1)
50
V
(2)
42
V
(1)
630
V
(2)Output Short
Ta:25℃
P
D200 Rated
SF10SC6 : 60V
3
Clamp Diode Peak Voltage
10 A
D1 Rated
I/P:High-Line +3V = 267 V
HER308 : 1000V 3A
O/P: (1) Dynamic Load
90%Duty/1KHz
P
Ta:25℃
4
Input Capacitor Voltage
C5 Rated
: 100u / 400V/ 105℃
I/P:High-Line +3V = 267 V
(1)
384
V
O/P: (1)Full Load Turn on /Off
(2)
388
V
(2) Min load Turn on /Off
(3)
388
V
P
(3)Full Load /Min load Change
Ta:25℃
5
Control IC Voltage Test
U1 Rated
I/P:High-Line +3V = 267 V
(1)
12.84
V
1230D165R2G : 18V
O/P: (1)Full Load Turn on /Off
(2)
11.69
V
(2) Min load Turn on /Off
(3)
12.84
V
P
(3)Full Load /Min load Change
Ta:25℃
DATE
2006/9/5
2006/11/14
2007/5/7
SAMPLE
TEST RESULT
TESTER
APPROVAL
RD SAMPLE
PASS
VINCENT TSENG
MAX LIN
PASS
VINCENT TSENG
MAX LIN
PASS
VINCENT TSENG
MAX LIN
PRODUCT SAMPLE
W0610A20
PRODUCT SAMPLE
W0703A31
2003/12/12 A50-F023
Model: RPD-75A
Test Report
5 / 5